Since the first proposal by Caves1 that squeezed light could be used to enhance the sensitivity of a gravitational wave interferometric antenna, squeezed states have been demonstrated in the laboratory but not yet used in an actual measurement system. This paper describes a polarization interferometer experiment, operated with a squeezed light source, designed to improve the interferometer sensitivity beyond the standard quantum limit. The interferometer consists of two polarization beam splitters, whose axes are oriented 45° with respect to each other. Linearly polarized incident laser light is transmitted by the first polarizer and split by the second one, which also acts as the splitter for a balanced homodyne detector. A rotation of the polarization between the two polarizers can be detected with quantum-limited sensitivity (e.g., ±1 μrad). The squeezed light entering the other port of the interferometer is produced using a χ2 crystal (KTP) in an optical cavity, pumped by a frequency-doubled cw ring Nd:YAG laser. The interferometer sensitivity can thus be increased by a factor depending on the degree of squeezing achieved and on the detector quantum efficiency. Sensitivity increases by factors between 2 and 3 are predicted, and current performance is described.
Since the first proposal by Caves1 that squeezed light could be used to enhance the sensitivity of a gravitational wave interferometric antenna, squeezed states have been demonstrated in the laboratory but not yet used in an actual measurement system. This paper describes a polarization interferometer experiment, operated with a squeezed light source, designed to improve the interferometer sensitivity beyond the standard quantum limit. The interferometer consists of two polarization beam splitters, whose axes are oriented 45° with respect to each other. Linearly polarized incident laser light is transmitted by the first polarizer and split by the second one, which also acts as the splitter for a balanced homodyne detector. A rotation of the polarization between the two polarizers can be detected with quantum-limited sensitivity (e.g., ±1 μrad). The squeezed light entering the other port of the interferometer is produced using a χ2 crystal (KTP) in an optical cavity, pumped by a frequency-doubled cw ring Nd:YAG laser. The interferometer sensitivity can thus be increased by a factor depending on the degree of squeezing achieved and on the detector quantum efficiency. Sensitivity increases by factors between 2 and 3 are predicted, and current performance is described.
An experimental study of enhancement of interferometric sensitivity using squeezed vacuum fluctuations is described. By operating an interferometer at the dark fringe, large laser powers can be used for high sensitivity. It is possible to enhance the interferometric sensitivity beyond the shot-noise limit in the dark fringe arrangement by using an electrooptic modulator in the interferometer to vary the phase periodically away from the dark fringe. A detector at the dark output port can then operate at the shot-noise limit and detect phase imbalance at the sensitivity determined by power level at the bright output port.1 This shot-noise limit can be reduced further by illuminating the dark input port with squeezed vacuum, as demonstrated recently in balanced output port experiments.2,3 However, there are interesting limits to the sensitivity improvement expected in the dark fringe case.1 The experimental arrangement includes a polarization interferometer illuminated with 100 mW of 1.064-μm wavelength light from a Nd:YAG cw laser. Squeezed vacuum is generated by a KTP optical parametric amplifier. Experimental progress is reported at sensitivities in the nanoradian polarization rotation range.
The ability to directly detect gravitational waves will open a completely new branch of astronomy to view the Universe, one that is inaccessible to electromagnetic-based astronomy. First generation ground-based interferometric gravitational-wave detectors have achieved strain sensitivities of order one part in ten to the twenty-one, at 100 Hz detection frequency. A new generation of detectors are under construction, designed to improve on the sensitivity of the first-generation detectors by a factor of 10. The quantum nature of light will broadly limit the sensitivity of these new instruments. This quantum noise will originate from the quantum vacuum fluctuations that enter the unused port of the interferometer. One of the most promising options for reducing the quantum noise impact and further increasing the sensitivity is applying quantum squeezed vacuum states. These squeezed states have lower noise in one quadrature than the vacuum state. By replacing the quantum vacuum fluctuations entering the interferometer with squeezed vacuum states, the quantum noise impact is reduced. This thesis firstly details the development of a squeezed light source that produces squeezed states applicable for enhancing interferometric gravitational-wave detectors. A doubly-resonant, travelling-wave bow-tie cavity squeezed light source is presented. For the first time, greater than 10 dB of quantum noise suppression across the gravitational-wave detection band is directly observed, and 11.6 dB of quantum noise suppression is observed above 200 Hz. This squeezing cavity design also has benefits with intrinsic isolation to backscattered light. Experiments that quantify this isolation are reported. The properties affecting squeezing magnitude and low-frequency squeezing measurement are discussed. In addition, a modified squeezing-ellipse-phase control technique for squeezed vacuum states is presented. This thesis secondly presents results from the LIGO Squeezed Light Injection Experiment, undertaken to test squeezed light injection into a 4 km interferometric gravitational-wave detector. The results of the experiment show the first measurement of squeezing-enhancement in a 4 km gravitational-wave detector, with 2.15 dB measured above 250 Hz. This represents the best sensitivity to gravitational waves yet achieved at these frequencies by any single gravitational-wave detector to date. An unknown area was whether the addition of a squeezer would introduce noise couplings that degrade the crucial low frequency sensitivity. The results demonstrate that injected squeezed states are compatible with low frequency gravitational-wave measurement. The characterisation of squeezing-injection optical losses and fluctuations of the squeezing angle are also reported. The knowledge and processes gained, from both the squeezed light source development work and the LIGO Squeezed Light Injection Experiment, will inform the design, planning and implementation of squeezed states in future gravitational-wave detectors.
