arXiv:2201.06936 · Applied Physics Letters

A highly drift-stable atomic magnetometer for fundamental physics experiments

Martin Rosner, BECK Douglas howard, P. Fierlinger, Hanno Filter, Christopher Klau, F. Kuchler, P. Rößner, M. Sturm, D. Wurm, Zhiyin Sun

Published Apr 18, 2022• 30 citations
arXivHTMLPDF

Abstract

We report the design and performance of a nonmagnetic drift stable optically pumped cesium magnetometer with a measured sensitivity of 35 fT at 200 s integration time and stability below 50 fT between 70 and 600 s. The sensor is based on the nonlinear magneto-optical rotation effect: in a Bell–Bloom configuration, a higher order polarization moment (alignment) of Cs atoms is created with a pump laser beam in an anti-relaxation coated Pyrex cell under vacuum, filled with Cs vapor at room temperature. The polarization plane of light passing through the cell is modulated due the precession of the atoms in an external magnetic field of 2.1 μT, used to optically determine the Larmor precession frequency. Operation is based on a sequence of optical pumping and observation of freely precessing spins at a repetition rate of 8 Hz. This free precession decay readout scheme separates optical pumping and probing and, thus, ensures a systematically highly clean measurement. Due to the residual offset of the sensor of <15 pT together with negligible crosstalk of adjacent sensors, this device is uniquely suitable for a variety of experiments in low-energy particle physics with extreme precision, here as a highly stable and systematically clean reference probe in search for time-reversal symmetry violating electric dipole moments.

DOI
10.1063/5.0083854

Related papers

Sensitive optical atomic magnetometer based on nonlinear magneto-optical rotation

Chris Hovde, Brian Patton, Eric Corsini +2 more

A self-oscillating magnetometer based on nonlinear magneto-optical rotation using amplitude-modulated pump light and unmodulated probe light (AM-NMOR) in 87Rb has been constructed and tested towards a goal of airborne detection of magnetic anomalies. In AM-NMOR, stroboscopic optical pumping via amplitude modulation of the pump beam creates alignment of the ground electronic state of the rubidium atoms. The Larmor precession causes an ac rotation of the polarization of a separate probe beam; the polarization rotation frequency provides a measure of the magnetic field. An anti-relaxation coating on the walls of the atomic vapor cell results in a long lifetime of 56 ms for the alignment, which enables precise measurement of the precession frequency. Light is delivered to the magnetometer by polarization-maintaining optical fibers. Tests of the sensitivity include directly measuring the beat frequency between the magnetometer and a commercial instrument and measurements of Earth's field under magnetically quiet conditions, indicating a sensitivity of at least 5 pT/νHz. Rotating the sensor indicates a heading error of less than 1 nT, limited in part by residual magnetism of the sensor.

arXiv:1003.1531•Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE•Apr 23, 2010•6 citations

Highly stable atomic vector magnetometer based on free spin precession

S. Afach, G. Ban, G. Bison +31 more

We present a magnetometer based on optically pumped Cs atoms that measures the magnitude and direction of a 1 μT magnetic field. Multiple circularly polarized laser beams were used to probe the free spin precession of the Cs atoms. The design was optimized for long-time stability and achieves a scalar resolution better than 300 fT for integration times ranging from 80 ms to 1000 s. The best scalar resolution of less than 80 fT was reached with integration times of 1.6 to 6 s. We were able to measure the magnetic field direction with a resolution better than 10 μrad for integration times from 10 s up to 2000 s.

arXiv:1507.08523•Optics Express•Aug 13, 2015•43 citations

Realization and Optimization of all-optical Cs atom magnetometer based on Bell-Bloom structure

Zhang Jun-ha

In order to availably solve the crosstalk of the traditionally optically pumped magnetometer,in which an oscillating magnetic field drives transitions between Zeeman magnetic sublevels,an all-optical cesium(Cs)atomic magnetometer based on Bell-Bloom structure has been investigated and demonstrated experimentally in this paper.When there is an external magnetic field,by analyzing the Larmor precession spectrum of the polarized ground state atoms induced by a modulated pump laser,the atomic magnetometer with high sensitivity is achieved.First,we present the operating principle of the magnetometer with amplitude modulation and set up the experimental system by means of a three-layer magnetic shield.Then in the case of 100 nT field by optimizing the experimental parameters,i.e.,the frequency and intensity of the pump laser,we have obtained the atomic precession spectrum with line width(full width at half maximum)of 22 Hz and signal noise ratio(SNR)of 3.2×104 per unit hertz band width,and then have evaluated the magnetometer with an intrinsic sensitivity of 100fT/Hz1/2 and spatial resolution smaller than 1.2cm,while the experimentally closed loop test shows that the cesium atomic magnetometer would have the sensitivity of 120fT/Hz1/2 and response time of 300 ms.

