Stabilisation of a Terahertz Hot-Electron Bolometer mixer with microwave feedback control
Рябчун Сергей Александрович, Edward Tong, R. Blundell, Robert Kimberk, Gregory N Gol'tsman
Abstract
We report on implementation of microwave feedback control loop to stabilise the performance of an HEB mixer receiver. It is shown that the receiver sensitivity increases by a factor of 4 over a 16-minute scan, and the corresponding Allan time increases up to 10 seconds, as opposed to an open loop value of 1 second. Our experiments also demonstrate that the receiver sensitivity is limited by the intermediate frequency chain.