Development of a transportable cold atom gradiometer

Andrew Hinton

Published Dec 1, 2016• 1 citation
OpenAlex

Abstract

This work describes the technology developed for the transportable gravity gradiometer, ''GGtop", constructed at the University of Birmingham. The device aims to simultaneously interrogate two physically separated, free-falling cold atom clouds using the technique of Raman interferometry to perform differential gravity measurements. Such a technique will suppress common-mode noise sources, such as vibrations, that otherwise limit classical gravimeters and cold atom interferometers based on single-cloud measurements. The technique is expected to improve speed and sensitivity for field measurements greatly. \nUsing a combination of commercially available components and novel designs, intended to enhance portability and robustness, allowed for demonstration of atomic interference with the apparatus via Ramsey's method of separated oscillatory fields. The achieved fringe contrast of ~2%, defined as the difference in the number of atoms detected in IF = 2>, was limited by drifts stemming from some of the novel designs which prompted continued optimisation of the underpinning subsystems. To address performance issues parts of the experiment were redesigned with the goal of improving reliability at the expense of some portability. Using the retrofitted experiment, interference was once again achieved with fitted fringe spacing of 134.7±2.0 uS in good agreement with the 133.9 uS defined by the experimental control. A factor of 10 improvement in contrast was found with the central fringe demonstrating 18% of the atoms detected in the IF = 2> state when normalised to the total 3D MOT number. The 3.71 ± 0.01 kHz fitted linewidth of the central fringe gives a frequency uncertainty of 5.43 ± 0.01 x 10 - 7. This result leaves the experiment in a good position to begin making measurements of gravity.

Related papers

Construction of a Rubidium Fountain Atomic Interferometer for Gravity Gradiometry

JN Tinsley

Atomic interferometry is a highly precise metrological technique of interest to both fundamental and applied physics. One field in which atomic interferometry is of particular interest is gravimetry, where instruments based upon the controlled manipulation of atomic states have superseded the performance of classical devices. In such devices, a standing wave derived from laser beams with well-defined phase difference are used to impart momentum to atoms undergoing free fall, generating spatial superpositions along the axis of gravity, before returning the atoms to an overlapped final state for recombination. This thesis presents the design and experimental progress towards a gravity gradiometer based upon the interference of rubidium-87 atoms, cooled and launched in a fountain configuration. The fountain is leveraged from a decommissioned frequency standard, which has been re-purposed and redesigned for use as a combined absolute gravimeter and gravity gradiometer package. It is capable of loading 100 billion atoms in 100 ms and then cooling and launching these atoms in a moving molasses configuration, with a final temperature of 5 μK. Furthermore, the process can be quickly repeated, with a second cloud subsequently launched, reaching its apogee at the same time as the first cloud, allowing for the possibility of simultaneous measurement at different heights and, in the future, the inference of the gravity gradient. Control of the fountain is provided by a stable and miniaturised optical setup, allowing for use with a future transportable or field-based device. A Raman beam system to perform the interferometry sequence has been constructed and characterised, likewise designed in a minaturised and stable manner. The Raman beams simultaneously generated Ramsey interference fringes on two independent clouds of atoms -- one launched to a height of 0.7 m and the other to 1 m. Such fringes were clearly observable after 200 ms of evolution time. Mach-Zehnder interference fringes were also demonstrated for dropped and launched atoms, providing the capability for gravimetric measurements to be made once the system has been fully characterised. The principle causes of noise in the Mach-Zehnder cases were identified as arriving from a lack of atoms and the phase noise of the Raman beams. To overcome these challenges, a new Raman beam system has been preliminary designed and tested and a new iteration of the fountain, moving towards quantum degeneracy, has been planned.

•Jan 1, 2019•2 citations

Towards mobile quantum sensors for gravity surveys

Alexander Niggebaum

Design and first characterisations of the mobile sensor head in the Gravity Gradient Technologies and Opportunities Programme (GGtop) are presented. The aim of the project is the development of a mobile gravity gradiometer and faces the challenge of condensing a normally lab filling experiment in a portable and robust package. A fibre network replaced free space optics light distribution and fibre based components free space equivalents. Although stable against misalignment, the systems performance is limited by polarization changes of the guided light which arise from birefringence fluctuations along the length of the network due to external temperature fluctuations and mechanical disturbances. These instabilities limit the achievable temperature of the trapped rubidium 87 cloud to approx. 18μK. In preparation for gravity measurements, Rabi oscillations and Ramsey fringes with a λ/2 time of 7.4μs were successfully demonstrated in a co-propagating Raman beam configuration. The atom cloud was launched as a first step towards gravity gradiometry. As the system was designed to be portable, the complete system fits into a 1.5m x 2m x 0.5m package, plus a 14u rack of support electronics.

