The mechanism of light-shift-related uncertainty in an atom-interferometry gravimeter is investigated systematically, which arises from the inhomogeneous intensity of Raman beams and the Gaussian distribution of an atomic cloud. Experiments using atom interferometry are performed to verify this systematic error by directly modulating the integration time $T$ and the intensity ratio of Raman laser beams. Using our method, the uncertainty of gravity measurements contributed by the single-photon light shift in an atom gravimeter can be corrected to a $\mathrm{sub}\ensuremath{-}\ensuremath{\mu}\mathrm{Gal}\phantom{\rule{0.28em}{0ex}}(1\phantom{\rule{0.28em}{0ex}}\ensuremath{\mu}\mathrm{Gal}=1\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}8}\phantom{\rule{0.28em}{0ex}}\mathrm{m}/{\mathrm{s}}^{2})$ level.