Challenges
Datasets
Workspaces
Papers
Discussions
Leaderboard
Log inSign up
Challenges
Datasets
Workspaces
Papers
Discussions
Leaderboard
Blog
Job Board
Q3AS

© 2026 Aqora Quantum S.A.S.

TermsPrivacyLegal Notice

IEEE International Conference on Electronic Measurement & Instruments

Sinusoidal Noise Suppression in SERF Co-magnetometers with Allan Variance Validation

Z X Liu, Yifan Yan, Lele Ma, Haoying Pang, Zhuo Wang, Jixi Lu

Published Aug 22, 2025• 0 citations
Publisher

Abstract

Spin-exchange relaxation-free (SERF) co-magnetometers, characterized by high sensitivity, compactness, and low drift, show great potential in quantum precision measurements based on atomic ensembles. However, periodic fluctuation noise—originating from laser intensity instability, heating signal crosstalk, and electronic interference—induces oscillatory artifacts in Allan variance analysis, degrading measurement accuracy and long-term stability. This study presents a periodic noise suppression scheme based on jointly optimized finite impulse response (FIR) notch filters, implemented for the first time in a SERF co-magnetometer. By precisely shaping the amplitude response through least-squares optimization, the scheme achieves targeted attenuation of interference frequencies. Experimental results confirm that the proposed method effectively suppresses periodic noise, yielding smoother Allan variance curves and significantly enhancing measurement stability. The proposed approach offers a practical and generalizable solution for improving the precision of atomic sensors and provides technical guidance for noise mitigation in broader precision measurement applications.

DOI
10.1109/icemi66537.2025.11306403

Related papers

A Noise Diagnostic and Suppression Methodology for Improving Measurement Stability of SERF Atomic Sensors

Jiaxin Liu, Puchen Gong, Xusheng Lei +1 more

Achieving high measurement stability in spin exchange relaxation free (SERF) atomic sensors is a central challenge in quantum instrumentation, yet the intrinsic noise mechanisms that ultimately limit measurement performance remain insufficiently characterized. This article proposes a model-based noise diagnostic and suppression methodology that systematically links microscopic spin fluctuations to macroscopic measurement noise. The Langevin stochastic terms are introduced into the coupled Bloch equations to form a noise characterization framework that captures stochastic excitation, dissipation, and finite correlation of the coupled electronic and nuclear spin modes, enabling identification and quantification of the noise sources responsible for spectral coloring and low-frequency accumulation in the measurement output. Two previously overlooked noise amplification pathways in the measurement system are diagnosed: projection leakage caused by nonorthogonal optical readout geometry and stochastic magnetic injection introduced by electromagnetic coupling in the temperature control subsystem. For each pathway, the targeted instrumentation optimization procedures are derived and implemented. The experimental evaluation on a SERF atomic gyroscope demonstrates that joint optimization of the heating and optical subsystems reduces the amplitude spectral density at 1 Hz from$7.66 \times 10^{-6}$to$3.12 \times 10^{{-6} {\circ }} $/s$/\sqrt {\mathrm {Hz}}$, a 41.6% improvement, and suppresses the nonwhite hump in the Allan deviation over$\tau \approx 0.1$–10 s. The proposed methodology provides a reusable, physics-informed route for diagnosing, attributing, and mitigating noise in high-stability quantum measurement systems.

IEEE Transactions on Instrumentation and Measurement•Jan 1, 2026•0 citations

Suppression of thermal coupling noise in the SERF atomic co-magnetometer

Jiasen Ruan, Lihong Duan, Yang Fu +4 more

Heating the alkali metal vapor cell to obtain high atomic density is a prerequisite for realizing the high-precision measurement for spin-exchange relaxation-free (SERF) co-magnetometer. However, the thermo-magnetic and thermo-optical coupling noise caused by the temperature diffusion from the alkali vapor cell is one of the main factors restricting the co-magnetometer performance at present. Here the influence of heating on the SERF co-magnetometer was comprehensively studied, and the thermal error model of the system was established. A novel heat insulation structure for the co-magnetometer was designed, and the effectiveness was verified through monitoring of characteristic temperature points. Compared to the conventional oven support, the average maximum temperature difference within 12 h of the optical path and magnetic shielding systems (OPMSS) decreased by 52%, the bias instability of the co-magnetometer decreased by 24%, and the noise of the co-magnetometer at 1 Hz decreased by 33%. The experimental results demonstrate that this method reduces the temperature fluctuation of the OPMSS, effectively suppresses the thermal coupling noise, and thus improves the performance of the SERF co-magnetometer. Moreover, this method is also applicable to other quantum sensors based on alkali metal vapor cell.

