Performance analysis of an optically pumped magnetometer in Earth’s magnetic field

G. Oelsner, V. Schultze, R.P.J. IJsselsteijn, Ronny Stolz

Published Dec 1, 2019• 27 citations
arXivHTMLPDF

Abstract

Abstract We experimentally investigate the influence of the orientation of optically pumped magnetometers in Earth’s magnetic field. We focus our analysis to an operational mode that promises femtotesla field resolutions at such field strengths. For this so-called light-shift dispersed $M_{z}$ Mz (LSD-Mz) regime, we focus on the key parameters defining its performance. That are the reconstructed Larmor frequency, the transfer function between output signal and magnetic field amplitude as well as the shot noise limited field resolution. We demonstrate that due to the use of two well balanced laser beams for optical pumping with different helicities the heading error as well as the field sensitivity of a detector both are only weakly influenced by the heading in a large orientation angle range.

DOI
10.1140/epjqt/s40507-019-0076-9

Related papers

An Optically Pumped Magnetometer Working in the Light-Shift Dispersed Mz Mode

V. Schultze, Bastian Schillig, R.P.J. IJsselsteijn +3 more

We present an optically pumped magnetometer working in a new operational mode-the light-shift dispersed Mz (LSD-Mz) mode. It is realized combining various features; (1) high power off-resonant optical pumping; (2) Mz configuration, where pumping light and magnetic field of interest are oriented parallel to each other; (3) use of small alkali metal vapor cells of identical properties in integrated array structures, where two such cells are pumped by circularly polarized light of opposite helicity; and (4) subtraction of the Mz signals of these two cells. The LSD-Mz magnetometer's performance depends on the inherent and very complex interplay of input parameters. In order to find the configuration of optimal magnetometer resolution, a sensitivity analysis of the input parameters by means of Latin Hypercube Sampling was carried out. The resulting datasets of the multi-dimensional parameter space exploration were assessed by a subsequent physically reasonable interpretation. Finally, the best shot-noise limited magnetic field resolution was determined within that parameter space. As the result, using two 50 mm3 integrated vapor cells a magnetic field resolution below 10 fT/√Hz at Earth's magnetic field strength is possible.

Sensors•Mar 10, 2017•67 citations

Sources of heading errors in optically pumped magnetometers operated in the Earth's magnetic field

G. Oelsner, V. Schultze, R.P.J. IJsselsteijn +2 more

When optically pumped magnetometers are aimed for use in the Earth's magnetic field, the orientation of the sensor to the field direction is of special importance to achieve accurate measurement results. Measurement errors and inaccuracies related to the heading of the sensor can be an even more severe problem in the case of special operational configurations, such as the use of strong off-resonant pumping. We systematically study the main contributions to the heading error in systems that promise high magnetic field resolutions at the Earth's magnetic field strengths, namely the nonlinear Zeeman splitting and the orientation-dependent light shift. The good correspondence of our theoretical analysis to experimental data demonstrates that both of these effects are related to a heading-dependent modification of the interaction between the laser light and the dipole moment of the atoms. Also, our results promise a compensation of both effects using a combination of clockwise and counterclockwise circular polarization.

arXiv:1809.00627•Physical Review A•Jan 16, 2019•49 citations

Asymmetric Heading Errors and its Suppression in Atomic Magnetometers

Yue Chang, Yuhao Guo, Shuangai Wan +1 more

Precession frequencies measured by optically pumped scalar magnetometers are dependent on the relative angle between the sensor and the external magnetic field. This dependence is known to be induced mainly by the nonlinear Zeeman effect and the orientation-dependent light shift, resulting in the so-called heading errors if the magnetic field orientation is not well known or is not stable. In this work, we find that the linear nuclear Zeeman effect has also a significant impact on the heading errors. It not only shifts the precession frequency but causes asymmetry: the heading error for sensors orienting in the upper-half plane with respect to the external field is different from the case when the sensors work in the lower-half plane. This heading error also depends on the relative direction of the probe laser to the driving magnetic field. With a left-handed circularly-polarized pump laser, when the probe laser is parallel to the driving field, the angular dependence of the precession frequency is smaller when the sensor is in the upper plane. Otherwise, when they are perpendicular to each other, the heading error is smaller when the sensor is in the lower plane. Furthermore, to suppress the heading error, we propose to utilize a small magnetic field along the propagation direction of the pump laser. By tuning the magnitude of this auxiliary field, the heading-error curve is flattened around different angles, which can increase the accuracy in practice when the magnetometer works around a certain orientation angle.