•Jul 1, 2016•4 citations

A Transportable Gravity Gradiometer Based on Atom Interferometry

Nan Yu, Robert J. Thompson, James R. Kellogg +3 more

A transportable atom interferometer-based gravity gradiometer has been developed at JPL to carry out measurements of Earth's gravity field at ever finer spatial resolutions, and to facilitate high-resolution monitoring of temporal variations in the gravity field from ground- and flight-based platforms. Existing satellite-based gravity missions such as CHAMP and GRACE measure the gravity field via precise monitoring of the motion of the satellites; i.e. the satellites themselves function as test masses. JPL's quantum gravity gradiometer employs a quantum phase measurement technique, similar to that employed in atomic clocks, made possible by recent advances in laser cooling and manipulation of atoms. This measurement technique is based on atomwave interferometry, and individual laser-cooled atoms are used as drag-free test masses. The quantum gravity gradiometer employs two identical atom interferometers as precision accelerometers to measure the difference in gravitational acceleration between two points (Figure 1). By using the same lasers for the manipulation of atoms in both interferometers, the accelerometers have a common reference frame and non-inertial accelerations are effectively rejected as common mode noise in the differential measurement of the gravity gradient. As a result, the dual atom interferometer-based gravity gradiometer allows gravity measurements on a moving platform, while achieving the same long-term stability of the best atomic clocks. In the laboratory-based prototype (Figure 2), the cesium atoms used in each atom interferometer are initially collected and cooled in two separate magneto-optic traps (MOTs). Each MOT, consisting of three orthogonal pairs of counter-propagating laser beams centered on a quadrupole magnetic field, collects up to 10(exp 9) atoms. These atoms are then launched vertically as in an atom fountain by switching off the magnetic field and introducing a slight frequency shift between pairs of lasers to create a moving rest frame for the trapped atoms. While still in this moving-frame molasses, the laser frequencies are further detuned from the atomic resonance (while maintaining this relative frequency shift) to cool the atom cloud's temperature to 2 K or below, corresponding to an rms velocity of less than 2 cm/s. After launch, the cold atoms undergo further state and velocity selection to prepare for atom interferometry. The atom interferometers are then realized using laser-induced stimulated Raman transitions to perform the necessary manipulations of each atom, and the resulting interferometer phase is measured using laser-induced fluorescence for state-normalized detection. More than 20 laser beams with independent controls of frequency, phase, and intensity are required for this measurement sequence. This instrument can facilitate the study of Earth's gravitational field from surface and air vehicles, as well as from space by allowing gravity mapping from a low-cost, single spacecraft mission. In addition, the operation of atom interferometer-based instruments in space offers greater sensitivity than is possible in terrestrial instruments due to the much longer interrogation times available in the microgravity environment. A space-based quantum gravity gradiometer has the potential to achieve sensitivities similar to the GRACE mission at long spatial wavelengths, and will also have resolution similar to GOCE for measurement at shorter length scales.

•May 1, 2010•2 citations

High-Sensitivity Absolute Atomic Gravimeter

Christine Guerlin, Tristan Farah, Anne Louchet-Chauvet +2 more

Cold atom free-fall interferometers now realize state-of-the-art precision inertial sensors, opening perspectives towards embedded devices for applications and fundamental physics tests on earth and in space. The acceleration of gravity can be measured in a stimulated Raman transition interferometer where the acceleration under gravity of a cloud of freely falling atoms is sensed by three Raman pulses allowing to split, reflect and interfere atomic wave packets [1]. The gravimeter developped at SYRTE allows to measure the acceleration of gravity at state-of-the-art performances compared to the best classical absolute gravimeters, with a sensitivity of 2×10−8g over 1s and an exactitude of 5×10−9 g [2]. This exactitude is currently limited by wavefront aberrations of the lasers, sampled by residual transverse trajectories of the molasse-cooled cloud over the free fall. Unbalanced transverse trajectories will also bias the measurement due to Coriolis effect [3]. We will investigate the use of a colder atom source to reduce these systematic effects, by implementing Bose-Einstein condensation of the initial cloud, allowing both better position control and better focusing of our source.

