Metrologia

Radiometric characterization of a triggered narrow-bandwidth single-photon source and its use for the calibration of silicon single-photon avalanche detectors

Hristina Georgieva, Marco López, Helmuth Hofer, Justus Christinck, Beatrice Rodiek, Peter Schnauber, Arsenty Kaganskiy, Tobias Heindel, Sven Rodt, Stephan Reitzenstein, S. Kück

Published Sep 7, 2020• 15 citations
Publisher

Abstract

The traceability of measurements of the parameters characterizing single-photon sources, such as photon flux and optical power, paves the way towards their reliable comparison and quantitative evaluation. In this paper, we present an absolute measurement of the optical power of a single-photon source based on an InGaAs quantum dot under pulsed excitation with a calibrated single-photon avalanche diode (SPAD) detector. For this purpose, a single excitonic line of the quantum dot emission with a bandwidth below 0.1 nm was spectrally filtered by using two tilted interference filters. Since high count rates are essential for many metrological applications, we optimized the setup efficiency by combining high overall transmission of the optical components with a geometrical enhancement of the extraction efficiency of a single quantum dot by a monolithic microlens to reach photon fluxes up to 3.7 ⋅ 10 5 photons per second at the SPADs. A relative calibration of two SPAD detectors with a relative standard uncertainty of 0.7% was carried out and verified by the standard calibration method using an attenuated laser. Finally, an Allan deviation analysis was performed giving an optimal averaging time of 92 s for the photon flux.

DOI
10.1088/1681-7575/ab9db6

Related papers

Metrological characterization of a commercial single-photon source with high photon flux emission

Hristina Georgieva, Petr Stepanov, Leonora Sewidan +3 more

We present a comprehensive metrological characterization of a commercial single-photon source with high photon flux emission for use in radiometry. The source is based on an InGaAs quantum dot in a micropillar. A comparative analysis of two excitation schemes—phonon-assisted excitation and two-photon excitation—explores differences in excitation power dependence, temporal stability and single-photon purity. The commercial source exhibits excellent properties for the field of quantum radiometry, achieving simultaneously a photon flux of (17.19 ± 0.09) million photons/s for a pulse repetition rate of 79.4 MHz, and a single-photon purity of 98%. Its optical power of (3.68 ± 0.02) pW is directly determined with a traceably calibrated low-noise photodiode. The ability to directly compare the photocurrent in a low-noise photodiode with the count rate at a single-photon avalanche detector allows for a seamless transition between the classical and quantum realizations of optical power. Therefore, we were able to build another bridge between classical and quantum radiometry by using a deterministic single-photon source.

Metrologia•Jul 23, 2024•4 citations

A germanium-vacancy center in diamond as single-photon source for radiometric application

Justus Christinck, F. Hirt, H. Hofer +7 more

We present the metrological characterization of a single-photon source based on a germanium-vacancy center in diamond under a solid immersion lens in a confocal microscope setup at room temperature. It was characterized in terms of the emission’s spectral distribution, single-photon purity, temporal stability and the emitter’s excited state lifetime and saturation behavior. An Allan deviation analysis was performed on the emission of the single-photon source to determine the optimal averaging time of the photon flux. The single-photon source was used for the relative calibration of the detection efficiency of two single-photon avalanche diode detectors. The results were compared with measurements using attenuated laser light for the calibration of the detectors.

Journal of Physics Conference Series•Oct 1, 2024•0 citations

Radiometric calibration of single photon detectors by a single photon source based on NV-centers in diamond

Waldemar Schmunk, M. Rodenberger, Silke Peters +2 more

A widespread use of various relative calibration techniques is established in order to realize reliable and low uncertainty measurements of the detection efficiency, which is one key parameter characterizing single photon detectors. In the following paper we will present an approach to evaluate the relative detection efficiency of single photon avalanche photo diode (SPAD) detectors compared to a standard detector. This calibration technique is based upon the fiber-coupled relative efficiency calibration of analogue detectors, used in fiber-optic communication. For the first time, to our knowledge, an intrinsic single photon source based on the nitrogen-vacancy center in diamond was used for this purpose. Furthermore, the possible influence of different photon statistics, arising from different irradiation sources like thermal sources or lasers on the calibration results for the fiber exchange method has been theoretically studied.

