Physical Review Research
Recoil frequency measurement with ppb-level uncertainty by Li6 atom interferometer
Yang Rui, Liang Zhang, Rui Li, Xuemei Liu, Chunyang Duan, Pengyue Liu, Yuelong Wu, Haibin Wu
Abstract
We report a first realization of cold atom interferometers with the lightest alkali metal $^{6}\mathrm{Li}$ atoms and precisely measure the recoil frequency. The atoms are cooled to approach the half of the photon-recoil limit by using a narrow $2S\ensuremath{\rightarrow}3P$ ultraviolet transition. A ``magic'' magnetically insensitive state, a high-efficiency state preparation, and a frequency-insensitive Raman coupling are developed to overcome the challenges from the half-integer spin and hyperfine interaction of atoms. A conjugated $\text{Ramsey-Bord}\stackrel{\ifmmode \acute{}\else \'{}\fi{}}{\mathrm{e}}$ interferometer with crossed Raman beams is realized with a coherence time longer than 2.3 ms. The four sets of interferometers developed using geometric relations greatly eliminate the error from the angle between Raman beams. The measured recoil frequency is ${\ensuremath{\omega}}_{r}=2\ensuremath{\pi}\ifmmode\times\else\texttimes\fi{}73\phantom{\rule{0.16em}{0ex}}672.789\phantom{\rule{0.16em}{0ex}}(36)$ Hz, representing the most precise measurement of the recoil frequency of $^{6}\mathrm{Li}$ by using the atom interferometer so far. The realized $^{6}\mathrm{Li}$ cold atom interferometer is not only an effective supplement to the existing interferometers, but also provides a good candidate in precision measurement owing to its low mass and high recoil frequency.