•Feb 5, 2021•0 citations

Multi-Parameter Optimization of Rubidium Laser Optically Pumped Magnetometers with Geomagnetic Field Intensity

Kun Xu, Xiuyan Ren, Yujie Xiang +7 more

Rubidium laser optically pumped magnetometers (OPMs) are widely used magnetic sensors based on the Zeeman effect, laser pumping, and magnetic resonance principles. They measure the magnetic field by measuring the magnetic resonance signal passing through a rubidium atomic gas cell. The quality of the magnetic resonance signal is a necessary condition for a magnetometer to achieve high sensitivity. In this research, to obtain the best magnetic resonance signal of rubidium laser OPMs in the Earth’s magnetic field intensity, the experiment system of rubidium laser OPMs is built with a rubidium atomic gas cell as the core component. The linewidth and amplitude ratio (LAR) of magnetic resonance signals is utilized as the optimization objective function. The magnetic resonance signals of the magnetometer experiment system are experimentally measured for different laser frequencies, radio frequency (RF) intensities, laser powers, and atomic gas cell temperatures in a background magnetic field of 50,765 nT. The experimental results indicate that optimizing these parameters can reduce the LAR by one order of magnitude. This shows that the optimal parameter combination can effectively improve the sensitivity of the magnetometer. The sensitivity defined using the noise spectral density measured under optimal experimental parameters is 1.5 pT/Hz1/2@1 Hz. This work will provide key technical support for rubidium laser OPMs’ product development.

Sensors•Nov 2, 2023•6 citations

Optimization and vectorization of an Mz-type optically pumped rubidium magnetometer

Zhengyu Su, Yang Li, Yongbiao Yang +3 more

Optically pumped magnetometers (OPMs) have demonstrated significant potential in weak magnetic field detection due to their high sensitivity. In this study, we developed an Mz-type optically pumped rubidium magnetometer using a paraffin-coated anti-relaxation vapor cell. The system optimization and performance characterization were conducted inside a magnetic shield. Specifically, the pump light intensity and radio-frequency (RF) magnetic field were jointly optimized by using the linewidth-amplitude ratio as the core metric. Based on the frequency-domain noise spectrum, the sensitivity in open-loop mode was measured to be approximately 30.8pT/Hz 1/2 . Furthermore, a closed-loop feedback locking technique was applied, reducing the measured noise floor under the tested conditions and improving the sensitivity to 22.9pT/Hz 1/2 , with a measured −3dB bandwidth of 123 Hz. The dynamic characteristics were evaluated via magnetic-field step response, showing that the system could track magnetic-field changes stably under closed-loop operation. Finally, by using tri-axial modulation and frequency-domain demodulation, we overcame the scalar measurement limitation of traditional Mz magnetometers. This work realizes vector magnetic field detection and provides a technical basis for applications such as geomagnetic navigation and magnetic anomaly detection.

Journal of the Optical Society of America B•Jun 24, 2026•1 citation

Optimization and vectorization of a Mz-type optically-pumped Rubidium magnetometer

