Optics Express

Repumping atomic media for an enhanced sensitivity atomic magnetometer

Rujie Li, Christopher Perrella, Andre Nicholas Luiten

Published Aug 3, 2022• 19 citations
Publisher

Abstract

Atomic vapour magnetometers sense the local magnetic field strength by measuring the resulting precession rate of a well-defined quantum state. An essential prerequisite for this approach is a requirement to drive the media into this quantum state, which is frequently achieved via optical pumping. In real-world alkali-metal atoms, with their multiplicity of ground states, the optical pumping process is necessarily lossy, with a large fraction of the atoms being lost to quantum states that do not contribute to the useful magnetically sensitive signal. This consequently reduces the sensitivity of all optically-pumped atomic sensors. Here we theoretically and experimentally study the population changes of the quantum ground states of 87 Rb during optical pumping. We use this understanding to develop a repumping scheme that allows us to increase the number of atoms that are contributing to the useful magnetic sensing output. Unlike prior schemes, our approach delivers this improved sensitivity with significantly suppressed fictitious magnetic fields associated with the repumping, which would otherwise reduce the accuracy of the sensor. When operated at Earth’s field strength (∼50 µ T), the repumped sensor shows a magnetic sensitivity of 200 fT/ Hz , that is nearly three times higher than the non-repumped version.

DOI
10.1364/oe.467513

Related papers

Improving the sensitivity of optically pumped magnetometers by hyperfine repumping

V. Schultze, Theo Scholtes, R.P.J. IJsselsteijn +1 more

Most optically pumped magnetometers based on alkali atom vapor cells and pumped by a single narrow-band laser suffer from a loss of signal since atoms become trapped in the ground hyperfine states that are not coupled to the laser beam. This can be counteracted by additional optical repumping of these ground-state levels. We study hyperfine repumping using cesium vapor cells with partly overlapped ground-state splitting due to their nitrogen buffer-gas filling. We implement two methods of repumping and compare them to the conventional case of F=4 pumping: F=3 repumping with an additional repumper laser and combined pumping/repumping in the light-narrowing mode, where a single high-power laser is tuned near the F=3 transitions. All these modes are investigated for two different methods of spin-phase synchronization: the Mx and intensity modulation methods. For both methods, any kind of repumping results in a clear improvement of the magnetometer sensitivity compared to no repumping, but in the Mx mode it is more pronounced (about 50 versus 200 fT/√Hz for a 50 mm3 Cs vapor cell). The mechanisms responsible for the distinct results in the different working modes are discussed.

Journal of the Optical Society of America B•Apr 2, 2015•20 citations

Improving the sensitivity of an optically pumped rubidium atomic magnetometer by using of a repumping laser beam

Lulu Zhang, Le-Le Bai, Yu-Lin Yang +5 more

For the experimental implementation of an optically pumped atomic magnetometer, the magnetic resonance signal with a narrow linewidth and a high signal-to-noise ratio (SNR) is required for achieving a high sensitivity. Using 795-nm laser as both the pumping and the probe laser, we compare the magnetic resonance signals from different rubidium atomic vapor cells and investigate the variations of magnetic resonance signals with temperature. Optimized magnetic resonance signal is achieved with a paraffin-coated rubidium atomic vapor cell. Then the 780-nm laser at rubidium D2 line is introduced as a repumping laser, and we explore the changes of linewidth and SNR of the magnetic resonance signal under different power of the pumping laser and the repumping laser. Owing to the 780-nm repumping laser beam, the signal amplitude of rubidium-85 magnetic resonance signal is improved remarkably because more rubidium-85 atoms are spin- polarized by the 795-nm pumping laser beam. At the same time, the linewidth of rubidium-85 magnetic resonance signal is roughly not broadened anymore. We realize a closed-loop optically pumped rubidium-85 atomic magnetometer with a bandwidth of ~1.2 kHz, and the sensitivity is calibrated to be ~245.5 pT/Hz1/2 only with the 795-nm pumping laser beam. Owing to the employment of the 780-nm repumping laser beam, the sensitivity is improved to be ~26.4 pT/Hz1/2 which is improved roughly by one order of magnitude. We also calibrate the measurement accuracy and deviation of a commercial fluxgate magnetometer by using the enhanced rubidium magnetic resonance signal.

