Design and Readout of Large MKID Arrays for Submillimeter Astronomy

James A. Schlaerth, Omid Noroozian, Nicole G. Czakon, Peter K. Day, T. P. Downes, Ran P. Duan, J. Gao, J. Glenn, S. R. Golwala, Matthew I. Hollister, Henry George LeDuc, P. R. Maloney, Ben Mazin, H. T. Nguyen, Jack Sayers, Seth R. Siegel, John E. Vaillancourt, Anastasios Vayonakis, Jonas Stasys Zmuidzinas

Published Mar 1, 2010• 0 citations
OpenAlex

Abstract

Microwave Kinetic Inductance Detectors (MKIDs) are superconducting microresonators useful as detectors for sub/millimeter and far-infrared cameras, as well as instruments in optical and x-ray wavelengths. A probe sine wave at the resonance frequency can detect a change in frequency and microwave loss due to submillimeter optical load as a phase or amplitude change. These detectors may be coupled to light through on-chip, phased-array antennas. A single antenna may feed several detectors with different bandpasses defined by on-chip lumped-element filters. The resonators are easily frequency multiplexed by tuning the resonance frequencies, so hundreds of detectors can be read out while coupled to a single feedline. The deployment and operation of ground-based, background-limited detectors in close-packed arrays present significant challenges in both readout strategy and array design. The initial designs consisted of simple coplanar waveguide microwave resonators with aluminum absorptive sections. The first change has been to substitute the capacitive portion of the resonator with a large interdigitated capacitor (IDC) to reduce its noise contribution. While their low-noise characteristics have been demonstrated, these detectors had not yet been tested on array-scale devices. We present results from arrays of detectors designed for sub/millimeter imaging, tested for their optical response characteristics. In particular, we address challenges of microwave inter-resonator coupling, out-of-band resonator optical sensitivity, and noise reduction in the latest device designs. We show the results of stepped-impedance filters in preventing direct pickup of wide-band radiation, and new frequency spacing to limit crosstalk between resonators. We show direct comparison between olderdesign coplanar waveguide resonators and new IDC resonators under optical load, demonstrating the noise reduction and comparable responsivity. In addition, we show results of the probe signal readout power that maximizes signal-to-noise.

Related papers

Superconducting Microwave Resonator Arrays for Submillimeter/Far-infrared Imaging

Omid Noroozian

Superconducting microwave resonators have the potential to revolutionize submillimeter and far-infrared astronomy, and with it our understanding of the universe. The field of low-temperature detector technology has reached a point where extremely sensitive devices like transition-edge sensors are now capable of detecting radiation limited by the background noise of the universe. However, the size of these detector arrays are limited to only a few thousand pixels. This is because of the cost and complexity of fabricating large-scale arrays of these detectors that can reach up to 10 lithographic levels on chip, and the complicated SQUID-based multiplexing circuitry and wiring for readout of each detector. In order to make substantial progress, next- generation ground-based telescopes such as CCAT or future space telescopes require focal planes with large-scale detector arrays of 104–106 pixels. Arrays using microwave kinetic inductance detectors (MKID) are a potential solution. These arrays can be easily made with a single layer of superconducting metal film deposited on a silicon substrate and pattered using conventional optical lithography. Furthermore, MKIDs are inherently multiplexable in the frequency domain, allowing ∼ 103 detectors to be read out using a single coaxial transmission line and cryogenic amplifier, drastically reducing cost and complexity. An MKID uses the change in the microwave surface impedance of a superconducting thin-film microresonator to detect photons. Absorption of photons in the superconductor breaks Cooper pairs into quasiparticles, changing the complex surface impedance, which results in a perturbation of resonator frequency and quality factor. For excitation and readout, the resonator is weakly coupled to a transmission line. The complex amplitude of a microwave probe signal tuned on-resonance and transmitted on the feedline past the resonator is perturbed as photons are absorbed in the superconductor. The perturbation can be detected using a cryogenic amplifier and subsequent homodyne mixing at room temperature. In an array of MKIDs, all the resonators are coupled to a shared feedline and are tuned to slightly different frequencies. They can be read out simultaneously using a comb of frequencies generated and measured using digital techniques. This thesis documents an effort to demonstrate the basic operation of ∼ 256 pixel arrays of lumped-element MKIDs made from superconducting TiNx on silicon. The resonators are designed and simulated for optimum operation. Various properties of the resonators and arrays are measured and compared to theoretical expectations. A particularly exciting observation is the extremely high quality factors (∼ 3 × 107) of our TiNx resonators which is essential for ultra-high sensitivity. The arrays are tightly packed both in space and in frequency which is desirable for larger full-size arrays. However, this can cause a serious problem in terms of microwave crosstalk between neighboring pixels. We show that by properly designing the resonator geometry, crosstalk can be eliminated; this is supported by our measurement results. We also tackle the problem of excess frequency noise in MKIDs. Intrinsic noise in the form of an excess resonance frequency jitter exists in planar superconducting resonators that are made on dielectric substrates. We conclusively show that this noise is due to fluctuations of the resonator capacitance. In turn, the capacitance fluctuations are thought to be driven by two-level system (TLS) fluctuators in a thin layer on the surface of the device. With a modified resonator design we demonstrate with measurements that this noise can be substantially reduced. An optimized version of this resonator was designed for the multiwavelength submillimeter kinetic inductance camera (MUSIC) instrument for the Caltech Submillimeter Observatory.

