Theory of a He4 parametric-resonance magnetometer based on atomic alignment

François Beato, Élie Belorizky, Étienne Labyt, Matthieu Le Prado, Agustin Palacios-Laloy

Published Nov 26, 2018• 43 citations
Publisher

Abstract

In parametric-resonance magnetometers (PRMs), optically pumped atoms subject to rf fields allow one to measure the components of a very low magnetic field. Here, instead of using a circularly polarized light for creating atomic orientation, as in previous works, a linearly polarized light is used for creating atomic alignment. A dressed-atom formalism introduced by Polonsky and Cohen-Tannoudji is extended and allows one to map this situation to a simpler Hanle effect on aligned atoms. Analytical expressions for all alignment tensor components and photodetection signals are obtained for both the Hanle magnetometer and PRMs. A two-rf field alignment PRM is shown to display several improvements from its orientation counterpart, providing a secular third axis sensitivity and a lower degradation from the one-axis sensitivity obtained by using a single rf field. These specificities are shown to result from the apparent depolarization of the pumping light acting on the atom dressed by two-rf fields. Experimental measurements showing a good agreement with the theoretical predictions are presented.

DOI
10.1103/physreva.98.053431

Related papers

Zero-field magnetometry based on the combination of atomic orientation and alignment

Gwenael Le Gal, Agustin Palacios-Laloy

Optically pumped magnetometers usually rely on optical pumping using circularly or linearly polarized light. We study here zero-field magnetometers pumped with elliptically polarized light, preparing both atomic orientation and alignment with complementary geometries. We start by extending the three-step approach for elliptically polarized pumping. This allows us to study the Hanle effect in elliptical polarization by comparing the analytical absorption signals with experiments made on $^{4}\mathrm{He}$ metastable state. We then study parametric resonance magnetometers based on elliptical polarization by using the dressed-atom formalism with one and two radio-frequency fields. The results show good agreement with the experimental measurements and interesting perspectives for magnetometry where symmetry breaking by pumping light is mitigated.

arXiv:2112.04262•Physical Review A•Apr 25, 2022•13 citations

Three-axis isotropic-sensitivity He4 magnetometer with alignment-based longitudinal parametric resonance

Bowen Wang, Xiang Peng, Wei Xiao +1 more

For measurements in near-zero magnetic fields, vector atomic magnetometers are promising alternatives to scalar versions, but generally suffer from inconsistent performance over three axes. We present a single-beam vector magnetometer with three-axis isotropic sensitivity based on $^{4}\mathrm{He}$ atomic alignment. Longitudinal parametric resonance with atoms pumped by linearly polarized light is proposed for the $z$-axis measurement, while the Hanle effect in transverse geometry is used for $x$- and $y$-axis measurements. The evolution of atomic alignment multipole components and expressions of the three-axis magnetic responses are analyzed in detail. We describe the experimental operation of the proposed magnetometer with a magnetic-field noise floor of $150\phantom{\rule{0.16em}{0ex}}\mathrm{fT}/{\mathrm{Hz}}^{1/2}$ for the three axes, which is attractive for biomagnetism requiring vector characterization and source reconstruction.

Physical Review A•Dec 12, 2022•6 citations

Second-order effects in parametric-resonance magnetometers based on atomic alignment

François Beato, Agustin Palacios-Laloy

Optically-pumped magnetometers (OPM) based on parametric resonance allow real-time tri-axial measurement of very small magnetic fields with a single optical access to the gas cell. Most of these magnetometers rely on circularly polarized pumping light. We focus here on the ones relying on linearly polarized light, yielding atomic alignment. For these magnetometers we investigate three second order effects which appear in the usual regimes of operation, so to clarify if they translate to metrological problems like systematic errors or increased noise. The first of these effects is the breakdown of the three-step approach when the optical beam has a large intensity. The second one is the breakdown of the rotating wave approximation when the frequencies of the RF fields are not much larger than the rates of other atomic processes. The third one is the tensor light-shift which appears when the light is slightly detuned from resonance. This work should help to clarify the accuracy reachable with OPM, which is an important question notably for medical imaging applications.

