Journal of Physics Conference Series
Sensitivity of RF atomic magnetometer affected by the gain of balanced photodetector
Kwanruthai Sibsiri, Piyachat Wungmool, Phusit Nualpijit, Puchong Kijamnajsuk, Jarin Osaklung, Malee Sutthiopad, Chaiya Luengviriya
Abstract
Abstract An atomic magnetometer (AM) is very sensitive to magnetic fields and useful in many applications. AM works by optical interrogation of orientation of alkali-atom spins in a vapor cell. Since the sensitivity characterizes the performance of AM, it should be optimized. We investigate the effect of the gain of a balanced photodetector on the sensitivity of an RF atomic magnetometer setup. While the amplitude of RF signal (BRF) varies between 0.1-10 nT, the frequency is fixed at 100 kHz. The gain of the balanced photodetector varies between 103 - 107 V/A. the noise density and AM signal are measured to calculate the signal-to-noise ratios (SNR) and the AM sensitivity. The results show that noise density increases with the gain value. The AM signal increases with both BRF as well as the gain value. SNR increases with BRF while the AM sensitivity is independent on BRF. When the gain is increased from 103 - 105 V/A, SNR increases and reaches a maximum while the AM sensitivity decreases and reaches a minimum value of 0.4 pT/√Hz. However, SNR and the AM sensitivity are almost unchanged for the gain value of 105 - 107 V/A. Therefore, the gain of balanced photodetector in our AM setup should be set sufficiently high, at least 105 V/A.