Journal of Optics

Proposal for calibration of a single-photon counting detector without the need of input photon flux calibration

Thomas Gerrits

Published Apr 4, 2016• 4 citations
Publisher

Abstract

Abstract A method for calibration of single-photon detectors without the need of input photon flux calibration is presented. The method relies on the use of waveguide-coupled single photon detectors and a series of photon-counting measurements using a single-photon source. It is shown that the method can yield relative uncertainties of less than 1% including counting statistics and fiber splice loss uncertainties with a total measurement time of about 1 h under the assumptions that the fractional losses of the waveguide-coupled detectors are known or zero and the loss along the waveguide is constant. It is also shown that the fractional losses of the waveguide-coupled detectors can be determined if they are equal. However, in this case an input photon flux calibration is required.

DOI
10.1088/2040-8978/18/5/054014

Related papers

Proposal for calibration of a single-photon counting detector without the need of input photon flux calibration

Thomas Gerrits

A method for calibration of single-photon detectors without the need of input photon flux calibration is presented. The method relies on the use of waveguide-coupled single photon detectors and a series of photon-counting measurements using a single-photon source. It is shown that the method can yield relative uncertainties of less than 1% including counting statistics and fiber splice loss uncertainties with a total measurement time of about 1 h under the assumptions that the fractional losses of the waveguide-coupled detectors are known or zero and the loss along the waveguide is constant. It is also shown that the fractional losses of the waveguide-coupled detectors can be determined if they are equal. However, in this case an input photon flux calibration is required.

Optics Express•Apr 4, 2016•0 citations

Calibration of a single-photon counting detectors without the need of input photon flux calibration (Conference Presentation)

Thomas Gerrits

Calibration of fiber-coupled single-photon detectors usually requires knowledge of the input photon flux inside the fiber and/or knowledge of the linearity of a reference power meter. Many approaches have been presented in the past to accurately measure the photon detection probability of a single photon detector [1-6]. Under certain assumptions, one can utilize waveguide-coupled single photon detectors and a series of photon-counting measurements and a single-photon source to calibrate the detection efficiency of a single photon detector without the need of a reference power meter and the knowledge of the incoming photon flux. Here, this method is presented. Furthermore, if a reference detector is used, the detection efficiency of all evanescently coupled waveguide detectors can be measured, and the measurement outcome does not depend on splicing or fiber connection losses within in the setup, i.e., the measurement is setup-independent. In addition, the method, when using a reference detector, can be utilized to measure and distinguish between the absorption of a waveguide-coupled single photon detector and its internal detection efficiency. [1] A. J. Miller et al, Opt. Express 19, 9102–9110 (2011) [2] I. Muller et al., Metrologia 51, S329 (2014). [3] A. L. Migdall, Instrumentation and Measurement, IEEE Transactions on 50, 478–481 (2001). [4] S. V. Polyakov, A. L. Migdall, Optics Express 15, 1390–1407 (2007). [5] A. Avella et al., Optics Express 19, 23249–23257 (2011). [6] T. Lunghi et al., Opt. Express 22, 18078-18092 (2014)

•Jun 9, 2017•0 citations

Absolute calibration of single-photon and multiplexed photon-number-resolving detectors

Lior Cohen, Y. Pilnyak, D. Istrati +3 more

Single-photon detectors are widely used in modern quantum optics experiments and applications. Like all detectors, it is important for these devices to be accurately calibrated. A single-photon detector is calibrated by determining its detection efficiency; the common method to measure this quantity requires comparison to another detector. Here, we suggest a method to measure the detection efficiency of a single-photon detector without requiring an external reference detector. Our method is valid for individual single-photon detectors as well as multiplexed detectors, which are known to be photon number resolving. The method exploits the even-number photon-statistics of a nonlinear source, as well as the nonlinear loss of a single-photon detector that occurs when multiple photons are incident simultaneously. We have analytically modeled multiplexed detectors and used the results to experimentally demonstrate the calibration of a single-photon detector without the need for an external reference detector.

