Performance Characterization and Optimization of a Miniaturized SERF Atomic Magnetometer via Tunable Laser Power

Peng Shi, Chen Zuo, Qisong Li, Shulin Zhang

Published Mar 23, 2026• 0 citations
Publisher

Abstract

Spin-exchange relaxation-free (SERF) atomic magnetometers have emerged as highly promising candidates for ultra-weak magnetic field detection, particularly in biomagnetic imaging, owing to their exceptional sensitivity, amenability to miniaturization, and near-room-temperature operation. While current miniaturized magnetometers typically employ laser chips with fixed optical power, the quantitative impact of laser power on critical performance metrics remains to be fully elucidated. This study systematically investigates the influence of laser power on sensitivity, bandwidth, and dynamic range by incorporating considerations of power broadening, saturation absorption, and noise constraints. A miniaturized probe, integrated with an actively controlled vertical-cavity surface-emitting laser (VCSEL), was developed for experimental validation. Theoretical and experimental results consistently demonstrate that as optical power increases, sensitivity exhibits a non-monotonic dependence, whereas both bandwidth and dynamic range manifest a monotonic upward trend, aligning well with theoretical simulations. The optimized sensor achieved a peak sensitivity of 16 fT/√Hz at 300 μW, while the bandwidth and dynamic range reached 230 Hz and ±5.4 nT at 500 μW, respectively. This work establishes a robust theoretical and experimental framework for the comprehensive performance optimization of laser-integrated miniaturized atomic magnetometers.

DOI
10.3390/s26062000

Related papers

A High Sensitivity Closed-Loop Spin-Exchange Relaxation-Free Atomic Magnetometer With Broad Bandwidth

Qingqian Guo, Tao Hu, Chun-qiao Chen +6 more

Spin-exchange relaxation-free (SERF) atomic magnetometers are the promising non-cryogenic sensors to potentially replace the superconducting quantum interference device (SQUID) in the ultra-weak magnetic-field measurements. Although the SERF magnetometers have achieved a sensitivity of sub-femtotesla, the bandwidth is limited at a narrow range. In this work, we developed a closed-loop SERF magnetometer under proportional-integral (PI) control, the bandwidth could be adaptively tuned and extended above 1.2 kHz. A nearly 40-fold enhancement of the typical bandwidth was achieved in comparison with the open-loop system and the sensitivity was maintained at 15 fT/Hz1/2from 30 Hz to 750 Hz. Moreover, the magnetic field dynamic range could also be greatly improved. This proposed atomic magnetometer with higher bandwidth, sensitivity, and dynamic range has a good prospect in developing a compact, portable, and non-cryogenic magnetic field sensor and may find more applications in biomedical, nuclear magnetic resonance, and fundamental physics research.

IEEE Sensors Journal•Aug 3, 2021•36 citations

Bandwidth Extension Enabled by Modulated VCSEL Pumping for Quantum Sensing Application

Peng Zhou, Wei Quan, Feilong Xu +6 more

Emerging quantum sensing devices namely atomic magnetometers have become promising candidate for future biomagnetism imaging. However, the limitation of detection bandwidth exists a major drawback for devices of this kind. Conventional approaches, including transverse magnetic field modulation and close-loop control, have seen challenging tradeoffs between size, resolution and bandwidths. In this study, a wide bandwidth zero-field magnetometer with a miniaturized light source has been developed for alternating current (AC) magnetic field detection. A dual-beam Spin-Exchange Relaxation-Free (SERF) magnetometer employing a 4×4×4mm3 alkali metal vapor cell achieves a bandwidth of 188 Hz. This represents a threefold increase in bandwidth compared to non-modulated mode. The sensitivity under modulated mode demonstrated a sensitivity of 55fT/Hz1/2within the frequency range of 10-188Hz. Our results also demonstrated that the sensitivity is constrained by relaxation induced by the modulation of pumping source, so possible tradeoffs can be made based on various applications. The proposed method provides novel approach for bandwidth extension of atomic magnetometers which further pave the way for future applications including magnetoencephalography (MEG), particularly in reducing magnetic noise crosstalk and enhancing spatial resolution.