EPJ Quantum Technology•May 7, 2020•17 citations

Zero-field optical magnetometer based on spin alignment

Adil Meraki, L. Elson, N.J. Ho +4 more

Optically pumped magnetometers are important instruments for imaging biological magnetic signals without the need for cryogenic cooling. These magnetometers are presently available in the commercial market and utilize the principles of atomic alignment or orientation, enabling remarkable sensitivity and precision in the measurement of magnetic fields. This research focuses on utilizing a spin-aligned atomic ensemble for magnetometry at zero field. An approach is introduced which involves evaluating how the linear polarization of light rotates as it passes through the atomic vapor to null the magnetic field. Analytical expressions are derived for the resulting spin alignment and photodetection signals. Experimental results are provided, demonstrating good agreement with the theoretical predictions. The sensitivity and bandwidth of the magnetometer are characterized based on the detected polarization rotation signal. Lastly, the practical utility of the magnetometer for medical applications is demonstrated by successfully detecting a synthetic cardiac signal.

arXiv:2308.12233•Physical Review A•Dec 11, 2023•20 citations

Light-induced line narrowing of the magnetic resonance in a parametric resonance optically pumped rubidium magnetometer

Yongbiao Yang, Yang Li, Zhengyu Su +4 more

Optically pumped atomic magnetometers are fundamentally limited by spin-exchange relaxation in finite magnetic fields. In conventional parametric resonance magnetometers (PRMs), this relaxation dominates the resonance linewidth, producing broad profiles that degrade sensitivity. Here, we suppress this relaxation by incorporating light narrowing via optical pumping into spin-stretched states in a longitudinally modulated PRM, which directly enables dual-axis vector detection of transverse magnetic fields through lock-in demodulation. Under optimized conditions, the magnetometer achieves transverse sensitivities of $210\phantom{\rule{4pt}{0ex}}\mathrm{fT}/\sqrt{\mathrm{Hz}}$ and $290\phantom{\rule{4pt}{0ex}}\mathrm{fT}/\sqrt{\mathrm{Hz}}$ along two orthogonal axes, a 3-dB bandwidth of about $500\phantom{\rule{4pt}{0ex}}\mathrm{Hz}$, and excellent cross-axis discrimination---an order-of-magnitude improvement over conventional PRMs. This work demonstrates that light narrowing effectively overcomes the spin-exchange limitation in finite-field operation, providing a promising route for high-sensitivity vector magnetometry in applications such as atomic gyroscopes and active magnetic compensation.

Physical Review A•Oct 28, 2026•0 citations

Laser pumped 4He magnetometer with light shift suppression

Zaisheng Lin, He Wang, Xiang Peng +2 more

We report a laser-pumped 4He atomic magnetometer with light shift suppression through the atomic sensor itself. A linearly polarized light is used to optically align the 4He metastable atoms and we monitor the magneto-optical double resonance (MODR) signals produced by the left- and right-circularly orthogonal components. It is shown that light shift leads to the atomic alignment to orientation conversion effect, and thus, the difference between the two MODR signals. One of these two MODR signals is locked at the Larmor frequency and is used to measure the ambient magnetic field, while the differential signal is, simultaneously, fed back to suppress the light shift. The scheme could be of the advantage to the design of compact magnetometers by reducing the systematic errors due to light shift.

