arXiv:2302.14305

Ultra-high vacuum pressure measurement using cold atoms

S. Supakar, Vivek Singh, Tiwari, V. B., S. R. Mishra

Published Feb 28, 2023• 0 citations
arXivHTMLPDF

Abstract

In this work, we have measured the background pressure in an ultra-high vacuum (UHV) chamber by measuring the collisional loss rates in a Rb atom magneto-optical trap (MOT) on an atom chip. The loss rate due to non-Rb gases in the background has been estimated by measuring the MOT loss rate in low Rb pressure regime. These results can be useful for development of cold-atoms based UHV pressure standards.

DOI
10.48550/arxiv.2302.14305

Related papers

Ultrahigh vacuum pressure measurement using magneto-optical trap on atom chip

S. Supakar, Vivek Singh, Vibhuti Bhushan Tiwari +1 more

In this work, the UHV background pressure as low as 1.1×10−10 Torr has been measured using the loss rate characteristics of a vapor-loaded magneto-optical trap (MOT) formed on an atom chip in a UHV chamber. The loss rate due to non-rubidium (Rb) gases in the background in the chamber has been estimated by operating the MOT in low cooling beam intensities and low Rb pressure regimes simultaneously. Using this approach, we minimized the contributions of the intra-trap collisions as well as background MOT species collisions to the measured trap loss rate in our setup. These results can be useful for development of cold-atom based UHV pressure standards.

Journal of Applied Physics•Jul 10, 2023•6 citations

Vacuum pressure measurement based on 6Li cold atoms in a magneto-optical trap

Suzhao Zhang, Wenjun Sun, Meng Dong +5 more

Ultra-high vacuum measurement and extremely high vacuum (UHV/XHV) measurement play an important role in high-tech fields such as deep space exploration, particle accelerators, and nanoscience; with the continuous extension of the lower limit of measurement, especially when it reaches the order of 10–10 Pa, higher requirements are placed on the accuracy of the measurement. At present, in the field of UHV/XHV measurement, ionization gauges based on the principle of neutral gas ionization are commonly applied to the vacuum measurement. However, traditional ionization vacuum gauges during use can create electronic excitation desorption effects, soft X-rays, and the effect of hot cathode outgassing, thereby affecting the accuracy of measurement and limiting the lower limit of measurement. Compared with the traditional measurement technology, this method uses the relationship between the loss rate and pressure caused by the collision of cold atoms trapped in the trap depth with the background gas to calculate the gas density and inversely calculate the vacuum pressure. Based on the intrinsic quantum mechanical properties of cold atom collisions, this method is expected to be developed into a new vacuum traceability standard. In this paper, based on the small-angle approximation and impulse approximation under the quantum scattering theory, the loss rate coefficient of the collision of 6Li cold atoms with background gas molecules is calculated. According to the ideal gas equation, the pressure inversion formula is obtained. The collision loss rate is extracted by accurately fitting the loss curve of the cold atom. In order to improve the accuracy of vacuum inversion and reduce the influence of quantum diffractive collision on loss rate measurement, the trap depth under the conditions of a certain cooling laser intensity, detuning, and magnetic field gradient is determined by the photoassociation method. Finally, in a range of 1 × 10–8–5 × 10–6 Pa, the inverted pressure value is compared with the measured value of the ionization meter, proving that this method has good accuracy and reliability in the inversion of vacuum pressure. At present, the main factor restricting the improvement of accuracy is the influence of the collision between the excited atoms in the magneto-optical trap and the background gas on the loss rate measurement. In the future, with the proportion of excited atoms and the excited state C6 coefficient to be precisely determined, the uncertainty of vacuum pressure measurement can be further reduced.

Acta Physica Sinica•Jan 1, 2022•4 citations

On the continuous loading of a U-magneto-optical trap on an atom-chip in an ultra high vacuum

Vivek Singh, Vibhuti Bhushan Tiwari, Satya Ram Mishra

Abstract Here, we report our studies on the continuous loading of a U-magneto-optical trap (U-MOT) on an atom-chip with a rubidium-(Rb)-vapor-generated background using a dispenser source in an ultra high vacuum environment. Using the U-MOT loading curves, the partial pressure due to the Rb vapor and the pressure due to the background gas have been estimated near the MOT cloud position. The estimated pressure due to the Rb vapor increased from ∼ Torr to Torr as the Rb-dispenser current was increased from 2.8 to 3.4 A. The increase in dispenser current also resulted in a decrease in loading time as well as the lifetime of the MOT cloud. This study will be useful for magnetic trapping experiments where accurate information of the pressure in a chamber is important for the lifetime of the magnetic trap.