A potential application of squeezed-state light is to improve the sensitivity of high-precision interferometers whose performance is generally limited by photon shot noise or photon-counting statistics. Such interferometers include passive laser gyroscopes and the very sensitive interferometers being developed to detect cosmic gravitational waves. One usually operates a high-precision interferometer in a differential mode, so that noise from the (laser) light source is in principle irrelevant. Operation in a differential mode means that there is a second, normally unused input port. Vacuum noise entering this second port is responsible for the irreducible photon shot noise at the interferometer’s output. Injecting squeezed-state light into the normally unused input port can improve the sensitivity. I discuss the use of squeezed-state light in ring interferometers, delayline Michelson interferometers, and Fabry-Perot interferometers. Included in the discussion are real degrading effects, such as losses, inefficient photodetectors, and laser noise in the presence of nonideal fringe contrast. I explore the compatibility of the squeezed-state technique with other exotic techniques for improving sensitivity, such as Drever's idea for power recycling. Sensitivity estimates for various configurations and techniques are presented.
Nearly a century after Einstein first predicted the existence of gravitational waves, a global network of Earth-based gravitational wave observatories 1-4 is seeking to directly detect this faint radiation using precision laser interferometry. Photon shot noise, due to the quantum nature of light, imposes a fundamental limit on the attometre-level sensitivity of the kilometre-scale Michelson interferometers deployed for this task. Here, we inject squeezed states to improve the performance of one of the detectors of the Laser Interferometer Gravitational-Wave Observatory (LIGO) beyond the quantum noise limit, most notably in the frequency region down to 150 Hz, critically important for several astrophysical sources, with no deterioration of performance observed at any frequency. With the injection of squeezed states, this LIGO detector demonstrated the best broadband sensitivity to gravitational waves ever achieved, with important implications for observing the gravitational-wave Universe with unprecedented sensitivity.
Quantum noise is one of the limiting factors in laser-interferometric gravitational-wave (GW) detectors. The application of squeezed states in these interferometers allows the reduction of quantum noise in one quadrature. Due to opto-mechanical coupling in a GW detector the squeezed quadrature needs to be rotated within the spectrum to achieve a broadband noise reduction. So far, the implementation of additional filter cavities is considered that allow for the optimal, frequency-dependent rotation of the squeezed quadrature. However, these cavities need to have low loss, a length in the order of \unit[100]{m} and must be situated in the vacuum system, making them cost-intensive. In 2017, Ma and coworkers proposed a scheme for the broadband quantum-noise reduction without the need of additional filter cavities. It was shown by Brown et al. that a similar scheme can be used to broadband-enhance interferometers with a detuned signal-recycling cavity. Here, we performed a proof-of-principle experiment of the proposal on a table-top-scale. Squeezed states were produced detuned to the carrier field of a \unit[2.5]{m}-linear cavity and read out in a bichromatic homodyne detection. The frequencies of the lower and upper local oscillator were at entangled sidebands of the squeezed field. Depending on the relations between the involved frequencies, we can address both variants of the proposal. We show, that the frequency-dependences of the resulting noise spectra fit to a theoretical model we derived from the theory used by Ma et al. With this work we set the path towards an implementation of these schemes in a GW-detector prototype, where the compatibility of the approach with a low-frequency suspended interferometer can be tested. Moreover, we used the same setup to show nonclassical interferometer enhancement at low frequencies by high-frequency squeezed states. Here, a heterodyne readout scheme was implemented to avoid limiting noises at low frequencies. The application of squeezed states centered around the local oscillator frequency yielded an improvement in signal-to-noise ratio of $\unit[3.4]{dB}\pm\unit[0.3]{dB}$. Additionally, I designed, built and characterized a compact source of squeezed vacuum-states at \unit[1064]{nm} with a footprint of just $\unit[0.8]{m^2}$. I show measurements of squeezed states from this source with a reduction of quantum noise of $\unit[10.7]{dB}\pm\unit[0.2]{dB}$ below the vacuum noise and present a noise reduction in the frequency range from \unit[70]{kHz} to \unit[65]{MHz}.
Direct detection of gravitational waves will require
earth based detectors to measure strains of the order 10-21, at
frequencies of 100 Hz, a sensitivity that has been accomplished
with the initial generation of LIGO interferometric gravitational
wave detectors. A new generation of detectors currently under
construction is designed improve on the sensitivity of the initial
detectors by about a factor of 10. The quantum nature of light will
limit the sensitivity of these Advanced LIGO interferometers at
most frequencies; new approaches to reducing the quantum noise will
be needed to improve the sensitivity further. This quantum noise
originates from the vacuum fluctuations that enter the unused port
of the interferometer and interfere with the laser light. Vacuum
fluctuations have the minimum noise allowed by Heisenberg's
uncertainty principle, [Delta]X1 [Delta]X2 >/=1, where the two
quadratures X1 and X2 are non-commuting observables responsible for
the two forms of quantum noise, shot noise and radiation pressure
noise. By replacing the vacuum fluctuations entering the
interferometer with squeezed states, which have lower noise in one
quadrature than the vacuum state, we have reduced the shot noise of
a LIGO interferometer. The sensitivity to gravitational waves
measured during this experiment represents the best sensitivity
achieved to date at frequencies above 200 Hz, and possibly the
first time that squeezing has been measured in an interferometer at
frequencies below 700 Hz. The possibility that injection of
squeezed states could introduce environmental noise couplings that
would degrade the crucial but fragile low frequency sensitivity of
a LIGO interferometer has been a major concern in planning to
implement squeezing as part of baseline interferometer operations.
These results demonstrate that squeezing is compatible with the low
frequency sensitivity of a full scale gravitational wave
interferometer. We also investigated the limits to the level of
squeezing observed, including optical losses and fluctuations of
the squeezing angle. The lessons learned should allow for
responsible planning to implement squeezing in Advanced LIGO,
either as an alternative to high power operation or an early
upgrade to improve the sensitivity. This thesis is available at
DSpace@MIT and has LIGO document number P1300006.