arXiv:1410.4350 · Physical Review A

Performance predictions for a laser-intensified thermal beam for use in high-resolution focused-ion-beam instruments

Steinar H. W. Wouters, G. ten Haaf, Remy Notermans, Nicola Debernardi, Peter H. A. Mutsaers, O.J. Luiten, Edgar J. D. Vredenbregt

Published Dec 15, 2014• 17 citations
arXivHTMLPDF

Abstract

Photoionization of a laser-cooled and compressed atomic beam from a high-flux thermal source can be used to create a high-brightness ion beam for use in focused-ion-beam instruments. Here we show using calculations and Doppler cooling simulations that an atomic rubidium beam with an equivalent brightness of $2.1\ifmmode\times\else\texttimes\fi{}{10}^{7} \mathrm{A}/(\mathrm{m}{}^{2} \mathrm{sr} \mathrm{eV})$ can be created using a compact 5 cm long two-dimensional magneto-optical compressor. If this can be conserved during the photoionization process, this leads to an ion beam brightness an order of magnitude higher than produced by a liquid metal ion source. The source is also capable of producing a flux of $6.2\ifmmode\times\else\texttimes\fi{}{10}^{9} \mathrm{s}{}^{\ensuremath{-}1}$ that results in a substantial beam current of 1 nA once fully ionized.

DOI
10.1103/physreva.90.063817

Related papers

A high brightness laser-cooled atomic beam for application in high resolution FIB

Shw Steinar Wouters, S.B. van der Geer, G. ten Haaf +2 more

A new type of high-brightness ion source is under development which employs transverse laser cooling and compression of a thermal atomic rubidium beam, followed by in-field photo-ionization. When attached to a focusing column, this Focused Ion Beam (FIB) has the advantage of supplying a higher current in a smaller spot compared to conventional LMIS-based FIBs, thus increasing both the resolution and the speed of the FIB. Furthermore, different types of ion species can be used, broadening the range of applications of the FIB. Simulations using a 10 cm long laser cooling and compression stage and a realistic ionization and acceleration structure, predict an achievable brightness for 87 Rb + of order 10 7 A/m 2 sr eV at an energy spread of less than 1 eV and a current of tens of pA. This would lead to a spot size below 5 nm. Simulations and modeling on the ionization process have led to a better understanding of stochastic heating. Experimental realization of the compact ion source has recently started with the development of an efficient high-flux atom source and a 2D laser cooler and compressor. Progress on simulations and experimental results will be reported.

•Jan 1, 2013•0 citations

Application of laser-cooling to achieve an ultra-cold ion beam for FIB

Steinar H. W. Wouters, Janssen, B, G. ten Haaf +4 more

A new type of ultra-cold ion source is under development which employs transverse laser cooling and compression of a thermal atomic rubidium beam followed by photo-ionization. The resulting ultra-cold plasma is focused to a nanometer-sized spot using an existing Focused Ion Beam column and this spot can be used for the fabrication of nano-structures. Simulations of a 10 cm long laser-cooling stage and of disorder-induced heating of the resulting ion beam, predict an achievable brightness for87Rb+ of order 107 A/m2 sr eV at an longitudinal energy spread of less than 1 eV and a current of tens of pA, which is substantially better than conventional ion sources. Experimental realization of the compact ion source has recently started with the development of an efficient high-flux atom source and a 2D laser cooler. Progress on these items will be reported.

•Jan 1, 2013•0 citations

Using laser-cooled atoms as a focused ion beam source

J. L. Hanssen, E. A. Dakin, Jabez J. McClelland +1 more

The authors describe a new method for creating a high quality focused ion beam using laser-cooled neutral atoms in a magneto-optical trap as an ion source. They show that this new technique can provide spot resolutions and brightness values that are better than the state of the art in focused ion beams. The source can be used with a range of different ionic species and can be combined with laser cooling techniques to exert unprecedented control over the ion emission, for example, producing single ions “on demand.” The beam quality is a result of a high brightness and a narrow energy distribution, both of which stem from the cold temperature (≈100μK) of the atoms. The ions are produced by subjecting the cold neutral atoms to a photoionization laser, after which they become a compact source of nearly monoenergetic ions. With the application of a potential gradient, the ions form a beam that can be focused via standard ion optical techniques. They discuss estimations based on the initial size of the ion cloud and the energy distribution and show that the resulting beam has a low emittance.