Journal of Optoelectronics·laser•Jan 1, 2015•0 citations

All-optical vector atomic magnetometer

E. Zhivun, Brian Patton, Chris Hovde +1 more

We demonstrate an all-optical magnetometer capable of measuring the magnitude and direction of a magnetic field using nonlinear magneto-optical rotation in a cesium vapor. Vector capability is added by effective modulation of the field along orthogonal axes and subsequent demodulation of the magnetic-resonance frequency. This modulation is provided by the AC Stark shift induced by circularly polarized laser beams. The sensor exhibits a demonstrated rms noise floor of 50 fT/ √ Hz in measurement of the field magnitude and 0.5 mrad/ √ Hz in the field direction; elimination of technical noise would improve these sensitivities to 12 fT/ √ Hz and 5 μrad/ √ Hz, respectively. Applications for a precise all-optical vector magnetometer would include magnetically sensitive fundamental physics experiments, such as the search for a permanent electric dipole moment of the neutron.

Bulletin of the American Physical Society•Jun 5, 2014•21 citations

Rubidium atomic magnetometer based on pump-probe nonlinear magneto-optical rotation

Pei-Xian Miao, Shiyu Yang, Wang Jian-Xiang +4 more

We report a rubidium atomic magnetometer based on pump-probe nonlinear magneto-optical rotation. The rubidium vapor cell is placed in a five-layer magnetic shield with inner coils that can generate uniform magnetic fields along the direction of pump beam, and the cell is also placed in the center of a Helmholtz coil that can generate an oscillating magnetic field perpendicular to the direction of pump beam. The atoms are optically pumped by circularly polarized pump beam along a constant magnetic field in a period of time, then the pump beam is turned off and a /2 pulse of oscillating magnetic field for 87Rb atoms is applied. After the above process, the individual atomic magnetic moments become phase coherent, resulting in a transverse magnetization vector precessing at the Larmor frequency in the magnetic field. The linearly polarized probing beam is perpendicular to the direction of magnetic field, and can be seen as a superposition of the left and right circularly polarized light. Because of the different absorptions and dispersions of the left and right circularly polarized light by rubidium atoms, the polarization direction of probing beam rotates when probing beam passes through rubidium vapor cell. The rotation of the polarization is subsequently converted into an electric signal through a polarizing beam splitter. Finally, the decay signal related to the transverse magnetization vector is measured. The Larmor frequency proportional to magnetic field is obtained by the Fourier transform of the decay signal. The value of magnetic field is calculated from the formula:B=(2/) f, where and f are the gyromagnetic ratio and Larmor frequency, respectively. In order to measure the magnetic field in a wide range, the tracking lock mode is proposed and tested. The atomic magnetometer can track the magnetic field jump of 1000 nT or 10000 nT, indicating that the atomic magnetometer has strong locking ability and can be easily locked after start-up. The main performances in different magnetic fields are tested. The results show that the measurement range of the atomic magnetometer is from 100 nT to 100000 nT, the extreme sensitivity is 0.2 pT/Hz1/2, and the magnetic field resolution is 0.1 pT. The transverse relaxation times of the transverse magnetization vector in different magnetic fields are obtained, and the relaxation time decreases with the increase of the magnetic field. When the measurement range is from 5000 nT to 100000 nT, the magnetic field sampling rate of the atomic magnetometer can be adjusted in a range from 1 Hz to 1000 Hz. The atomic magnetometer in high sampling rate can measure weak alternating magnetic field at low frequency. This paper provides an important reference for developing the atomic magnetometer with large measurement range, high sensitivity and high sampling rate.

