Abstract To address the interference of environmental magnetic field fluctuations on the diamond Nitrogen-Vacancy (NV) center quantum frequency standard, this paper proposes and implements a composite frequency stabilization scheme based on dual-frequency locking and active magnetic field compensation. Distinct from traditional single-frequency locking schemes, this approach utilizes dual microwave sources to simultaneously lock onto two magnetic resonance transition frequencies of the NV center ground state; one locked peak serves as the frequency standard output, while the other is utilized for magnetic field monitoring. Experimental results demonstrate that this scheme effectively eliminates the influence of ambient stray magnetic fields. At an integration time of 𝜏 = 30 s, the frequency stability at 2.87 GHz reaches 2.53 × 10 −6 , providing a technical reference for the development of robust solid-state quantum frequency standards.
Abstract To address the interference of environmental magnetic field fluctuations on the diamond Nitrogen-Vacancy (NV) center quantum frequency standard, this paper proposes and implements a composite frequency stabilization scheme based on dual-frequency locking and active magnetic field compensation. Distinct from traditional single-frequency locking schemes, this approach utilizes dual microwave sources to simultaneously lock onto two magnetic resonance transition frequencies of the NV center ground state; one locked peak serves as the frequency standard output, while the other is utilized for magnetic field monitoring. Experimental results demonstrate that this scheme effectively eliminates the influence of ambient stray magnetic fields. At an integration time of 𝜏 = 30 s, the frequency stability at 2.87 GHz reaches 2.53 × 10 −6 , providing a technical reference for the development of robust solid-state quantum frequency standards.
High-resolution wide-spectrum measurement techniques have important applications in fields such as astronomy, wireless communication, and medical imaging. Nitrogen-vacancy (NV) center in diamond is well known for its high stability, high sensitivity, real-time monitoring, single-point detection, and suitability for long-term measurement, and has an outstanding choice for spectrum analyzers. Currently, spectrum analyzers based on NV centers as detectors can perform real-time spectrum analysis in the range of several tens of gigahertz, but their frequency resolution is limited to a MHz level. In this study, we construct a quantum diamond microwave spectrum imaging system by combining continuous wave-mixing techniques. According to the spin-related properties of the NV center in diamond, we implement optical pumping by 532 nm green laser light illuminating the diamond NV center. A spherical magnet is used to produce a magnetic field gradient along the direction of the diamond crystal. By adjusting the size and direction of the magnetic field gradient, spatial encoding of the resonance frequency of the NV center is achieved. The magnetic field gradient induces the Zeeman effect on the diamond surface at different positions, generating corresponding ODMR signals. Through accurate programming, we coordinate the frequency scanning step size of the microwave source with the camera exposure and image storage time, and synchronize them circularly according to the order of image acquisition. Ultimately, after algorithmic processing, we successfully obtain comprehensive spectrum data in a range from 900 MHz to 6.0 GHz. Within the measurable spectrum range, the system employs continuous wave-mixing, simultaneously applying resonant microwaves and slightly detuning auxiliary microwaves to effectively excite the NV center. This method triggers off microwave interference effects, disrupting the balance between laser-induced polarization and microwave-induced spontaneous relaxation. Specifically, microwave interference causes the phase and amplitude of the fluorescence signal to change, leading to the generation of alternating current fluorescence signals. This further enhances the response of the NV magnetometer to weak microwave signals. The method enables the system to achieve a frequency resolution of 1 Hz in the measurable spectrum range, and it can separately measure the frequency resolution of multiple frequency points with a frequency step size of 1 MHz. The research results indicate that the wide-spectrum measurement based on NV centers can achieve sub-hertz frequency resolution, providing robust technical support for future spectrum analysis and applications.
Nitrogen-vacancy (NV) centers in diamonds play a large role in advanced quantum sensing with solid-state spins for potential miniaturized and portable application scenarios. With the temperature sensitivity of NV centers, the temperature fluctuations caused by the unknown environment and the system itself will mix with the magnetic field measurement. In this research, the temperature-sensitive characteristics of different diamonds, alongside the temperature noise generated by a measurement system, were tested and analyzed with a homemade NV magnetometer in a fiber-optic scheme. In this work, a multi-frequency synchronous manipulation method for resonating with the NV centers in all axial directions was proposed to compensate for the temperature fluctuations in a fibered NV magnetic field sensing scheme. The symmetrical features of the resonance lines of the NV centers, the common-mode fluctuations including temperature fluctuations, underwent effective compensation and elimination. The fluorescence change was reduced to 1.0% by multi-frequency synchronous manipulation from 5.5% of the single-frequency manipulation within a ±2 °C temperature range. Additionally, the multi-frequency synchronous manipulation improved the fluorescence contrast and the magnetic field measurement SNR through an omnidirectional manipulation scheme. It was very important to compensate for the temperature fluctuations, caused by both internal and external factors, to make use of the NV magnetometer in fiber-optic schemes' practicality. This work will promote the rapid development and widespread applications of quantum sensing based on various systems and principles.