•Jun 12, 2012•11 citations

Antenna-coupled Microwave Kinetic Inductance detectors (MKIDs) for mm and submm imaging arrays

Anastasios Vayonakis, James A. Schlaerth, S. Kumar +9 more

We present results from completely lithographic antenna-coupled Microwave Kinetic Inductance detectors (MKIDs). MKIDs are superconducting resonators with resonant frequency and quality-factor which are highly sensitive to changes in the density of the quasiparticle population which occurs when photons above the superconducting gap energy are absorbed. The resonators are coupled to submm light through on-chip phased-array slot antennas. Each planar antenna consists of an array of N long slots which are fed along their length at M points. The resulting N*M feed points are combined in-phase using a binary summing tree made of low-loss superconducting microstrip lines. Due to its large size, the resulting planar antenna produces a narrow beam pattern and can therefore be used without additional optical coupling elements such as feedhorns or substrate lenses. The output of the antenna is a single superconducting thin-film microstrip line which can be efficiently coupled to one (or more) kinetic inductance coplanar waveguide resonators to produce a single (or multi-color) pixel in an imaging focal plane array, using in-line lumped element lithographic band-pass filters. Such highly integrated architectures can be easily fabricated on a single substrate, and many detectors can be frequency multiplexed through coupling to a single feedline. Microwave readout provides a lot of bandwidth per detector, allowing a large number of pixels to be read using a single cryogenic microwave amplifier and warm readout electronics.

Softwaretechnik-Trends•Apr 1, 2008•1 citation

Performance of hybrid NbTiN-Al microwave kinetic inductance detectors as direct detectors for sub-millimeter astronomy

R. M. J. Janssen, J. J. A. Baselmans, Akira Endo +4 more

In the next decades millimeter and sub-mm astronomy requires large format imaging arrays and broad-band spectrometers to complement the high spatial and spectral resolution of the Atacama Large Millimeter/submillimeter Array. The desired sensors for these instruments should have a background limited sensitivity and a high optical efficiency and enable arrays thousands of pixels in size. Hybrid microwave kinetic inductance detectors consisting of NbTiN and Al have shown to satisfy these requirements. We present the second generation hybrid NbTiN-Al MKIDs, which are photon noise limited in both phase and amplitude readout for loading levels P850GHz < 10 fW. Thanks to the increased responsivity, the photon noise level achieved in phase allows us to simultaneously read out approximately 8000 pixels using state-of-the-art electronics. In addition, the choice of superconducting materials and the use of a Si lens in combination with a planar antenna gives these resonators the flexibility to operate within the frequency range 0:09 < v < 1:1 THz. Given these specifications, hybrid NbTiN-Al MKIDs will enable astronomically usable kilopixel arrays for sub-mm imaging and moderate resolution spectroscopy.

