We demonstrate an optically pumped magnetometer (OPM) operated in a free-induction-decay (FID) configuration that is capable of tracking oscillating magnetic signals in the presence of a 50 μT static field. Excellent waveform reconstruction is demonstrated for low frequency modulations with respect to the Nyquist limited bandwidth. A 100 pT oscillation was successfully reconstructed using signal averaging, and an optimum sensitivity of 3.9 pT/Hz was measured from the spectrum of the residuals relative to the sinusoidal fit. The impact of the pump-probe repetition rate and spin depolarization on the frequency response of the sensor is investigated in detail using miniaturized vapor cell technology, with the (-3 dB) bandwidths residing beyond the Nyquist limit in each case. We also discuss technical limitations associated with the magnetometer when exposed to oscillating fields of sufficiently high amplitude or frequency. This is discussed in the context of potential distortions arising in the reproduced signals, induced by frequency modulation (FM) and aliasing artefacts.
We demonstrate an optically pumped magnetometer (OPM) operated in a free-induction-decay (FID) configuration that is capable of tracking oscillating magnetic signals in the presence of a 50 μT static field. Excellent waveform reconstruction is demonstrated for low frequency modulations with respect to the Nyquist limited bandwidth. A 100 pT oscillation was successfully reconstructed using signal averaging, and an optimum sensitivity of 3.9 pT/Hz was measured from the spectrum of the residuals relative to the sinusoidal fit. The impact of the pump-probe repetition rate and spin depolarization on the frequency response of the sensor is investigated in detail using miniaturized vapor cell technology, with the (-3 dB) bandwidths residing beyond the Nyquist limit in each case. We also discuss technical limitations associated with the magnetometer when exposed to oscillating fields of sufficiently high amplitude or frequency. This is discussed in the context of potential distortions arising in the reproduced signals, induced by frequency modulation (FM) and aliasing artefacts.
Magnetic field imaging is a valuable resource for signal source localization and characterization. This work reports an optically pumped magnetometer (OPM) based on the free-induction-decay (FID) protocol, that implements microfabricated cesium (Cs) vapor cell technology to visualize the magnetic field distributions resulting from various magnetic sources placed close to the cell. The slow diffusion of Cs atoms in the presence of a nitrogen (N 2 ) buffer gas enables spatially independent measurements to be made within the same vapor cell by translating a 175 μ m diameter probe beam over the sensing area. For example, the OPM was used to record temporal and spatial information to reconstruct magnetic field distributions in one and two dimensions. The optimal magnetometer sensitivity was estimated to be 0.43 pT/Hz within a Nyquist limited bandwidth of 500 Hz. Furthermore, the sensor’s dynamic range exceeds the Earth’s field of approximately 50 μ T, which provides a framework for magnetic field imaging in unshielded environments.
Optically pumped magnetometers (OPMs) based on the free-induction-decay (FID) configuration have recently attracted interest owing to their advantages, such as being calibration-free, easy to operate, and showing reduced light shifts induced by the pump light. Currently, alkali-metal vapor cells have been employed in most of the FID magnetometers. However, the relatively narrow linewidth of the magnetic resonance signal and nonlinear Zeeman (NLZ) effects generate bandwidth limitation and heading errors of these magnetometers in the Earth's field, respectively. In this work, we propose a 4He-based FID magnetometer, which not only has the advantages of being NLZ-free and showing a high bandwidth, but also maintains an enhanced sensitivity assisted by a multipass cell. This magnetometer demonstrates a magnetic field noise floor of 0.34pT/Hz1/2 with a Nyquist-limited bandwidth of 5 kHz, which opens another route for atomic magnetometry using FID signals, and exhibits potential for applications in high-frequency magnetic field detection, such as biomagnetic measurements and magnetic communications in the geomagnetic field.
