Journal of Applied Physics
Thermal noise-limited sensitivity of the pulse-biased SQUID magnetometer
Norio Fujimaki, Hirotaka Tamura, Takeshi Imamura, S. Hasuo
Abstract
We numerically calculated and obtained an approximate expression for the thermal noise-limited sensitivity of the pulse-biased SQUID sensor used in the integrated magnetometer. We found that the sensitivity is inversely proportional to the bias frequency and inversely proportional to the critical current of the Josephson junction to the 4/3 power. The sensitivity is also proportional to the temperature to the 4/3 power, and is nearly independent of the bias pulse duration. The sensitivity expression is similar to that of the rf SQUID, however, the coefficient is more than one order of magnitude smaller. The best sensitivity has the same order of magnitude as that of the dc SQUID. The quantum limit is attained at a bias frequency of 0.9 GHz at 4.2 K with the typical junction parameters. This sensitivity is available for the SQUID magnetometer operated in the digital feedback loop.