Zhengyu Su, Yang Li, Yongbiao Yang +3 more

Optically pumped magnetometers (OPMs) have demonstrated significant potential in weak magnetic field detection due to their high sensitivity. In this study, we developed an Mz-type optically pumped rubidium magnetometer using a paraffin-coated anti-relaxation vapor cell. The system optimization and performance characterization were conducted inside a magnetic shield. Specifically, the pump light intensity and radio-frequency (RF) magnetic field were jointly optimized by using the linewidth-amplitude ratio as the core metric. Based on the frequency-domain noise spectrum, the sensitivity in open-loop mode was measured to be approximately 30.8 pT/Hz^{1/2}. Furthermore, a closed-loop feedback locking technique was applied, reducing the measured noise floor under the tested conditions and improving the sensitivity to 22.9 pT/Hz^{1/2}, with a measured -3 dB bandwidth of 123 Hz. The dynamic characteristics were evaluated via magnetic-field step response, showing that the system could track magnetic-field changes stably under closed-loop operation. Finally, by using tri-axial modulation and frequency-domain demodulation, we overcame the scalar measurement limitation of traditional Mz magnetometers. This work realizes vector magnetic field detection and provides a technical basis for applications such as geomagnetic navigation and magnetic anomaly detection.

arXiv:2604.02884•Apr 3, 2026•0 citations
Abstract We experimentally investigate the influence of the orientation of optically pumped magnetometers in Earth’s magnetic field. We focus our analysis to an operational mode that promises femtotesla field resolutions at such field strengths. For this so-called light-shift dispersed MzM_{z}Mz​ Mz (LSD-Mz) regime, we focus on the key parameters defining its performance. That are the reconstructed Larmor frequency, the transfer function between output signal and magnetic field amplitude as well as the shot noise limited field resolution. We demonstrate that due to the use of two well balanced laser beams for optical pumping with different helicities the heading error as well as the field sensitivity of a detector both are only weakly influenced by the heading in a large orientation angle range.
We present an optically pumped magnetometer working in a new operational mode-the light-shift dispersed Mz (LSD-Mz) mode. It is realized combining various features; (1) high power off-resonant optical pumping; (2) Mz configuration, where pumping light and magnetic field of interest are oriented parallel to each other; (3) use of small alkali metal vapor cells of identical properties in integrated array structures, where two such cells are pumped by circularly polarized light of opposite helicity; and (4) subtraction of the Mz signals of these two cells. The LSD-Mz magnetometer's performance depends on the inherent and very complex interplay of input parameters. In order to find the configuration of optimal magnetometer resolution, a sensitivity analysis of the input parameters by means of Latin Hypercube Sampling was carried out. The resulting datasets of the multi-dimensional parameter space exploration were assessed by a subsequent physically reasonable interpretation. Finally, the best shot-noise limited magnetic field resolution was determined within that parameter space. As the result, using two 50 mm3 integrated vapor cells a magnetic field resolution below 10 fT/√Hz at Earth's magnetic field strength is possible.
When optically pumped magnetometers are aimed for use in the Earth's magnetic field, the orientation of the sensor to the field direction is of special importance to achieve accurate measurement results. Measurement errors and inaccuracies related to the heading of the sensor can be an even more severe problem in the case of special operational configurations, such as the use of strong off-resonant pumping. We systematically study the main contributions to the heading error in systems that promise high magnetic field resolutions at the Earth's magnetic field strengths, namely the nonlinear Zeeman splitting and the orientation-dependent light shift. The good correspondence of our theoretical analysis to experimental data demonstrates that both of these effects are related to a heading-dependent modification of the interaction between the laser light and the dipole moment of the atoms. Also, our results promise a compensation of both effects using a combination of clockwise and counterclockwise circular polarization.