A free-induction-decay (FID) type optically-pumped rubidium atomic magnetometer driven by a radio-frequency (RF) magnetic field is presented in this paper. Influences of parameters, such as the temperature of rubidium vapor cell, the power of pump beam, and the strength of RF magnetic field and static magnetic field on the amplitude and the full width at half maximum (FWHM) of the FID signal, have been investigated in the time domain and frequency domain. At the same time, the sensitivities of the magnetometer for the single-pass and the triple-pass probe beam cases have been compared by changing the optical path of the interaction between probe beam and atomic ensemble. Compared with the sensitivity of ∼21.2 pT/Hz1/2 in the case of the single-pass probe beam, the amplitude of FID signal in the case of the triple-pass probe beam has been significantly enhanced, and the sensitivity has been improved to ∼13.4 pT/Hz1/2. The research in this paper provids a reference for the subsequent study of influence of different buffer gas pressure on the FWHM and also a foundation for further improving the sensitivity of FID rubidium atomic magnetometer by employing a polarization-squeezed light as probe beam, to achieve a sensitivity beyond the photo-shot-noise level.
Free-induction-decay (FID) magnetometers have evolved as simple magnetic sensors for sensitive detection of unknown magnetic fields. However, these magnetometers suffer from a fundamental problem known as a “dead zone,” making them insensitive to certain magnetic field directions. Here, we demonstrate a simple experimental scheme for the dead-zone-free operation of a FID atomic magnetometer. Using a single laser beam containing equal strength of linear- and circular-polarization components and amplitude modulation at a low-duty cycle, we have synchronously pumped the rubidium (87Rb) atoms with both first- and second-order frequency harmonics. Such a pumping scheme has enabled us to observe the free Larmor precession of atomic spins at a frequency of ΩL (orientation) and/or 2ΩL (alignment) in a single FID signal, depending on the direction of the external magnetic field. We observed that the amplitude of the FID signal does not go to zero for any magnetic field direction, proving the absence of dead zones in the magnetometer. The magnetometer has a sensitivity in the range of 3.2–8.4 pT/Hz in all directions. Our experimental scheme can be crucial in developing miniaturized atomic magnetometers for various practical applications, including geomagnetic applications.
Optically pumped magnetometers (OPMs) have demonstrated significant potential in weak magnetic field detection due to their high sensitivity. In this study, we developed an Mz-type optically pumped rubidium magnetometer using a paraffin-coated anti-relaxation vapor cell. The system optimization and performance characterization were conducted inside a magnetic shield. Specifically, the pump light intensity and radio-frequency (RF) magnetic field were jointly optimized by using the linewidth-amplitude ratio as the core metric. Based on the frequency-domain noise spectrum, the sensitivity in open-loop mode was measured to be approximately 30.8 pT/Hz^{1/2}. Furthermore, a closed-loop feedback locking technique was applied, reducing the measured noise floor under the tested conditions and improving the sensitivity to 22.9 pT/Hz^{1/2}, with a measured -3 dB bandwidth of 123 Hz. The dynamic characteristics were evaluated via magnetic-field step response, showing that the system could track magnetic-field changes stably under closed-loop operation. Finally, by using tri-axial modulation and frequency-domain demodulation, we overcame the scalar measurement limitation of traditional Mz magnetometers. This work realizes vector magnetic field detection and provides a technical basis for applications such as geomagnetic navigation and magnetic anomaly detection.
Robust calibration of vector optically pumped magnetometers (OPMs) is a nontrivial task, but increasingly important for applications requiring high-accuracy such as magnetic navigation, geophysics research, and space exploration. Here, we showcase a vector OPM that utilizes Rabi oscillations driven between the hyperfine manifolds of 87 Rb to measure the direction of a DC magnetic field against the polarization ellipse structure of a microwave field. By relying solely on atomic measurements—free-induction decay (FID) signals and Rabi measurements across multiple atomic transitions—this sensor can detect drift in the microwave vector reference and compensate for systematic shifts caused by off-resonant driving, nonlinear Zeeman (NLZ) effects, and buffer gas collisions. To facilitate deadzone-free operation, we also introduce a Rabi measurement that utilizes dressed-state resonances that appear during simultaneous Larmor precession and Rabi driving (SPaR). These measurements, performed within a microfabricated vapor cell platform, achieve an average vector accuracy of 0.46 mrad and vector sensitivities down to 11µrad/Hz for geomagnetic field strengths near 50 µT. This performance surpasses the challenging 1-deg (17 mrad) accuracy threshold of several contemporary OPM methods utilizing atomic vapors with an electromagnetic vector reference.