Design of the Ion Optical Components in a Time-of-flight Mass Spectrometer For The Study of Ultracold Chemistry

Davis Holmes

Published Jan 1, 2009• 0 citations
OpenAlex

Abstract

The design of ion optical components for a time-of-flight mass spectrometer (TOF-MS) was motivated by the recent interest in the study of chemical reaction resonances. An optimal setting to study these resonances would have reactants with very well defined kinetic energies. The system proposed to accomplish this uses a magneto-optical trap (MOT) to localize and cool atoms to temperatures below 100µK, which act as targets for a beam of cold molecules created using a hexapole Stark guide. The reaction products would then be ionized through resonance-enhanced multiphoton ionization (REMPI) and analyzed with a TOF-MS. Many obstacles were presented by the high voltage hexapole’s close proximity with the TOF-MS electrodes. However, they were overcome by a careful selection of dimensions and active shielding. Adding the MOT brought about other constraints: off-axis ion detection and open aperature extraction plates. The issues that these generated were solved by employing a set of parallel deflection plates and an Einzel lens. Engineering solutions were tested by simulating ion trajectories in SIMION 3D, a commonly used commerical program in mass spectrometer design. To keep solutions financially feasible, a budget constraint of $100,000 was put in place.

Related papers

An ion trap time-of-flight mass spectrometer with high mass resolution for cold trapped ion experiments

Philipp Christoph Schmid, J. M. Greenberg, Mikhail I. Miller +2 more

Trapping molecular ions that have been sympathetically cooled with laser-cooled atomic ions is a useful platform for exploring cold ion chemistry. We designed and characterized a new experimental apparatus for probing chemical reaction dynamics between molecular cations and neutral radicals at temperatures below 1 K. The ions are trapped in a linear quadrupole radio-frequency trap and sympathetically cooled by co-trapped, laser-cooled, atomic ions. The ion trap is coupled to a time-of-flight mass spectrometer to readily identify product ion species and to accurately determine trapped ion numbers. We discuss, and present in detail, the design of this ion trap time-of-flight mass spectrometer and the electronics required for driving the trap and mass spectrometer. Furthermore, we measure the performance of this system, which yields mass resolutions of m/Δm ≥ 1100 over a wide mass range, and discuss its relevance for future measurements in chemical reaction kinetics and dynamics.

arXiv:1707.07036•Review of Scientific Instruments•Dec 1, 2017•48 citations

Measurement of the isotope selectivity of the strontium magneto-optical trap by a time-of-flight mass spectrometer

Kwang-Hoon Ko, You-Kyoung Lim, Do-Young Jeong +4 more

The isotope selectivity of the strontium (Sr) magneto-optical trap (MOT) is measured by a time-of-flight (TOF) mass spectrometer. The thermally generated Sr atoms are slowed down by the Zeeman slowing technique and trapped by the magneto-optical method. The trapped atoms are photoionized by the nonresonant two-photon ionization, and the generated ions are analyzed by the TOF mass spectrometer. The isotope selectivity of the MOT is up to approximately 2000 for 86Sr, and the overall selectivity of the system, which consists of the MOT and the mass spectrometer is also discussed.

