Review of Scientific Instruments

A dead-zone free 4He atomic magnetometer with intensity-modulated linearly polarized light and a liquid crystal polarization rotator

Teng Wu, Xiang Peng, Zaisheng Lin, Hong Mei Guo

Published Oct 1, 2015• 25 citations
Publisher

Abstract

We demonstrate an all-optical (4)He atomic magnetometer experimental scheme based on an original Bell-Bloom configuration. A single intensity-modulated linearly polarized laser beam is used both for generating spin polarization within a single (4)He vapor and probing the spin precessing under a static magnetic field. The transmitted light signal from the vapor is then phase-sensitively detected at the modulation frequency and its harmonics, which lead to the atomic magnetic resonance signals. Based on this structure, a liquid crystal is added in our magnetometer system and constitutes a polarization rotator. By controlling the voltage applied on the liquid crystal, the light linear polarization vector can be kept perpendicular with the ambient magnetic field direction, which in turn provides the maximum resonance signal amplitude. Moreover, the system exhibits a magnetic-field noise floor of about 2pT/√Hz, which is not degraded due to the presence of the liquid crystal and varying magnetic field direction. The experiment results prove that our method can eliminate the dead-zone effect, improve the system spatial isotropy, and thus be suitable in mobile applications.

DOI
10.1063/1.4932528

Related papers

Dead-zone-free single-beam atomic magnetometer based on free-induction-decay of Rb atoms

Shrey Mehta, G. K. Samanta, Raghwinder Singh Grewal

Free-induction-decay (FID) magnetometers have evolved as simple magnetic sensors for sensitive detection of unknown magnetic fields. However, these magnetometers suffer from a fundamental problem known as a “dead zone,” making them insensitive to certain magnetic field directions. Here, we demonstrate a simple experimental scheme for the dead-zone-free operation of a FID atomic magnetometer. Using a single laser beam containing equal strength of linear- and circular-polarization components and amplitude modulation at a low-duty cycle, we have synchronously pumped the rubidium (87Rb) atoms with both first- and second-order frequency harmonics. Such a pumping scheme has enabled us to observe the free Larmor precession of atomic spins at a frequency of ΩL (orientation) and/or 2ΩL (alignment) in a single FID signal, depending on the direction of the external magnetic field. We observed that the amplitude of the FID signal does not go to zero for any magnetic field direction, proving the absence of dead zones in the magnetometer. The magnetometer has a sensitivity in the range of 3.2–8.4 pT/Hz in all directions. Our experimental scheme can be crucial in developing miniaturized atomic magnetometers for various practical applications, including geomagnetic applications.

arXiv:2410.15663•Applied Physics Letters•Jan 27, 2025•8 citations

Dead-zone-free orientation-based pulsed atomic magnetometer

Jinghong Xu, Liwei Jiang, Yuntian Zou +6 more

The dead zones of atomic magnetometers bring great challenges for the mobile applications in high-sensitivity magnetic field measurement. To solve this fundamental problem, we present a dead-zone-free orientation-based atomic magnetometer employing free-induction-decay (FID) scheme. A modulation scheme combining synchronous optical pumping and radio frequency pulse manipulation is utilized to achieve the free precession of spin polarization in all magnetic field orientations, and the FID signals are detected in two directions with an orthogonal pump-probe light configuration. We perform a comprehensive theoretical analysis of spin polarization evolution for arbitrary magnetic field orientations, and both simulation and experimental results demonstrate that the fusion of information from the pump and probe beams enables the dead-zone-free measurement of the magnetometer, with the sensor exhibiting a sensitivity in the range of 0.6 – 1.4 pT / Hz 1 / 2 across all directions. The proposed scheme is of great significance for the advancement of pulsed atomic magnetometers and their various applications in mobile platforms.