arXiv:1711.03594•Physical Review A•Jul 5, 2018•27 citations

Calibration of free-space and fiber-coupled single-photon detectors *

Thomas Gerrits, Alan L. Migdall, Joshua C Bienfang +5 more

Abstract We measure the detection efficiency of single-photon detectors at wavelengths near 851 nm and 1533.6 nm. We investigate the spatial uniformity of one free-space-coupled single-photon avalanche diode and present a comparison between fusion-spliced and connectorized fiber-coupled single-photon detectors. We find that our expanded relative uncertainty for a single measurement of the detection efficiency is as low as 0.70% for fiber-coupled measurements at 1533.6 nm and as high as 1.78% for our free-space characterization at 851.7 nm. The detection-efficiency determination includes corrections for afterpulsing, dark count, and count-rate effects of the single-photon detector with the detection efficiency interpolated to operation at a specified detected count rate.

arXiv:1906.02258•Metrologia•Sep 18, 2019•58 citations

Measured uncertainty analysis of detection efficiency of single photon detectors with correlated photons calibration

Lin Yin, Youbo Hu, Xiaobing Zheng

Detection efficiency calibration based on correlated photons is a high precision calibration method at the single photon level. During the calibration process, measured time, photon count rate, and channel detection efficiency are important experimental parameters that affect the measured uncertainty of detection efficiency. This paper constructs simulation models of measured uncertainty with experimental parameters by the correlated photons method. Our experiment measures and predicts the uncertainty when parameters change. This research provides a quantitative basis for experimental parameter settings of calibration of single photon detectors using correlated photons technology.

Applied Optics•Jan 19, 2022•6 citations

Single-photon quantum metrology (Conference Presentation)

Thomas Gerrits

We will present our current efforts on single photon quantum metrology using high-efficiency superconducting detectors and high-efficiency single-photon sources based on spontaneous parametric downconversion. Optical power measurements based on single photon counting could establish a quantum standard for optical power calibration in the future. At NIST we are pursuing the development of single photon sources and single photon detectors for metrology, quantum and classical applications. As part of these efforts, we are pursuing the establishment of a measurement service for the calibration of single photon detectors. We present how our calibration is tied to the calibration of our transfer standard optical fiber power meters. Using the beamsplitter method, we have implemented a fiber-coupled and free-space measurement system. Also, we have developed testbeds and measurement protocols for the characterization of single photon sources and single photon detectors. We will review several methods, which allow for the characterization of the spatial and spectral degree of freedom in spontaneous parametric downconversion.