IEEE Transactions on Instrumentation and Measurement•Jan 1, 2025•2 citations

Femtotesla Spin-Exchange Relaxation-Free Atomic Magnetometer With a Multipass Cell

Yibo Qi, Bo Li, Shuying Wang +3 more

Miniaturized high-sensitivity atomic magnetometers play a crucial role in biomagnetic measurements. The utilization of multipass cells enhances the sensitivity of atomic magnetometers. This study explores the benefits of employing a multipass cell in spin-exchange relaxation-free (SERF) atomic magnetometers through theoretical and experimental analysis. A signal-to-noise ratio (SNR) model was established by accounting for the optimal detuning frequency of the probe light and low-frequency 1/f noise. This enabled the optimal number of probe light passes through the vapor cell to be determined. Additionally, compact and symmetric reflective optical paths were designed to mitigate polarization gradients in atomic magnetometers with high atomic number density. Experimental results confirmed that the magnetic field sensitivity of the triple-pass configuration (1.8 fT/Hz$^{1/2}$) was 1.89 times that of the single-pass configuration (3.4 fT/Hz$^{1/2}$), consistent with the theoretical analysis. This study provides valuable theoretical guidance for analyzing the SNR performance of SERF atomic magnetometers and demonstrates the potential of multipass cells in enhancing their magnetic field sensitivity. The utilization of multipass cells presents a viable approach toward the realization of highly sensitive and portable miniaturized SERF atomic magnetometers.

IEEE Transactions on Instrumentation and Measurement•Jan 1, 2024•18 citations

Ultrasensitive SERF atomic magnetometer with a miniaturized hybrid vapor cell

Yintao Ma, Yao Chen, Mingzhi Yu +8 more

Abstract The chip-scale hybrid optical pumping spin-exchange relaxation-free (SERF) atomic magnetometer with a single-beam arrangement has prominent applications in biomagnetic measurements because of its outstanding features, including ultrahigh sensitivity, an enhanced signal-to-noise ratio, homogeneous spin polarization and a much simpler optical configuration than other devices. In this work, a miniaturized single-beam hybrid optical pumping SERF atomic magnetometer based on a microfabricated atomic vapor cell is demonstrated. Although the optically thin Cs atoms are spin-polarized, the dense Rb atoms determine the experimental results. The enhanced signal strength and narrowed resonance linewidth are experimentally proven, which shows the superiority of the proposed magnetometer scheme. By using a differential detection scheme, we effectively suppress optical noise with an approximate five-fold improvement. Moreover, the cell temperature markedly affects the performance of the magnetometer. We systematically investigate the effects of temperature on the magnetometer parameters. The theoretical basis for these effects is explained in detail. The developed miniaturized magnetometer has an optimal magnetic sensitivity of 20 fT/Hz 1/2 . The presented work provides a foundation for the chip-scale integration of ultrahighly sensitive quantum magnetometers that can be used for forward-looking magnetocardiography (MCG) and magnetoencephalography (MEG) applications.

Microsystems & Nanoengineering•Aug 30, 2024•62 citations

Development of spin-exchange relaxation free magnetometer with a compact heating system

Hyun Joon Lee, Kiwoong Kim, Seong-Joo Lee +4 more

Atomic magnetometers based on spin-exchange relaxation free (SERF) regime typically operate with the high optical density and low magnetic fields. We present SERF magnetometer with a compact heating system for compact and efficient temperature control. A resistive heating system with high frequency alternating current was used to achieve high optical density. The current frequency was carefully chosen to avoid resonance with the atomic magnetometer. This resulted in a compact heating system that also allows independent operation of multiple heating systems. Optical rotation signal was obtained at 200 oC and width was about 3 nT with pump laser power of 55 mW. Bandwidth of the magnetometer with the compact heater was 40 Hz. Consequently, the compact resistive heating system proved to be a good alternative to the air-flow heating system. Additionally, residual field effect and signal-to-noise yielded a sensitivity of 50 fT/Hz1/2 at 10 Hz.

