Review of Scientific Instruments
Design, structural stability, and performance of a cesium magneto-optical trap with large optical access using additively manufactured magnetic field coil mounts
Kamalkant Kamalkant, Poonam Yadav, Arnab Ghosh, Neha Singh, Shweta, Hemant Yadav, Deepshikha Singh, Bodhaditya Santra
Abstract
Cold atoms form the backbone of many quantum technology applications, including quantum computation, sensing, and networking. These experiments require a robust cold atom source, precise control of beam parameters, and large optical access. We demonstrate a compact cesium magneto-optical trap (MOT) directly loaded from background vapor, with a footprint of 34 × 56 cm2. The quadrupole and compensation coils are mounted on additively manufactured components, eliminating the need for water cooling and minimizing eddy currents, which enables fast magnetic field switching with a measured settling time of 240 μs. Finite element analysis confirms the mechanical robustness and thermal stability of the coil mounts. The MOT achieves 5 × 106 atoms at a density of 109 cm-3, with a temperature of 76 ± 2 μK, demonstrating a reliable and compact platform suitable for quantum technology experiments.