Journal of Vacuum Science & Technology B Microelectronics and Nanometer Structures Processing Measurement and Phenomena•Nov 1, 2006•18 citations

Direct Magneto-Optical Compression of an Effusive Atomic Beam for Application in a High-Resolution Focused Ion Beam

G. ten Haaf, T. C. H. de Raadt, G. P. Offermans +4 more

Photoionizing a beam of ultracold atoms can provide a high-brightness ion beam, to improve resolution in focused-ion-beam nanofabrication. With this aim, the authors use laser cooling and compression to give a ${}^{85}$Rb atomic beam an equivalent brightness better than the current standard for the semiconductor industry. This could represent an important step toward smaller feature sizes in integrated circuits.

arXiv:1612.01883•Physical Review Applied•May 22, 2017•13 citations

Cavity-enhanced photoionization of an ultracold rubidium beam for application in focused ion beams

G. ten Haaf, Steinar H. W. Wouters, Peter H. A. Mutsaers +1 more

A two-step photoionization strategy of an ultracold rubidium beam for application in a focused ion beam instrument is analyzed and implemented. In this strategy the atomic beam is partly selected with an aperture after which the transmitted atoms are ionized in the overlap of a tightly cylindrically focused excitation laser beam and an ionization laser beam whose power is enhanced in a build-up cavity. The advantage of this strategy, as compared to without the use of a build-up cavity, is that higher ionization degrees can be reached at higher currents. Optical Bloch equations including the photoionization process are used to calculate what ionization degree and ionization position distribution can be reached. Furthermore, the ionization strategy is tested on an ultracold beam of $^{85}\mathrm{Rb}$ atoms. The beam current is measured as a function of the excitation and ionization laser beam intensity and the selection aperture size. Although details are different, the global trends of the measurements agree well with the calculation. With a selection aperture diameter of $52\phantom{\rule{0.28em}{0ex}}\ensuremath{\mu}\mathrm{m}$, a current of $\left(170\ifmmode\pm\else\textpm\fi{}4\right)$ pA is measured, which according to calculations is 63% of the current equivalent of the transmitted atomic flux. Taking into account the ionization degree the ion beam peak reduced brightness is estimated at $1\ifmmode\times\else\texttimes\fi{}{10}^{7}$ A/(${\mathrm{m}}^{2}\phantom{\rule{0.16em}{0ex}}\mathrm{sr}\phantom{\rule{0.16em}{0ex}}\mathrm{eV}$).

arXiv:1708.01933•Physical Review A•Nov 13, 2017•10 citations

Application of laser-cooling and compression to create a high resolution focused ion beam

Steinar H. W. Wouters, G. ten Haaf, Peter H. A. Mutsaers +2 more

Ultra-low temperature (1 mK) ion beams can be created by photo-ionization of a laser cooled and compressed thermal atomic beam. Full device simulations show that for rubidium ions, spot sizes down to 1 nm can be achieved at a current of 1 pA.

