International Conference on Measurement, Information and Control

A high stability magnetic field applied in atomic magnetometer system

Huang Qiang, Huang Zongjun, Qiang Liu, Zhang Junhai, Sun Weimin

Published May 1, 2012• 1 citation
Publisher

Abstract

A pair of Helmholtz coils producing a magnetic field with super-high stability in atomic magnetometer experiment has been designed and realized. To drive this coils, we developed a DC constant current source, which can provide a near 1 mA DC current with short-term stability of 5.8 parts in 107in experimentally. Analysis shows that the stability of center magnetic field is about 0.06 pT.

DOI
10.1109/mic.2012.6273303

Related papers

Design of compact coils and a big stable current driver for Feshbach resonance researches

Shuai Chen

Atomic Feshbach resonance is one of the important tools in quantum gases experiments.The position of a resonance,B0,typically occurs at a few hundred gauss for alkali atoms.The width of a resonance Δ can be relatively smaller,thus requires a stable magnetic field.A set of compact Helmholtz coils were designed,which are capable of carrying a high current for the strong magnetic field while maintaining the required stability.The negative integral feedback loop was used to regulate a current stability up to 10-5.

•Jan 1, 2012•0 citations

Direct control of high magnetic fields for cold atom experiments based on NV centers

Alexander Hesse, Kerim Köster, Jakob M. Steiner +5 more

Abstract In ultracold quantum gases, the interactions between the individual atoms can be controlled by applying magnetic bias fields. As magnetic field fluctuations limit the precision here, typically a feedback loop needs to be employed to regulate the current through a pair of Helmholtz coils. No commercially available magnetic field sensor allows to measure large fields directly with high enough precision, leading to many unsatisfactory solutions being used in experiments. Here, we demonstrate a direct magnetic field stabilization in a regime previously not accessible, using NV centers as the magnetic field sensor. This allows us to measure and stabilize fields of 4.66 mT down to 12 nT RMS noise over the course of 24 h, measured on a 1 Hz bandwidth. We achieve a control of better than 1 ppm after 20 min of integration time, ensuring high long-term stability for experiments. This approach extends direct magnetic field control to strong magnetic fields, which could enable new precise quantum simulations in this regime.

arXiv:2003.08101•New Journal of Physics•Feb 1, 2021•8 citations

Magnetic field sensing based on multi-order resonances of atomic spins

Hongying Yang, Qian Wang, Binbin Zhao +4 more

Broad-dynamic-range magnetometers are demanded in practical applications and fundamental research. We experimentally demonstrate a parametrically modulated atomic magnetometer with a large dynamic range by taking advantage of the high-order resonance effects. With the increase of the strength of the modulation field, both low-order and high-order resonances are well resolved and used to measure the DC or AC magnetic fields. The experimentally demonstrated sensitivity of the magnetometer based on the zeroth-order resonance is 1.5 pT/Hz, and those based on the high-order resonances are below 3 pT/Hz, making the measurement of high magnetic fields feasible under an open-loop operation. Moreover, we also demonstrated the measurement of high-frequency large AC magnetic field with the high-order resonances, and the sensitivity for the AC magnetic field based on the first-order resonance is 7 pT/Hz. Our scheme provides a new path for the development of broad-dynamic-range and miniaturized atomic magnetometers.

Optics Express•Nov 22, 2021•6 citations

Magnetic field stabilization system for atomic physics experiments

B. Merkel, K. Thirumalai, J. E. Tarlton +4 more

Atomic physics experiments commonly use millitesla-scale magnetic fields to provide a quantization axis. As atomic transition frequencies depend on the magnitude of this field, many experiments require a stable absolute field. Most setups use electromagnets, which require a power supply stability not usually met by commercially available units. We demonstrate the stabilization of a field of 14.6 mT to 4.3 nT rms noise (0.29 ppm), compared to noise of >100 nT without any stabilization. The rms noise is measured using a field-dependent hyperfine transition in a single 43Ca+ ion held in a Paul trap at the center of the magnetic field coils. For the 43Ca+ “atomic clock” qubit transition at 14.6 mT, which depends on the field only in second order, this would yield a projected coherence time of many hours. Our system consists of a feedback loop and a feedforward circuit that control the current through the field coils and could easily be adapted to other field amplitudes, making it suitable for other applications such as neutral atom traps.

