arXiv:2409.09885 · Optica
Accurate vector optically pumped magnetometer with microwave-driven Rabi frequency measurements
Christopher Kiehl, Thanmay S. Menon, Svenja A. Knappe, Tobias Thiele, Cindy A. Regal
Abstract
Robust calibration of vector optically pumped magnetometers (OPMs) is a nontrivial task, but increasingly important for applications requiring high-accuracy such as magnetic navigation, geophysics research, and space exploration. Here, we showcase a vector OPM that utilizes Rabi oscillations driven between the hyperfine manifolds of 87 Rb to measure the direction of a DC magnetic field against the polarization ellipse structure of a microwave field. By relying solely on atomic measurements—free-induction decay (FID) signals and Rabi measurements across multiple atomic transitions—this sensor can detect drift in the microwave vector reference and compensate for systematic shifts caused by off-resonant driving, nonlinear Zeeman (NLZ) effects, and buffer gas collisions. To facilitate deadzone-free operation, we also introduce a Rabi measurement that utilizes dressed-state resonances that appear during simultaneous Larmor precession and Rabi driving (SPaR). These measurements, performed within a microfabricated vapor cell platform, achieve an average vector accuracy of 0.46 mrad and vector sensitivities down to 11µrad/Hz for geomagnetic field strengths near 50 µT. This performance surpasses the challenging 1-deg (17 mrad) accuracy threshold of several contemporary OPM methods utilizing atomic vapors with an electromagnetic vector reference.