Results in Physics•Sep 22, 2023•14 citations

Decoupling and suppression of opto-thermal drifts in SERF atomic co-magnetometers via precise null-point characterization

Xiaoping Li, Wenfeng Fan, Shimiao Fan +2 more

The long-term bias stability of spin-exchange relaxation-free atomic co-magnetometers is significantly constrained by coupled opto-thermal fluctuations. Traditionally, random systemic offsets at the magnetic compensation point lead to inconsistent opto-thermal coupling, compromising the reproducibility of sensor performance. In this work, we propose a method to simultaneously decouple and suppress these primary drifts by precisely characterizing the physical null-point of the atomic ensemble. A combined compensation scheme, joining residual rotation angle alignment and transverse magnetic field adjustment, is implemented to isolate the sensor output from fluctuations in probe laser intensity and vapor cell temperature. Experimental results demonstrate that the sensitivities to temperature drifts and probe power fluctuations are suppressed simultaneously at the identified null-point by factors of 72 and 880, respectively. This approach establishes a stable zero point, which is expected to significantly enhance sensor reproducibility and stability without sacrificing fundamental sensing performance.

Applied Physics Letters•Apr 20, 2026•5 citations

Comprehensive analysis of the impact of optical magnetic misalignment on the performance of the SERF co-magnetometer

Ze Yu Cai, Lihong Duan, Shimiao Fan +2 more

The ultra-sensitive atomic co-magnetometer based on the spin-exchange relaxation-free (SERF) regime is a valuable tool for inertial measurements and fundamental physics research. However, the optical magnetic misalignment effect is a significant source of error in SERF co-magnetometers, limiting their magnetic field suppression ability and long-term stability. This paper establishes a response model of angular velocity and magnetic field for the SERF co-magnetometer. The impact of optical magnetic misalignment on the scale factor, magnetic field amplitude-frequency response, and angular velocity errors caused by magnetic noise is systematically investigated. The results indicate that the scale factor and magnetic field suppression ability decrease as the optical magnetic misalignment angle increases, while the angular velocity errors caused by magnetic noise increase as the misalignment angle increases. Finally, experimental results demonstrate that the optical-magnetic misalignment effect significantly compromises the long-term stability of SERF co-magnetometers, underscoring its substantial influence on the overall performance. This work contributes to understanding the mechanism of optical magnetic misalignment effects on the SERF co-magnetometer and enhances its performance.

Optics Express•Apr 24, 2025•4 citations

Effects of Stochastic Magnetic Noise on the SERF Co‐Magnetometer

Feng Li, Zhuo Wang, Wenfeng Fan +4 more

Stochastic magnetic field noise has a non‐negligible negative impact on the performance of quantum precision measurement instruments such as spin‐exchange relaxation‐free (SERF) co‐magnetometer in terms of sensitivity and stability. Conventional magnetic field error analysis and suppression methods are not applicable to stochastic magnetic field noise. In this paper, a stochastic model of the SERF co‐magnetometer is proposed to analyze the mechanism of stochastic magnetic field noise. The mean and covariance propagation model of the SERF co‐magnetometer is given based on the stochasticity model and statistical principles. The analysis of simulation experiments reveals that reducing the longitudinal electron spin polarization can somewhat reduce the standard deviation of the system output caused by stochastic magnetic field noise. And the atomic ensembles are most resistant to stochastic magnetic field noise when the electron spin polarization strongly couples with the nuclear spin polarization. Finally, the experiments verified the above conclusions. In conclusion, this paper provides theoretical and experimental support for analyzing and suppressing the effect of stochastic magnetic field noise on the SERF co‐magnetometer, which is of great significance for improving the sensitivity and stability of the measurement.

