arXiv:2203.15527 · Applied Physics Letters

Scanning nitrogen-vacancy magnetometry down to 350 mK

Patrick Scheidegger, S. Diesch, Marius L. Palm, Christian L. Degen

Published May 30, 2022• 36 citations
arXivHTMLPDF

Abstract

We report on the implementation of a scanning nitrogen-vacancy (NV) magnetometer in a dry dilution refrigerator. Using pulsed optically detected magnetic resonance combined with efficient microwave delivery through a co-planar waveguide, we reach a base temperature of 350 mK, limited by experimental heat load and thermalization of the probe. We demonstrate scanning NV magnetometry by imaging superconducting vortices in a 50-nm-thin aluminum microstructure. The sensitivity of our measurements is approximately 3 μT per square root Hz. Our work demonstrates the feasibility for performing noninvasive magnetic field imaging with scanning NV centers at sub-Kelvin temperatures.

DOI
10.1063/5.0093548

Related papers

Wide-field NV magnetometry under simultaneous high-pressure and high-temperature conditions

Masahiro Ohkuma, Eikichi Kimura, Shumpei Ohyama +7 more

We demonstrate wide-field optically detected magnetic resonance (ODMR) under simultaneous high-pressure and high-temperature conditions using nitrogen-vacancy (NV) centers. Although NV-center magnetometry has been widely used for spatially resolved magnetic-field imaging, its application to extreme environments combining pressure and temperature remains challenging. In this work, we show that ODMR can be observed at 5 GPa and 500 K, demonstrating the feasibility of NV spin readout under such combined extreme conditions. We further perform wide-field ODMR of iron at 7 GPa and 500 K, where the stray magnetic field from the sample is spatially visualized through the pressure cell. These results establish NV-center magnetometry as a promising platform for imaging magnetic phenomena in materials under high-pressure and high-temperature environments.

physics.app-phCross-listedarXiv:2606.25378•Jun 25, 2026

Wideband microwave resonator for NV center vector magnetometry

S. V. Bolshedvorskii, E. A. Koreshin, V. V. Soshenko +10 more

The nitrogen-vacancy (NV) color center in diamond has found applications in magnetometry across various geometrical scales. In particular, vector magnetometry using hand-held devices with low power consumption is of interest for mobile and geological applications. However, such a device requires efficient driving of the four different orientations of the NV center, thus placing demands on a broadband microwave resonator that provides a uniform and highly insensitive microwave magnetic field at the resonance frequencies of the NV centers. Here, we propose a unique planar resonator design that enables efficient driving of all four orientations of NV color centers over a 100 MHz range with only 0.5 W of microwave input power and 250 mW of laser power, allowing vector magnetic field measurements. Using the planar resonator to drive NV centers in diamond, a compact vector magnetometer was fabricated and experimentally studied. The sensitivity of the device is about \[{1.2\text{ nT}}/{\sqrt{\text{Hz}}}\;\]. While this is not a record sensitivity for diamond-based magnetometers, this device occupies its own niche as a compact hand-held magnetometer.

quant-pharXiv:2610.13379•Oct 12, 2026

High-contrast absorption magnetometry in the visible to near-infrared range with nitrogen-vacancy ensembles

Florian Schall, Felix A. Hahl, Lukas Lindner +6 more

Magnetometry with nitrogen-vacancy (NV) centers in diamond has so far been measured via emission of light from NV centers or via absorption at the singlet transition at 1042 nm. Here, we demonstrate a phenomenon of broadband optical absorption by the NV centers starting in the emission wavelength and reaching up to 1000 nm. The measurements are enabled by a high-finesse cavity, which is used for room-temperature continuous wave pump-probe experiments. The red to infrared probe beam shows the typical optically detected magnetic resonance (ODMR) signal of the NV spin with contrasts up to 42 %. This broadband optical absorption has not yet been reported in terms of NV magnetometry. We argue that the lower level of the absorbing transition could be the energetically lower NV singlet state, based on the increased optical absorption for a resonant microwave field and the spectral behavior. Investigations of the photon-shot-noise-limited sensitivity show improvements with increasing probe wavelength, reaching an optimum of 7.5pT/Hz. The results show significantly improved ODMR contrast compared to emission-based magnetometry. This opens a new detection wavelength regime with coherent laser signal detection for high-sensitivity NV magnetometry.