Journal of the Optical Society of America B•Nov 29, 2006•5 citations

Digital ion trap mass spectrometry for cold ion-molecule chemistry

Laura L. Pollum

A promising new approach for studying cold ion-molecule chemical reactions is the combination of laser- or sympathetically-cooled trapped ions and slow-moving molecules from a cold molecule source, such as a quadrupole velocity selector or a Stark decelerator. Previous reaction studies using trapped atomic ions and slow molecules from a quadrupole velocity selector were able to reach average collision energies as low as 1 K. However, the guided molecules had an approximately room temperature rotational energy distribution, so the reactions studied were not truly cold. Thus, a new molecular source for producing translationally and rotationally cold molecules utilizing buffer gas cooling and quadrupole velocity selection was constructed by K. Twyman and characterized for use in cold reaction studies. This new source of cold molecules is referred to as the buffer gas guide. A new ion trap has been designed and built for use with the existing buffer gas guide. The new ion trap apparatus is compact and mechanically compatible with this new guide. It uses a linear Paul ion trap with cylindrical electrodes to trap ions. Two optical axes (one axial and one radial) enable efficient cooling of small ion crystals. A field-free time-of-flight tube and ion detection assembly are also incorporated into the apparatus. A new technique for determining the mass and quantity of trapped ions has also been developed, termed digital ion trap mass spectrometry. The new technique uses a digital RF waveform to trap ions before ejecting the ions radially from the trap using an ejection pulse applied to the trap electrodes. The ions are then detected after free flight along a time-of-flight tube. This technique was characterized by ejecting crystals of various sizes and compositions: Ca+ only, Ca+/CaF +, Ca+/CaOH +/CaOD+, and Ca+/NH +3 /NH +4 /H3O+. A linear relationship between the number of ions ejected (determined by comparing experimental and simulated crystal images) and the integral of the time-of-flight peak was observed for Ca+ and Ca+/CaF +. All mass peaks were resolved. Simulations of the trapped ions and their trajectories through the time-of-flight tube were also performed, and excellent agreement between the simulated and experimental mass resolution was observed. Progress towards combining the buffer gas guide with the previously independent ion trap is also presented. It is anticipated that the combined buffer gas guide ion trap apparatus will enable the study of ion-molecule reactions at low temperatures with translationally and rotationally cold molecules. It is anticipated that the new digital ion trap mass spectrometry technique will simplify the study of reactions when multiple product ions whose masses are separated by only 1 AMU are formed. A new ion trap has been designed and built for use with the existing buffer gas guide. The new ion trap apparatus is compact and mechanically compatible with this new guide. It uses a linear Paul ion trap with cylindrical electrodes to trap ions. Two optical axes (one axial and one radial) enable efficient cooling of small ion crystals. A field-free time-of-flight tube and ion detection assembly are also incorporated into the apparatus. A new technique for determining the mass and quantity of trapped ions has also been developed, termed digital ion trap mass spectrometry. The new technique uses a digital RF waveform to trap ions before ejecting the ions radially from the trap using an ejection pulse applied to the trap electrodes. The ions are then detected after free flight along a time-of-flight tube. This technique was characterized by ejecting crystals of various sizes and compositions: Ca+ only, Ca+/CaF+, Ca+/CaOH+/CaOD+, and Ca+/NH+3/NH+4/H3O+. A linear relationship between the number of ions ejected (determined by comparing experimental and simulated crystal images) and the integral of the time-of-flight peak was observed for Ca+ and Ca+/CaF+. All mass peaks were resolved. Simulations of the trapped ions and their trajectories through the time-of-flight tube were also performed, and excellent agreement between the simulated and experimental mass resolution was observed. Progress towards combining the buffer gas guide with the previously independent ion trap is also presented. It is anticipated that the combined buffer gas guide ion trap apparatus will enable the study of ion-molecule reactions at low temperatures with translationally and rotationally cold molecules. It is anticipated that the new digital ion trap mass spectrometry technique will simplify the study of reactions when multiple product ions whose masses are separated by only 1 AMU are formed.

•Jan 1, 2015•0 citations

A hybrid ion-atom trap with integrated high resolution mass spectrometer

S. Jyothi, Kisra N. Egodapitiya, Brad Bondurant +5 more

In this article, we describe the design, construction, and implementation of our ion-atom hybrid system incorporating a high resolution time of flight mass spectrometer (TOFMS). Potassium atoms (39K) in a magneto optical trap and laser cooled calcium ions (40Ca+) in a linear Paul trap are spatially overlapped, and the combined trap is integrated with a TOFMS for radial extraction and detection of reaction products. We also present some experimental results showing interactions between 39K+ and 39K, 40Ca+ and 39K+, as well as 40Ca+ and 39K pairs. Finally, we discuss prospects for cooling CaH+ molecular ions in the hybrid ion-atom system.

arXiv:1907.10455•Review of Scientific Instruments•Oct 1, 2019•18 citations

Electron Collisions in a Magneto-Optical Trap

Jeffery P. Dech

Measurements of the multiple ionization cross section ratios of Cesium were performed with ion time-of-flight (TOF) spectroscopy with a magneto-optical trap (MOT) apparatus, updating the previous measurement which dates back almost a century. Results are presented for collisions at energies of 50 eV to 120 eV. With a MOT, experiments can be performed with trapped, cold atomic targets which allow for unparalleled accuracy and experiments with signicant excited state target fractions above those achievable in most atomic beam experiments. A basic overview of optical cooling trapping, electron collision and atomic phenomena are presented. Experimental studies of electrons with Argon and Cesium targets were performed, measuring the multiple ionization ratios with ion TOF spectroscopy. The experimental apparatus and analysis methods are described in detail. Results are compared with previous measurements of multiple ionization ratios for both targets. Agreement within experimental error is found with the results of Tate and Smith across the energy range.

