IEEE Transactions on Applied Superconductivity

Microwave Stabilization of a HEB Mixer in a Pulse-Tube Cryocooler

Alexander S. Shurakov, Edward Tong, R. Blundell, N. Kaurova, B. Voronov, Gregory N Gol'tsman

Published Jan 21, 2013• 4 citations
Publisher

Abstract

We report the results of our study of the stability of an 800 GHz hot electron bolometer (HEB) mixer cooled with a pulse-tube cryocooler. Pulse-tube cryocoolers introduce temperature fluctuations as well as mechanical vibrations at a frequency of$\sim$1 Hz, both of which can cause receiver gain fluctuations at that frequency. In our system, the motor of the cryocooler was separated from the cryostat to minimize mechanical vibrations, leaving thermal effects as the dominant source of the receiver gain fluctuations. We measured root mean square temperature variations of the 4 K stage of$\sim$7 mK. The HEB mixer was pumped by a solid state local oscillator at 810 GHz. The root mean square current fluctuations at the low noise operating point (1.50 mV, 56.5$\mu\hbox{A}$) were$\sim 0.12\ \mu\hbox{A}$, and were predominantly due to thermal fluctuations. To stabilize the bias current, microwave radiation was injected to the HEB mixer. The injected power level was set by a proportional-integral-derivative controller, which completely compensates for the bias current oscillations induced by the pulse-tube cryocooler. Significant improvement in the Allan variance of the receiver output power was obtained, and an Allan time of 5 s was measured.

DOI
10.1109/tasc.2013.2241591

Related papers

Probing the Stability of HEB Mixers With Microwave Injection

Andrey Trifonov, Edward Tong, R. Blundell +2 more

Using a microwave probe as a tool, we have performed experiments aimed at understanding the origin of the output-power fluctuations in hot-electron-bolometer (HEB) mixers. We use a probe frequency of 1.5 GHz. The microwave probe picks up impedance changes of the HEB, which are examined upon demodulation of the reflected wave outside the cryostat. This study shows that the HEB mixer operates in two different regimes under a terahertz pump. At a low pumping level, strong pulse modulation is observed, as the device switches between the superconducting state and the normal state at a rate of a few megahertz. When pumped much harder, to approximate the low-noise mixer operating point, residual modulation can still be observed, showing that the HEB mixer is intrinsically unstable even in the resistive state. Based on these observations, we introduced a low-frequency termination to the HEB mixer. By terminating the device in a 50-Ω resistor in the megahertz frequency range, we have been able to improve the output-power Allan time of our HEB receiver by a factor of four to about 10 s for a detection bandwidth of 15 MHz, with a corresponding gain fluctuation of about 0.035%.

IEEE Transactions on Applied Superconductivity•Nov 25, 2014•4 citations

Stability of a Quasi-Optical Superconducting NbTiN Hot-Electron Bolometer Mixer at 1.5 THz Frequency Band

Hiroyuki Maezawa, T. Yamakura, T. Shiino +10 more

We are developing quasi-optical superconducting hot-electron bolometer (HEB) mixer receivers for astronomical and atmospheric remote sensing applications. The microbridge of the HEB mixer was fabricated at room temperature from a 6.8-nm-thick niobium titanium nitride (NbTiN) film deposited on a 20-nm-thick aluminum nitride (AlN) buffer layer, using a helicon sputtering technique at a slow deposition rate. The mixer was cooled to 4.2 K using a closed-cycled mechanical 4 K pulse-tube cryocooler with a temperature fluctuation of$\pm$1.6 mK. The stability of a large-volume NbTiN HEB mixer was studied at 1.47 THz by changing local oscillator (LO) power with the mixer bias voltage fixed. The intermediate frequency (IF) signal measured at 1.5 GHz had a maximum peak at a certain mixer bias current. The receiver noise temperature was lowest at around the IF maximum peak. It was also found that the IF signal was most stable at around the IF maximum peak under conditions in which the instability of LO pumping level, induced by small mechanical vibrations of the cryostat, remained in the optical system.