Acta Physica Sinica•Jan 1, 2021•5 citations

Quantum sensitivity limits and dynamical correlations in alkali-vapor sensors

K. Mouloudakis, V. Koutrouli, I.K. Kominis +2 more

For spin-polarized ensembles of alkali-metal atoms with hyperfine structure and spin greater than 1/2, the interplay between spin dynamics and quantum noise mechanisms remains largely unexplored, especially in the spin-exchange-relaxation-free (SERF) regime. In particular, the theoretical understanding of the dynamics for the mean values of the spin operators was established shortly after the advent of optical pumping. In contrast, a corresponding theory for spin fluctuations has not yet been developed, and is addressed in this work. Here, we develop such a first-principles theory of spin fluctuations in alkalimetal vapors. The nonlinearity bottleneck of spin-exchange collisions, which dominate the physics of such vapors, is overcome by linearizing the dynamics around the equilibrium density matrix. The result is a precise calculation of the absolute magnetic sensing limit of optically pumped atomic magnetometers, which can precisely benchmark quantum limited and quantum enhanced protocols, in contrast to the traditional approach shown to be adequate only within a factor of 2. Further, we elucidate how signal and noise in atomic magnetometers are distributed among the spin eigenmodes defined under continuous optical pumping and probing via the paramagnetic Faraday rotation. Notably, in the SERF regime, alongside the magnetic resonance narrowing, we find a reduction of spin noise at the spin-precession frequency as a consequence of strong hyperfine correlations redistributing spin noise, when both hyperfine manifolds are simultaneously probed. These insights can lead to improved quantum devices utilizing alkali-metal atoms or similar spin-based sensing platforms.

Physical Review Applied•Dec 11, 2025•3 citations

Research on the influence of alkali polarization on an all-optical atomic spin magnetometer based on hybrid pumping potassium and rubidium

Hong Zhang, Sheng Zou, Xiyuan Chen +1 more

An all-optical atomic spin magnetometer using a hybrid cell of K and Rb atoms is demonstrated in this paper. Via deriving and solving the Bloch equations, we established the models for the polarization rate of Rb and the magnetometer’s output signal, respectively. In order to achieve the maximum sensitivity, we carried out research on the influence of the pump power on the polarization rate and magnetometer’s output signal. The magnetometer’s sensitivity was found to be proportional to the polarization rate of Rb, which was confined by the pump power. In a certain context, theoretical predictions indicated that the larger the pump power was, the higher magnetometer’s sensitivity that could be realized. However, the theoretical estimate deviated from the experimental results. An analysis of the errors between the actual value and the theoretical value showed that the different molar ratio between the two species of alkali-metal atoms was the primary reason causing the deviation. By comparing the experimental measurement result with simulation result, we found the difference in the molar ratio to be approximate 4 times. After various parameters had been optimized, a sensitivity of 18 fT/Hz 1/2 was achieved by analyzing the signal-to-noise ratio.

Journal of the Korean Physical Society•Apr 1, 2015•9 citations

Broadband Ultra-Sensitive Adiabatic Magnetometer

Igor Savukov, Young Jin Kim

We introduce a new ultra-sensitive adiabatic magnetometer that has a broad bandwidth and can operate in the presence of magnetic fields and gradients. It follows conceptually typical implementations of atomic magnetometers based on alkali-metal vapor cells and lasers for optical pumping and optical Faraday effect detection, while its unique feature is a measurement of an oscillating magnetic field along the probe beam direction at frequencies lower than the resonant frequency, proportional to a static magnetic field along the pump beam direction. The bandwidth of the adiabatic magnetometer scales as the strength of the field along the pump beam. From our theoretical studies it is expected that the adiabatic magnetometer can reach 1 fT sensitivity with a bandwidth of 10 kHz, which any type of atomic magnetometers cannot achieve. Among anticipated various applications of this adiabatic magnetometer are biomagnetic sensing, nuclear magnetic resonance detection, and alkali-metal density measurements. We experimentally conducted alkali-metal density measurements, as an example of applications.