•Jan 1, 2015•0 citations

Controlled quantum-state-resolved chemistry to enable quantum logic with molecular ions

Michael Charles Mills

This dissertation describes experiments performed in the MOTion trap, a hybrid atom--ion trap comprised of a magneto-optical trap (MOT) and a linear Paul trap with an integrated time-of-flight mass spectrometer (ToF-MS).We aim to sympathetically cool the vibrational and rotational degrees of freedom of a BaCl$^+$ molecular ion by overlapping a cold ($\\sim4$~mK) Ca MOT.\nThis sympathetic cooling can prepare a molecular ion in its rovibrational quantum ground state, providing a method for state preparation for applications in molecular ion quantum logic.\nWhile collisions between the molecular ion and the cold atoms can cool the internal degrees of freedom, chemical reactions are also possible for electronically excited atoms.\nAs such, we study these excited-state atom--ion reactions at low temperature $<1$~K and develop tools to control these collisions.\nWe make the first experimental observation of reaction blockading, an effect that suppresses excited-state reactions at low temperature, and develop a method to reverse this suppression if desired with the addition of a catalyst laser, providing a means to optically control such reactions.Beyond its intended purpose of studying and achieving sympathetic cooling, we find that this apparatus is a powerful tool for chemical studies.By imaging the ions, we can observe reactions occurring one atom at a time.\nWe can precisely tune the collision energy or temperature down to the mK regime to as high as $\\sim$10~K using techniques described in this dissertation, allowing controlled interrogation of these low-temperature reactions.\nFurther, using lasers to address electronic transitions, we can precisely control the quantum state of the reactants, and we can measure the reaction products with the ToF-MS.\nA testament to the value of this apparatus is the study of BaOCa$^+$ described in this dissertation.\nUsing the tools of the MOTion trap, we were able to identify the production of BaOCa$^+$, the first observation of a mixed hypermetallic oxide, and determine the path of reaction as BaOCH$_3^+$ + Ca($4s4p$~$^3P_J$) $\\rightarrow$ BaOCa$^+$ + CH$_3$.In addition to developing control of these atom--ion interactions towards sympathetic cooling of molecular ions, we also work towards further developing new quantum logic schemes for molecular ions, such as dipolar quantum logic, dipole--phonon quantum logic (DPQL), and electric-field gradient gates (EGGs).Specifically, we propose a class of molecules particularly well-suited for DPQL, the alkaline earth monoxide cations, that can also be added to existing atomic ion experiments with minimal additional experimental complexity.\nWe further develop the techniques for DPQL and propose methods for single- and two-qubit gates, outlining a path towards universal quantum computation with molecular ions.