arXiv:1912.07934•Laser Physics Letters•Jan 28, 2020•11 citations

Measurement of vacuum pressure with a magneto-optical trap: A pressure-rise method

R. Moore, Lucie A. Lee, Elizabeth A. Findlay +3 more

The lifetime of an atom trap is often limited by the presence of residual background gases in the vacuum chamber. This leads to the lifetime being inversely proportional to the pressure. Here, we use this dependence to estimate the pressure and to obtain pressure rate-of-rise curves, which are commonly used in vacuum science to evaluate the performance of a system. We observe different rates of pressure increase in response to different levels of outgassing in our system. Therefore, we suggest that this is a sensitive method which will find useful applications in cold atom systems, in particular, where the inclusion of a standard vacuum gauge is impractical.

arXiv:1401.7949•Review of Scientific Instruments•Sep 1, 2015•39 citations

Progress towards a primary pressure standard with cold atoms

Kais Jooya

This thesis describes a method of using an ultra-cold ensemble of atoms confined in a trap as an atomic primary pressure standard. The development of the standard and its current status are described in detail. This standard uses a 3D MOT to trap ⁸⁷Rb and then transfer them to a quadrupole magnetic trap where the atoms undergoes collisions with a background gas. By measuring the number of atoms left in the magnetic trap as a function of time one extract a loss rate and from this rate determine the background gas density. This loss rate is a product of the density of the background, multiplied by the loss cross section averaged over the velocity distribution of the background gas. By computing the average loss cross section in the magnetic trap and measuring the loss rate, the density of the background gas can be determined. This gives a calibration free measurement of density of a background gas in the UHV range (10-⁶ ‑ 10-⁹) Torr or (10-⁴ - 10-⁷) Pa which allows for it to be used as a standard. In conjunction with this, preparation of the atoms prior to the loss rate measurement is investigated to ensure accuracy and reproducibility of the standard. Finally a comparison between UBC's atomic standard and NIST's (National Institute for Standards and Technology) orifice flow standard is conducted via an ionization gauge which employed as a transfer standard.All measurement are carried out using Argon gas as the background gas of study.

•Jan 1, 2017•0 citations

Pulsed loading of a magneto-optical trap on atom chip for fast pressure recovery in ultrahigh vacuum environment

Singh, Vivek, Vibhuti Bhushan Tiwari, A. Chaudhary +2 more

This study presents investigations on pulsed loading of a magneto-optical trap (MOT) on an atom chip in an UHV environment. Using three parallel resistively heated Rb-metal dispensers activated by pulsed current supply, approximately 3.0 $\times$ $10^{7}$ cold $^{87}Rb$ atoms were loaded into the MOT. A current pulse of $\sim$ 24 A with duration of $\sim$ 10 s raised the pressure in the chamber from 2.0 $\times$ $10^{-10}$ Torr to 3.3 $\times$ $10^{-10}$ Torr. Remarkably, the pressure recovery time after switching off the dispensers current was found to be $\sim$ 600 ms, making a significant advancement in achieving fast recovery of UHV environment surrounding the MOT region. This study is very useful for laser cooling and magnetic trapping / evaporative cooling of atoms on atom chip in the same UHV chamber.

