Setup of Dry Cryostat for Measurements on Superconducting Microwave Resonators

Nicholas Cockton

Published Sep 20, 2019• 0 citations
OpenAlex

Abstract

A cryogen-free cryostat was set up in order to perform measurements at ultra low temperatures. The setup involved the removal of an adiabatic demagnetization refrigerator (ADR) and replacing it with a dilution refrigerator system. The modified cryostat is able to provide continuous cooling power, enabling greater flexibility over the ADR in the type of low temperature experiments that can be performed. Condensation stages, main flow impedance and radiation baffles were designed, constructed and installed into the cryostat. Several general cryostat design considerations are also discussed. Condensation of the incoming 3He - 4He mixture utilizes only the cooling power of the two-stage pulse tube refrigerator. For a circulation rate of 120 μmols/sec, the base temperature of the cryostat is 70 mK with a cooling power of 33 μW of heat at 100 mK. \nWith this cryostat, a measurement system is developed for studying superconducting microwave resonators. Residual losses in these devices continue to impede their performance in applications such as ultra sensitive detectors and quantum computing. A microwave circuit consisting of coaxial cables, attenuators and amplifiers was installed on the cryostat to measure losses in these devices. Data analysis and fitting procedures were written in Python programming language to extract key resonator properties. Characterizing and understanding loss mechanisms for these devices will lead to improved materials and fabrication methods. This will ultimately result in losses that approach the fundamental limit set by the Mattis-Bardeen theory.

Related papers

Dry Dilution Refrigerator for Experiments on Quantum Effects in the Microwave Regime

Achim Marx, J. Hoess, Kurt Uhlig

At the Walther-Meißner-Institut (WMI), a new cryogen-free 3He/4He dilution refrigerator (DR) has been completed; the cryostat will be employed to cool experiments on superconducting quantum circuits for quantum information technology and quantum simulations. All major components have been made at the WMI. The DR offers lots of space at the various stages of the apparatus for microwave components and cables. E. g., the usable space at the mixing chamber has a height of more than 60 cm and a diameter of 30 cm (mixing chamber mounting plate). To cool cables and cold amplifiers, the DR is equipped with a separate 4He-1K-loop which offers a cooling power of up to 100 mW near 1K. The refrigeration power of the still is 18 mW at 0.9 K; the diameter of its mounting plate is 35 cm. The cryostat rests in an aluminum trestle on air springs to attenuate building vibrations. It is precooled by a Cryomech PT410-RM pulse tube cryocooler (PTC) which is mechanically decoupled from the vacuum can of the cryostat by a bellows assembly. The two stages of the PTC are thermally connected to the DR via copper ropes. There are no nitrogen cooled traps with this DR to purify the gas streams of the 3He and 4He loops; instead, charcoal traps are mounted inside the DR at the first stage of the PTC. The dilution unit has three heat exchangers; its base temperature is 11 mK and its cooling power is 300 μW at 100 mK.

arXiv:1412.3619•Dec 11, 2014•2 citations

Setup and Design of a Cryogenic Testing Station with Focus on Characterization of Superconducting Devices

Clifford E. Plesha

The goal of my thesis was to construct a general purpose cryostat, capable of measuring multiple devices at 100mK. This included modifying an existing SHASTA model 106 Adiabatic Demagnetization Refrigerator (ADR) by adding a new physical support structure to allow for more weight to be added to the cryostat’s testing stages, as well as custom pieces of electronics, such as a room temperature programmable current source. This allowed for computer programs to be written so we could automate the data acquisition process. \nWe successfully measured multiple superconducting quantum interference devices (SQUIDs), with the goal of proving that our custom electronics were able to control and record the output of the SQUID. This included measuring the biasing conditions of the SQUID with \nthe automated current sources, as well as measuring the noise temperature of a SQUID down to 300mK. Furthermore, we measured the thermal conductivity of our new sup- port structure for the cryostat and were able to prove that the material used was new \nand novel approach for high strength, low thermal conductivity in cryogenic research. We also investigated a new technique in permanently modifying the operational frequency of a SQUID, and were able to show that this new method of laser ablating the input coil could successfully shift the frequency. While the focus of cryogenic testing was on SQUIDs the cryostat itself was designed in such a way such that many types of low temperature devices could be tested within the system.

