IEEE Transactions on Microwave Theory and Techniques
An Atomic Sensor for Direct Detection of Weak Microwave Signals
Vladislav P. Gerginov, Fabio C. S. da Silva, Archita Hati, Craig W. Nelson
Abstract
This paper demonstrates direct detection of weak signals at microwave frequencies based on parametric frequency conversion. The atomic medium is optically pumped by a resonant light field and prepared in a coherent atomic superposition by a weak microwave field. The coherent atomic superposition causes a parametric modulation of the polarization of a probe light field at the microwave field frequency. An upper limit magnetic field component sensitivity of 1.2(1.0) pT/Hz1/2, corresponding to 3.7(3.1)-μV/cm/Hz1/2electric field component sensitivity, is achieved at ~6.835 GHz with a 33-mm3vapor cell.