Abstract
Geomagnetic measurement relies on high-sensitivity sensors, where coherent population trapping (CPT) magnetometers show promise due to advantages such as no dead zone and miniaturization capability. However, their sensitivity in geomagnetic environments is limited by the nonlinear Zeeman effect. This article establishes a magnetic field measurement model that suppresses this effect by utilizing the linear relationship between the external magnetic field and the frequency difference of the two peaks resulting from the resplitting of the zeroth-order CPT resonance. Theoretical analysis confirms that, at geomagnetic field strength, the frequency difference maintains a strict linear proportionality of 55.713 kHz/mT, whereas traditional methods fail due to higher order energy shifts. To enhance the weak CPT resonance, a closed-loop control system based on spontaneous fluorescence feedback was implemented, stabilizing the excited-state population via real-time laser power adjustment. Experimental results show that under closed-loop operation, the CPT signal amplitude increases approximately threefold with no significant linewidth broadening, theoretically improving sensitivity by 1.1 times, and the resplit double-peak structure is clearly observed in a geomagnetic environment. This approach provides a viable strategy for improving the sensitivity and accuracy of CPT magnetometers in geomagnetic field measurements.