Review of Scientific Instruments•Nov 1, 2016•6 citations
In parametric-resonance magnetometers (PRMs), optically pumped atoms subject to rf fields allow one to measure the components of a very low magnetic field. Here, instead of using a circularly polarized light for creating atomic orientation, as in previous works, a linearly polarized light is used for creating atomic alignment. A dressed-atom formalism introduced by Polonsky and Cohen-Tannoudji is extended and allows one to map this situation to a simpler Hanle effect on aligned atoms. Analytical expressions for all alignment tensor components and photodetection signals are obtained for both the Hanle magnetometer and PRMs. A two-rf field alignment PRM is shown to display several improvements from its orientation counterpart, providing a secular third axis sensitivity and a lower degradation from the one-axis sensitivity obtained by using a single rf field. These specificities are shown to result from the apparent depolarization of the pumping light acting on the atom dressed by two-rf fields. Experimental measurements showing a good agreement with the theoretical predictions are presented.
Optically pumped magnetometers usually rely on optical pumping using circularly or linearly polarized light. We study here zero-field magnetometers pumped with elliptically polarized light, preparing both atomic orientation and alignment with complementary geometries. We start by extending the three-step approach for elliptically polarized pumping. This allows us to study the Hanle effect in elliptical polarization by comparing the analytical absorption signals with experiments made on 4He^{4}\mathrm{He}4He metastable state. We then study parametric resonance magnetometers based on elliptical polarization by using the dressed-atom formalism with one and two radio-frequency fields. The results show good agreement with the experimental measurements and interesting perspectives for magnetometry where symmetry breaking by pumping light is mitigated.
For measurements in near-zero magnetic fields, vector atomic magnetometers are promising alternatives to scalar versions, but generally suffer from inconsistent performance over three axes. We present a single-beam vector magnetometer with three-axis isotropic sensitivity based on 4He^{4}\mathrm{He}4He atomic alignment. Longitudinal parametric resonance with atoms pumped by linearly polarized light is proposed for the zzz-axis measurement, while the Hanle effect in transverse geometry is used for xxx- and yyy-axis measurements. The evolution of atomic alignment multipole components and expressions of the three-axis magnetic responses are analyzed in detail. We describe the experimental operation of the proposed magnetometer with a magnetic-field noise floor of 150fT/Hz1/2150\phantom{\rule{0.16em}{0ex}}\mathrm{fT}/{\mathrm{Hz}}^{1/2}150fT/Hz1/2 for the three axes, which is attractive for biomagnetism requiring vector characterization and source reconstruction.
Optically-pumped magnetometers (OPM) based on parametric resonance allow real-time tri-axial measurement of very small magnetic fields with a single optical access to the gas cell. Most of these magnetometers rely on circularly polarized pumping light. We focus here on the ones relying on linearly polarized light, yielding atomic alignment. For these magnetometers we investigate three second order effects which appear in the usual regimes of operation, so to clarify if they translate to metrological problems like systematic errors or increased noise. The first of these effects is the breakdown of the three-step approach when the optical beam has a large intensity. The second one is the breakdown of the rotating wave approximation when the frequencies of the RF fields are not much larger than the rates of other atomic processes. The third one is the tensor light-shift which appears when the light is slightly detuned from resonance. This work should help to clarify the accuracy reachable with OPM, which is an important question notably for medical imaging applications.
Optically pumped magnetometers are important instruments for imaging biological magnetic signals without the need for cryogenic cooling. These magnetometers are presently available in the commercial market and utilize the principles of atomic alignment or orientation, enabling remarkable sensitivity and precision in the measurement of magnetic fields. This research focuses on utilizing a spin-aligned atomic ensemble for magnetometry at zero field. An approach is introduced which involves evaluating how the linear polarization of light rotates as it passes through the atomic vapor to null the magnetic field. Analytical expressions are derived for the resulting spin alignment and photodetection signals. Experimental results are provided, demonstrating good agreement with the theoretical predictions. The sensitivity and bandwidth of the magnetometer are characterized based on the detected polarization rotation signal. Lastly, the practical utility of the magnetometer for medical applications is demonstrated by successfully detecting a synthetic cardiac signal.
Optically pumped atomic magnetometers are fundamentally limited by spin-exchange relaxation in finite magnetic fields. In conventional parametric resonance magnetometers (PRMs), this relaxation dominates the resonance linewidth, producing broad profiles that degrade sensitivity. Here, we suppress this relaxation by incorporating light narrowing via optical pumping into spin-stretched states in a longitudinally modulated PRM, which directly enables dual-axis vector detection of transverse magnetic fields through lock-in demodulation. Under optimized conditions, the magnetometer achieves transverse sensitivities of 210fT/Hz210\phantom{\rule{4pt}{0ex}}\mathrm{fT}/\sqrt{\mathrm{Hz}}210fT/Hz​ and 290fT/Hz290\phantom{\rule{4pt}{0ex}}\mathrm{fT}/\sqrt{\mathrm{Hz}}290fT/Hz​ along two orthogonal axes, a 3-dB bandwidth of about 500Hz500\phantom{\rule{4pt}{0ex}}\mathrm{Hz}500Hz, and excellent cross-axis discrimination---an order-of-magnitude improvement over conventional PRMs. This work demonstrates that light narrowing effectively overcomes the spin-exchange limitation in finite-field operation, providing a promising route for high-sensitivity vector magnetometry in applications such as atomic gyroscopes and active magnetic compensation.
We report a laser-pumped 4He atomic magnetometer with light shift suppression through the atomic sensor itself. A linearly polarized light is used to optically align the 4He metastable atoms and we monitor the magneto-optical double resonance (MODR) signals produced by the left- and right-circularly orthogonal components. It is shown that light shift leads to the atomic alignment to orientation conversion effect, and thus, the difference between the two MODR signals. One of these two MODR signals is locked at the Larmor frequency and is used to measure the ambient magnetic field, while the differential signal is, simultaneously, fed back to suppress the light shift. The scheme could be of the advantage to the design of compact magnetometers by reducing the systematic errors due to light shift.