•Jan 1, 2020•0 citations
The design of ion optical components for a time-of-flight mass spectrometer (TOF-MS) was motivated by the recent interest in the study of chemical reaction resonances. An optimal setting to study these resonances would have reactants with very well defined kinetic energies. The system proposed to accomplish this uses a magneto-optical trap (MOT) to localize and cool atoms to temperatures below 100µK, which act as targets for a beam of cold molecules created using a hexapole Stark guide. The reaction products would then be ionized through resonance-enhanced multiphoton ionization (REMPI) and analyzed with a TOF-MS. Many obstacles were presented by the high voltage hexapole’s close proximity with the TOF-MS electrodes. However, they were overcome by a careful selection of dimensions and active shielding. Adding the MOT brought about other constraints: off-axis ion detection and open aperature extraction plates. The issues that these generated were solved by employing a set of parallel deflection plates and an Einzel lens. Engineering solutions were tested by simulating ion trajectories in SIMION 3D, a commonly used commerical program in mass spectrometer design. To keep solutions financially feasible, a budget constraint of $100,000 was put in place.
Trapping molecular ions that have been sympathetically cooled with laser-cooled atomic ions is a useful platform for exploring cold ion chemistry. We designed and characterized a new experimental apparatus for probing chemical reaction dynamics between molecular cations and neutral radicals at temperatures below 1 K. The ions are trapped in a linear quadrupole radio-frequency trap and sympathetically cooled by co-trapped, laser-cooled, atomic ions. The ion trap is coupled to a time-of-flight mass spectrometer to readily identify product ion species and to accurately determine trapped ion numbers. We discuss, and present in detail, the design of this ion trap time-of-flight mass spectrometer and the electronics required for driving the trap and mass spectrometer. Furthermore, we measure the performance of this system, which yields mass resolutions of m/Δm ≥ 1100 over a wide mass range, and discuss its relevance for future measurements in chemical reaction kinetics and dynamics.
The isotope selectivity of the strontium (Sr) magneto-optical trap (MOT) is measured by a time-of-flight (TOF) mass spectrometer. The thermally generated Sr atoms are slowed down by the Zeeman slowing technique and trapped by the magneto-optical method. The trapped atoms are photoionized by the nonresonant two-photon ionization, and the generated ions are analyzed by the TOF mass spectrometer. The isotope selectivity of the MOT is up to approximately 2000 for 86Sr, and the overall selectivity of the system, which consists of the MOT and the mass spectrometer is also discussed.
A promising new approach for studying cold ion-molecule chemical reactions is the combination of laser- or sympathetically-cooled trapped ions and slow-moving molecules from a cold molecule source, such as a quadrupole velocity selector or a Stark decelerator. Previous reaction studies using trapped atomic ions and slow molecules from a quadrupole velocity selector were able to reach average collision energies as low as 1 K. However, the guided molecules had an approximately room temperature rotational energy distribution, so the reactions studied were not truly cold. Thus, a new molecular source for producing translationally and rotationally cold molecules utilizing buffer gas cooling and quadrupole velocity selection was constructed by K. Twyman and characterized for use in cold reaction studies. This new source of cold molecules is referred to as the buffer gas guide. A new ion trap has been designed and built for use with the existing buffer gas guide. The new ion trap apparatus is compact and mechanically compatible with this new guide. It uses a linear Paul ion trap with cylindrical electrodes to trap ions. Two optical axes (one axial and one radial) enable efficient cooling of small ion crystals. A field-free time-of-flight tube and ion detection assembly are also incorporated into the apparatus. A new technique for determining the mass and quantity of trapped ions has also been developed, termed digital ion trap mass spectrometry. The new technique uses a digital RF waveform to trap ions before ejecting the ions radially from the trap using an ejection pulse applied to the trap electrodes. The ions are then detected after free flight along a time-of-flight tube. This technique was characterized by ejecting crystals of various sizes and compositions: Ca+ only, Ca+/CaF +, Ca+/CaOH +/CaOD+, and Ca+/NH +3 /NH +4 /H3O+. A linear relationship between the number of ions ejected (determined by comparing experimental and simulated crystal images) and the integral of the time-of-flight peak was observed for Ca+ and Ca+/CaF +. All mass peaks were resolved. Simulations of the trapped ions and their trajectories through the time-of-flight tube were also performed, and excellent agreement between the simulated and experimental mass resolution was observed. Progress towards combining the buffer gas guide with the previously independent ion trap is also presented. It is anticipated that the combined buffer gas guide ion trap apparatus will enable the study of ion-molecule reactions at low temperatures with translationally and rotationally cold molecules. It is anticipated that the new digital ion trap mass spectrometry technique will simplify the study of reactions when multiple product ions whose masses are separated by only 1 AMU are formed. A new ion trap has been designed and built for use with the existing buffer gas guide. The new ion trap apparatus is compact and mechanically compatible with this new guide. It uses a linear Paul ion trap with cylindrical electrodes to trap ions. Two optical axes (one axial and one radial) enable efficient cooling of small ion crystals. A field-free time-of-flight tube and ion detection assembly are also