Revue de Physique Appliquée
Heterodyne laser radiation detection at 891 GHz using josephson point contacts
T. G. Blaney
Abstract
Experiments were carried out to investigate the performance of a heterodyne detector at 891 GHz using a cw hydrogen cyanide laser as the local oscillator and Nb-Nb point contact Josephson junctions as mixing devices. Using a second off-tuned HCN laser as the signal, intermediate frequencies (IF's) up to a few MHz could be studied. Both pre-set and adjustable types of junction were used, and the latter, arranged so that radiation could be directly focussed onto them gave the best ultimate sensitivity. Dependence of the IF power on dc current bias and local oscillator power is described, and compared with simple theory. IF power varied linearly with signal power for signals less than 10-5 to 10-6 W. The mixer conversion efficiency was independent of IF. The sensitivity of the detector was a strong function of junction characteristics and ranged over several orders of magnitude in the junctions studied. Noise at the output was dominated by the room temperature preamplifier with a noise temperature of about 500 K. The smallest observed value for conversion loss of the junction was about 29 dB. This figure includes the signal input matching loss. The best sensitivity of the whole receiving system was equivalent to 1 × 10-17 W/Hz.