•Jan 1, 2015•0 citations
Photoionization of a laser-cooled and compressed atomic beam from a high-flux thermal source can be used to create a high-brightness ion beam for use in focused-ion-beam instruments. Here we show using calculations and Doppler cooling simulations that an atomic rubidium beam with an equivalent brightness of 2.1\ifmmode×\else\texttimes\fi107A/(m2sreV)2.1\ifmmode\times\else\texttimes\fi{}{10}^{7} \mathrm{A}/(\mathrm{m}{}^{2} \mathrm{sr} \mathrm{eV})2.1\ifmmode×\else\texttimes\fi107A/(m2sreV) can be created using a compact 5 cm long two-dimensional magneto-optical compressor. If this can be conserved during the photoionization process, this leads to an ion beam brightness an order of magnitude higher than produced by a liquid metal ion source. The source is also capable of producing a flux of 6.2\ifmmode×\else\texttimes\fi109s\ensuremath−16.2\ifmmode\times\else\texttimes\fi{}{10}^{9} \mathrm{s}{}^{\ensuremath{-}1}6.2\ifmmode×\else\texttimes\fi109s\ensuremath−1 that results in a substantial beam current of 1 nA once fully ionized.
A new type of high-brightness ion source is under development which employs transverse laser cooling and compression of a thermal atomic rubidium beam, followed by in-field photo-ionization. When attached to a focusing column, this Focused Ion Beam (FIB) has the advantage of supplying a higher current in a smaller spot compared to conventional LMIS-based FIBs, thus increasing both the resolution and the speed of the FIB. Furthermore, different types of ion species can be used, broadening the range of applications of the FIB. Simulations using a 10 cm long laser cooling and compression stage and a realistic ionization and acceleration structure, predict an achievable brightness for 87 Rb + of order 10 7 A/m 2 sr eV at an energy spread of less than 1 eV and a current of tens of pA. This would lead to a spot size below 5 nm. Simulations and modeling on the ionization process have led to a better understanding of stochastic heating. Experimental realization of the compact ion source has recently started with the development of an efficient high-flux atom source and a 2D laser cooler and compressor. Progress on simulations and experimental results will be reported.
A new type of ultra-cold ion source is under development which employs transverse laser cooling and compression of a thermal atomic rubidium beam followed by photo-ionization. The resulting ultra-cold plasma is focused to a nanometer-sized spot using an existing Focused Ion Beam column and this spot can be used for the fabrication of nano-structures. Simulations of a 10 cm long laser-cooling stage and of disorder-induced heating of the resulting ion beam, predict an achievable brightness for87Rb+ of order 107 A/m2 sr eV at an longitudinal energy spread of less than 1 eV and a current of tens of pA, which is substantially better than conventional ion sources. Experimental realization of the compact ion source has recently started with the development of an efficient high-flux atom source and a 2D laser cooler. Progress on these items will be reported.
The authors describe a new method for creating a high quality focused ion beam using laser-cooled neutral atoms in a magneto-optical trap as an ion source. They show that this new technique can provide spot resolutions and brightness values that are better than the state of the art in focused ion beams. The source can be used with a range of different ionic species and can be combined with laser cooling techniques to exert unprecedented control over the ion emission, for example, producing single ions “on demand.” The beam quality is a result of a high brightness and a narrow energy distribution, both of which stem from the cold temperature (≈100μK) of the atoms. The ions are produced by subjecting the cold neutral atoms to a photoionization laser, after which they become a compact source of nearly monoenergetic ions. With the application of a potential gradient, the ions form a beam that can be focused via standard ion optical techniques. They discuss estimations based on the initial size of the ion cloud and the energy distribution and show that the resulting beam has a low emittance.
Photoionizing a beam of ultracold atoms can provide a high-brightness ion beam, to improve resolution in focused-ion-beam nanofabrication. With this aim, the authors use laser cooling and compression to give a 85{}^{85}85Rb atomic beam an equivalent brightness better than the current standard for the semiconductor industry. This could represent an important step toward smaller feature sizes in integrated circuits.
A two-step photoionization strategy of an ultracold rubidium beam for application in a focused ion beam instrument is analyzed and implemented. In this strategy the atomic beam is partly selected with an aperture after which the transmitted atoms are ionized in the overlap of a tightly cylindrically focused excitation laser beam and an ionization laser beam whose power is enhanced in a build-up cavity. The advantage of this strategy, as compared to without the use of a build-up cavity, is that higher ionization degrees can be reached at higher currents. Optical Bloch equations including the photoionization process are used to calculate what ionization degree and ionization position distribution can be reached. Furthermore, the ionization strategy is tested on an ultracold beam of 85Rb^{85}\mathrm{Rb}85Rb atoms. The beam current is measured as a function of the excitation and ionization laser beam intensity and the selection aperture size. Although details are different, the global trends of the measurements agree well with the calculation. With a selection aperture diameter of $52\phantom{\rule{0.28em}{0ex}}\ensuremath{\mu}\mathrm{m}$, a current of $\left(170\ifmmode\pm\else\textpm\fi{}4\right)$ pA is measured, which according to calculations is 63% of the current equivalent of the transmitted atomic flux. Taking into account the ionization degree the ion beam peak reduced brightness is estimated at $1\ifmmode\times\else\texttimes\fi{}{10}^{7}$ A/(m2sreV{\mathrm{m}}^{2}\phantom{\rule{0.16em}{0ex}}\mathrm{sr}\phantom{\rule{0.16em}{0ex}}\mathrm{eV}m2sreV).