arXiv:1408.3314•Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE•Jul 23, 2014•14 citations

Optimization of MKID noise performance via readout technique for astronomical applications

Nicole G. Czakon, James A. Schlaerth, Peter K. Day +17 more

Detectors employing superconducting microwave kinetic inductance detectors (MKIDs) can be read out by measuring changes in either the resonator frequency or dissipation. We will discuss the pros and cons of both methods, in particular, the readout method strategies being explored for the Multiwavelength Sub/millimeter Inductance Camera (MUSIC) to be commissioned at the CSO in 2010. As predicted theoretically and observed experimentally, the frequency responsivity is larger than the dissipation responsivity, by a factor of 2-4 under typical conditions. In the absence of any other noise contributions, it should be easier to overcome amplifier noise by simply using frequency readout. The resonators, however, exhibit excess frequency noise which has been ascribed to a surface distribution of two-level fluctuators sensitive to specific device geometries and fabrication techniques. Impressive dark noise performance has been achieved using modified resonator geometries employing interdigitated capacitors (IDCs). To date, our noise measurement and modeling efforts have assumed an onresonance readout, with the carrier power set well below the nonlinear regime. Several experimental indicators suggested to us that the optimal readout technique may in fact require a higher readout power, with the carrier tuned somewhat off resonance, and that a careful systematic study of the optimal readout conditions was needed. We will present the results of such a study, and discuss the optimum readout conditions as well as the performance that can be achieved relative to BLIP.

Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE•Jul 8, 2010•5 citations

Crosstalk Reduction for Superconducting Microwave Resonator Arrays

Omid Noroozian, Peter K. Day, Byeong Ho Eom +2 more

Large-scale arrays of microwave kinetic inductance detectors (MKIDs) are attractive candidates for use in imaging instruments for next generation submillimeter-wave telescopes such as CCAT. We have designed and fabricated tightly packed ~250-pixel MKID arrays using lumped-element resonators etched from a thin layer of superconducting TiNxdeposited on a silicon substrate. The high pixel packing density in our initial design resulted in large microwave crosstalk due to electromagnetic coupling between the resonators. Our second design eliminates this problem by adding a grounding shield and using a double-wound geometry for the meander inductor to allow conductors with opposite polarity to be in close proximity. In addition, the resonator frequencies are distributed in a checkerboard pattern across the array. We present details for the two resonator and array designs and describe a circuit model for the full array that predicts the distribution of resonator frequencies and the crosstalk level. We also show results from a new experimental technique that conveniently measures crosstalk without the need for an optical setup. Our results reveal an improvement in crosstalk from 57% in the initial design down to ≤2% in the second design. The general procedure and design guidelines in this work are applicable to future large arrays employing microwave resonators.

arXiv:1206.5571•IEEE Transactions on Microwave Theory and Techniques•Mar 8, 2012•81 citations

Submicron Kinetic Inductance Detectors for SAFARI: Improving MKID sensitivity through width reduction

R. M. J. Janssen

Spectroscopic observations of a large number of galaxies in the far-infrared are critical in the study of galaxy evolution. These observations are difficult at the moment, because of the lack of good far-infrared (FIR) detector technology. Microwave Kinetic Inductance Detectors (MKIDs) are showing promising results to fill this void. A common MKID design is the antenna-coupled coplanar waveguide (CPW) resonator patterned in a superconducting film. These resonators are extremely sensitive to changes in the Cooper pair density. FIR photons are energetic enough to break Cooper pairs. This means a superconducting resonator can be used as a pixel for a FIR camera. The inherent multiplexing advantage of MKIDs means they can easily be fabricated in arrays of up to 5000 pixels. By putting such an array in the SAFARI instrument on the cooled SPICA telescope a blind spectroscopic survey that can detect Milky Way sized galaxies out to redshift z = 2 will be possible. However, to achieve background limited performance on a cooled telescope like SPICA the sensivity of MKIDs needs to be improved by at least an order of magnitude. This project aims to investigate if this can be achieved by reducing the width of the CPW to much below what has typically been made using optical lithography (> 1 ?m). CPW resonators with a central line width as narrow as 300 nm were made in NbTiN using electron beam lithography and reactive ion etching. Using this fabrication method NbTiN resonators with a central line width varying between 0.3 ?m and 3 ?m were made. These resonators were used to do a systematic study of the width dependence of the kinetic inductance fraction, quality factor, responsivity, read-out power, noise and NEP. It is shown that the width dependence of all these properties does not change down to 300 nm. Encouraged by these results two prototype submicron pixel arrays were made in aluminum. These showed sensitivities comparable to resonators with a few micron wide CPW. The main reason for this lack of sensitivity improvement is the lower maximum read-out power of submicron wide aluminum resonators. Based on these results power handling seems a key issue that will determine if submicron resonators are a viable route to improve the MKIDs sensitivity. These results of this project indicate that aluminum has a different width dependence of its power handling compared to NbTiN. Therefore, a systematic study of submicron resonators in aluminum, which focuses on the power handling, is required.