ACTA IMEKO•Aug 16, 2013•3 citations

Measurement of Relaxation Mechanism for the Spin-Exchange Relaxation-Free Magnetometer

Jin Peng, An‐Ning Xu, Yiran Yin +1 more

The spin-exchange relaxation-free (SERF) atomic magnetometer (AM) has vital importance in advanced research and has been widely applied in different fields. To improve the sensitivity and robustness of SERF magnetometers, understanding the relaxation mechanism is a significant approach. By controlling the amplitude and direction of external magnetic fields, and switching statuses of the pump and probe laser beam, we measured both the transverse and longitudinal relaxation rate in two experimental methods, respectively, and obtained the Larmor precession frequency, the spin-exchange relaxation rate, the spin-destruction relaxation rate, and the relaxation rate due to collisions between atoms and vapor cell walls. All experimental results are verified with the theoretical calculations. The final longitudinal relaxation rate of our magnetometer is 53 Hz. Furthermore, we measured the sensitivity of our SERF magnetometer reaching 5 fT/Hz1/2 from 5 to 70 Hz after differential measurement with the 3-dB bandwidth of 70 Hz. Our work helps to improve the detection capability of SERF magnetometers, providing performance assurance for searching for exotic spin-dependent interactions and applications, including magnetoencephalograms and magnetocardiograms.

