ARFTG Microwave Measurement Conference

A Cryogenic Quantum-Based RF Source

Justus A. Brevik, Alírio Boaventura, Manuel A. Castellanos-Beltran, Christine A. Donnelly, Nathan E. Flowers-Jacobs, Anna E. Fox, P. F. Hopkins, Paul D. Dresselhaus, Dylan Forrest Williams, Samuel P. Benz

Published Aug 4, 2020• 2 citations
Publisher

Abstract

We performed a preliminary calibrated measurement of the output power of a Josephson arbitrary waveform synthesizer up to 1 GHz. We present the results and measurement procedure for generating quantum-based signals using an array of Josephson junctions operating at cryogenic temperature and calibrating those signals to transfer the on-wafer quantum-based accuracy to room temperature.

DOI
10.1109/arftg47271.2020.9241378

Related papers

Cryogenic Calibration of the RF Josephson Arbitrary Waveform Synthesizer

Justus A. Brevik, Alírio Boaventura, Akim A. Babenko +7 more

We performed a preliminary calibrated measurement of the output power of a Josephson arbitrary waveform synthesizer up to 1 GHz. We present the results and measurement procedure for generating quantum-based signals using an array of Josephson junctions operating at cryogenic temperature and calibrating those signals to transfer the on-wafer quantum-based accuracy to room temperature.

Conference on Precision Electromagnetic Measurements•Aug 1, 2020•12 citations

Cryogenic Calibration of a Quantum-based Radio Frequency Source

Alírio Boaventura, Justus A. Brevik, Dylan Forrest Williams +5 more

We report on the calibration of quantum-based radio frequency waveforms generated by a Josephson arbitrary waveform synthesizer system. We measure these waveforms using a vector network analyzer and calibrate them at 4 K using a custom-designed cryogenic on-wafer multi-line thru-reflect-line calibration kit and a two-tier calibration procedure. The signals tested in this work can be used as reference signals to calibrate measurement instruments with potential benefits in terms of accuracy and flexibility compared to the conventional methods.

ARFTG Microwave Measurement Conference•Aug 4, 2020•5 citations

Traceable Validation of Calculable Quantum-Based RF Waveforms

Jeremy N. Thomas, Nathan E. Flowers-Jacobs, Anna E. Fox +2 more

We compare the calculated, quantum-based output of a radio-frequency Josephson arbitrary waveform synthesizer (RF-JAWS) with its measured output that is calibrated through the established NIST RF power traceability path, a necessary first step towards quantum-based RF metrology. An array of 1,500 Josephson junctions (JJs) is used to synthesize sinusoids from 50MHz to 3GHz, each with a nominal output power of −49.67dBm and a demonstrated quantum-locking range (QLR) exceeding 1mA. We perform a two-tier calibration of two-port vector network analyzer (VNA) measurements of the RF-JAWS waveforms with uncertainty analysis: (1) the VNA wave parameters are referenced to a NIST-calibrated coaxial power sensor, and (2) a cryogenic probe station, together with on-chip superconducting calibration standards, establishes an on-chip reference plane at the JJ-array. The calibrated measurement of the forward- and backward-propagating power is compared with the expected, calculated outputs, and, after accounting for high-frequency roll-off, we observe agreement within the expanded (k=2) uncertainty limits for most frequencies.

Conference on Precision Electromagnetic Measurements•Sep 1, 2026•0 citations

Cryogenic Characterization of a Superconductor Quantum-Based Microwave Reference Source for Communications and Quantum Information

Alírio Boaventura, Justus A. Brevik, Dylan Forrest Williams +6 more

We are developing a new instrument, the RF Josephson arbitrary waveform synthesizer (RF-JAWS), for communications metrology and quantum information applications. An important aspect of the RF-JAWS design is the accurate and traceable characterization of its superconducting devices. In this article, we present a procedure for characterizing microwave superconducting devices in a cryogenic RF probe station via a vector network analyzer (VNA) calibrated with a custom cryogenic calibration kit colocated with the superconducting device under test (DUT) in the cryogenic environment. By de-embedding lossy and dispersive RF interconnects linking the superconducting DUT to the measurement apparatus at room temperature, we characterize the DUT exactly at the cryogenic on-wafer reference plane. More importantly, we operate our VNA with an external modulated source and our procedure features metrology-grade multiline thru-reflect-line calibration and absolute power and phase corrections, as opposed to the more common relative scattering-parameter correction. In addition, we apply anX-parameter model to account for impedance mismatch in cryogenic Josephson microwave sources. Our techniques are also suitable for cryogenic characterization of microwave superconducting devices for solid-state quantum computers and could help to optimize the quantum-classical interfaces in these systems.

