URSI General Assembly and Scientific Symposium
Improved measurement of the40Ca+clock transition frequency towards an uncertainty of 10−15level
Kensuke Matsubara, Kazuhiro Hayasaka, Ying Li, Shigeo Nagano, Hiroyuki Ito, Reiko Kojima, Masatoshi Kajita, Yuko Hanado, Yasuhiro Koyama, Mizuhiko Hosokawa
Abstract
We report improved frequency measurement of a single40Ca+optical clock towards the 10-15uncertainty level. In contrast with our previous measurement, in which the accuracy was limited by the magnetic field fluctuation, the use of a magnetic shield significantly reduced the observed linewidth of its2S1/2-2D5/2clock transition. From frequency comparison between the40Ca+clock and an87Sr lattice clock, an Allan deviation of 7 × 10-16at an averaging time of one thousand seconds was measured. We now measure the absolute frequency of the40Ca+clock transition with respect to the SI definition of the second.