Recovery technique for saturated masers
Günter Wessel
Abstract
The practical application of masers is often hindered by saturation effects of the maser material. Any strong signal frequency (e.g., the transmitter pulse of a radar system leaking into the maser) will cause saturation. This paralyzes the maser for a period of time, too long for most intended applications. In this report, a method is described which has been used in desaturating a four-level ruby maser. The saturation is caused by equalizing the populations of levels one, two, and three of the maser crystal by the leakage of the transmitter pulse, in addition to the action of the maser pump. The application of a desaturation pulse between levels one and four will then desaturate the maser restoring the excess population density between levels two and one. In a cavity maser, full recovery of the amplification capabilities has been achieved within less than 1 msec after the maser has been saturated. Repetition rates up to 120 cps are tolerable.