•May 23, 2019•1 citation

Materials loss measurements using superconducting microwave resonators

C. R. H. McRae, H. Wang, J. Gao +5 more

The performance of superconducting circuits for quantum computing is limited by materials losses. In particular, coherence times are typically bounded by two-level system (TLS) losses at single photon powers and millikelvin temperatures. The identification of low loss fabrication techniques, materials, and thin film dielectrics is critical to achieving scalable architectures for superconducting quantum computing. Superconducting microwave resonators provide a convenient qubit proxy for assessing performance and studying TLS loss and other mechanisms relevant to superconducting circuits such as non-equilibrium quasiparticles and magnetic flux vortices. In this review article, we provide an overview of considerations for designing accurate resonator experiments to characterize loss, including applicable types of losses, cryogenic setup, device design, and methods for extracting material and interface losses, summarizing techniques that have been evolving for over two decades. Results from measurements of a wide variety of materials and processes are also summarized. Finally, we present recommendations for the reporting of loss data from superconducting microwave resonators to facilitate materials comparisons across the field.

arXiv:2006.04718•Review of Scientific Instruments•Sep 1, 2020•257 citations

Fabrication and Characterization of Superconducting Resonators

Giuseppe Cataldo, Emily M. Barrentine, Ari-David Brown +3 more

Superconducting microwave resonators are of interest for a wide range of applications, including for their use as microwave kinetic inductance detectors (MKIDs) for the detection of faint astrophysical signatures, as well as for quantum computing applications and materials characterization. In this paper, procedures are presented for the fabrication and characterization of thin-film superconducting microwave resonators. The fabrication methodology allows for the realization of superconducting transmission-line resonators with features on both sides of an atomically smooth single-crystal silicon dielectric. This work describes the procedure for the installation of resonator devices into a cryogenic microwave testbed and for cool-down below the superconducting transition temperature. The set-up of the cryogenic microwave testbed allows one to do careful measurements of the complex microwave transmission of these resonator devices, enabling the extraction of the properties of the superconducting lines and dielectric substrate (e.g., internal quality factors, loss and kinetic inductance fractions), which are important for device design and performance.

Journal of Visualized Experiments•May 21, 2016•1 citation

Insertable system for fast turnaround time microwave experiments in a dilution refrigerator

Florian R. Ong, Jean-Luc Orgiazzi, A. de Waard +2 more

Microwave experiments in dilution refrigerators are a central tool in the field of superconducting quantum circuits and other research areas. This type of experiments relied so far on attaching a device to the mixing chamber of a dilution refrigerator. The minimum turnaround time in this case is a few days as required by cooling down and warming up the entire refrigerator. We developed a new approach, in which a suitable sample holder is attached to a cold-insertable probe and brought in contact with transmission lines permanently mounted inside the cryostat. The total turnaround time is 8 h if the target temperature is 80 mK. The lowest attainable temperature is 30 mK. Our system can accommodate up to six transmission lines, with a measurement bandwidth tested from zero frequency to 12 GHz. This bandwidth is limited by low-pass components in the setup; we expect the intrinsic bandwidth to be at least 18 GHz. We present our setup, discuss the experimental procedure, and give examples of experiments enabled by this system. This new measurement method will have a major impact on systematic ultra-low temperature studies using microwave signals, including those requiring quantum coherence.

arXiv:1209.1039•Review of Scientific Instruments•Sep 1, 2012•6 citations

Traceable In Situ Microwave Power Measurement at the Cryogenic Device Plane in a Dilution Refrigerator

Paolo Panetta, Andrea Celotto, Alessandro Alocco +7 more

Accurate knowledge of the microwave power delivered to a cryogenic device under test (DUT) is essential for the characterization and operation of superconducting quantum circuits. However, this information is difficult to obtain inside dilution refrigerators because of distributed attenuation, impedance mismatch, switch-path repeatability, and temperature-dependent microwave components. This paper presents an in situ measurement method for RF power at the cryogenic device plane. The method uses a custom variable temperature stage (VTS) as a cryogenic thermal-transfer element. The TVS is alternately heated by a four-wire DC heater and by microwave power dissipated in a 20 dB pass-through attenuator. By fitting the thermal transients and comparing the corresponding steady-state temperatures, the absorbed microwave power is inferred from a directly measured DC electrical power through an AC/DC substitution procedure. The finite reflection and transmission of the attenuator are then accounted for by cryogenic two-port scattering-parameter measurements based on a switch-assisted Short--Open--Load--Reciprocal calibration, so that the result is referred to the DUT reference plane. The system is demonstrated in a dilution refrigerator with powers between -43 and -58 dBm at the DUT input plane. The demonstrated relative standard uncertainty ranges from about 2% at -43.9 dBm to about 40% at -57.6 dBm. The proposed approach combines thermal RF power transfer, cryogenic S-parameter correction, and uncertainty evaluation in a measurement architecture compatible with quantum-device experiments, providing a practical route toward traceable microwave-power calibration at millikelvin stages.