Acta Physica Sinica•Jan 1, 2017•11 citations

High precision measurement of light polarization using a Cs atomic magnetometer configuration

Xinxin Peng, Yinmin Zhou, Lianhua Li +3 more

Abstract The polarization of light affected by the birefringence of optical elements, such as vapor cell walls, mirrors, and windows, may introduce systematic errors that can be minimized by precisely measuring and correcting the light polarization with a highly sensitive atomic polarimeter. In high precision measurements, it is necessary to measure the light polarization with a high precision where the light interacts with atoms. However, it is inconvenient or even impossible to use the presented method to measure the light polarization, for example, in a cell or cavity. We present a precise and relatively straightforward method to measure the light polarization based on a cesium (Cs) atomic magnetometer configuration. The pump-probe scheme is involved in our experiment. The pump light polarizes atomic spins and the probe light measures the spins polarization. The optical rotation of the probe light significantly depends on the pump light ellipticity, which can be used to measure the light polarization with a high precision. In our work, we present a method to measure the pump light ellipticity with a sensitivity of 2.1 × 10 −8 rad Hz −1/2 for linearly polarized light, which has approximately an order of magnitude improvement than previous research results.

Journal of Physics B Atomic Molecular and Optical Physics•Apr 29, 2021•3 citations
We report the design and performance of a nonmagnetic drift stable optically pumped cesium magnetometer with a measured sensitivity of 35 fT at 200 s integration time and stability below 50 fT between 70 and 600 s. The sensor is based on the nonlinear magneto-optical rotation effect: in a Bell–Bloom configuration, a higher order polarization moment (alignment) of Cs atoms is created with a pump laser beam in an anti-relaxation coated Pyrex cell under vacuum, filled with Cs vapor at room temperature. The polarization plane of light passing through the cell is modulated due the precession of the atoms in an external magnetic field of 2.1 μT, used to optically determine the Larmor precession frequency. Operation is based on a sequence of optical pumping and observation of freely precessing spins at a repetition rate of 8 Hz. This free precession decay readout scheme separates optical pumping and probing and, thus, ensures a systematically highly clean measurement. Due to the residual offset of the sensor of <15 pT together with negligible crosstalk of adjacent sensors, this device is uniquely suitable for a variety of experiments in low-energy particle physics with extreme precision, here as a highly stable and systematically clean reference probe in search for time-reversal symmetry violating electric dipole moments.
A self-oscillating magnetometer based on nonlinear magneto-optical rotation using amplitude-modulated pump light and unmodulated probe light (AM-NMOR) in 87Rb has been constructed and tested towards a goal of airborne detection of magnetic anomalies. In AM-NMOR, stroboscopic optical pumping via amplitude modulation of the pump beam creates alignment of the ground electronic state of the rubidium atoms. The Larmor precession causes an ac rotation of the polarization of a separate probe beam; the polarization rotation frequency provides a measure of the magnetic field. An anti-relaxation coating on the walls of the atomic vapor cell results in a long lifetime of 56 ms for the alignment, which enables precise measurement of the precession frequency. Light is delivered to the magnetometer by polarization-maintaining optical fibers. Tests of the sensitivity include directly measuring the beat frequency between the magnetometer and a commercial instrument and measurements of Earth's field under magnetically quiet conditions, indicating a sensitivity of at least 5 pT/νHz. Rotating the sensor indicates a heading error of less than 1 nT, limited in part by residual magnetism of the sensor.
We present a magnetometer based on optically pumped Cs atoms that measures the magnitude and direction of a 1 μT magnetic field. Multiple circularly polarized laser beams were used to probe the free spin precession of the Cs atoms. The design was optimized for long-time stability and achieves a scalar resolution better than 300 fT for integration times ranging from 80 ms to 1000 s. The best scalar resolution of less than 80 fT was reached with integration times of 1.6 to 6 s. We were able to measure the magnetic field direction with a resolution better than 10 μrad for integration times from 10 s up to 2000 s.
In order to availably solve the crosstalk of the traditionally optically pumped magnetometer,in which an oscillating magnetic field drives transitions between Zeeman magnetic sublevels,an all-optical cesium(Cs)atomic magnetometer based on Bell-Bloom structure has been investigated and demonstrated experimentally in this paper.When there is an external magnetic field,by analyzing the Larmor precession spectrum of the polarized ground state atoms induced by a modulated pump laser,the atomic magnetometer with high sensitivity is achieved.First,we present the operating principle of the magnetometer with amplitude modulation and set up the experimental system by means of a three-layer magnetic shield.Then in the case of 100 nT field by optimizing the experimental parameters,i.e.,the frequency and intensity of the pump laser,we have obtained the atomic precession spectrum with line width(full width at half maximum)of 22 Hz and signal noise ratio(SNR)of 3.2×104 per unit hertz band width,and then have evaluated the magnetometer with an intrinsic sensitivity of 100fT/Hz1/2 and spatial resolution smaller than 1.2cm,while the experimentally closed loop test shows that the cesium atomic magnetometer would have the sensitivity of 120fT/Hz1/2 and response time of 300 ms.