incorporated into the apparatus. A new technique for determining the mass and quantity of trapped ions has also been developed, termed digital ion trap mass spectrometry. The new technique uses a digital RF waveform to trap ions before ejecting the ions radially from the trap using an ejection pulse applied to the trap electrodes. The ions are then detected after free flight along a time-of-flight tube. This technique was characterized by ejecting crystals of various sizes and compositions: Ca+ only, Ca+/CaF+, Ca+/CaOH+/CaOD+, and Ca+/NH+3/NH+4/H3O+. A linear relationship between the number of ions ejected (determined by comparing experimental and simulated crystal images) and the integral of the time-of-flight peak was observed for Ca+ and Ca+/CaF+. All mass peaks were resolved. Simulations of the trapped ions and their trajectories through the time-of-flight tube were also performed, and excellent agreement between the simulated and experimental mass resolution was observed. Progress towards combining the buffer gas guide with the previously independent ion trap is also presented. It is anticipated that the combined buffer gas guide ion trap apparatus will enable the study of ion-molecule reactions at low temperatures with translationally and rotationally cold molecules. It is anticipated that the new digital ion trap mass spectrometry technique will simplify the study of reactions when multiple product ions whose masses are separated by only 1 AMU are formed.
In this article, we describe the design, construction, and implementation of our ion-atom hybrid system incorporating a high resolution time of flight mass spectrometer (TOFMS). Potassium atoms (39K) in a magneto optical trap and laser cooled calcium ions (40Ca+) in a linear Paul trap are spatially overlapped, and the combined trap is integrated with a TOFMS for radial extraction and detection of reaction products. We also present some experimental results showing interactions between 39K+ and 39K, 40Ca+ and 39K+, as well as 40Ca+ and 39K pairs. Finally, we discuss prospects for cooling CaH+ molecular ions in the hybrid ion-atom system.
Measurements of the multiple ionization cross section ratios of Cesium were performed with ion time-of-flight (TOF) spectroscopy with a magneto-optical trap (MOT) apparatus, updating the previous measurement which dates back almost a century. Results are presented for collisions at energies of 50 eV to 120 eV. With a MOT, experiments can be performed with trapped, cold atomic targets which allow for unparalleled accuracy and experiments with signicant excited state target fractions above those achievable in most atomic beam experiments. A basic overview of optical cooling trapping, electron collision and atomic phenomena are presented. Experimental studies of electrons with Argon and Cesium targets were performed, measuring the multiple ionization ratios with ion TOF spectroscopy. The experimental apparatus and analysis methods are described in detail. Results are compared with previous measurements of multiple ionization ratios for both targets. Agreement within experimental error is found with the results of Tate and Smith across the energy range.
This dissertation describes experiments performed in the MOTion trap, a hybrid atom--ion trap comprised of a magneto-optical trap (MOT) and a linear Paul trap with an integrated time-of-flight mass spectrometer (ToF-MS).We aim to sympathetically cool the vibrational and rotational degrees of freedom of a BaCl+^++ molecular ion by overlapping a cold (sim4\\sim4sim4~mK) Ca MOT.\nThis sympathetic cooling can prepare a molecular ion in its rovibrational quantum ground state, providing a method for state preparation for applications in molecular ion quantum logic.\nWhile collisions between the molecular ion and the cold atoms can cool the internal degrees of freedom, chemical reactions are also possible for electronically excited atoms.\nAs such, we study these excited-state atom--ion reactions at low temperature <1<1<1~K and develop tools to control these collisions.\nWe make the first experimental observation of reaction blockading, an effect that suppresses excited-state reactions at low temperature, and develop a method to reverse this suppression if desired with the addition of a catalyst laser, providing a means to optically control such reactions.Beyond its intended purpose of studying and achieving sympathetic cooling, we find that this apparatus is a powerful tool for chemical studies.By imaging the ions, we can observe reactions occurring one atom at a time.\nWe can precisely tune the collision energy or temperature down to the mK regime to as high as $\\sim$10~K using techniques described in this dissertation, allowing controlled interrogation of these low-temperature reactions.\nFurther, using lasers to address electronic transitions, we can precisely control the quantum state of the reactants, and we can measure the reaction products with the ToF-MS.\nA testament to the value of this apparatus is the study of BaOCa$^+$ described in this dissertation.\nUsing the tools of the MOTion trap, we were able to identify the production of BaOCa+^++, the first observation of a mixed hypermetallic oxide, and determine the path of reaction as BaOCH3+_3^+3+​ + Ca(4s4p4s4p4s4p~3PJ^3P_J3PJ​) rightarrow\\rightarrowrightarrow BaOCa+^++ + CH3_33​.In addition to developing control of these atom--ion interactions towards sympathetic cooling of molecular ions, we also work towards further developing new quantum logic schemes for molecular ions, such as dipolar quantum logic, dipole--phonon quantum logic (DPQL), and electric-field gradient gates (EGGs).Specifically, we propose a class of molecules particularly well-suited for DPQL, the alkaline earth monoxide cations, that can also be added to existing atomic ion experiments with minimal additional experimental complexity.\nWe further develop the techniques for DPQL and propose methods for single- and two-qubit gates, outlining a path towards universal quantum computation with molecular ions.