Frequency references based on solid state spins promise simplicity, compactness, robustness, multifunctionality, ease of integration, and high densities of emitters. Nitrogen-vacancy (NV) centers in diamond are a natural candidate, but the electronic zero-field splitting exhibits a large fractional temperature dependence, which has precluded its use as a stable clock transition. Here we show that this limitation can be overcome by forming a composite frequency reference that combines measurements of the electronic splitting D with the nuclear quadrupole splitting of the 14N nuclear spin intrinsic to the NV center. We further benchmark this composite approach against alternative strategies for mitigating temperature sensitivity. By implementing a specially designed pulse sequence with an eight-phase control scheme that suppresses pulse imperfections, we interleave measurements of D and Q in a high-density NV ensemble and demonstrate a temperature-compensated composite frequency reference. The stability of this composite diamond clock is characterized over a 10-day period at room temperature through a comparison to a Rb vapor-cell clock, yielding a fractional instability below 5×10−9 for an averaging time of τ=200 s and below 1×10−8 at τ=2×105 s, corresponding to measured improvements by a factor of 4 and 200, respectively, over a clock based purely on the single frequency D for the same periods. By characterizing the residual sensitivity to magnetic fields, optical power, and radio-frequency drive amplitudes, we find that temperature is no longer the dominant source of instability. These results establish complementary electron- and nuclear-spin transitions in diamond as a viable route to thermally robust frequency metrology, providing a pathway toward compact, multifunctional solid-state clocks and quantum sensors.
Microwave control of the negatively charged nitrogen-vacancy (NV) defect in diamond is usually based on a two-level approximation of its ground state that is only valid in the regime of high bias magnetic fields or low microwave power. We present the NV-ERC technique, which has demonstrated efficient initialization and readout of the double quantum transition with no leakage to any third level, thanks to an effective Raman coupling. The protocol may be used in the low-field regime and for high-frequency sensing. Furthermore, we propose a scheme to perform fast single-qubit gates in the double quantum transition, consisting of pulse trains of constant frequency and intensity. We analyze its robustness with respect to pulse-timing errors resulting from inaccurate identification of system parameters or phase-control limitations. We also demonstrate that the technique can be implemented in the presence of unknown electric or strain fields.
Abstract Multiple-microwave-photon transition is a nonlinear process involving the simultaneous absorption of multiple microwave photons to reach a higher energy state. This phenomenon enables microwave frequency conversion, a crucial technique for spin-wave devices, frequency synthesizers and communication devices. However, it has not been observed in nitrogen-vacancy (NV) centers in diamonds, which serve as a key quantum platform linking flying qubits, microwave photons, and solid-state spins. In this work, we first observed multiple-microwave-photon transitions in NV centers in diamonds, using optically detected magnetic resonance (ODMR) methods with a single laser and microwave beam. We demonstrate that microwave frequencies, effective g-factor, and angular momentum exhibit a 1/n scaling law, but the linewidth exhibits a 1/n^2 scaling law for n-photon transitions between the ground states. Additionally, we detect the hyperfine structure of the ground states and achieve coherent spin manipulation under multi-photon excitation. Our findings offer a novel approach for the precise manipulation of NV center spin states, advancing quantum sensing and quantum information.
In this work we present a compact and portable tabletop magnetometer that utilizes negatively charged nitrogen-vacancy (NV) centers in diamond. The magnetometer is operated using a dual microwave resonance detection approach in combination with an optically detected magnetic resonance (ODMR) technique (mitigating drifts in results due to changes of the diamond temperature), capable of simultaneously exciting and registering two ODMR transitions. The experimentally measured magnetic field noise-floor is ≈2.3nTHz while the calculated shot-noise-limited magnetic field sensitivity is ≈585pTHz when excited with a continuous wave laser at 0.5~W.
These results pave the way for realizing a simple set-up magnetometer for precise single axis magnetic field measurements for example for accurate electric current measurements for stabilization purposes and magnetic communication applications.