Research in Optical Sciences•Jan 1, 2012•0 citations

Optimisation of a compact cold-atoms interferometer for gravimetry

Clemens Rammeloo

The work presented in this thesis focusses on the development of a transportable atom-interferometry experiment and a compact fibre laser system towards precision measurements of gravitational acceleration. Interference fringes are shown with clouds of cold 87Rb atoms using co-propagating laser beams to drive stimulated Raman transitions. This is demonstrated both inside and outside of laboratory environments for which an integrated and transportable experiment is constructed. Further improvements are presented that enable the generation of clouds containing 1.7 · 108 atoms at a rate of 2.5 Hz and having a temperature of (7 ± 1) µK. This is largely due to the development of a compact laser system based on all-fibre coupled components. It is demonstrated that the laser system designed here can achieve fast frequency sweeps over 1.8 GHz within 2 ms, making it widely applicable in compact atom-interferometry experiments with rubidium atoms. This is shown by creating a Mach–Zehnder type interferometer with counter-propagating Raman beams, thus enabling measurements of gravitational acceleration. Since the laser system uses only two lasers and one fibre amplifier, a significant reduction in size is achieved, as well as a decrease in the total power consumption of the overall experiment by a third to (162 ± 7) W.

•Jul 1, 2018•1 citation

Construction of a portable platform for cold atom interferometry

Jonathan Ian Malcolm

This thesis details the construction of a portable platform for cold atom experiments, as part of the European iSense project. This culminated in the demonstration of a working portable cold atom interferometer to the praise of the EC and reviewers during the final project meeting. Reductions in the size, weight and power consumption of all the crucial components for a cold atom system such as lasers, optics, magnetic field generation and electronics have been realised. These novel components have been integrated into a portable device that has been transported and operated by just two people. The completed device weighs 63 kg and uses 240 W power in a volume of just 61.1 litres. The system has generated a cloud of (6.7±0.7)x 10\(^6\) rubidium-87 atoms at a temperature of 4.4 μK in 2 seconds using a mirror-MOT setup. The device has successfully performed atom interferometry in the form of a Ramsey interferometer both at the University of Birmingham, UK and also at an office in Brussels, BE after a journey of 570 km. A measurement of the hyperfine splitting of the ground state of rubidium-87 was performed and gives an uncertainty of \(\delta\)\(\omega\)/\(\omega\) = 5.2 × 10\(^-\)\(^7\). The theoretical limit of the sensitivity to gravity of the iSense system is 3.9 μGal/√Hz.

