We have theoretically and experimentally studied the dispersive signal of the Rydberg atomic electromagnetically-induced transparency (EIT) Autler–Townes (AT) splitting spectra obtained using amplitude modulation of the microwave (MW) electric field. In addition to the two zero-crossing points interval Δf zeros, the dispersion signal has two positive maxima with an interval defined as the shoulder interval Δf sho, which is theoretically expected to be used to measure a much weaker MW electric field. The relationship of the MW field strength E MW and Δf sho is experimentally studied at the MW frequencies of 31.6 GHz and 9.2 GHz respectively. The results show that Δf sho can be used to characterize the much weaker E MW than that of Δf zeros and the traditional EIT–AT splitting interval Δf m; the minimum E MW measured by Δf sho is about 30 times smaller than that by Δf m. As an example, the minimum E MW at 9.2 GHz that can be characterized by Δf sho is 0.056 mV/cm, which is the minimum value characterized by the frequency interval using a vapor cell without adding any auxiliary fields. The proposed method can improve the weak limit and sensitivity of E MW measured by the spectral frequency interval, which is important in the direct measurement of weak E MW.
We have theoretically and experimentally studied the dispersive signal of the Rydberg atomic electromagnetically induced transparency (EIT) - Autler-Townes (AT) splitting spectra obtained using amplitude modulation of the microwave (MW) field. In addition to the two zero-crossing points, the dispersion signal has two positive maxima with an interval defined as the shoulder interval of the dispersion signal
Δfsho. The relationship of MW field strength
EMW and
Δfsho are studied at the MW frequencies of 31.6 GHz, 22.1 GHz, and 9.2 GHz respectively. The results show that
Δfsho can be used to character the much weaker
EMW than the interval of two zero-crossing points
Δfzeros and the traditional EIT-AT splitting interval
Δfm, the minimum
EMW measured by
Δfsho is about 30 times smaller than that by
Δfm. As an example, the minimum
EMW at 9.2 GHz that can be characterized by
Δfsho is 0.056 mV/cm, which is the minimum value characterized by frequency interval using vapour cell without adding any auxiliary fields. The proposed method can improve the weak limit and sensitivity of
EMW measured by spectral frequency interval, which is important in the direct measurement of weak
EMW.
The microwave (MW) field can be measured by the Autler–Townes (AT) splitting of the electromagnetically induced transparency (EIT) spectrum in the Rydberg atomic system; however, the EIT-AT splitting method fails in weak MW fields. We used the amplitude modulation of the MW field to resolve the EIT-AT splitting in weak MW fields. The EIT-AT splitting interval can be directly obtained, and the minimum detectable MW strength is improved by six times compared with the traditional EIT-AT splitting method. The proposed method is more intuitive and convenient for measuring the strength of weak MW fields in practical applications.
Abstract We theoretically and experimentally studied quantum microwave electrometry in a cold atomic system using Rydberg electromagnetic induction transparency (EIT) and Autler–Townes splitting (EIT-AT splitting). In cold atoms, a spectral linewidth of ∼500 kHz for EIT was achieved owing to a significant reduction in residual Doppler width, i.e. by at least an order of magnitude, compared to that in vapor cells at room temperature. Therefore, the minimum microwave electric field intensity (E MW) that can be measured is 430 µV cm−1, which is one order higher sensitivity in the EIT-AT regime than that in vapor cells at room temperature. Unlike microwave electrometry in atomic vapor cells, EIT-AT splitting cannot be observed if E MW is so large that the EIT-AT splitting interval Δ f m exceeds the absorption peak width of the cold atom while scanning the frequency of the probe laser (ω p ). Moreover, EIT-AT splitting can be observed if Δ f m exceeds the natural linewidth Γ eg of the intermediate states while scanning the coupling laser (ω c ) and maintains a high spectral resolution with a high signal-to-noise ratio. While scanning ω c , the upper microwave electric field intensity is limited by the scanning range of our setup. Using our system, we measure the maximum field to be 21.6 mV cm−1, nearly three times higher than that of 6.8 mV cm−1 while scanning ω p . The results indicate that the linear range of E MW measured using EIT-AT splitting considerably improves in cold Rydberg atoms.
Microwave electric fields are measured by using cold Rydberg atoms. We obtain spindle-shaped cold atomic clouds in a magneto-optical trap and then pump the cold atoms to quantum state 5S1/2, F = 2, mF = 2 by using an optical-pump laser. We obtain the Rydberg electromagnetic induction transparency (EIT) spectrum peak with narrow linewidth by the low temperature and small residual Doppler broadening. The results show that the typical EIT linewidth with 16 μK cold atoms is about 460 kHz which is 15 times narrower than that of 7 MHz obtained in the thermal vapor cell. The microwave electric field amplitude is measured by Autler-Townes splitting (EIT-AT splitting) in the cold atoms at frequencies of 9.2, 14.2 and 22.1 GHz, receptively. The results show that there is a good linear relationship between the EIT-AT splitting interval and the microwave electric field amplitude. The lower limit of the microwave electric field amplitude that can be measured in the linear region can reach as low as 222 μV/cm, which is about 22 times larger than the lower limit in the traditional thermal vapor cell about of 5 mV/cm. The improvement of the lower limit by EIT-AT splitting method is roughly proportional to the narrowing EIT line width by cold atom samples. This demonstrates that benefiting from the smaller residual Doppler effect and the narrower EIT linewidth in cold atoms, the cold atom system is more advantageous in the experimental measuring of the weak microwave electric field amplitude by using the EIT-AT splitting method. This is of great benefit to the absolute calibration of very weak microwave electric fields. Furthermore, the lower limit of the microwave electric field amplitude that can be measured is smaller than 1 μV/cm by using the change of transmittance of the prober laser at the EIT resonance, and the corresponding sensitivity can reach 1 μV·cm–1·Hz–1/2. These results demonstrate the advantages of cold atomic sample in microwave electric field measurement and its absolute calibration.
We herein developed and demonstrated a Zeeman frequency modulation scheme for improving the signal-to-noise ratio of microwave electric field measurement using Rydberg atoms. The spectra of the electromagnetically induced transparency (EIT) and Autler-Townes splitting of Rydberg atoms is frequency modulated by an alternating current magnetic field. The signal-to-noise ratio of the corresponding dispersive error signal is enhanced more than 10 times than that of the original spectrum. Furthermore, we show that the slope of the dispersive error signal near the resonance of the Rydberg EIT can be used to characterize the weak microwave electric field amplitudes. The more intuitive and simpler structure compared with other existing frequency modulation technologies greatly reduces the difficulties of experiments and experimental data analysis.
Abstract In this study, we investigated Rydberg atom-based microwave (MW) electrometry using polarization spectroscopy in a room-temperature vapor cell. By measuring Autler-Townes splitting in the electromagnetically induced transparency (EIT) spectrum, we determined that the minimum measurable MW electric field is approximately five times lower than conventional EIT techniques. The results are well reproduced by a full optical Bloch equation model, which takes into account all the hyperfine levels involved. Subsequently, the EIT setup was used to characterize a custom MW cylindrical lens, which increases the field at the focus by a factor of three, decreasing the minimum measurable MW electric field by the same amount. Our results indicate that the combination of polarization spectroscopy and a MW lens may enhance MW electrometry, and may allow its use as a secondary standard.