IEEE Transactions on Applied Superconductivity•Oct 13, 2021•9 citations

Quantum-Based Modulated Microwave Waveforms

Akim A. Babenko, Nathan E. Flowers-Jacobs, Anna E. Fox +3 more

This article presents a superconducting voltage source that generates modulated microwave waveforms with quantum-based stability. The voltage source—an RF Josephson Arbitrary Waveform Synthesizer (RF-JAWS)—uses a superconducting IC with an array of 4500 series-connected Josephson junctions (JJs) that are embedded in a coplanar waveguide (CPW) transmission line. The JJs are driven with 100-ps wide current pulses that force each JJ to generate voltage pulses with a quantized integrated area. A pulse sequence is created using a delta-sigma ($\Delta $–$\Sigma $) algorithm to generate a 101-tone waveform with a 40-MHz instantaneous bandwidth around 1005 MHz with a flat power distribution of −53 dBm and a Schroeder phase distribution. The JJ’s quantum-based nonlinearity produces an amplitude stability with respect to various drive and bias parameters of ±0.005 dB and a detrended phase stability of ±0.05°. The stability of the source is verified without the need for calibrated measurements. Another$\Delta $–$\Sigma $pulse sequence generates a 10-MHz quadrature-phase shift-keying (QPSK) waveform on a 1005-MHz carrier that has the same stability and the mean error vector magnitude (EVM) of −48 dB. This work is a step toward creating a quantum-based reference source for power and cross-frequency phase calibrations, as well as a reference-modulated signal source for calibrating telecommunication links.

IEEE Transactions on Microwave Theory and Techniques•Aug 4, 2023•6 citations

Investigating the Use of Super Conducting Josephson Circuits for Microwave Power Measurements at Cryogenic Temperatures

Tezgül Coşkun Öztürk, Oliver Kieler

In this paper, the use of Josephson Arbitrary Waveform Synthesizer System's superconducting integrated circuit consisting of two Josephson Junctions arrays, for microwave power measurements is investigated. The critical current, the temperature dependent phenomena of the Josephson Junctions, is used to compare DC power and MW power reaching the neighbor array. Validation of the method is roughly demonstrated by measuring the attenuation of the waveguide used within the stick used to immerse the circuit to cryogenic environment. The attenuation of the waveguide is measured with the novel sensor and the conventional method. The preliminary results show 0.2 dB consistence of the attenuation measurements between the conventional and the novel method at different power levels at 1 GHz, that suggest that the method can be useful. To improve the repeatability and accuracy by automation, which is essential for such measurements, new power sensor is designed and presented.