arXiv:1808.03310•Review of Scientific Instruments•Apr 1, 2019•59 citations

Investigation and improvement of the spin self-sustaining magnetometer

Q. Zhao, B. L. Fan, Shiguang Wang +1 more

The spin self-sustaining atomic magnetometer has the advantage of 1/τ measurement and great development potential in many applications. In this paper, we investigated the main elements that affect the stability and accuracy of the self-sustaining magnetometer and proposed the methods to improve its performance based on the measurement results. The correlation coefficient between fluctuations of the magnetic field generated by coils and the spin Larmor precession frequency is 0.97, which mainly dominates the stability in a short term. The accuracy of the magnetometer is affected by the power and frequency of the pump light. The Larmor precession frequency coefficient related to the pump light power is 26 mHz/mW, and the effect on the Larmor precession frequency is minimized when the pump light frequency is red detuned by 200 MHz from the 85Rb transition D1 line F = 3 to F′ = 3. The 1/τ measurement time after these corrections can be extended to 10 s, and the sensitivity achieved is 149 fT/Hz, which is close to the quantum projection noise limit of the system.

AIP Advances•Sep 1, 2020•0 citations

Comprehensive analysis of the influence of magnetic field gradients on single-beam SERF atomic magnetometer

Jiajie Li, Ying Liu, Renjie Li +3 more

Magnetic field gradients interfere with the coherence of the atomic ensemble and degrade the performance of the spin-exchange relaxation-free (SERF) atomic magnetometer. In this paper, the influence of magnetic field gradients is incorporated into the response model of single-beam atomic magnetometer, and the theoretical models of transverse relaxation rate, spin polarization and scale factor are established. Based on this model,we find that magnetic field gradients in different directions can cause varying degrees of degradation in the performance parameters of the magnetometer. The magnetic linewidths under different light power and magnetic field gradients are measured using the designed dedicated magnetic field gradients coils, and the spin polarization and magnetic field gradients relaxation rate are obtained through fitting. An unevenness and deviation of the curve are also observed in the experiments, indicating the presence of the magnetic field gradient of approximately 2 nT/cm in the magnetic shields. Furthermore, in a magnetic field gradient environment of 20 nT/cm, the mean of the long-term sensitivity decreased by 3.5 times and the standard deviation increased by 11 times. It shows that the magnetic field gradients will not only affect the signal-to-noise ratio, but also make the magnetometer more vulnerable to external interference when working. The sensitivity and stability of magnetometer will be greatly reduced. The research in this article provides a theoretical and experimental basis for eliminating the influence of magnetic field gradients and improving the accuracy of magnetic field measurements.

