Europhysics Letters (EPL)
An experimental study of the spin-exchange frequency shift in a laser-cooled cesium fountain frequency standard
S. Ghezali, Philippe Laurent, S.N. Lea, Andr Clairon
Abstract
The spin-exchange frequency shift due to cold collisions between ground-state atoms appears to be the systematic effect which could limit the accuracy of a laser-cooled cesium fountain operating as a frequency standard. We have investigated this shift for atoms distributed among all the Zeeman substates of either the F = 4 or F = 3 hyperfine level of the 6 S 1/2 ground state of cesium and for the case where most of the atoms are prepared in F = 3, M F = 0 sublevel. The measured relative frequency shifts are respectively − 1.0·10 −21 , − 0.6·10 −21 and − 2.0·10 −21 (at/cm 3 ) −1 . These results agree with the measurements performed by Gibble and Chu at higher atom densities with atoms prepared in F = 3, M F = 0 sublevel. However, the precision of our frequency measurements is much better and is only limited at 2·10 −15 by the H-maser used as a reference oscillator. Finally, we show that the relative frequency shift due to cold collisions can be reduced to 10 −15 , while preserving a short-term frequency stability of about 10 −13 τ −1/2 .