Joint Conference of the European Frequency and Time Forum and IEEE International Frequency Control Symposium (EFTF/IFCS)
Improvement in accuracy of a single40Ca+optical clock towards the 10−15level using a magnetic shield
Kensuke Matsubara, Ying Li, Shigeo Nagano, Hiroyuki Ito, Masatoshi Kajita, Reiko Kojima, Yuko Hanado, Kazuhiro Hayasaka, Mizuhiko Hosokawa
Abstract
We report a frequency measurement of a single40Ca+optical clock towards the 10−15uncertainty level. In contrast with our previous measurement, in which the accuracy was limited by line broadening due to the magnetic field fluctuation, the use of a magnetic shield significantly reduced the spectrum linewidth of the clock transition. By a frequency comparison to an87Sr lattice clock, an Allan deviation of 7 × 10−16was observed at an averaging time of 103seconds. We preliminarily measured the clock transition frequency to be 411 042 129 776 397 (3) Hz with respect to the SI definition of the second. Further measurements are executed aiming at an accuracy of low parts in the 10−15level.