•Dec 10, 2010•1 citation
Microwave Kinetic Inductance Detectors (MKIDs) are superconducting microresonators useful as detectors for sub/millimeter and far-infrared cameras, as well as instruments in optical and x-ray wavelengths. A probe sine wave at the resonance frequency can detect a change in frequency and microwave loss due to submillimeter optical load as a phase or amplitude change. These detectors may be coupled to light through on-chip, phased-array antennas. A single antenna may feed several detectors with different bandpasses defined by on-chip lumped-element filters. The resonators are easily frequency multiplexed by tuning the resonance frequencies, so hundreds of detectors can be read out while coupled to a single feedline. The deployment and operation of ground-based, background-limited detectors in close-packed arrays present significant challenges in both readout strategy and array design. The initial designs consisted of simple coplanar waveguide microwave resonators with aluminum absorptive sections. The first change has been to substitute the capacitive portion of the resonator with a large interdigitated capacitor (IDC) to reduce its noise contribution. While their low-noise characteristics have been demonstrated, these detectors had not yet been tested on array-scale devices. We present results from arrays of detectors designed for sub/millimeter imaging, tested for their optical response characteristics. In particular, we address challenges of microwave inter-resonator coupling, out-of-band resonator optical sensitivity, and noise reduction in the latest device designs. We show the results of stepped-impedance filters in preventing direct pickup of wide-band radiation, and new frequency spacing to limit crosstalk between resonators. We show direct comparison between olderdesign coplanar waveguide resonators and new IDC resonators under optical load, demonstrating the noise reduction and comparable responsivity. In addition, we show results of the probe signal readout power that maximizes signal-to-noise.
Superconducting microwave resonators have the potential to revolutionize submillimeter and far-infrared astronomy, and with it our understanding of the universe. The field of low-temperature detector technology has reached a point where extremely sensitive devices like transition-edge sensors are now capable of detecting radiation limited by the background noise of the universe. However, the size of these detector arrays are limited to only a few thousand pixels. This is because of the cost and complexity of fabricating large-scale arrays of these detectors that can reach up to 10 lithographic levels on chip, and the complicated SQUID-based multiplexing circuitry and wiring for readout of each detector. In order to make substantial progress, next- generation ground-based telescopes such as CCAT or future space telescopes require focal planes with large-scale detector arrays of 104–106 pixels. Arrays using microwave kinetic inductance detectors (MKID) are a potential solution. These arrays can be easily made with a single layer of superconducting metal film deposited on a silicon substrate and pattered using conventional optical lithography. Furthermore, MKIDs are inherently multiplexable in the frequency domain, allowing ∼ 103 detectors to be read out using a single coaxial transmission line and cryogenic amplifier, drastically reducing cost and complexity. An MKID uses the change in the microwave surface impedance of a superconducting thin-film microresonator to detect photons. Absorption of photons in the superconductor breaks Cooper pairs into quasiparticles, changing the complex surface impedance, which results in a perturbation of resonator frequency and quality factor. For excitation and readout, the resonator is weakly coupled to a transmission line. The complex amplitude of a microwave probe signal tuned on-resonance and transmitted on the feedline past the resonator is perturbed as photons are absorbed in the superconductor. The perturbation can be detected using a cryogenic amplifier and subsequent homodyne mixing at room temperature. In an array of MKIDs, all the resonators are coupled to a shared feedline and are tuned to slightly different frequencies. They can be read out simultaneously using a comb of frequencies generated and measured using digital techniques. This thesis documents an effort to demonstrate the basic operation of ∼ 256 pixel arrays of lumped-element MKIDs made from superconducting TiNx on silicon. The resonators are designed and simulated for optimum operation. Various properties of the resonators and arrays are measured and compared to theoretical expectations. A particularly exciting