In spin-exchange-relaxation-free (SERF) atomic magnetometers (AMs), the attenuation of circularly polarized pump beam introduces spatial-dependent artificial magnetic fields in atomic spin polarization and light shift. In general, such a spatial-dependent effective magnetic field, especially prevalent in single-cell magnetometer arrays, cannot be uniformly compensated over the entire excitation region, resulting in a complex distributed compensation mask effect. In addition, any external effective-field-compensation methods inevitably complicate the final analysis of the unknown magnetic field measured. Here, we report a new excitation scheme that can substantially suppress spatial nonuniformity due to pump beam attenuation. Using counter-propagating pump beams with symmetrically red- and blue-detuning-paired frequencies, we demonstrate a significant reduction of spatial inhomogeneity in atomic spin polarization and light shift in comparison with the usual single-frequency unidirection and counter-propagating pump schemes. Employing this red-/blue-detuning-paired pump scheme, we construct a single-cell, four-channel SERF magnetometer array with high spatial homogeneity. We demonstrate ~10-fT/Hz
1/2sensitivity in the 5–45-Hz band for all four channels, with a gradient sensitivity of 2 fT/Hz1/2
1/2/cm in the same band. This new scheme enables the development of single-vapor-cell, high-uniformity multichannel magnetic field mapping devices with potential applications in magnetoencephalography (MEG) and magnetocardiography (MCG), or in searching for exotic spin-dependent interactions.
Abstract The chip-scale hybrid optical pumping spin-exchange relaxation-free (SERF) atomic magnetometer with a single-beam arrangement has prominent applications in biomagnetic measurements because of its outstanding features, including ultrahigh sensitivity, an enhanced signal-to-noise ratio, homogeneous spin polarization and a much simpler optical configuration than other devices. In this work, a miniaturized single-beam hybrid optical pumping SERF atomic magnetometer based on a microfabricated atomic vapor cell is demonstrated. Although the optically thin Cs atoms are spin-polarized, the dense Rb atoms determine the experimental results. The enhanced signal strength and narrowed resonance linewidth are experimentally proven, which shows the superiority of the proposed magnetometer scheme. By using a differential detection scheme, we effectively suppress optical noise with an approximate five-fold improvement. Moreover, the cell temperature markedly affects the performance of the magnetometer. We systematically investigate the effects of temperature on the magnetometer parameters. The theoretical basis for these effects is explained in detail. The developed miniaturized magnetometer has an optimal magnetic sensitivity of 20 fT/Hz 1/2 . The presented work provides a foundation for the chip-scale integration of ultrahighly sensitive quantum magnetometers that can be used for forward-looking magnetocardiography (MCG) and magnetoencephalography (MEG) applications.
The ultrasensitive magnetometer has a vital importance in fundamental research and applications. Currently, the spin-exchange relaxation-free (SERF) atomic magnetometer has been reported with a sensitivity around the level of fT/Hz 1/2 . To enhance the sensitivity, a gradiometer configuration has usually been introduced to cancel the common-mode noise between two separate channels. However, the signal and response from different channels are not the same due to the attenuation of the pump beam. Here, we proposed a counter-propagating optical sideband pumping method to polarize the atoms, using the electro-optic modulator to modulate the single-pump beam, generating two symmetrically red- and blue-detuned sidebands of frequency. This scheme leads to a significant reduction of undesirable effects coming along with the optical pumping, such as light shifts and spatial inhomogeneity in atomic spin polarization. With the help of this pumping scheme, the two channels have the same magnetic response, and we have built a gradiometer atomic magnetometer with a sensitivity of 0.5 fT/Hz 1/2 ranging from 5 to 40 Hz. Our results propose the possibility of creating larger arrays of atomic magnetometers (AMs) with high sensitivity and spatial resolution based on single-vapor cells for magnetocardiography and magnetoencephalography imaging or searching for exotic spin-dependent interactions.