Conference on Precision Electromagnetic Measurements
Evaluation of Systematic Uncertainty Induced by Excess Micromotion in NIMT Optical Atomic Clock
R. Kaewuam, P. Phoonthong
Abstract
Excess micromotion is a systematic effect that can cause a significant frequency shift to an optical atomic clock based on trapped ions. It is a rapid driven motion that is typically caused by a stray electric field in the trap. This fast motion results in motion-induced shifts of the clock frequency. Here we present the measurement of excess micromotion in a linear Paul trap at National Institute of Metrology Thailand (NIMT). After the stray field compensation, the estimated second-order Doppler shift and scalar Stark shift of the 435-nm clock transition in171Yb+are –1.5(8) × 10–18and 1.3(7) × 10–17, respectively.