IEEE Transactions on Applied Superconductivity

Dynamic Range of High Temperature Superconducting Josephson Detecting Systems at Millimeter-Wave Band

J. Chen, P. Wu, Kiyotaka Nakajima, Takashi Yamashita

Published Jun 1, 2005• 6 citations
Publisher

Abstract

YBCO grain boundary Josephson junctions (GBJJ's), with the normal-state resistance (R/sub N/) around 1 /spl Omega/ and I/sub C/R/sub N/ product of less than 2 mV at 45 K, where I/sub C/ is the critical current of the junction, are used to study the dynamic range of high temperature superconducting (HTS) Josephson detecting systems at millimeter-wave band. A compact pulse tube cryocooler (PTC), or a liquid helium cryostat (LHC), or a liquid nitrogen cryostat (LNC) provides the necessary low temperature environment for the operation, while Gunn oscillators together with doubler generate signals at the frequencies of 50, 98, 196 GHz whose amplitude can be adjusted by a series of calibrated attenuators. By changing the junction parameters, operating frequencies, and the cryocooler or cryostat, we have carefully looked at the dynamic range of the system. For a measuring channel bandwidth of 1 Hz, the dynamic range is around 38 dB which is determined basically by the preamplifier. This indicates that the PTC does not introduce excess noise compared with other kinds of cryostats such as LHC's or LNC's. A measuring channel bandwidth up to 1 MHz can be helpful if fast measurements are needed.

DOI
10.1109/tasc.2005.849900

Related papers

A Millimeter Wave Detecting System Based on High Temperature Superconducting Device and Pulse Tube Cryocooler

J. Chen, Pan Wu, N. Ebisawa +2 more

A millimeter (mm) wave broadband video detecting system based on high temperature superconducting (HTS) junction and compact pulse tube cryocooler (PTC) has been studied. The lowest attainable temperature of the PTC is 42K and the operating temperature (T) can be adjusted by changing the pressure difference in the compressor. By measuring the linewidth of the Josephson oscillation as well as the dynamic range of the Josephson detector, it is found that the PTC has no excess noise compared with other kinds of cryostats such as liquid helium cryostats, and is very suitable for the applications in the mm wave detecting system. Furthermore, to improve the sensitivity of the system, the coupling efficiency of the system has been studied in detail. It is found that the coupling efficiency increases with the increase of RN linearly, and is better than 1% for RN of 1.7 Ω . A sensitivity of about 318 V/W has been obtained for the system based on the PTC and a junction with RN=1.7 Ω and ICRN = 1mV.

Chinese Journal of Physics•Aug 1, 2004•0 citations

Sensitivity of high temperature superconducting Josephson detectors at millimetre-wave band

Jianfang Chen, Y. Kurigata, Huabing Wang +3 more

Sensitivity of high temperature superconducting (HTS) Josephson detectors at millimetre-wave band has been studied in detail. YBCO grain boundary Josephson junctions with the normal-state resistance ( R N ) around 1 Ω and I C R N product of less than 2 mV have been fabricated for this study, where I C is the critical current of the junction. A Gunn oscillator at the frequency of 98 GHz together with a set of calibrated attenuators with a total attenuation of 90 dB are used as a signal source. A quasi-optical system, which consists of a wideband planar log-periodic antenna encompassing the junction, parabolic mirrors and a hyperhemispherical silicon lens, has been used as an effective coupling system. Sensitivity increases with I C R N . Using a junction with I C R N of 300 µV, a sensitivity of about 9000 V W −1 has been obtained, which is higher than that of the Schottky diode detectors at the same frequency band and demonstrates that the HTS Josephson detectors are very sensitive at millimetre-wave band.

Superconductor Science and Technology•Nov 5, 2003•8 citations

A high-temperature superconducting millimeter wave detecting system based on pulse tube cryocooler

Jian Chen, Peiheng Wu, Kensuke Nakajima +1 more

A millimeter (mm) wave broadband video detecting system using high temperature superconducting (HTS) junction and compact pulse tube cryocooler (PTC) has been studied. The lowest attainable temperature of the PTC is 42K and the operating temperature (T) can be adjusted by changing the pressure difference in the compressor. By measuring the linewidth of the Josephson oscillation as well as the dynamic range of the Josephson detector, it is found that the PTC has no excess noise compared with other kinds of cryostats such as liquid helium cryostats, and is very suitable for the applications in the mm wave detecting system. Furthermore, to improve the sensitivity of the system, the coupling efficiency of the system has been studied in detail. It is found that the coupling efficiency increases with the increase of RN linearly, and is better than 1% for RN of 1.7 Ohm. A sensitivity of about 318V/W has been obtained for the system based on the PTC and a junction with RN=1.7 Ohm and ICRN =1mV.

Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE•Oct 8, 2004•0 citations

Dynamic range of frequency-selective response of high-T/sub c/ Josephson detector to millimeter-wave radiation

V.V. Shirotov, Yu. Ya. Divin, Knut W. Urban

We have studied a voltage dependence of the response /spl Delta/I(V) of high-T/sub c/ Josephson detector to millimeter-wave radiation as a function of power of incident radiation. YBa/sub 2/Cu/sub 3/O/sub 7-x/ grain-boundary junctions with the resistances R/sub n/=0.5-1.5 Ohm and the I/sub c/R/sub n/-product in the range 0.16-0.25 mV at 80 K have been fabricated for this study. Gunn oscillator with the frequency f=86 GHz and a set of calibrated attenuators with total attenuation of 80 dB have been used for the measurements. The dynamic range, i.e. the range of power of electromagnetic radiation at which the response of the detector is directly proportional to radiation power, was found to be around 40 dB above the noise-equivalent power for frequency-selective response at voltages V near hf/2e and >45 dB for the broadband response at low voltages.

IEEE Transactions on Applied Superconductivity•Mar 1, 2001•4 citations

Millimeter-wave detection by GBJJ using high-T/sub c/ superconducting YBaCuO films

S. Kita, H. Tanabe, Takeshi Kobayashi

Millimeter-wave detection in the heterodyne mixing mode has been carried out using the high-T/sub c/ superconducting GBJJ (grain boundary Josephson junction). The GBJJ was fabricated from high T/sub c/ superconducting Y-Ba-Cu-O films, prepared on YSZ and SrTiO/sub 3/ substrates using an RF sputtering technique. The junction was patterned by chemical wet etching with phosphoric acid solution. The authors carried out experiments in the heterodyne mixing mode for a signal of 35 GHz with local oscillation at 34.97 GHz. The detected IF (intermediate frequency) output power was obtained at a bias voltage lying midway between the zeroth and first Shapiro steps. A mixer conversion efficiency of -15 dB was obtained with a dynamic range of about 20 dB at 35 GHz. This result strongly implies that high-T/sub c/ superconducting GBJJ works well as a detector at millimeter-wave frequencies.>

IEEE Transactions on Magnetics•Mar 1, 1989•16 citations

Wideband frequency metrology using high temperature superconducting Josephson junctions

Jian Chen, Y. Kurigata, Huabing Wang +3 more

We have studied the responses of high temperature superconducting (HTS) Josephson junctions to monochromatic and polychromatic electromagnetic irradiations from 50 GHz to 4.25 THz. The highest operating frequency and the maximum value of normalized response are shown to be proportional to the I/sub C/R/sub N/ product of the junction, where I/sub C/ and R/sub N/ are the critical current and normal resistance of the junction, respectively. Using a junction with I/sub C/R/sub N/ of 330 /spl mu/V, direct detections of monochromatic irradiations from 76 GHz to 3.1 THz have been obtained, indicating that fast detectors can be realized to cover around 50 : 1 frequency band by operating only one junction at one temperature. Also reported are the superposition of polychromatic signals, as well as some possible applications of the system.