IEEE Sensors Journal•Dec 7, 2023•11 citations
Spin-exchange relaxation-free (SERF) atomic magnetometers have emerged as highly promising candidates for ultra-weak magnetic field detection, particularly in biomagnetic imaging, owing to their exceptional sensitivity, amenability to miniaturization, and near-room-temperature operation. While current miniaturized magnetometers typically employ laser chips with fixed optical power, the quantitative impact of laser power on critical performance metrics remains to be fully elucidated. This study systematically investigates the influence of laser power on sensitivity, bandwidth, and dynamic range by incorporating considerations of power broadening, saturation absorption, and noise constraints. A miniaturized probe, integrated with an actively controlled vertical-cavity surface-emitting laser (VCSEL), was developed for experimental validation. Theoretical and experimental results consistently demonstrate that as optical power increases, sensitivity exhibits a non-monotonic dependence, whereas both bandwidth and dynamic range manifest a monotonic upward trend, aligning well with theoretical simulations. The optimized sensor achieved a peak sensitivity of 16 fT/√Hz at 300 μW, while the bandwidth and dynamic range reached 230 Hz and ±5.4 nT at 500 μW, respectively. This work establishes a robust theoretical and experimental framework for the comprehensive performance optimization of laser-integrated miniaturized atomic magnetometers.
Spin-exchange relaxation-free (SERF) atomic magnetometers are the promising non-cryogenic sensors to potentially replace the superconducting quantum interference device (SQUID) in the ultra-weak magnetic-field measurements. Although the SERF magnetometers have achieved a sensitivity of sub-femtotesla, the bandwidth is limited at a narrow range. In this work, we developed a closed-loop SERF magnetometer under proportional-integral (PI) control, the bandwidth could be adaptively tuned and extended above 1.2 kHz. A nearly 40-fold enhancement of the typical bandwidth was achieved in comparison with the open-loop system and the sensitivity was maintained at 15 fT/Hz1/2from 30 Hz to 750 Hz. Moreover, the magnetic field dynamic range could also be greatly improved. This proposed atomic magnetometer with higher bandwidth, sensitivity, and dynamic range has a good prospect in developing a compact, portable, and non-cryogenic magnetic field sensor and may find more applications in biomedical, nuclear magnetic resonance, and fundamental physics research.
Emerging quantum sensing devices namely atomic magnetometers have become promising candidate for future biomagnetism imaging. However, the limitation of detection bandwidth exists a major drawback for devices of this kind. Conventional approaches, including transverse magnetic field modulation and close-loop control, have seen challenging tradeoffs between size, resolution and bandwidths. In this study, a wide bandwidth zero-field magnetometer with a miniaturized light source has been developed for alternating current (AC) magnetic field detection. A dual-beam Spin-Exchange Relaxation-Free (SERF) magnetometer employing a 4×4×4mm3 alkali metal vapor cell achieves a bandwidth of 188 Hz. This represents a threefold increase in bandwidth compared to non-modulated mode. The sensitivity under modulated mode demonstrated a sensitivity of 55fT/Hz1/2within the frequency range of 10-188Hz. Our results also demonstrated that the sensitivity is constrained by relaxation induced by the modulation of pumping source, so possible tradeoffs can be made based on various applications. The proposed method provides novel approach for bandwidth extension of atomic magnetometers which further pave the way for future applications including magnetoencephalography (MEG), particularly in reducing magnetic noise crosstalk and enhancing spatial resolution.
Miniaturized high-sensitivity atomic magnetometers play a crucial role in biomagnetic measurements. The utilization of multipass cells enhances the sensitivity of atomic magnetometers. This study explores the benefits of employing a multipass cell in spin-exchange relaxation-free (SERF) atomic magnetometers through theoretical and experimental analysis. A signal-to-noise ratio (SNR) model was established by accounting for the optimal detuning frequency of the probe light and low-frequency 1/f noise. This enabled the optimal number of probe light passes through the vapor cell to be determined. Additionally, compact and symmetric reflective optical paths were designed to mitigate polarization gradients in atomic magnetometers with high atomic number density. Experimental results confirmed that the magnetic field sensitivity of the triple-pass configuration (1.8 fT/Hz1/2^{1/2}1/2) was 1.89 times that of the single-pass configuration (3.4 fT/Hz1/2^{1/2}1/2), consistent with the theoretical analysis. This study provides valuable theoretical guidance for analyzing the SNR performance of SERF atomic magnetometers and demonstrates the potential of multipass cells in enhancing their magnetic field sensitivity. The utilization of multipass cells presents a viable approach toward the realization of highly sensitive and portable miniaturized SERF atomic magnetometers.
Abstract The chip-scale hybrid optical pumping spin-exchange relaxation-free (SERF) atomic magnetometer with a single-beam arrangement has prominent applications in biomagnetic measurements because of its outstanding features, including ultrahigh sensitivity, an enhanced signal-to-noise ratio, homogeneous spin polarization and a much simpler optical configuration than other devices. In this work, a miniaturized single-beam hybrid optical pumping SERF atomic magnetometer based on a microfabricated atomic vapor cell is demonstrated. Although the optically thin Cs atoms are spin-polarized, the dense Rb atoms determine the experimental results. The enhanced signal strength and narrowed resonance linewidth are experimentally proven, which shows the superiority of the proposed magnetometer scheme. By using a differential detection scheme, we effectively suppress optical noise with an approximate five-fold improvement. Moreover, the cell temperature markedly affects the performance of the magnetometer. We systematically investigate the effects of temperature on the magnetometer parameters. The theoretical basis for these effects is explained in detail. The developed miniaturized magnetometer has an optimal magnetic sensitivity of 20 fT/Hz 1/2 . The presented work provides a foundation for the chip-scale integration of ultrahighly sensitive quantum magnetometers that can be used for forward-looking magnetocardiography (MCG) and magnetoencephalography (MEG) applications.
Atomic magnetometers based on spin-exchange relaxation free (SERF) regime typically operate with the high optical density and low magnetic fields. We present SERF magnetometer with a compact heating system for compact and efficient temperature control. A resistive heating system with high frequency alternating current was used to achieve high optical density. The current frequency was carefully chosen to avoid resonance with the atomic magnetometer. This resulted in a compact heating system that also allows independent operation of multiple heating systems. Optical rotation signal was obtained at 200 oC and width was about 3 nT with pump laser power of 55 mW. Bandwidth of the magnetometer with the compact heater was 40 Hz. Consequently, the compact resistive heating system proved to be a good alternative to the air-flow heating system. Additionally, residual field effect and signal-to-noise yielded a sensitivity of 50 fT/Hz1/2 at 10 Hz.
The spin-exchange relaxation-free (SERF) atomic magnetometer (AM) has vital importance in advanced research and has been widely applied in different fields. To improve the sensitivity and robustness of SERF magnetometers, understanding the relaxation mechanism is a significant approach. By controlling the amplitude and direction of external magnetic fields, and switching statuses of the pump and probe laser beam, we measured both the transverse and longitudinal relaxation rate in two experimental methods, respectively, and obtained the Larmor precession frequency, the spin-exchange relaxation rate, the spin-destruction relaxation rate, and the relaxation rate due to collisions between atoms and vapor cell walls. All experimental results are verified with the theoretical calculations. The final longitudinal relaxation rate of our magnetometer is 53 Hz. Furthermore, we measured the sensitivity of our SERF magnetometer reaching 5 fT/Hz1/2 from 5 to 70 Hz after differential measurement with the 3-dB bandwidth of 70 Hz. Our work helps to improve the detection capability of SERF magnetometers, providing performance assurance for searching for exotic spin-dependent interactions and applications, including magnetoencephalograms and magnetocardiograms.