observation is the extremely high quality factors (∼ 3 × 107) of our TiNx resonators which is essential for ultra-high sensitivity. The arrays are tightly packed both in space and in frequency which is desirable for larger full-size arrays. However, this can cause a serious problem in terms of microwave crosstalk between neighboring pixels. We show that by properly designing the resonator geometry, crosstalk can be eliminated; this is supported by our measurement results. We also tackle the problem of excess frequency noise in MKIDs. Intrinsic noise in the form of an excess resonance frequency jitter exists in planar superconducting resonators that are made on dielectric substrates. We conclusively show that this noise is due to fluctuations of the resonator capacitance. In turn, the capacitance fluctuations are thought to be driven by two-level system (TLS) fluctuators in a thin layer on the surface of the device. With a modified resonator design we demonstrate with measurements that this noise can be substantially reduced. An optimized version of this resonator was designed for the multiwavelength submillimeter kinetic inductance camera (MUSIC) instrument for the Caltech Submillimeter Observatory.
We present results from completely lithographic antenna-coupled Microwave Kinetic Inductance detectors (MKIDs). MKIDs are superconducting resonators with resonant frequency and quality-factor which are highly sensitive to changes in the density of the quasiparticle population which occurs when photons above the superconducting gap energy are absorbed. The resonators are coupled to submm light through on-chip phased-array slot antennas. Each planar antenna consists of an array of N long slots which are fed along their length at M points. The resulting N*M feed points are combined in-phase using a binary summing tree made of low-loss superconducting microstrip lines. Due to its large size, the resulting planar antenna produces a narrow beam pattern and can therefore be used without additional optical coupling elements such as feedhorns or substrate lenses. The output of the antenna is a single superconducting thin-film microstrip line which can be efficiently coupled to one (or more) kinetic inductance coplanar waveguide resonators to produce a single (or multi-color) pixel in an imaging focal plane array, using in-line lumped element lithographic band-pass filters. Such highly integrated architectures can be easily fabricated on a single substrate, and many detectors can be frequency multiplexed through coupling to a single feedline. Microwave readout provides a lot of bandwidth per detector, allowing a large number of pixels to be read using a single cryogenic microwave amplifier and warm readout electronics.
In the next decades millimeter and sub-mm astronomy requires large format imaging arrays and broad-band spectrometers to complement the high spatial and spectral resolution of the Atacama Large Millimeter/submillimeter Array. The desired sensors for these instruments should have a background limited sensitivity and a high optical efficiency and enable arrays thousands of pixels in size. Hybrid microwave kinetic inductance detectors consisting of NbTiN and Al have shown to satisfy these requirements. We present the second generation hybrid NbTiN-Al MKIDs, which are photon noise limited in both phase and amplitude readout for loading levels P850GHz < 10 fW. Thanks to the increased responsivity, the photon noise level achieved in phase allows us to simultaneously read out approximately 8000 pixels using state-of-the-art electronics. In addition, the choice of superconducting materials and the use of a Si lens in combination with a planar antenna gives these resonators the flexibility to operate within the frequency range 0:09 < v < 1:1 THz. Given these specifications, hybrid NbTiN-Al MKIDs will enable astronomically usable kilopixel arrays for sub-mm imaging and moderate resolution spectroscopy.
Detectors employing superconducting microwave kinetic inductance detectors (MKIDs) can be read out by measuring changes in either the resonator frequency or dissipation. We will discuss the pros and cons of both methods, in particular, the readout method strategies being explored for the Multiwavelength Sub/millimeter Inductance Camera (MUSIC) to be commissioned at the CSO in 2010. As predicted theoretically and observed experimentally, the frequency responsivity is larger than the dissipation responsivity, by a factor of 2-4 under typical conditions. In the absence of any other noise contributions, it should be easier to overcome amplifier noise by simply using frequency readout. The resonators, however, exhibit excess frequency noise which