Advanced Quantum Technologies•Apr 19, 2024•6 citations

Magnetic noise suppression based on mid-frequency magnetic field response in SERF atomic comagnetometers

Ze Yu Cai, Zhihong Wu, Feng Liu +4 more

The atomic comagnetometers (ACMs) based on the spin-exchange relaxation-free (SERF) regime exhibit extremely high sensitivity in angular velocity measurements, positioning them as leading candidates for next-generation precision inertial sensing. However, magnetic noise remains a significant factor that contributes significantly to long-term bias drift. Here, a comprehensive model in full-frequency band of the angular velocity measurement error induced by magnetic noise in ACMs is established. Based on the proposed model, what we believe to be a new method is proposed to reduce magnetic noise error by suppressing the magnetic field response in the mid-frequency range. After optimizing the temperature of the vapor cell to minimize the magnetic field response in the mid-frequency range, the angular velocity measurement error caused by magnetic noise was reduced by 20.4%, and the bias drift was improved by 34.8%. The theories and methods proposed in this study innovatively highlight the critical role of mid-frequency magnetic field responses in suppressing magnetic noise errors, and provide important guidance for improving bias drift in SERF ACMs.

Optics Express•Jul 31, 2025•3 citations
Spin-exchange relaxation-free (SERF) co-magnetometers, characterized by high sensitivity, compactness, and low drift, show great potential in quantum precision measurements based on atomic ensembles. However, periodic fluctuation noise—originating from laser intensity instability, heating signal crosstalk, and electronic interference—induces oscillatory artifacts in Allan variance analysis, degrading measurement accuracy and long-term stability. This study presents a periodic noise suppression scheme based on jointly optimized finite impulse response (FIR) notch filters, implemented for the first time in a SERF co-magnetometer. By precisely shaping the amplitude response through least-squares optimization, the scheme achieves targeted attenuation of interference frequencies. Experimental results confirm that the proposed method effectively suppresses periodic noise, yielding smoother Allan variance curves and significantly enhancing measurement stability. The proposed approach offers a practical and generalizable solution for improving the precision of atomic sensors and provides technical guidance for noise mitigation in broader precision measurement applications.
Achieving high measurement stability in spin exchange relaxation free (SERF) atomic sensors is a central challenge in quantum instrumentation, yet the intrinsic noise mechanisms that ultimately limit measurement performance remain insufficiently characterized. This article proposes a model-based noise diagnostic and suppression methodology that systematically links microscopic spin fluctuations to macroscopic measurement noise. The Langevin stochastic terms are introduced into the coupled Bloch equations to form a noise characterization framework that captures stochastic excitation, dissipation, and finite correlation of the coupled electronic and nuclear spin modes, enabling identification and quantification of the noise sources responsible for spectral coloring and low-frequency accumulation in the measurement output. Two previously overlooked noise amplification pathways in the measurement system are diagnosed: projection leakage caused by nonorthogonal optical readout geometry and stochastic magnetic injection introduced by electromagnetic coupling in the temperature control subsystem. For each pathway, the targeted instrumentation optimization procedures are derived and implemented. The experimental evaluation on a SERF atomic gyroscope demonstrates that joint optimization of the heating and optical subsystems reduces the amplitude spectral density at 1 Hz from7.66×10−67.66 \times 10^{-6}7.66×10−6to$3.12 \times 10^{{-6} {\circ }} /s/s/s/\sqrt {\mathrm {Hz}},a41.6, a 41.6% improvement, and suppresses the nonwhite hump in the Allan deviation over,a41.6\tau \approx 0.1$–10 s. The proposed methodology provides a reusable, physics-informed route for diagnosing, attributing, and mitigating noise in high-stability quantum measurement systems.
Heating the alkali metal vapor cell to obtain high atomic density is a prerequisite for realizing the high-precision measurement for spin-exchange relaxation-free (SERF) co-magnetometer. However, the thermo-magnetic and thermo-optical coupling noise caused by the temperature diffusion from the alkali vapor cell is one of the main factors restricting the co-magnetometer performance at present. Here the influence of heating on the SERF co-magnetometer was comprehensively studied, and the thermal error model of the system was established. A novel heat insulation structure for the co-magnetometer was designed, and the effectiveness was verified through monitoring of characteristic temperature points. Compared to the conventional oven support, the average maximum temperature difference within 12 h of the optical path and magnetic shielding systems (OPMSS) decreased by 52%, the bias instability of the co-magnetometer decreased by 24%, and the noise of the co-magnetometer at 1 Hz decreased by 33%. The experimental results demonstrate that this method reduces the temperature fluctuation of the OPMSS, effectively suppresses the thermal coupling noise, and thus improves the performance of the SERF co-magnetometer. Moreover, this method is also applicable to other quantum sensors based on alkali metal vapor cell.