•Mar 5, 2019•0 citations
Abstract A method for calibration of single-photon detectors without the need of input photon flux calibration is presented. The method relies on the use of waveguide-coupled single photon detectors and a series of photon-counting measurements using a single-photon source. It is shown that the method can yield relative uncertainties of less than 1% including counting statistics and fiber splice loss uncertainties with a total measurement time of about 1 h under the assumptions that the fractional losses of the waveguide-coupled detectors are known or zero and the loss along the waveguide is constant. It is also shown that the fractional losses of the waveguide-coupled detectors can be determined if they are equal. However, in this case an input photon flux calibration is required.
A method for calibration of single-photon detectors without the need of input photon flux calibration is presented. The method relies on the use of waveguide-coupled single photon detectors and a series of photon-counting measurements using a single-photon source. It is shown that the method can yield relative uncertainties of less than 1% including counting statistics and fiber splice loss uncertainties with a total measurement time of about 1 h under the assumptions that the fractional losses of the waveguide-coupled detectors are known or zero and the loss along the waveguide is constant. It is also shown that the fractional losses of the waveguide-coupled detectors can be determined if they are equal. However, in this case an input photon flux calibration is required.
Calibration of fiber-coupled single-photon detectors usually requires knowledge of the input photon flux inside the fiber and/or knowledge of the linearity of a reference power meter. Many approaches have been presented in the past to accurately measure the photon detection probability of a single photon detector [1-6]. Under certain assumptions, one can utilize waveguide-coupled single photon detectors and a series of photon-counting measurements and a single-photon source to calibrate the detection efficiency of a single photon detector without the need of a reference power meter and the knowledge of the incoming photon flux. Here, this method is presented. Furthermore, if a reference detector is used, the detection efficiency of all evanescently coupled waveguide detectors can be measured, and the measurement outcome does not depend on splicing or fiber connection losses within in the setup, i.e., the measurement is setup-independent. In addition, the method, when using a reference detector, can be utilized to measure and distinguish between the absorption of a waveguide-coupled single photon detector and its internal detection efficiency. [1] A. J. Miller et al, Opt. Express 19, 9102–9110 (2011) [2] I. Muller et al., Metrologia 51, S329 (2014). [3] A. L. Migdall, Instrumentation and Measurement, IEEE Transactions on 50, 478–481 (2001). [4] S. V. Polyakov, A. L. Migdall, Optics Express 15, 1390–1407 (2007). [5] A. Avella et al., Optics Express 19, 23249–23257 (2011). [6] T. Lunghi et al., Opt. Express 22, 18078-18092 (2014)
Single-photon detectors are widely used in modern quantum optics experiments and applications. Like all detectors, it is important for these devices to be accurately calibrated. A single-photon detector is calibrated by determining its detection efficiency; the common method to measure this quantity requires comparison to another detector. Here, we suggest a method to measure the detection efficiency of a single-photon detector without requiring an external reference detector. Our method is valid for individual single-photon detectors as well as multiplexed detectors, which are known to be photon number resolving. The method exploits the even-number photon-statistics of a nonlinear source, as well as the nonlinear loss of a single-photon detector that occurs when multiple photons are incident simultaneously. We have analytically modeled multiplexed detectors and used the results to experimentally demonstrate the calibration of a single-photon detector without the need for an external reference detector.
Abstract We measure the detection efficiency of single-photon detectors at wavelengths near 851 nm and 1533.6 nm. We investigate the spatial uniformity of one free-space-coupled single-photon avalanche diode and present a comparison between fusion-spliced and connectorized fiber-coupled single-photon detectors. We find that our expanded relative uncertainty for a single measurement of the detection efficiency is as low as 0.70% for fiber-coupled measurements at 1533.6 nm and as high as 1.78% for our free-space characterization at 851.7 nm. The detection-efficiency determination includes corrections for afterpulsing, dark count, and count-rate effects of the single-photon detector with the detection efficiency interpolated to operation at a specified detected count rate.
Detection efficiency calibration based on correlated photons is a high precision calibration method at the single photon level. During the calibration process, measured time, photon count rate, and channel detection efficiency are important experimental parameters that affect the measured uncertainty of detection efficiency. This paper constructs simulation models of measured uncertainty with experimental parameters by the correlated photons method. Our experiment measures and predicts the uncertainty when parameters change. This research provides a quantitative basis for experimental parameter settings of calibration of single photon detectors using correlated photons technology.
We will present our current efforts on single photon quantum metrology using high-efficiency superconducting detectors and high-efficiency single-photon sources based on spontaneous parametric downconversion. Optical power measurements based on single photon counting could establish a quantum standard for optical power calibration in the future. At NIST we are pursuing the development of single photon sources and single photon detectors for metrology, quantum and classical applications. As part of these efforts, we are pursuing the establishment of a measurement service for the calibration of single photon detectors. We present how our calibration is tied to the calibration of our transfer standard optical fiber power meters. Using the beamsplitter method, we have implemented a fiber-coupled and free-space measurement system. Also, we have developed testbeds and measurement protocols for the characterization of single photon sources and single photon detectors. We will review several methods, which allow for the characterization of the spatial and spectral degree of freedom in spontaneous parametric downconversion.