Results in Physics•Nov 30, 2023•11 citations
A pair of Helmholtz coils producing a magnetic field with super-high stability in atomic magnetometer experiment has been designed and realized. To drive this coils, we developed a DC constant current source, which can provide a near 1 mA DC current with short-term stability of 5.8 parts in 107in experimentally. Analysis shows that the stability of center magnetic field is about 0.06 pT.
Atomic Feshbach resonance is one of the important tools in quantum gases experiments.The position of a resonance,B0,typically occurs at a few hundred gauss for alkali atoms.The width of a resonance Δ can be relatively smaller,thus requires a stable magnetic field.A set of compact Helmholtz coils were designed,which are capable of carrying a high current for the strong magnetic field while maintaining the required stability.The negative integral feedback loop was used to regulate a current stability up to 10-5.
Abstract In ultracold quantum gases, the interactions between the individual atoms can be controlled by applying magnetic bias fields. As magnetic field fluctuations limit the precision here, typically a feedback loop needs to be employed to regulate the current through a pair of Helmholtz coils. No commercially available magnetic field sensor allows to measure large fields directly with high enough precision, leading to many unsatisfactory solutions being used in experiments. Here, we demonstrate a direct magnetic field stabilization in a regime previously not accessible, using NV centers as the magnetic field sensor. This allows us to measure and stabilize fields of 4.66 mT down to 12 nT RMS noise over the course of 24 h, measured on a 1 Hz bandwidth. We achieve a control of better than 1 ppm after 20 min of integration time, ensuring high long-term stability for experiments. This approach extends direct magnetic field control to strong magnetic fields, which could enable new precise quantum simulations in this regime.
Broad-dynamic-range magnetometers are demanded in practical applications and fundamental research. We experimentally demonstrate a parametrically modulated atomic magnetometer with a large dynamic range by taking advantage of the high-order resonance effects. With the increase of the strength of the modulation field, both low-order and high-order resonances are well resolved and used to measure the DC or AC magnetic fields. The experimentally demonstrated sensitivity of the magnetometer based on the zeroth-order resonance is 1.5 pT/Hz, and those based on the high-order resonances are below 3 pT/Hz, making the measurement of high magnetic fields feasible under an open-loop operation. Moreover, we also demonstrated the measurement of high-frequency large AC magnetic field with the high-order resonances, and the sensitivity for the AC magnetic field based on the first-order resonance is 7 pT/Hz. Our scheme provides a new path for the development of broad-dynamic-range and miniaturized atomic magnetometers.
Atomic physics experiments commonly use millitesla-scale magnetic fields to provide a quantization axis. As atomic transition frequencies depend on the magnitude of this field, many experiments require a stable absolute field. Most setups use electromagnets, which require a power supply stability not usually met by commercially available units. We demonstrate the stabilization of a field of 14.6 mT to 4.3 nT rms noise (0.29 ppm), compared to noise of >100 nT without any stabilization. The rms noise is measured using a field-dependent hyperfine transition in a single 43Ca+ ion held in a Paul trap at the center of the magnetic field coils. For the 43Ca+ “atomic clock” qubit transition at 14.6 mT, which depends on the field only in second order, this would yield a projected coherence time of many hours. Our system consists of a feedback loop and a feedforward circuit that control the current through the field coils and could easily be adapted to other field amplitudes, making it suitable for other applications such as neutral atom traps.
The spin self-sustaining atomic magnetometer has the advantage of 1/τ measurement and great development potential in many applications. In this paper, we investigated the main elements that affect the stability and accuracy of the self-sustaining magnetometer and proposed the methods to improve its performance based on the measurement results. The correlation coefficient between fluctuations of the magnetic field generated by coils and the spin Larmor precession frequency is 0.97, which mainly dominates the stability in a short term. The accuracy of the magnetometer is affected by the power and frequency of the pump light. The Larmor precession frequency coefficient related to the pump light power is 26 mHz/mW, and the effect on the Larmor precession frequency is minimized when the pump light frequency is red detuned by 200 MHz from the 85Rb transition D1 line F = 3 to F′ = 3. The 1/τ measurement time after these corrections can be extended to 10 s, and the sensitivity achieved is 149 fT/Hz, which is close to the quantum projection noise limit of the system.
Magnetic field gradients interfere with the coherence of the atomic ensemble and degrade the performance of the spin-exchange relaxation-free (SERF) atomic magnetometer. In this paper, the influence of magnetic field gradients is incorporated into the response model of single-beam atomic magnetometer, and the theoretical models of transverse relaxation rate, spin polarization and scale factor are established. Based on this model,we find that magnetic field gradients in different directions can cause varying degrees of degradation in the performance parameters of the magnetometer. The magnetic linewidths under different light power and magnetic field gradients are measured using the designed dedicated magnetic field gradients coils, and the spin polarization and magnetic field gradients relaxation rate are obtained through fitting. An unevenness and deviation of the curve are also observed in the experiments, indicating the presence of the magnetic field gradient of approximately 2 nT/cm in the magnetic shields. Furthermore, in a magnetic field gradient environment of 20 nT/cm, the mean of the long-term sensitivity decreased by 3.5 times and the standard deviation increased by 11 times. It shows that the magnetic field gradients will not only affect the signal-to-noise ratio, but also make the magnetometer more vulnerable to external interference when working. The sensitivity and stability of magnetometer will be greatly reduced. The research in this article provides a theoretical and experimental basis for eliminating the influence of magnetic field gradients and improving the accuracy of magnetic field measurements.