IEEE Sensors Applications Symposium (SAS)•Aug 23, 2021•1 citation

Sensitive optical atomic magnetometer based on nonlinear magneto-optical rotation

Chris Hovde, Brian Patton, Eric Corsini +2 more

A self-oscillating magnetometer based on nonlinear magneto-optical rotation using amplitude-modulated pump light and unmodulated probe light (AM-NMOR) in 87Rb has been constructed and tested towards a goal of airborne detection of magnetic anomalies. In AM-NMOR, stroboscopic optical pumping via amplitude modulation of the pump beam creates alignment of the ground electronic state of the rubidium atoms. The Larmor precession causes an ac rotation of the polarization of a separate probe beam; the polarization rotation frequency provides a measure of the magnetic field. An anti-relaxation coating on the walls of the atomic vapor cell results in a long lifetime of 56 ms for the alignment, which enables precise measurement of the precession frequency. Light is delivered to the magnetometer by polarization-maintaining optical fibers. Tests of the sensitivity include directly measuring the beat frequency between the magnetometer and a commercial instrument and measurements of Earth's field under magnetically quiet conditions, indicating a sensitivity of at least 5 pT/νHz. Rotating the sensor indicates a heading error of less than 1 nT, limited in part by residual magnetism of the sensor.

arXiv:1003.1531•Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE•Apr 23, 2010•6 citations
Atomic vapour magnetometers sense the local magnetic field strength by measuring the resulting precession rate of a well-defined quantum state. An essential prerequisite for this approach is a requirement to drive the media into this quantum state, which is frequently achieved via optical pumping. In real-world alkali-metal atoms, with their multiplicity of ground states, the optical pumping process is necessarily lossy, with a large fraction of the atoms being lost to quantum states that do not contribute to the useful magnetically sensitive signal. This consequently reduces the sensitivity of all optically-pumped atomic sensors. Here we theoretically and experimentally study the population changes of the quantum ground states of 87 Rb during optical pumping. We use this understanding to develop a repumping scheme that allows us to increase the number of atoms that are contributing to the useful magnetic sensing output. Unlike prior schemes, our approach delivers this improved sensitivity with significantly suppressed fictitious magnetic fields associated with the repumping, which would otherwise reduce the accuracy of the sensor. When operated at Earth’s field strength (∼50 µ T), the repumped sensor shows a magnetic sensitivity of 200 fT/ Hz , that is nearly three times higher than the non-repumped version.
Most optically pumped magnetometers based on alkali atom vapor cells and pumped by a single narrow-band laser suffer from a loss of signal since atoms become trapped in the ground hyperfine states that are not coupled to the laser beam. This can be counteracted by additional optical repumping of these ground-state levels. We study hyperfine repumping using cesium vapor cells with partly overlapped ground-state splitting due to their nitrogen buffer-gas filling. We implement two methods of repumping and compare them to the conventional case of F=4 pumping: F=3 repumping with an additional repumper laser and combined pumping/repumping in the light-narrowing mode, where a single high-power laser is tuned near the F=3 transitions. All these modes are investigated for two different methods of spin-phase synchronization: the Mx and intensity modulation methods. For both methods, any kind of repumping results in a clear improvement of the magnetometer sensitivity compared to no repumping, but in the Mx mode it is more pronounced (about 50 versus 200 fT/√Hz for a 50 mm3 Cs vapor cell). The mechanisms responsible for the distinct results in the different working modes are discussed.
For the experimental implementation of an optically pumped atomic magnetometer, the magnetic resonance signal with a narrow linewidth and a high signal-to-noise ratio (SNR) is required for achieving a high sensitivity. Using 795-nm laser as both the pumping and the probe laser, we compare the magnetic resonance signals from different rubidium atomic vapor cells and investigate the variations of magnetic resonance signals with temperature. Optimized magnetic resonance signal is achieved with a paraffin-coated rubidium atomic vapor cell. Then the 780-nm laser at rubidium D2 line is introduced as a repumping laser, and we explore the changes of linewidth and SNR of the magnetic resonance signal under different power of the pumping laser and the repumping laser. Owing to the 780-nm repumping laser beam, the signal amplitude of rubidium-85 magnetic resonance signal is improved remarkably because more rubidium-85 atoms are spin- polarized by the 795-nm pumping laser beam. At the same time, the linewidth of rubidium-85 magnetic resonance signal is roughly not broadened anymore. We realize a closed-loop optically pumped rubidium-85 atomic magnetometer with a bandwidth of ~1.2 kHz, and the sensitivity is calibrated to be ~245.5 pT/Hz1/2 only with the 795-nm pumping laser beam. Owing to the employment of the 780-nm repumping laser beam, the sensitivity is improved to be ~26.4 pT/Hz1/2 which is improved roughly by one order of magnitude. We also calibrate the measurement accuracy and deviation of a commercial fluxgate magnetometer by using the enhanced rubidium magnetic resonance signal.