Precession frequencies measured by optically pumped scalar magnetometers are dependent on the relative angle between the sensor and the external magnetic field. This dependence is known to be induced mainly by the nonlinear Zeeman effect and the orientation-dependent light shift, resulting in the so-called heading errors if the magnetic field orientation is not well known or is not stable. In this work, we find that the linear nuclear Zeeman effect has also a significant impact on the heading errors. It not only shifts the precession frequency but causes asymmetry: the heading error for sensors orienting in the upper-half plane with respect to the external field is different from the case when the sensors work in the lower-half plane. This heading error also depends on the relative direction of the probe laser to the driving magnetic field. With a left-handed circularly-polarized pump laser, when the probe laser is parallel to the driving field, the angular dependence of the precession frequency is smaller when the sensor is in the upper plane. Otherwise, when they are perpendicular to each other, the heading error is smaller when the sensor is in the lower plane. Furthermore, to suppress the heading error, we propose to utilize a small magnetic field along the propagation direction of the pump laser. By tuning the magnitude of this auxiliary field, the heading-error curve is flattened around different angles, which can increase the accuracy in practice when the magnetometer works around a certain orientation angle.
Rubidium laser optically pumped magnetometers (OPMs) are widely used magnetic sensors based on the Zeeman effect, laser pumping, and magnetic resonance principles. They measure the magnetic field by measuring the magnetic resonance signal passing through a rubidium atomic gas cell. The quality of the magnetic resonance signal is a necessary condition for a magnetometer to achieve high sensitivity. In this research, to obtain the best magnetic resonance signal of rubidium laser OPMs in the Earth’s magnetic field intensity, the experiment system of rubidium laser OPMs is built with a rubidium atomic gas cell as the core component. The linewidth and amplitude ratio (LAR) of magnetic resonance signals is utilized as the optimization objective function. The magnetic resonance signals of the magnetometer experiment system are experimentally measured for different laser frequencies, radio frequency (RF) intensities, laser powers, and atomic gas cell temperatures in a background magnetic field of 50,765 nT. The experimental results indicate that optimizing these parameters can reduce the LAR by one order of magnitude. This shows that the optimal parameter combination can effectively improve the sensitivity of the magnetometer. The sensitivity defined using the noise spectral density measured under optimal experimental parameters is 1.5 pT/Hz1/2@1 Hz. This work will provide key technical support for rubidium laser OPMs’ product development.
Optically pumped magnetometers (OPMs) have demonstrated significant potential in weak magnetic field detection due to their high sensitivity. In this study, we developed an Mz-type optically pumped rubidium magnetometer using a paraffin-coated anti-relaxation vapor cell. The system optimization and performance characterization were conducted inside a magnetic shield. Specifically, the pump light intensity and radio-frequency (RF) magnetic field were jointly optimized by using the linewidth-amplitude ratio as the core metric. Based on the frequency-domain noise spectrum, the sensitivity in open-loop mode was measured to be approximately 30.8pT/Hz 1/2 . Furthermore, a closed-loop feedback locking technique was applied, reducing the measured noise floor under the tested conditions and improving the sensitivity to 22.9pT/Hz 1/2 , with a measured −3dB bandwidth of 123 Hz. The dynamic characteristics were evaluated via magnetic-field step response, showing that the system could track magnetic-field changes stably under closed-loop operation. Finally, by using tri-axial modulation and frequency-domain demodulation, we overcame the scalar measurement limitation of traditional Mz magnetometers. This work realizes vector magnetic field detection and provides a technical basis for applications such as geomagnetic navigation and magnetic anomaly detection.
Optically pumped magnetometers (OPMs) have demonstrated significant potential in weak magnetic field detection due to their high sensitivity. In this study, we developed an Mz-type optically pumped rubidium magnetometer using a paraffin-coated anti-relaxation vapor cell. The system optimization and performance characterization were conducted inside a magnetic shield. Specifically, the pump light intensity and radio-frequency (RF) magnetic field were jointly optimized by using the linewidth-amplitude ratio as the core metric. Based on the frequency-domain noise spectrum, the sensitivity in open-loop mode was measured to be approximately 30.8 pT/Hz^{1/2}. Furthermore, a closed-loop feedback locking technique was applied, reducing the measured noise floor under the tested conditions and improving the sensitivity to 22.9 pT/Hz^{1/2}, with a measured -3 dB bandwidth of 123 Hz. The dynamic characteristics were evaluated via magnetic-field step response, showing that the system could track magnetic-field changes stably under closed-loop operation. Finally, by using tri-axial modulation and frequency-domain demodulation, we overcame the scalar measurement limitation of traditional Mz magnetometers. This work realizes vector magnetic field detection and provides a technical basis for applications such as geomagnetic navigation and magnetic anomaly detection.