•Jul 1, 2016•6 citations
This work describes the technology developed for the transportable gravity gradiometer, ''GGtop", constructed at the University of Birmingham. The device aims to simultaneously interrogate two physically separated, free-falling cold atom clouds using the technique of Raman interferometry to perform differential gravity measurements. Such a technique will suppress common-mode noise sources, such as vibrations, that otherwise limit classical gravimeters and cold atom interferometers based on single-cloud measurements. The technique is expected to improve speed and sensitivity for field measurements greatly. \nUsing a combination of commercially available components and novel designs, intended to enhance portability and robustness, allowed for demonstration of atomic interference with the apparatus via Ramsey's method of separated oscillatory fields. The achieved fringe contrast of ~2%, defined as the difference in the number of atoms detected in IF = 2>, was limited by drifts stemming from some of the novel designs which prompted continued optimisation of the underpinning subsystems. To address performance issues parts of the experiment were redesigned with the goal of improving reliability at the expense of some portability. Using the retrofitted experiment, interference was once again achieved with fitted fringe spacing of 134.7±2.0 uS in good agreement with the 133.9 uS defined by the experimental control. A factor of 10 improvement in contrast was found with the central fringe demonstrating 18% of the atoms detected in the IF = 2> state when normalised to the total 3D MOT number. The 3.71 ± 0.01 kHz fitted linewidth of the central fringe gives a frequency uncertainty of 5.43 ± 0.01 x 10 - 7. This result leaves the experiment in a good position to begin making measurements of gravity.
Atomic interferometry is a highly precise metrological technique of interest to both fundamental and applied physics. One field in which atomic interferometry is of particular interest is gravimetry, where instruments based upon the controlled manipulation of atomic states have superseded the performance of classical devices. In such devices, a standing wave derived from laser beams with well-defined phase difference are used to impart momentum to atoms undergoing free fall, generating spatial superpositions along the axis of gravity, before returning the atoms to an overlapped final state for recombination. This thesis presents the design and experimental progress towards a gravity gradiometer based upon the interference of rubidium-87 atoms, cooled and launched in a fountain configuration. The fountain is leveraged from a decommissioned frequency standard, which has been re-purposed and redesigned for use as a combined absolute gravimeter and gravity gradiometer package. It is capable of loading 100 billion atoms in 100 ms and then cooling and launching these atoms in a moving molasses configuration, with a final temperature of 5 μK. Furthermore, the process can be quickly repeated, with a second cloud subsequently launched, reaching its apogee at the same time as the first cloud, allowing for the possibility of simultaneous measurement at different heights and, in the future, the inference of the gravity gradient. Control of the fountain is provided by a stable and miniaturised optical setup, allowing for use with a future transportable or field-based device. A Raman beam system to perform the interferometry sequence has been constructed and characterised, likewise designed in a minaturised and stable manner. The Raman beams simultaneously generated Ramsey interference fringes on two independent clouds of atoms -- one launched to a height of 0.7 m and the other to 1 m. Such fringes were clearly observable after 200 ms of evolution time. Mach-Zehnder interference fringes were also demonstrated for dropped and launched atoms, providing the capability for gravimetric measurements to be made once the system has been fully characterised. The principle causes of noise in the Mach-Zehnder cases were identified as arriving from a lack of atoms and the phase noise of the Raman beams. To overcome these challenges, a new Raman beam system has been preliminary designed and tested and a new iteration of the fountain, moving towards quantum degeneracy, has been planned.
Design and first characterisations of the mobile sensor head in the Gravity Gradient Technologies and Opportunities Programme (GGtop) are presented. The aim of the project is the development of a mobile gravity gradiometer and faces the challenge of condensing a normally lab filling experiment in a portable and robust package. A fibre network replaced free space optics light distribution and fibre based components free space equivalents. Although stable against misalignment, the systems performance is limited by polarization changes of the guided light which arise from birefringence fluctuations along the length of the network due to external temperature fluctuations and mechanical disturbances. These instabilities limit the achievable temperature of the trapped rubidium 87 cloud to approx. 18μK. In preparation for gravity measurements, Rabi oscillations and Ramsey fringes with a λ/2 time of 7.4μs were successfully demonstrated in a co-propagating Raman beam configuration. The atom cloud was launched as a first step towards gravity gradiometry. As the system was designed to be portable, the complete system fits into a 1.5m x 2m x 0.5m package, plus a 14u rack of support electronics.