We demonstrate an all-optical magnetometer capable of measuring the magnitude and direction of a magnetic field using nonlinear magneto-optical rotation in a cesium vapor. Vector capability is added by effective modulation of the field along orthogonal axes and subsequent demodulation of the magnetic-resonance frequency. This modulation is provided by the AC Stark shift induced by circularly polarized laser beams. The sensor exhibits a demonstrated rms noise floor of 50 fT/ √ Hz in measurement of the field magnitude and 0.5 mrad/ √ Hz in the field direction; elimination of technical noise would improve these sensitivities to 12 fT/ √ Hz and 5 μrad/ √ Hz, respectively. Applications for a precise all-optical vector magnetometer would include magnetically sensitive fundamental physics experiments, such as the search for a permanent electric dipole moment of the neutron.
We report a rubidium atomic magnetometer based on pump-probe nonlinear magneto-optical rotation. The rubidium vapor cell is placed in a five-layer magnetic shield with inner coils that can generate uniform magnetic fields along the direction of pump beam, and the cell is also placed in the center of a Helmholtz coil that can generate an oscillating magnetic field perpendicular to the direction of pump beam. The atoms are optically pumped by circularly polarized pump beam along a constant magnetic field in a period of time, then the pump beam is turned off and a /2 pulse of oscillating magnetic field for 87Rb atoms is applied. After the above process, the individual atomic magnetic moments become phase coherent, resulting in a transverse magnetization vector precessing at the Larmor frequency in the magnetic field. The linearly polarized probing beam is perpendicular to the direction of magnetic field, and can be seen as a superposition of the left and right circularly polarized light. Because of the different absorptions and dispersions of the left and right circularly polarized light by rubidium atoms, the polarization direction of probing beam rotates when probing beam passes through rubidium vapor cell. The rotation of the polarization is subsequently converted into an electric signal through a polarizing beam splitter. Finally, the decay signal related to the transverse magnetization vector is measured. The Larmor frequency proportional to magnetic field is obtained by the Fourier transform of the decay signal. The value of magnetic field is calculated from the formula:B=(2/) f, where and f are the gyromagnetic ratio and Larmor frequency, respectively. In order to measure the magnetic field in a wide range, the tracking lock mode is proposed and tested. The atomic magnetometer can track the magnetic field jump of 1000 nT or 10000 nT, indicating that the atomic magnetometer has strong locking ability and can be easily locked after start-up. The main performances in different magnetic fields are tested. The results show that the measurement range of the atomic magnetometer is from 100 nT to 100000 nT, the extreme sensitivity is 0.2 pT/Hz1/2, and the magnetic field resolution is 0.1 pT. The transverse relaxation times of the transverse magnetization vector in different magnetic fields are obtained, and the relaxation time decreases with the increase of the magnetic field. When the measurement range is from 5000 nT to 100000 nT, the magnetic field sampling rate of the atomic magnetometer can be adjusted in a range from 1 Hz to 1000 Hz. The atomic magnetometer in high sampling rate can measure weak alternating magnetic field at low frequency. This paper provides an important reference for developing the atomic magnetometer with large measurement range, high sensitivity and high sampling rate.
Abstract The polarization of light affected by the birefringence of optical elements, such as vapor cell walls, mirrors, and windows, may introduce systematic errors that can be minimized by precisely measuring and correcting the light polarization with a highly sensitive atomic polarimeter. In high precision measurements, it is necessary to measure the light polarization with a high precision where the light interacts with atoms. However, it is inconvenient or even impossible to use the presented method to measure the light polarization, for example, in a cell or cavity. We present a precise and relatively straightforward method to measure the light polarization based on a cesium (Cs) atomic magnetometer configuration. The pump-probe scheme is involved in our experiment. The pump light polarizes atomic spins and the probe light measures the spins polarization. The optical rotation of the probe light significantly depends on the pump light ellipticity, which can be used to measure the light polarization with a high precision. In our work, we present a method to measure the pump light ellipticity with a sensitivity of 2.1 × 10 −8 rad Hz −1/2 for linearly polarized light, which has approximately an order of magnitude improvement than previous research results.