arXiv:2403.11458•Mar 18, 2024•0 citations
In this work, we have measured the background pressure in an ultra-high vacuum (UHV) chamber by measuring the collisional loss rates in a Rb atom magneto-optical trap (MOT) on an atom chip. The loss rate due to non-Rb gases in the background has been estimated by measuring the MOT loss rate in low Rb pressure regime. These results can be useful for development of cold-atoms based UHV pressure standards.
In this work, the UHV background pressure as low as 1.1×10−10 Torr has been measured using the loss rate characteristics of a vapor-loaded magneto-optical trap (MOT) formed on an atom chip in a UHV chamber. The loss rate due to non-rubidium (Rb) gases in the background in the chamber has been estimated by operating the MOT in low cooling beam intensities and low Rb pressure regimes simultaneously. Using this approach, we minimized the contributions of the intra-trap collisions as well as background MOT species collisions to the measured trap loss rate in our setup. These results can be useful for development of cold-atom based UHV pressure standards.
Ultra-high vacuum measurement and extremely high vacuum (UHV/XHV) measurement play an important role in high-tech fields such as deep space exploration, particle accelerators, and nanoscience; with the continuous extension of the lower limit of measurement, especially when it reaches the order of 10–10 Pa, higher requirements are placed on the accuracy of the measurement. At present, in the field of UHV/XHV measurement, ionization gauges based on the principle of neutral gas ionization are commonly applied to the vacuum measurement. However, traditional ionization vacuum gauges during use can create electronic excitation desorption effects, soft X-rays, and the effect of hot cathode outgassing, thereby affecting the accuracy of measurement and limiting the lower limit of measurement. Compared with the traditional measurement technology, this method uses the relationship between the loss rate and pressure caused by the collision of cold atoms trapped in the trap depth with the background gas to calculate the gas density and inversely calculate the vacuum pressure. Based on the intrinsic quantum mechanical properties of cold atom collisions, this method is expected to be developed into a new vacuum traceability standard. In this paper, based on the small-angle approximation and impulse approximation under the quantum scattering theory, the loss rate coefficient of the collision of 6Li cold atoms with background gas molecules is calculated. According to the ideal gas equation, the pressure inversion formula is obtained. The collision loss rate is extracted by accurately fitting the loss curve of the cold atom. In order to improve the accuracy of vacuum inversion and reduce the influence of quantum diffractive collision on loss rate measurement, the trap depth under the conditions of a certain cooling laser intensity, detuning, and magnetic field gradient is determined by the photoassociation method. Finally, in a range of 1 × 10–8–5 × 10–6 Pa, the inverted pressure value is compared with the measured value of the ionization meter, proving that this method has good accuracy and reliability in the inversion of vacuum pressure. At present, the main factor restricting the improvement of accuracy is the influence of the collision between the excited atoms in the magneto-optical trap and the background gas on the loss rate measurement. In the future, with the proportion of excited atoms and the excited state C6 coefficient to be precisely determined, the uncertainty of vacuum pressure measurement can be further reduced.
Abstract Here, we report our studies on the continuous loading of a U-magneto-optical trap (U-MOT) on an atom-chip with a rubidium-(Rb)-vapor-generated background using a dispenser source in an ultra high vacuum environment. Using the U-MOT loading curves, the partial pressure due to the Rb vapor and the pressure due to the background gas have been estimated near the MOT cloud position. The estimated pressure due to the Rb vapor increased from ∼ Torr to Torr as the Rb-dispenser current was increased from 2.8 to 3.4 A. The increase in dispenser current also resulted in a decrease in loading time as well as the lifetime of the MOT cloud. This study will be useful for magnetic trapping experiments where accurate information of the pressure in a chamber is important for the lifetime of the magnetic trap.
The lifetime of an atom trap is often limited by the presence of residual background gases in the vacuum chamber. This leads to the lifetime being inversely proportional to the pressure. Here, we use this dependence to estimate the pressure and to obtain pressure rate-of-rise curves, which are commonly used in vacuum science to evaluate the performance of a system. We observe different rates of pressure increase in response to different levels of outgassing in our system. Therefore, we suggest that this is a sensitive method which will find useful applications in cold atom systems, in particular, where the inclusion of a standard vacuum gauge is impractical.
This thesis describes a method of using an ultra-cold ensemble of atoms confined in a trap as an atomic primary pressure standard. The development of the standard and its current status are described in detail. This standard uses a 3D MOT to trap ⁸⁷Rb and then transfer them to a quadrupole magnetic trap where the atoms undergoes collisions with a background gas. By measuring the number of atoms left in the magnetic trap as a function of time one extract a loss rate and from this rate determine the background gas density. This loss rate is a product of the density of the background, multiplied by the loss cross section averaged over the velocity distribution of the background gas. By computing the average loss cross section in the magnetic trap and measuring the loss rate, the density of the background gas can be determined. This gives a calibration free measurement of density of a background gas in the UHV range (10-⁶ ‑ 10-⁹) Torr or (10-⁴ - 10-⁷) Pa which allows for it to be used as a standard. In conjunction with this, preparation of the atoms prior to the loss rate measurement is investigated to ensure accuracy and reproducibility of the standard. Finally a comparison between UBC's atomic standard and NIST's (National Institute for Standards and Technology) orifice flow standard is conducted via an ionization gauge which employed as a transfer standard.All measurement are carried out using Argon gas as the background gas of study.
This study presents investigations on pulsed loading of a magneto-optical trap (MOT) on an atom chip in an UHV environment. Using three parallel resistively heated Rb-metal dispensers activated by pulsed current supply, approximately 3.0 ×\times× 10710^{7}107 cold 87Rb^{87}Rb87Rb atoms were loaded into the MOT. A current pulse of ∼\sim∼ 24 A with duration of ∼\sim∼ 10 s raised the pressure in the chamber from 2.0 ×\times× 10−1010^{-10}10−10 Torr to 3.3 ×\times× 10−1010^{-10}10−10 Torr. Remarkably, the pressure recovery time after switching off the dispensers current was found to be ∼\sim∼ 600 ms, making a significant advancement in achieving fast recovery of UHV environment surrounding the MOT region. This study is very useful for laser cooling and magnetic trapping / evaporative cooling of atoms on atom chip in the same UHV chamber.