A transportable atom interferometer-based gravity gradiometer has been developed at JPL to carry out measurements of Earth's gravity field at ever finer spatial resolutions, and to facilitate high-resolution monitoring of temporal variations in the gravity field from ground- and flight-based platforms. Existing satellite-based gravity missions such as CHAMP and GRACE measure the gravity field via precise monitoring of the motion of the satellites; i.e. the satellites themselves function as test masses. JPL's quantum gravity gradiometer employs a quantum phase measurement technique, similar to that employed in atomic clocks, made possible by recent advances in laser cooling and manipulation of atoms. This measurement technique is based on atomwave interferometry, and individual laser-cooled atoms are used as drag-free test masses. The quantum gravity gradiometer employs two identical atom interferometers as precision accelerometers to measure the difference in gravitational acceleration between two points (Figure 1). By using the same lasers for the manipulation of atoms in both interferometers, the accelerometers have a common reference frame and non-inertial accelerations are effectively rejected as common mode noise in the differential measurement of the gravity gradient. As a result, the dual atom interferometer-based gravity gradiometer allows gravity measurements on a moving platform, while achieving the same long-term stability of the best atomic clocks. In the laboratory-based prototype (Figure 2), the cesium atoms used in each atom interferometer are initially collected and cooled in two separate magneto-optic traps (MOTs). Each MOT, consisting of three orthogonal pairs of counter-propagating laser beams centered on a quadrupole magnetic field, collects up to 10(exp 9) atoms. These atoms are then launched vertically as in an atom fountain by switching off the magnetic field and introducing a slight frequency shift between pairs of lasers to create a moving rest frame for the trapped atoms. While still in this moving-frame molasses, the laser frequencies are further detuned from the atomic resonance (while maintaining this relative frequency shift) to cool the atom cloud's temperature to 2 K or below, corresponding to an rms velocity of less than 2 cm/s. After launch, the cold atoms undergo further state and velocity selection to prepare for atom interferometry. The atom interferometers are then realized using laser-induced stimulated Raman transitions to perform the necessary manipulations of each atom, and the resulting interferometer phase is measured using laser-induced fluorescence for state-normalized detection. More than 20 laser beams with independent controls of frequency, phase, and intensity are required for this measurement sequence. This instrument can facilitate the study of Earth's gravitational field from surface and air vehicles, as well as from space by allowing gravity mapping from a low-cost, single spacecraft mission. In addition, the operation of atom interferometer-based instruments in space offers greater sensitivity than is possible in terrestrial instruments due to the much longer interrogation times available in the microgravity environment. A space-based quantum gravity gradiometer has the potential to achieve sensitivities similar to the GRACE mission at long spatial wavelengths, and will also have resolution similar to GOCE for measurement at shorter length scales.
Cold atom free-fall interferometers now realize state-of-the-art precision inertial sensors, opening perspectives towards embedded devices for applications and fundamental physics tests on earth and in space. The acceleration of gravity can be measured in a stimulated Raman transition interferometer where the acceleration under gravity of a cloud of freely falling atoms is sensed by three Raman pulses allowing to split, reflect and interfere atomic wave packets [1]. The gravimeter developped at SYRTE allows to measure the acceleration of gravity at state-of-the-art performances compared to the best classical absolute gravimeters, with a sensitivity of 2×10−8g over 1s and an exactitude of 5×10−9 g [2]. This exactitude is currently limited by wavefront aberrations of the lasers, sampled by residual transverse trajectories of the molasse-cooled cloud over the free fall. Unbalanced transverse trajectories will also bias the measurement due to Coriolis effect [3]. We will investigate the use of a colder atom source to reduce these systematic effects, by implementing Bose-Einstein condensation of the initial cloud, allowing both better position control and better focusing of our source.
The work presented in this thesis focusses on the development of a transportable atom-interferometry experiment and a compact fibre laser system towards precision measurements of gravitational acceleration. Interference fringes are shown with clouds of cold 87Rb atoms using co-propagating laser beams to drive stimulated Raman transitions. This is demonstrated both inside and outside of laboratory environments for which an integrated and transportable experiment is constructed. Further improvements are presented that enable the generation of clouds containing 1.7 · 108 atoms at a rate of 2.5 Hz and having a temperature of (7 ± 1) µK. This is largely due to the development of a compact laser system based on all-fibre coupled components. It is demonstrated that the laser system designed here can achieve fast frequency sweeps over 1.8 GHz within 2 ms, making it widely applicable in compact atom-interferometry experiments with rubidium atoms. This is shown by creating a Mach–Zehnder type interferometer with counter-propagating Raman beams, thus enabling measurements of gravitational acceleration. Since the laser system uses only two lasers and one fibre amplifier, a significant reduction in size is achieved, as well as a decrease in the total power consumption of the overall experiment by a third to (162 ± 7) W.
This thesis details the construction of a portable platform for cold atom experiments, as part of the European iSense project. This culminated in the demonstration of a working portable cold atom interferometer to the praise of the EC and reviewers during the final project meeting. Reductions in the size, weight and power consumption of all the crucial components for a cold atom system such as lasers, optics, magnetic field generation and electronics have been realised. These novel components have been integrated into a portable device that has been transported and operated by just two people. The completed device weighs 63 kg and uses 240 W power in a volume of just 61.1 litres. The system has generated a cloud of (6.7±0.7)x 106^66 rubidium-87 atoms at a temperature of 4.4 μK in 2 seconds using a mirror-MOT setup. The device has successfully performed atom interferometry in the form of a Ramsey interferometer both at the University of Birmingham, UK and also at an office in Brussels, BE after a journey of 570 km. A measurement of the hyperfine splitting of the ground state of rubidium-87 was performed and gives an uncertainty of δ\deltaδω\omegaω/ω\omegaω = 5.2 × 10−^-−7^77. The theoretical limit of the sensitivity to gravity of the iSense system is 3.9 μGal/√Hz.