IEEE Transactions on Applied Superconductivity•Jan 24, 2011•6 citations

Effect of microwave radiation on the stability of terahertz hot-electron bolometer mixers

Рябчун Сергей Александрович, Edward Tong, R. Blundell +2 more

We report our studies of the effect of microwave radiation, with a frequency much lower than that corresponding to the energy gap of the superconductor, on the performance of the NbN hot-electron bolometer (HEB) mixer incorporated into a THz heterodyne receiver. It is shown that exposing the HEB mixer to microwave radiation does not result in a significant rise of the receiver noise temperature and degradation of the mixer conversion gain so long as the level of microwave power is small compared to the local oscillator drive. Hence the injection of a small, but controlled amount of microwave radiation enables active compensation of local oscillator power and coupling fluctuations which can significantly degrade the stability of HEB mixer receivers.

Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE•Oct 18, 2006•2 citations

Stabilization Scheme for Hot-Electron Bolometer Receivers Using Microwave Radiation

Рябчун Сергей Александрович, Edward Tong, R. Blundell +1 more

We present the results of a stabilization scheme for terahertz receivers based on NbN hot-electron bolometer (HEB) mixers that uses microwave radiation with a frequency much lower than the gap frequency of NbN to compensate for mixer current fluctuations. A feedback control loop, which actively controls the power level of the injected microwave radiation, has successfully been implemented to stabilize the operating point of the HEB mixer. This allows us to increase the receiver Allan time to 10 s and also improve the temperature resolution of the receiver by about 30% in the total power mode of operation.

IEEE Transactions on Applied Superconductivity•Dec 18, 2008•12 citations

Microwave stabilization of HEB mixer by a microchip controller

Alexander S. Shurakov, Edward Tong, R. Blundell +1 more

The stability of a Hot Electron Bolometer (HEB) mixer can be improved by the use of microwave injection. In this article we report a refinement of this approach. We introduce a microchip controller to facilitate the implementation of the stabilization scheme, and demonstrate that the feedback loop effectively suppresses drifts in the HEB bias current, leading to an improvement in the receiver stability. The measured Allan time of the mixer's IF output power is increased to > 10 s.

IEEE MTT-S International Microwave Symposium digest•Jun 1, 2012•3 citations

Characterization of a quasi-optical NbN superconducting HEB mixer

Ling Jiang, Wei Miao, Wen Zhang +10 more

In this paper, the performance of a quasi-optical NbN superconducting hot-electron bolometer (HEB) mixer, cryogenically cooled by a close-cycled 4-K refrigerator, is thoroughly investigated at 300, 500, and 850 GHz. The lowest receiver noise temperatures measured at the respective three frequencies are 1400, 900, and 1350 K, which can go down to 659, 413, and 529 K, respectively, after correcting the loss and associated noise contribution of the quasi-optical system before the measured superconducting HEB mixer. The stability of the quasi-optical superconducting HEB mixer is also investigated here. The Allan variance time measured with a local oscillator pumping at 500 GHz and an IF bandwidth of 110 MHz is 1.5 s at the dc-bias voltage exhibiting the lowest noise temperature and increases to 2.5 s at a dc bias twice that voltage.