physics.ins-detCross-listedarXiv:2607.12751•Jul 15, 2026•0 citations
A cryogen-free cryostat was set up in order to perform measurements at ultra low temperatures. The setup involved the removal of an adiabatic demagnetization refrigerator (ADR) and replacing it with a dilution refrigerator system. The modified cryostat is able to provide continuous cooling power, enabling greater flexibility over the ADR in the type of low temperature experiments that can be performed. Condensation stages, main flow impedance and radiation baffles were designed, constructed and installed into the cryostat. Several general cryostat design considerations are also discussed. Condensation of the incoming 3He - 4He mixture utilizes only the cooling power of the two-stage pulse tube refrigerator. For a circulation rate of 120 μmols/sec, the base temperature of the cryostat is 70 mK with a cooling power of 33 μW of heat at 100 mK. \nWith this cryostat, a measurement system is developed for studying superconducting microwave resonators. Residual losses in these devices continue to impede their performance in applications such as ultra sensitive detectors and quantum computing. A microwave circuit consisting of coaxial cables, attenuators and amplifiers was installed on the cryostat to measure losses in these devices. Data analysis and fitting procedures were written in Python programming language to extract key resonator properties. Characterizing and understanding loss mechanisms for these devices will lead to improved materials and fabrication methods. This will ultimately result in losses that approach the fundamental limit set by the Mattis-Bardeen theory.
At the Walther-Meißner-Institut (WMI), a new cryogen-free 3He/4He dilution refrigerator (DR) has been completed; the cryostat will be employed to cool experiments on superconducting quantum circuits for quantum information technology and quantum simulations. All major components have been made at the WMI. The DR offers lots of space at the various stages of the apparatus for microwave components and cables. E. g., the usable space at the mixing chamber has a height of more than 60 cm and a diameter of 30 cm (mixing chamber mounting plate). To cool cables and cold amplifiers, the DR is equipped with a separate 4He-1K-loop which offers a cooling power of up to 100 mW near 1K. The refrigeration power of the still is 18 mW at 0.9 K; the diameter of its mounting plate is 35 cm. The cryostat rests in an aluminum trestle on air springs to attenuate building vibrations. It is precooled by a Cryomech PT410-RM pulse tube cryocooler (PTC) which is mechanically decoupled from the vacuum can of the cryostat by a bellows assembly. The two stages of the PTC are thermally connected to the DR via copper ropes. There are no nitrogen cooled traps with this DR to purify the gas streams of the 3He and 4He loops; instead, charcoal traps are mounted inside the DR at the first stage of the PTC. The dilution unit has three heat exchangers; its base temperature is 11 mK and its cooling power is 300 μW at 100 mK.
The goal of my thesis was to construct a general purpose cryostat, capable of measuring multiple devices at 100mK. This included modifying an existing SHASTA model 106 Adiabatic Demagnetization Refrigerator (ADR) by adding a new physical support structure to allow for more weight to be added to the cryostat’s testing stages, as well as custom pieces of electronics, such as a room temperature programmable current source. This allowed for computer programs to be written so we could automate the data acquisition process. \nWe successfully measured multiple superconducting quantum interference devices (SQUIDs), with the goal of proving that our custom electronics were able to control and record the output of the SQUID. This included measuring the biasing conditions of the SQUID with \nthe automated current sources, as well as measuring the noise temperature of a SQUID down to 300mK. Furthermore, we measured the thermal conductivity of our new sup- port structure for the cryostat and were able to prove that the material used was new \nand novel approach for high strength, low thermal conductivity in cryogenic research. We also investigated a new technique in permanently modifying the operational frequency of a SQUID, and were able to show that this new method of laser ablating the input coil could successfully shift the frequency. While the focus of cryogenic testing was on SQUIDs the cryostat itself was designed in such a way such that many types of low temperature devices could be tested within the system.