For spin-polarized ensembles of alkali-metal atoms with hyperfine structure and spin greater than 1/2, the interplay between spin dynamics and quantum noise mechanisms remains largely unexplored, especially in the spin-exchange-relaxation-free (SERF) regime. In particular, the theoretical understanding of the dynamics for the mean values of the spin operators was established shortly after the advent of optical pumping. In contrast, a corresponding theory for spin fluctuations has not yet been developed, and is addressed in this work. Here, we develop such a first-principles theory of spin fluctuations in alkalimetal vapors. The nonlinearity bottleneck of spin-exchange collisions, which dominate the physics of such vapors, is overcome by linearizing the dynamics around the equilibrium density matrix. The result is a precise calculation of the absolute magnetic sensing limit of optically pumped atomic magnetometers, which can precisely benchmark quantum limited and quantum enhanced protocols, in contrast to the traditional approach shown to be adequate only within a factor of 2. Further, we elucidate how signal and noise in atomic magnetometers are distributed among the spin eigenmodes defined under continuous optical pumping and probing via the paramagnetic Faraday rotation. Notably, in the SERF regime, alongside the magnetic resonance narrowing, we find a reduction of spin noise at the spin-precession frequency as a consequence of strong hyperfine correlations redistributing spin noise, when both hyperfine manifolds are simultaneously probed. These insights can lead to improved quantum devices utilizing alkali-metal atoms or similar spin-based sensing platforms.
An all-optical atomic spin magnetometer using a hybrid cell of K and Rb atoms is demonstrated in this paper. Via deriving and solving the Bloch equations, we established the models for the polarization rate of Rb and the magnetometer’s output signal, respectively. In order to achieve the maximum sensitivity, we carried out research on the influence of the pump power on the polarization rate and magnetometer’s output signal. The magnetometer’s sensitivity was found to be proportional to the polarization rate of Rb, which was confined by the pump power. In a certain context, theoretical predictions indicated that the larger the pump power was, the higher magnetometer’s sensitivity that could be realized. However, the theoretical estimate deviated from the experimental results. An analysis of the errors between the actual value and the theoretical value showed that the different molar ratio between the two species of alkali-metal atoms was the primary reason causing the deviation. By comparing the experimental measurement result with simulation result, we found the difference in the molar ratio to be approximate 4 times. After various parameters had been optimized, a sensitivity of 18 fT/Hz 1/2 was achieved by analyzing the signal-to-noise ratio.
We introduce a new ultra-sensitive adiabatic magnetometer that has a broad bandwidth and can operate in the presence of magnetic fields and gradients. It follows conceptually typical implementations of atomic magnetometers based on alkali-metal vapor cells and lasers for optical pumping and optical Faraday effect detection, while its unique feature is a measurement of an oscillating magnetic field along the probe beam direction at frequencies lower than the resonant frequency, proportional to a static magnetic field along the pump beam direction. The bandwidth of the adiabatic magnetometer scales as the strength of the field along the pump beam. From our theoretical studies it is expected that the adiabatic magnetometer can reach 1 fT sensitivity with a bandwidth of 10 kHz, which any type of atomic magnetometers cannot achieve. Among anticipated various applications of this adiabatic magnetometer are biomagnetic sensing, nuclear magnetic resonance detection, and alkali-metal density measurements. We experimentally conducted alkali-metal density measurements, as an example of applications.
A self-oscillating magnetometer based on nonlinear magneto-optical rotation using amplitude-modulated pump light and unmodulated probe light (AM-NMOR) in 87Rb has been constructed and tested towards a goal of airborne detection of magnetic anomalies. In AM-NMOR, stroboscopic optical pumping via amplitude modulation of the pump beam creates alignment of the ground electronic state of the rubidium atoms. The Larmor precession causes an ac rotation of the polarization of a separate probe beam; the polarization rotation frequency provides a measure of the magnetic field. An anti-relaxation coating on the walls of the atomic vapor cell results in a long lifetime of 56 ms for the alignment, which enables precise measurement of the precession frequency. Light is delivered to the magnetometer by polarization-maintaining optical fibers. Tests of the sensitivity include directly measuring the beat frequency between the magnetometer and a commercial instrument and measurements of Earth's field under magnetically quiet conditions, indicating a sensitivity of at least 5 pT/νHz. Rotating the sensor indicates a heading error of less than 1 nT, limited in part by residual magnetism of the sensor.