IEEE Transactions on Microwave Theory and Techniques•Jul 1, 2006•19 citations
We report the results of our study of the stability of an 800 GHz hot electron bolometer (HEB) mixer cooled with a pulse-tube cryocooler. Pulse-tube cryocoolers introduce temperature fluctuations as well as mechanical vibrations at a frequency of∼\sim∼1 Hz, both of which can cause receiver gain fluctuations at that frequency. In our system, the motor of the cryocooler was separated from the cryostat to minimize mechanical vibrations, leaving thermal effects as the dominant source of the receiver gain fluctuations. We measured root mean square temperature variations of the 4 K stage of∼\sim∼7 mK. The HEB mixer was pumped by a solid state local oscillator at 810 GHz. The root mean square current fluctuations at the low noise operating point (1.50 mV, 56.5μA\mu\hbox{A}μA) were∼0.12 μA\sim 0.12\ \mu\hbox{A}∼0.12 μA, and were predominantly due to thermal fluctuations. To stabilize the bias current, microwave radiation was injected to the HEB mixer. The injected power level was set by a proportional-integral-derivative controller, which completely compensates for the bias current oscillations induced by the pulse-tube cryocooler. Significant improvement in the Allan variance of the receiver output power was obtained, and an Allan time of 5 s was measured.
Using a microwave probe as a tool, we have performed experiments aimed at understanding the origin of the output-power fluctuations in hot-electron-bolometer (HEB) mixers. We use a probe frequency of 1.5 GHz. The microwave probe picks up impedance changes of the HEB, which are examined upon demodulation of the reflected wave outside the cryostat. This study shows that the HEB mixer operates in two different regimes under a terahertz pump. At a low pumping level, strong pulse modulation is observed, as the device switches between the superconducting state and the normal state at a rate of a few megahertz. When pumped much harder, to approximate the low-noise mixer operating point, residual modulation can still be observed, showing that the HEB mixer is intrinsically unstable even in the resistive state. Based on these observations, we introduced a low-frequency termination to the HEB mixer. By terminating the device in a 50-Ω resistor in the megahertz frequency range, we have been able to improve the output-power Allan time of our HEB receiver by a factor of four to about 10 s for a detection bandwidth of 15 MHz, with a corresponding gain fluctuation of about 0.035%.
We are developing quasi-optical superconducting hot-electron bolometer (HEB) mixer receivers for astronomical and atmospheric remote sensing applications. The microbridge of the HEB mixer was fabricated at room temperature from a 6.8-nm-thick niobium titanium nitride (NbTiN) film deposited on a 20-nm-thick aluminum nitride (AlN) buffer layer, using a helicon sputtering technique at a slow deposition rate. The mixer was cooled to 4.2 K using a closed-cycled mechanical 4 K pulse-tube cryocooler with a temperature fluctuation of$\pm$1.6 mK. The stability of a large-volume NbTiN HEB mixer was studied at 1.47 THz by changing local oscillator (LO) power with the mixer bias voltage fixed. The intermediate frequency (IF) signal measured at 1.5 GHz had a maximum peak at a certain mixer bias current. The receiver noise temperature was lowest at around the IF maximum peak. It was also found that the IF signal was most stable at around the IF maximum peak under conditions in which the instability of LO pumping level, induced by small mechanical vibrations of the cryostat, remained in the optical system.
We report our studies of the effect of microwave radiation, with a frequency much lower than that corresponding to the energy gap of the superconductor, on the performance of the NbN hot-electron bolometer (HEB) mixer incorporated into a THz heterodyne receiver. It is shown that exposing the HEB mixer to microwave radiation does not result in a significant rise of the receiver noise temperature and degradation of the mixer conversion gain so long as the level of microwave power is small compared to the local oscillator drive. Hence the injection of a small, but controlled amount of microwave radiation enables active compensation of local oscillator power and coupling fluctuations which can significantly degrade the stability of HEB mixer receivers.
We present the results of a stabilization scheme for terahertz receivers based on NbN hot-electron bolometer (HEB) mixers that uses microwave radiation with a frequency much lower than the gap frequency of NbN to compensate for mixer current fluctuations. A feedback control loop, which actively controls the power level of the injected microwave radiation, has successfully been implemented to stabilize the operating point of the HEB mixer. This allows us to increase the receiver Allan time to 10 s and also improve the temperature resolution of the receiver by about 30% in the total power mode of operation.
In this paper, the performance of a quasi-optical NbN superconducting hot-electron bolometer (HEB) mixer, cryogenically cooled by a close-cycled 4-K refrigerator, is thoroughly investigated at 300, 500, and 850 GHz. The lowest receiver noise temperatures measured at the respective three frequencies are 1400, 900, and 1350 K, which can go down to 659, 413, and 529 K, respectively, after correcting the loss and associated noise contribution of the quasi-optical system before the measured superconducting HEB mixer. The stability of the quasi-optical superconducting HEB mixer is also investigated here. The Allan variance time measured with a local oscillator pumping at 500 GHz and an IF bandwidth of 110 MHz is 1.5 s at the dc-bias voltage exhibiting the lowest noise temperature and increases to 2.5 s at a dc bias twice that voltage.