has been ascribed to a surface distribution of two-level fluctuators sensitive to specific device geometries and fabrication techniques. Impressive dark noise performance has been achieved using modified resonator geometries employing interdigitated capacitors (IDCs). To date, our noise measurement and modeling efforts have assumed an onresonance readout, with the carrier power set well below the nonlinear regime. Several experimental indicators suggested to us that the optimal readout technique may in fact require a higher readout power, with the carrier tuned somewhat off resonance, and that a careful systematic study of the optimal readout conditions was needed. We will present the results of such a study, and discuss the optimum readout conditions as well as the performance that can be achieved relative to BLIP.
Large-scale arrays of microwave kinetic inductance detectors (MKIDs) are attractive candidates for use in imaging instruments for next generation submillimeter-wave telescopes such as CCAT. We have designed and fabricated tightly packed ~250-pixel MKID arrays using lumped-element resonators etched from a thin layer of superconducting TiNxdeposited on a silicon substrate. The high pixel packing density in our initial design resulted in large microwave crosstalk due to electromagnetic coupling between the resonators. Our second design eliminates this problem by adding a grounding shield and using a double-wound geometry for the meander inductor to allow conductors with opposite polarity to be in close proximity. In addition, the resonator frequencies are distributed in a checkerboard pattern across the array. We present details for the two resonator and array designs and describe a circuit model for the full array that predicts the distribution of resonator frequencies and the crosstalk level. We also show results from a new experimental technique that conveniently measures crosstalk without the need for an optical setup. Our results reveal an improvement in crosstalk from 57% in the initial design down to ≤2% in the second design. The general procedure and design guidelines in this work are applicable to future large arrays employing microwave resonators.
Spectroscopic observations of a large number of galaxies in the far-infrared are critical in the study of galaxy evolution. These observations are difficult at the moment, because of the lack of good far-infrared (FIR) detector technology. Microwave Kinetic Inductance Detectors (MKIDs) are showing promising results to fill this void. A common MKID design is the antenna-coupled coplanar waveguide (CPW) resonator patterned in a superconducting film. These resonators are extremely sensitive to changes in the Cooper pair density. FIR photons are energetic enough to break Cooper pairs. This means a superconducting resonator can be used as a pixel for a FIR camera. The inherent multiplexing advantage of MKIDs means they can easily be fabricated in arrays of up to 5000 pixels. By putting such an array in the SAFARI instrument on the cooled SPICA telescope a blind spectroscopic survey that can detect Milky Way sized galaxies out to redshift z = 2 will be possible. However, to achieve background limited performance on a cooled telescope like SPICA the sensivity of MKIDs needs to be improved by at least an order of magnitude. This project aims to investigate if this can be achieved by reducing the width of the CPW to much below what has typically been made using optical lithography (> 1 ?m). CPW resonators with a central line width as narrow as 300 nm were made in NbTiN using electron beam lithography and reactive ion etching. Using this fabrication method NbTiN resonators with a central line width varying between 0.3 ?m and 3 ?m were made. These resonators were used to do a systematic study of the width dependence of the kinetic inductance fraction, quality factor, responsivity, read-out power, noise and NEP. It is shown that the width dependence of all these properties does not change down to 300 nm. Encouraged by these results two prototype submicron pixel arrays were made in aluminum. These showed sensitivities comparable to resonators with a few micron wide CPW. The main reason for this lack of sensitivity improvement is the lower maximum read-out power of submicron wide aluminum resonators. Based on these results power handling seems a key issue that will determine if submicron resonators are a viable route to improve the MKIDs sensitivity. These results of this project indicate that aluminum has a different width dependence of its power handling compared to NbTiN. Therefore, a systematic study of submicron resonators in aluminum, which focuses on the power handling, is required.