The performance of superconducting circuits for quantum computing is limited by materials losses. In particular, coherence times are typically bounded by two-level system (TLS) losses at single photon powers and millikelvin temperatures. The identification of low loss fabrication techniques, materials, and thin film dielectrics is critical to achieving scalable architectures for superconducting quantum computing. Superconducting microwave resonators provide a convenient qubit proxy for assessing performance and studying TLS loss and other mechanisms relevant to superconducting circuits such as non-equilibrium quasiparticles and magnetic flux vortices. In this review article, we provide an overview of considerations for designing accurate resonator experiments to characterize loss, including applicable types of losses, cryogenic setup, device design, and methods for extracting material and interface losses, summarizing techniques that have been evolving for over two decades. Results from measurements of a wide variety of materials and processes are also summarized. Finally, we present recommendations for the reporting of loss data from superconducting microwave resonators to facilitate materials comparisons across the field.
Superconducting microwave resonators are of interest for a wide range of applications, including for their use as microwave kinetic inductance detectors (MKIDs) for the detection of faint astrophysical signatures, as well as for quantum computing applications and materials characterization. In this paper, procedures are presented for the fabrication and characterization of thin-film superconducting microwave resonators. The fabrication methodology allows for the realization of superconducting transmission-line resonators with features on both sides of an atomically smooth single-crystal silicon dielectric. This work describes the procedure for the installation of resonator devices into a cryogenic microwave testbed and for cool-down below the superconducting transition temperature. The set-up of the cryogenic microwave testbed allows one to do careful measurements of the complex microwave transmission of these resonator devices, enabling the extraction of the properties of the superconducting lines and dielectric substrate (e.g., internal quality factors, loss and kinetic inductance fractions), which are important for device design and performance.
Microwave experiments in dilution refrigerators are a central tool in the field of superconducting quantum circuits and other research areas. This type of experiments relied so far on attaching a device to the mixing chamber of a dilution refrigerator. The minimum turnaround time in this case is a few days as required by cooling down and warming up the entire refrigerator. We developed a new approach, in which a suitable sample holder is attached to a cold-insertable probe and brought in contact with transmission lines permanently mounted inside the cryostat. The total turnaround time is 8 h if the target temperature is 80 mK. The lowest attainable temperature is 30 mK. Our system can accommodate up to six transmission lines, with a measurement bandwidth tested from zero frequency to 12 GHz. This bandwidth is limited by low-pass components in the setup; we expect the intrinsic bandwidth to be at least 18 GHz. We present our setup, discuss the experimental procedure, and give examples of experiments enabled by this system. This new measurement method will have a major impact on systematic ultra-low temperature studies using microwave signals, including those requiring quantum coherence.
Accurate knowledge of the microwave power delivered to a cryogenic device under test (DUT) is essential for the characterization and operation of superconducting quantum circuits. However, this information is difficult to obtain inside dilution refrigerators because of distributed attenuation, impedance mismatch, switch-path repeatability, and temperature-dependent microwave components. This paper presents an in situ measurement method for RF power at the cryogenic device plane. The method uses a custom variable temperature stage (VTS) as a cryogenic thermal-transfer element. The TVS is alternately heated by a four-wire DC heater and by microwave power dissipated in a 20 dB pass-through attenuator. By fitting the thermal transients and comparing the corresponding steady-state temperatures, the absorbed microwave power is inferred from a directly measured DC electrical power through an AC/DC substitution procedure. The finite reflection and transmission of the attenuator are then accounted for by cryogenic two-port scattering-parameter measurements based on a switch-assisted Short--Open--Load--Reciprocal calibration, so that the result is referred to the DUT reference plane. The system is demonstrated in a dilution refrigerator with powers between -43 and -58 dBm at the DUT input plane. The demonstrated relative standard uncertainty ranges from about 2% at -43.9 dBm to about 40% at -57.6 dBm. The proposed approach combines thermal RF power transfer, cryogenic S-parameter correction, and uncertainty evaluation in a measurement architecture compatible with quantum-device experiments, providing a practical route toward traceable microwave-power calibration at millikelvin stages.