Optics Express•Jan 23, 2025•2 citations

High-resolution magnetic field imaging with a nitrogen-vacancy diamond sensor integrated with a photonic-crystal fiber

И. В. Федотов, Sean M. Blakley, Evgenii E. Serebryannikov +3 more

We demonstrate high-resolution magnetic field imaging with a scanning fiber-optic probe which couples nitrogen-vacancy (NV) centers in diamond to a high-numerical-aperture photonic-crystal fiber integrated with a two-wire microwave transmission line. Magnetic resonance excitation of NV centers driven by the microwave field is read out through optical interrogation through the photonic-crystal fiber to enable high-speed, high-sensitivity magnetic field imaging with sub 30 μm spatial resolution.

Optics Letters•Jan 22, 2016•50 citations

Microwave-free magnetometry based on cross-relaxation resonances in diamond nitrogen-vacancy centers

Ахмеджанов Ринат Абдулхаевич, Л. А. Гущин, Н. А. Низов +4 more

We study cross-relaxation resonances between differently oriented groups of nitrogen-vacancy (NV) centers at small external magnetic fields. These resonances are observed using resonant optical pumping on the NV-center zero phonon line. We suggest a magnetometry protocol that is based on measuring the positions of cross-relaxation resonances on the fluorescence scan. This protocol does not require microwave radiation and can be used to find all components of the unknown magnetic field vector. A room-temperature variant of the protocol with off-resonant optical pumping is also briefly discussed.

Physical Review A•Jul 6, 2017•34 citations

Using NV centers in diamond to detect DC to very-low frequency magnetic fields

Valts Krumins, Ivars Krastins, Oskars Rudzitis +3 more

In this work we present a compact and portable tabletop magnetometer that utilizes negatively charged nitrogen-vacancy (NV) centers in diamond. The magnetometer is operated using a dual microwave resonance detection approach in combination with an optically detected magnetic resonance (ODMR) technique (mitigating drifts in results due to changes of the diamond temperature), capable of simultaneously exciting and registering two ODMR transitions. The experimentally measured magnetic field noise-floor is $\approx 2.3~\textrm{nT}\sqrt{\textrm{Hz}}$ while the calculated shot-noise-limited magnetic field sensitivity is $\approx 585~\textrm{pT}\sqrt{\textrm{Hz}}$ when excited with a continuous wave laser at 0.5~W. These results pave the way for realizing a simple set-up magnetometer for precise single axis magnetic field measurements for example for accurate electric current measurements for stabilization purposes and magnetic communication applications.

