Journal of Applied Physics
Double relaxation oscillation superconducting quantum interference devices with gradiometric layout
Michiel J. van Duuren, G.C.S. Brons, Derk Jan Adelerhof, J. Flokstra, Horst Rogalla
Abstract
Double relaxation oscillation superconducting quantum interference devices (DROSs) with a gradiometric signal SQUID and either a reference SQUID or a reference junction will be presented in this article. The devices are user friendly, particularly those with a reference junction. Because of the large flux-to-voltage transfer of ∂V/∂Φ=0.7–1 mV/Φ0, the devices can be operated in a flux locked loop based on direct voltage readout without loss of sensitivity. The typical white flux noise of the DROSs amounts to √SΦ=5–6μΦ0/√Hz, which corresponds to an energy resolution ε=SΦ/2Lsq≃200 h. Coupled to an external planar first-order gradiometer, a white magnetic field sensitivity of √SB<2 fT/√Hz was measured inside a magnetically shielded room.