IEEE Transactions on Applied Superconductivity•Jun 1, 2003•21 citations
YBCO grain boundary Josephson junctions (GBJJ's), with the normal-state resistance (R/sub N/) around 1 /spl Omega/ and I/sub C/R/sub N/ product of less than 2 mV at 45 K, where I/sub C/ is the critical current of the junction, are used to study the dynamic range of high temperature superconducting (HTS) Josephson detecting systems at millimeter-wave band. A compact pulse tube cryocooler (PTC), or a liquid helium cryostat (LHC), or a liquid nitrogen cryostat (LNC) provides the necessary low temperature environment for the operation, while Gunn oscillators together with doubler generate signals at the frequencies of 50, 98, 196 GHz whose amplitude can be adjusted by a series of calibrated attenuators. By changing the junction parameters, operating frequencies, and the cryocooler or cryostat, we have carefully looked at the dynamic range of the system. For a measuring channel bandwidth of 1 Hz, the dynamic range is around 38 dB which is determined basically by the preamplifier. This indicates that the PTC does not introduce excess noise compared with other kinds of cryostats such as LHC's or LNC's. A measuring channel bandwidth up to 1 MHz can be helpful if fast measurements are needed.
A millimeter (mm) wave broadband video detecting system based on high temperature superconducting (HTS) junction and compact pulse tube cryocooler (PTC) has been studied. The lowest attainable temperature of the PTC is 42K and the operating temperature (T) can be adjusted by changing the pressure difference in the compressor. By measuring the linewidth of the Josephson oscillation as well as the dynamic range of the Josephson detector, it is found that the PTC has no excess noise compared with other kinds of cryostats such as liquid helium cryostats, and is very suitable for the applications in the mm wave detecting system. Furthermore, to improve the sensitivity of the system, the coupling efficiency of the system has been studied in detail. It is found that the coupling efficiency increases with the increase of RN linearly, and is better than 1% for RN of 1.7 Ω . A sensitivity of about 318 V/W has been obtained for the system based on the PTC and a junction with RN=1.7 Ω and ICRN = 1mV.
Sensitivity of high temperature superconducting (HTS) Josephson detectors at millimetre-wave band has been studied in detail. YBCO grain boundary Josephson junctions with the normal-state resistance ( R N ) around 1 Ω and I C R N product of less than 2 mV have been fabricated for this study, where I C is the critical current of the junction. A Gunn oscillator at the frequency of 98 GHz together with a set of calibrated attenuators with a total attenuation of 90 dB are used as a signal source. A quasi-optical system, which consists of a wideband planar log-periodic antenna encompassing the junction, parabolic mirrors and a hyperhemispherical silicon lens, has been used as an effective coupling system. Sensitivity increases with I C R N . Using a junction with I C R N of 300 µV, a sensitivity of about 9000 V W −1 has been obtained, which is higher than that of the Schottky diode detectors at the same frequency band and demonstrates that the HTS Josephson detectors are very sensitive at millimetre-wave band.
A millimeter (mm) wave broadband video detecting system using high temperature superconducting (HTS) junction and compact pulse tube cryocooler (PTC) has been studied. The lowest attainable temperature of the PTC is 42K and the operating temperature (T) can be adjusted by changing the pressure difference in the compressor. By measuring the linewidth of the Josephson oscillation as well as the dynamic range of the Josephson detector, it is found that the PTC has no excess noise compared with other kinds of cryostats such as liquid helium cryostats, and is very suitable for the applications in the mm wave detecting system. Furthermore, to improve the sensitivity of the system, the coupling efficiency of the system has been studied in detail. It is found that the coupling efficiency increases with the increase of RN linearly, and is better than 1% for RN of 1.7 Ohm. A sensitivity of about 318V/W has been obtained for the system based on the PTC and a junction with RN=1.7 Ohm and ICRN =1mV.
We have studied a voltage dependence of the response /spl Delta/I(V) of high-T/sub c/ Josephson detector to millimeter-wave radiation as a function of power of incident radiation. YBa/sub 2/Cu/sub 3/O/sub 7-x/ grain-boundary junctions with the resistances R/sub n/=0.5-1.5 Ohm and the I/sub c/R/sub n/-product in the range 0.16-0.25 mV at 80 K have been fabricated for this study. Gunn oscillator with the frequency f=86 GHz and a set of calibrated attenuators with total attenuation of 80 dB have been used for the measurements. The dynamic range, i.e. the range of power of electromagnetic radiation at which the response of the detector is directly proportional to radiation power, was found to be around 40 dB above the noise-equivalent power for frequency-selective response at voltages V near hf/2e and >45 dB for the broadband response at low voltages.
Millimeter-wave detection in the heterodyne mixing mode has been carried out using the high-T/sub c/ superconducting GBJJ (grain boundary Josephson junction). The GBJJ was fabricated from high T/sub c/ superconducting Y-Ba-Cu-O films, prepared on YSZ and SrTiO/sub 3/ substrates using an RF sputtering technique. The junction was patterned by chemical wet etching with phosphoric acid solution. The authors carried out experiments in the heterodyne mixing mode for a signal of 35 GHz with local oscillation at 34.97 GHz. The detected IF (intermediate frequency) output power was obtained at a bias voltage lying midway between the zeroth and first Shapiro steps. A mixer conversion efficiency of -15 dB was obtained with a dynamic range of about 20 dB at 35 GHz. This result strongly implies that high-T/sub c/ superconducting GBJJ works well as a detector at millimeter-wave frequencies.>
We have studied the responses of high temperature superconducting (HTS) Josephson junctions to monochromatic and polychromatic electromagnetic irradiations from 50 GHz to 4.25 THz. The highest operating frequency and the maximum value of normalized response are shown to be proportional to the I/sub C/R/sub N/ product of the junction, where I/sub C/ and R/sub N/ are the critical current and normal resistance of the junction, respectively. Using a junction with I/sub C/R/sub N/ of 330 /spl mu/V, direct detections of monochromatic irradiations from 76 GHz to 3.1 THz have been obtained, indicating that fast detectors can be realized to cover around 50 : 1 frequency band by operating only one junction at one temperature. Also reported are the superposition of polychromatic signals, as well as some possible applications of the system.