The long-term bias stability of spin-exchange relaxation-free atomic co-magnetometers is significantly constrained by coupled opto-thermal fluctuations. Traditionally, random systemic offsets at the magnetic compensation point lead to inconsistent opto-thermal coupling, compromising the reproducibility of sensor performance. In this work, we propose a method to simultaneously decouple and suppress these primary drifts by precisely characterizing the physical null-point of the atomic ensemble. A combined compensation scheme, joining residual rotation angle alignment and transverse magnetic field adjustment, is implemented to isolate the sensor output from fluctuations in probe laser intensity and vapor cell temperature. Experimental results demonstrate that the sensitivities to temperature drifts and probe power fluctuations are suppressed simultaneously at the identified null-point by factors of 72 and 880, respectively. This approach establishes a stable zero point, which is expected to significantly enhance sensor reproducibility and stability without sacrificing fundamental sensing performance.
The ultra-sensitive atomic co-magnetometer based on the spin-exchange relaxation-free (SERF) regime is a valuable tool for inertial measurements and fundamental physics research. However, the optical magnetic misalignment effect is a significant source of error in SERF co-magnetometers, limiting their magnetic field suppression ability and long-term stability. This paper establishes a response model of angular velocity and magnetic field for the SERF co-magnetometer. The impact of optical magnetic misalignment on the scale factor, magnetic field amplitude-frequency response, and angular velocity errors caused by magnetic noise is systematically investigated. The results indicate that the scale factor and magnetic field suppression ability decrease as the optical magnetic misalignment angle increases, while the angular velocity errors caused by magnetic noise increase as the misalignment angle increases. Finally, experimental results demonstrate that the optical-magnetic misalignment effect significantly compromises the long-term stability of SERF co-magnetometers, underscoring its substantial influence on the overall performance. This work contributes to understanding the mechanism of optical magnetic misalignment effects on the SERF co-magnetometer and enhances its performance.
Stochastic magnetic field noise has a non‐negligible negative impact on the performance of quantum precision measurement instruments such as spin‐exchange relaxation‐free (SERF) co‐magnetometer in terms of sensitivity and stability. Conventional magnetic field error analysis and suppression methods are not applicable to stochastic magnetic field noise. In this paper, a stochastic model of the SERF co‐magnetometer is proposed to analyze the mechanism of stochastic magnetic field noise. The mean and covariance propagation model of the SERF co‐magnetometer is given based on the stochasticity model and statistical principles. The analysis of simulation experiments reveals that reducing the longitudinal electron spin polarization can somewhat reduce the standard deviation of the system output caused by stochastic magnetic field noise. And the atomic ensembles are most resistant to stochastic magnetic field noise when the electron spin polarization strongly couples with the nuclear spin polarization. Finally, the experiments verified the above conclusions. In conclusion, this paper provides theoretical and experimental support for analyzing and suppressing the effect of stochastic magnetic field noise on the SERF co‐magnetometer, which is of great significance for improving the sensitivity and stability of the measurement.
The atomic comagnetometers (ACMs) based on the spin-exchange relaxation-free (SERF) regime exhibit extremely high sensitivity in angular velocity measurements, positioning them as leading candidates for next-generation precision inertial sensing. However, magnetic noise remains a significant factor that contributes significantly to long-term bias drift. Here, a comprehensive model in full-frequency band of the angular velocity measurement error induced by magnetic noise in ACMs is established. Based on the proposed model, what we believe to be a new method is proposed to reduce magnetic noise error by suppressing the magnetic field response in the mid-frequency range. After optimizing the temperature of the vapor cell to minimize the magnetic field response in the mid-frequency range, the angular velocity measurement error caused by magnetic noise was reduced by 20.4%, and the bias drift was improved by 34.8%. The theories and methods proposed in this study innovatively highlight the critical role of mid-frequency magnetic field responses in suppressing magnetic noise errors, and provide important guidance for improving bias drift in SERF ACMs.