National Conference on Electrical and Electronics Engineering (ELECO)•Nov 30, 2023•3 citations
We performed a preliminary calibrated measurement of the output power of a Josephson arbitrary waveform synthesizer up to 1 GHz. We present the results and measurement procedure for generating quantum-based signals using an array of Josephson junctions operating at cryogenic temperature and calibrating those signals to transfer the on-wafer quantum-based accuracy to room temperature.
We report on the calibration of quantum-based radio frequency waveforms generated by a Josephson arbitrary waveform synthesizer system. We measure these waveforms using a vector network analyzer and calibrate them at 4 K using a custom-designed cryogenic on-wafer multi-line thru-reflect-line calibration kit and a two-tier calibration procedure. The signals tested in this work can be used as reference signals to calibrate measurement instruments with potential benefits in terms of accuracy and flexibility compared to the conventional methods.
We compare the calculated, quantum-based output of a radio-frequency Josephson arbitrary waveform synthesizer (RF-JAWS) with its measured output that is calibrated through the established NIST RF power traceability path, a necessary first step towards quantum-based RF metrology. An array of 1,500 Josephson junctions (JJs) is used to synthesize sinusoids from 50MHz to 3GHz, each with a nominal output power of −49.67dBm and a demonstrated quantum-locking range (QLR) exceeding 1mA. We perform a two-tier calibration of two-port vector network analyzer (VNA) measurements of the RF-JAWS waveforms with uncertainty analysis: (1) the VNA wave parameters are referenced to a NIST-calibrated coaxial power sensor, and (2) a cryogenic probe station, together with on-chip superconducting calibration standards, establishes an on-chip reference plane at the JJ-array. The calibrated measurement of the forward- and backward-propagating power is compared with the expected, calculated outputs, and, after accounting for high-frequency roll-off, we observe agreement within the expanded (k=2) uncertainty limits for most frequencies.
We are developing a new instrument, the RF Josephson arbitrary waveform synthesizer (RF-JAWS), for communications metrology and quantum information applications. An important aspect of the RF-JAWS design is the accurate and traceable characterization of its superconducting devices. In this article, we present a procedure for characterizing microwave superconducting devices in a cryogenic RF probe station via a vector network analyzer (VNA) calibrated with a custom cryogenic calibration kit colocated with the superconducting device under test (DUT) in the cryogenic environment. By de-embedding lossy and dispersive RF interconnects linking the superconducting DUT to the measurement apparatus at room temperature, we characterize the DUT exactly at the cryogenic on-wafer reference plane. More importantly, we operate our VNA with an external modulated source and our procedure features metrology-grade multiline thru-reflect-line calibration and absolute power and phase corrections, as opposed to the more common relative scattering-parameter correction. In addition, we apply anX-parameter model to account for impedance mismatch in cryogenic Josephson microwave sources. Our techniques are also suitable for cryogenic characterization of microwave superconducting devices for solid-state quantum computers and could help to optimize the quantum-classical interfaces in these systems.
This article presents a superconducting voltage source that generates modulated microwave waveforms with quantum-based stability. The voltage source—an RF Josephson Arbitrary Waveform Synthesizer (RF-JAWS)—uses a superconducting IC with an array of 4500 series-connected Josephson junctions (JJs) that are embedded in a coplanar waveguide (CPW) transmission line. The JJs are driven with 100-ps wide current pulses that force each JJ to generate voltage pulses with a quantized integrated area. A pulse sequence is created using a delta-sigma ($\Delta –––\Sigma )algorithmtogeneratea101−tonewaveformwitha40−MHzinstantaneousbandwidtharound1005MHzwithaflatpowerdistributionof−53dBmandaSchroederphasedistribution.TheJJ’squantum−basednonlinearityproducesanamplitudestabilitywithrespecttovariousdriveandbiasparametersof±0.005dBandadetrendedphasestabilityof±0.05°.Thestabilityofthesourceisverifiedwithouttheneedforcalibratedmeasurements.Another) algorithm to generate a 101-tone waveform with a 40-MHz instantaneous bandwidth around 1005 MHz with a flat power distribution of −53 dBm and a Schroeder phase distribution. The JJ’s quantum-based nonlinearity produces an amplitude stability with respect to various drive and bias parameters of ±0.005 dB and a detrended phase stability of ±0.05°. The stability of the source is verified without the need for calibrated measurements. Another)algorithmtogeneratea101−tonewaveformwitha40−MHzinstantaneousbandwidtharound1005MHzwithaflatpowerdistributionof−53dBmandaSchroederphasedistribution.TheJJ’squantum−basednonlinearityproducesanamplitudestabilitywithrespecttovariousdriveandbiasparametersof±0.005dBandadetrendedphasestabilityof±0.05°.Thestabilityofthesourceisverifiedwithouttheneedforcalibratedmeasurements.Another\Delta –––\Sigma $pulse sequence generates a 10-MHz quadrature-phase shift-keying (QPSK) waveform on a 1005-MHz carrier that has the same stability and the mean error vector magnitude (EVM) of −48 dB. This work is a step toward creating a quantum-based reference source for power and cross-frequency phase calibrations, as well as a reference-modulated signal source for calibrating telecommunication links.
In this paper, the use of Josephson Arbitrary Waveform Synthesizer System's superconducting integrated circuit consisting of two Josephson Junctions arrays, for microwave power measurements is investigated. The critical current, the temperature dependent phenomena of the Josephson Junctions, is used to compare DC power and MW power reaching the neighbor array. Validation of the method is roughly demonstrated by measuring the attenuation of the waveguide used within the stick used to immerse the circuit to cryogenic environment. The attenuation of the waveguide is measured with the novel sensor and the conventional method. The preliminary results show 0.2 dB consistence of the attenuation measurements between the conventional and the novel method at different power levels at 1 GHz, that suggest that the method can be useful. To improve the repeatability and accuracy by automation, which is essential for such measurements, new power sensor is designed and presented.