Journal of Modern Optics•Aug 4, 2011•35 citations

Experimental realization of an absolute single-photon source based on a single nitrogen vacancy center in a nanodiamond

Beatrice Rodiek, Marco López, Helmuth Hofer +8 more

We report on the experimental realization of an absolute single-photon source based on a single nitrogen vacancy (NV) center in a nanodiamond at room temperature and on the calculation of its absolute spectral photon flux from experimental data. The single-photon source was calibrated with respect to its photon flux and its spectral photon rate density. The photon flux was measured with a low-noise silicon photodiode traceable to the primary standard for optical flux, taking into account the absolute spectral power distribution using a calibrated spectroradiometer. The optical radiant flux is adjustable from 55 fW, which is almost the lowest detection limit for the silicon photodiode, and 75 fW, which is the saturation power of the NV center. These fluxes correspond to total photon flux rates between 190,000 photons per second and 260,000 photons per second, respectively. The single-photon emission purity is indicated by a g((2))(0) value, which is between 0.10 and 0.23, depending on the excitation power. To our knowledge, this is the first single-photon source absolutely calibrated with respect to its absolute optical radiant flux and spectral power distribution, traceable to the corresponding national standards via an unbroken traceability chain. The prospects for its application, e.g., for the detection efficiency calibration of single-photon detectors as well as for use as a standard photon source in the low photon flux regime, are promising. (C) 2017 Optical Society of America

Optica•Jan 6, 2017•92 citations

A Molecule‐Based Single‐Photon Source Applied in Quantum Radiometry

Pietro Lombardi, Marco Trapuzzano, Maja Colautti +5 more

Single‐photon sources (SPSs) based on quantum emitters hold promise in quantum radiometry as metrology standard for photon fluxes at the low light level. Ideally this requires control over the photon flux in a wide dynamic range, sub‐Poissonian photon statistics, and narrow‐band emission spectrum. In this work, a monochromatic SPS based on an organic dye molecule is presented, whose photon flux is traceably measured to be adjustable between 144 000 and 1320 000 photons per second at a wavelength of (785.6 ± 0.1) nm, corresponding to an optical radiant flux between 36.5 and 334 fW. The high purity of the single‐photon stream is verified, with a second‐order autocorrelation function at zero time delay below 0.1 throughout the whole range. Such molecule‐based SPS is hence used for the calibration of a single‐photon avalanche detector against a low‐noise analog photodiode traceable to the primary standard for optical radiant flux (i.e., the cryogenic radiometer). Due to the narrow bandwidth of the source, corrections to the detector efficiency arising from the spectral power distribution are negligible. With this major advantage, the developed device may finally realize a low‐photon‐flux standard source for quantum radiometry.

arXiv:1908.00616•Advanced Quantum Technologies•Oct 22, 2019•44 citations

A study to develop a robust method for measuring the detection efficiency of free-running InGaAs/InP single-photon detectors

Marco López, Alice Meda, Geiland Porrovecchio +7 more

The challenges faced in a comparison of measuring the detection efficiency of free-running InGaAs/InP single-photon avalanche detectors (InGaAs/InP SPAD) were studied by four European National Metrology Institutes (NMIs) meeting at a single laboratory. The main purpose of this study is to develop a trustable measurement technique and to provide a snapshot of the methods used by the four NMIs for measuring such photon-counting detectors at telecom wavelengths in order to establish proper procedures for characterising such devices. The detection efficiency measurements were performed using different experimental setups and reference standards with independent traceability chains at the wavelength of 1550 nm. A dedicated model to correct the dead time and dark count effects on the SPAD’s free-running counting process was developed, allowing the correct value of the photon rate impinging on the detector to be recovered from simple ratemeter measurements. The detection efficiency was measured for mean photon number per pulse between 0.01 and 2.4, corresponding to photon rates between approximately 1100 photon/s and 193,000 photon/s, respectively. We found that the measured values reported by the participants are all consistent within the stated uncertainties, proving the consistency of the measurement approach developed.

