Sensors and Actuators A Physical

In-situ magnetic field compensation for zero-field NMOR atomic magnetometer

Changhao Zhang, Jiali Liu, Xin Hui Zhao, Junlin Chen, Jiaqi Yang, Jianli Li, Liwei Jiang

Published Mar 10, 2025• 9 citations
Publisher
DOI
10.1016/j.sna.2025.116385

Related papers

Simultaneous in-situ compensation method of residual magnetic fields for the dual-beam SERF atomic magnetometer

Kun Wang, Danyue Ma, Siran Li +2 more

Sensors and Actuators A Physical•Nov 29, 2022•23 citations

Unification In-Situ Compensation Method Of Residual Magnetic Fields For The Dual-Beam Serf Atomic Magnetometer

Kun Wang, Danyue Ma, Siran Li +2 more

•Jan 1, 2022•1 citation

Modeling and analysis of phase characteristics in NMOR atomic magnetometers

Yanchao Chai, Liwei Jiang, Mengnan Tian +6 more

Sensors and Actuators A Physical•Mar 20, 2025•7 citations

Magnetic field compensation method of dual-beam SERF atomic magnetometer under probing-axis bias

Xinda Song, Di Wu, Tengyue Long +2 more

Measurement•Jun 21, 2025•6 citations

Active magnetic compensation of spin-exchange-relaxation-free atomic magnetometer

楚中毅 CHU Zhong-yi, 孙晓光 SUN Xiao-guang, 万双爱 WAN Shuang-ai +1 more

Optics and Precision Engineering•Jan 1, 2014•2 citations

Rapid atomic-signal-based in-situ magnetic compensation for unshielded spin-exchange relaxation-free atomic magnetometer

Jianjun Li, Haidong Wang, Teng Wu +2 more

We propose a novel method, which leverages the atomic signal itself, to achieve rapid in situ magnetic field compensation in an unknown geomagnetic environment for an unshielded spin-exchange relaxation-free atomic magnetometer. This method employs a strong bias magnetic field aligned with the pump beam direction to maximize the strength of the magnetic field resonance signal. The compensation process is divided into three sequential stages, each focusing on one of the x, y, and z axes by adjusting the magnitude of the bias magnetic field. As a result, the zero magnetic field is identified within 11 seconds, with compensation resolution smaller than 3.0 nT, primarily constrained by the laser power stability. This approach not only enhances the applicability of atomic magnetometers in unshielded environment but also proves effective for in situ magnetic field compensation in shielded environment, such as those used in magnetoencephalography, magnetocardiography, and space science.

Review of Scientific Instruments•Feb 1, 2026•0 citations
We propose a novel method, which leverages the atomic signal itself, to achieve rapid in situ magnetic field compensation in an unknown geomagnetic environment for an unshielded spin-exchange relaxation-free atomic magnetometer. This method employs a strong bias magnetic field aligned with the pump beam direction to maximize the strength of the magnetic field resonance signal. The compensation process is divided into three sequential stages, each focusing on one of the x, y, and z axes by adjusting the magnitude of the bias magnetic field. As a result, the zero magnetic field is identified within 11 seconds, with compensation resolution smaller than 3.0 nT, primarily constrained by the laser power stability. This approach not only enhances the applicability of atomic magnetometers in unshielded environment but also proves effective for in situ magnetic field compensation in shielded environment, such as those used in magnetoencephalography, magnetocardiography, and space science.