quant-pharXiv:2512.24273•Dec 31, 2025
We report on the implementation of a scanning nitrogen-vacancy (NV) magnetometer in a dry dilution refrigerator. Using pulsed optically detected magnetic resonance combined with efficient microwave delivery through a co-planar waveguide, we reach a base temperature of 350 mK, limited by experimental heat load and thermalization of the probe. We demonstrate scanning NV magnetometry by imaging superconducting vortices in a 50-nm-thin aluminum microstructure. The sensitivity of our measurements is approximately 3 μT per square root Hz. Our work demonstrates the feasibility for performing noninvasive magnetic field imaging with scanning NV centers at sub-Kelvin temperatures.
We demonstrate wide-field optically detected magnetic resonance (ODMR) under simultaneous high-pressure and high-temperature conditions using nitrogen-vacancy (NV) centers. Although NV-center magnetometry has been widely used for spatially resolved magnetic-field imaging, its application to extreme environments combining pressure and temperature remains challenging. In this work, we show that ODMR can be observed at 5 GPa and 500 K, demonstrating the feasibility of NV spin readout under such combined extreme conditions. We further perform wide-field ODMR of iron at 7 GPa and 500 K, where the stray magnetic field from the sample is spatially visualized through the pressure cell. These results establish NV-center magnetometry as a promising platform for imaging magnetic phenomena in materials under high-pressure and high-temperature environments.
The nitrogen-vacancy (NV) color center in diamond has found applications in magnetometry across various geometrical scales. In particular, vector magnetometry using hand-held devices with low power consumption is of interest for mobile and geological applications. However, such a device requires efficient driving of the four different orientations of the NV center, thus placing demands on a broadband microwave resonator that provides a uniform and highly insensitive microwave magnetic field at the resonance frequencies of the NV centers. Here, we propose a unique planar resonator design that enables efficient driving of all four orientations of NV color centers over a 100 MHz range with only 0.5 W of microwave input power and 250 mW of laser power, allowing vector magnetic field measurements. Using the planar resonator to drive NV centers in diamond, a compact vector magnetometer was fabricated and experimentally studied. The sensitivity of the device is about 1.2 nT/Hz  {1.2\text{ nT}}/{\sqrt{\text{Hz}}}\;1.2 nT/Hz​. While this is not a record sensitivity for diamond-based magnetometers, this device occupies its own niche as a compact hand-held magnetometer.
Magnetometry with nitrogen-vacancy (NV) centers in diamond has so far been measured via emission of light from NV centers or via absorption at the singlet transition at 1042 nm. Here, we demonstrate a phenomenon of broadband optical absorption by the NV centers starting in the emission wavelength and reaching up to 1000 nm. The measurements are enabled by a high-finesse cavity, which is used for room-temperature continuous wave pump-probe experiments. The red to infrared probe beam shows the typical optically detected magnetic resonance (ODMR) signal of the NV spin with contrasts up to 42 %. This broadband optical absorption has not yet been reported in terms of NV magnetometry. We argue that the lower level of the absorbing transition could be the energetically lower NV singlet state, based on the increased optical absorption for a resonant microwave field and the spectral behavior. Investigations of the photon-shot-noise-limited sensitivity show improvements with increasing probe wavelength, reaching an optimum of 7.5pT/Hz. The results show significantly improved ODMR contrast compared to emission-based magnetometry. This opens a new detection wavelength regime with coherent laser signal detection for high-sensitivity NV magnetometry.
We study cross-relaxation resonances between differently oriented groups of nitrogen-vacancy (NV) centers at small external magnetic fields. These resonances are observed using resonant optical pumping on the NV-center zero phonon line. We suggest a magnetometry protocol that is based on measuring the positions of cross-relaxation resonances on the fluorescence scan. This protocol does not require microwave radiation and can be used to find all components of the unknown magnetic field vector. A room-temperature variant of the protocol with off-resonant optical pumping is also briefly discussed.
In this work we present a compact and portable tabletop magnetometer that utilizes negatively charged nitrogen-vacancy (NV) centers in diamond. The magnetometer is operated using a dual microwave resonance detection approach in combination with an optically detected magnetic resonance (ODMR) technique (mitigating drifts in results due to changes of the diamond temperature), capable of simultaneously exciting and registering two ODMR transitions. The experimentally measured magnetic field noise-floor is ≈2.3 nTHz\approx 2.3~\textrm{nT}\sqrt{\textrm{Hz}}≈2.3 nTHz​ while the calculated shot-noise-limited magnetic field sensitivity is ≈585 pTHz\approx 585~\textrm{pT}\sqrt{\textrm{Hz}}≈585 pTHz​ when excited with a continuous wave laser at 0.5~W. These results pave the way for realizing a simple set-up magnetometer for precise single axis magnetic field measurements for example for accurate electric current measurements for stabilization purposes and magnetic communication applications.