EPJ Quantum Technology•Nov 25, 2020•15 citations
The traceability of measurements of the parameters characterizing single-photon sources, such as photon flux and optical power, paves the way towards their reliable comparison and quantitative evaluation. In this paper, we present an absolute measurement of the optical power of a single-photon source based on an InGaAs quantum dot under pulsed excitation with a calibrated single-photon avalanche diode (SPAD) detector. For this purpose, a single excitonic line of the quantum dot emission with a bandwidth below 0.1 nm was spectrally filtered by using two tilted interference filters. Since high count rates are essential for many metrological applications, we optimized the setup efficiency by combining high overall transmission of the optical components with a geometrical enhancement of the extraction efficiency of a single quantum dot by a monolithic microlens to reach photon fluxes up to 3.7 ⋅ 10 5 photons per second at the SPADs. A relative calibration of two SPAD detectors with a relative standard uncertainty of 0.7% was carried out and verified by the standard calibration method using an attenuated laser. Finally, an Allan deviation analysis was performed giving an optimal averaging time of 92 s for the photon flux.
We present a comprehensive metrological characterization of a commercial single-photon source with high photon flux emission for use in radiometry. The source is based on an InGaAs quantum dot in a micropillar. A comparative analysis of two excitation schemes—phonon-assisted excitation and two-photon excitation—explores differences in excitation power dependence, temporal stability and single-photon purity. The commercial source exhibits excellent properties for the field of quantum radiometry, achieving simultaneously a photon flux of (17.19 ± 0.09) million photons/s for a pulse repetition rate of 79.4 MHz, and a single-photon purity of 98%. Its optical power of (3.68 ± 0.02) pW is directly determined with a traceably calibrated low-noise photodiode. The ability to directly compare the photocurrent in a low-noise photodiode with the count rate at a single-photon avalanche detector allows for a seamless transition between the classical and quantum realizations of optical power. Therefore, we were able to build another bridge between classical and quantum radiometry by using a deterministic single-photon source.
We present the metrological characterization of a single-photon source based on a germanium-vacancy center in diamond under a solid immersion lens in a confocal microscope setup at room temperature. It was characterized in terms of the emission’s spectral distribution, single-photon purity, temporal stability and the emitter’s excited state lifetime and saturation behavior. An Allan deviation analysis was performed on the emission of the single-photon source to determine the optimal averaging time of the photon flux. The single-photon source was used for the relative calibration of the detection efficiency of two single-photon avalanche diode detectors. The results were compared with measurements using attenuated laser light for the calibration of the detectors.
A widespread use of various relative calibration techniques is established in order to realize reliable and low uncertainty measurements of the detection efficiency, which is one key parameter characterizing single photon detectors. In the following paper we will present an approach to evaluate the relative detection efficiency of single photon avalanche photo diode (SPAD) detectors compared to a standard detector. This calibration technique is based upon the fiber-coupled relative efficiency calibration of analogue detectors, used in fiber-optic communication. For the first time, to our knowledge, an intrinsic single photon source based on the nitrogen-vacancy center in diamond was used for this purpose. Furthermore, the possible influence of different photon statistics, arising from different irradiation sources like thermal sources or lasers on the calibration results for the fiber exchange method has been theoretically studied.
We report on the experimental realization of an absolute single-photon source based on a single nitrogen vacancy (NV) center in a nanodiamond at room temperature and on the calculation of its absolute spectral photon flux from experimental data. The single-photon source was calibrated with respect to its photon flux and its spectral photon rate density. The photon flux was measured with a low-noise silicon photodiode traceable to the primary standard for optical flux, taking into account the absolute spectral power distribution using a calibrated spectroradiometer. The optical radiant flux is adjustable from 55 fW, which is almost the lowest detection limit for the silicon photodiode, and 75 fW, which is the saturation power of the NV center. These fluxes correspond to total photon flux rates between 190,000 photons per second and 260,000 photons per second, respectively. The single-photon emission purity is indicated by a g((2))(0) value, which is between 0.10 and 0.23, depending on the excitation power. To our knowledge, this is the first single-photon source absolutely calibrated with respect to its absolute optical radiant flux and spectral power distribution, traceable to the corresponding national standards via an unbroken traceability chain. The prospects for its application, e.g., for the detection efficiency calibration of single-photon detectors as well as for use as a standard photon source in the low photon flux regime, are promising. (C) 2017 Optical Society of America
Single‐photon sources (SPSs) based on quantum emitters hold promise in quantum radiometry as metrology standard for photon fluxes at the low light level. Ideally this requires control over the photon flux in a wide dynamic range, sub‐Poissonian photon statistics, and narrow‐band emission spectrum. In this work, a monochromatic SPS based on an organic dye molecule is presented, whose photon flux is traceably measured to be adjustable between 144 000 and 1320 000 photons per second at a wavelength of (785.6 ± 0.1) nm, corresponding to an optical radiant flux between 36.5 and 334 fW. The high purity of the single‐photon stream is verified, with a second‐order autocorrelation function at zero time delay below 0.1 throughout the whole range. Such molecule‐based SPS is hence used for the calibration of a single‐photon avalanche detector against a low‐noise analog photodiode traceable to the primary standard for optical radiant flux (i.e., the cryogenic radiometer). Due to the narrow bandwidth of the source, corrections to the detector efficiency arising from the spectral power distribution are negligible. With this major advantage, the developed device may finally realize a low‐photon‐flux standard source for quantum radiometry.
The challenges faced in a comparison of measuring the detection efficiency of free-running InGaAs/InP single-photon avalanche detectors (InGaAs/InP SPAD) were studied by four European National Metrology Institutes (NMIs) meeting at a single laboratory. The main purpose of this study is to develop a trustable measurement technique and to provide a snapshot of the methods used by the four NMIs for measuring such photon-counting detectors at telecom wavelengths in order to establish proper procedures for characterising such devices. The detection efficiency measurements were performed using different experimental setups and reference standards with independent traceability chains at the wavelength of 1550 nm. A dedicated model to correct the dead time and dark count effects on the SPAD’s free-running counting process was developed, allowing the correct value of the photon rate impinging on the detector to be recovered from simple ratemeter measurements. The detection efficiency was measured for mean photon number per pulse between 0.01 and 2.4, corresponding to photon rates between approximately 1100 photon/s and 193,000 photon/s, respectively. We found that the measured values reported by the participants are all consistent within the stated uncertainties, proving the consistency of the measurement approach developed.