Physical Review Research•Apr 6, 2026•0 citations

Total-field atomic magnetometry using an elliptically-polarized frequency-modulated light beam for geomagnetic-field applications

Denis V. Brazhnikov, A. O. Makarov, K. S. Kozlova +10 more

We investigate an all-optical scheme for high-sensitivity measurements of the Earth's scale magnetic field ($B$). The scheme is based on the Bell-Bloom technique with frequency-modulated light. A single elliptically polarized light beam is used both for pumping the alkali-metal atoms and probing Larmor precession of their spins under the external magnetic field. In contrast to many other high-sensitivity magnetometry schemes based on nonlinear polarization rotation of the light, the proposed approach enables the observation of magnetic resonance via changes in the light-beam ellipticity parameter. The approach has been validated in experiments with a vertical-cavity surface-emitting laser, irradiating a $0.125$ cm$^3$ cesium vapor cell filled with a buffer gas. In the current experimental conditions, the achieved sensitivity is estimated at $220$ fT/$\surd$Hz under $B$$\,\approx\,$$50$ $μ$T, while the shot-noise floor corresponds to $\approx\,$$20$ fT/$\surd$Hz. The bandwidth is estimated at $\approx\,$$1$ kHz. The proposed simple and robust single-beam scheme is well suited for miniaturization and is therefore particularly promising for the development of compact, highly sensitive, low-power-consumption atomic magnetometers for a wide range of applications in the geomagnetic field.

arXiv:2610.08031•Oct 6, 2026•0 citations

Dead-zone-free atomic magnetometer based on hybrid Poincaré beams

Ke Tian, Weifeng Ding, Zhaoying Wang

In this paper, we present the experiment and the theory scheme of light-atom interaction in atomic magnetometers by using a hybrid Poincaré beam (HPB) to solve an annoying problem, named “dead zone.” This kind of magnetometer can be sensitive to arbitrary directions of external magnetic fields. The HPB has a complex polarization distribution, consisting of a vector radially polarized beam and a scalar circularly polarized beam in our experiment. These two kinds of beams have different directions of dead zones of external magnetic fields; thereby, the atomic magnetometer with an HPB can avoid the non-signal area when the direction of the external magnetic field is in the plane perpendicular to the light polarization plane. Furthermore, the optical magnetic resonance (OMR) signal using an HPB still has no dead zones even when the direction of the external magnetic field is in the plane parallel to the polarization plane in our scheme. Our work has the potential to simplify and optimize dead-zone-free atomic magnetometers.

Photonics Research•Mar 4, 2024•16 citations

Bell–Bloom magnetometer based on square optical path

Weipeng Li, Yue Chang, Xiaonan Zhang +3 more

We present a single-beam Bell-Bloom magnetometer using a compact, square optical path in a vapor cell with intensity modulation of the light beam. This design achieves dead-zone-free three-axis magnetic field measurements and, additionally, can determine the magnitude and direction of the magnetic field gradient. Experimental results demonstrate the dead-zone-free operation, with sensitivity in each axis scaling with the effective optical path length along that direction. Strong magnetic field gradients induce distinct resonance peak splitting, whereby the splitting width is proportional to the gradient magnitude. By applying a reference gradient, the system can infer the direction of an external gradient, effectively functioning as a dual-axis gradiometer. Rotating the vapor cell allows optimization of the optical path to minimize gradient-induced measurement distortions by aligning it perpendicular to the gradient direction.

Optics Express•Jul 18, 2025•0 citations

All-optical vector atomic magnetometer

E. Zhivun, Brian Patton, Chris Hovde +1 more

We demonstrate an all-optical magnetometer capable of measuring the magnitude and direction of a magnetic field using nonlinear magneto-optical rotation in a cesium vapor. Vector capability is added by effective modulation of the field along orthogonal axes and subsequent demodulation of the magnetic-resonance frequency. This modulation is provided by the AC Stark shift induced by circularly polarized laser beams. The sensor exhibits a demonstrated rms noise floor of 50 fT/ √ Hz in measurement of the field magnitude and 0.5 mrad/ √ Hz in the field direction; elimination of technical noise would improve these sensitivities to 12 fT/ √ Hz and 5 μrad/ √ Hz, respectively. Applications for a precise all-optical vector magnetometer would include magnetically sensitive fundamental physics experiments, such as the search for a permanent electric dipole moment of the neutron.

Bulletin of the American Physical Society•Jun 5, 2014•21 citations
We demonstrate an all-optical (4)He atomic magnetometer experimental scheme based on an original Bell-Bloom configuration. A single intensity-modulated linearly polarized laser beam is used both for generating spin polarization within a single (4)He vapor and probing the spin precessing under a static magnetic field. The transmitted light signal from the vapor is then phase-sensitively detected at the modulation frequency and its harmonics, which lead to the atomic magnetic resonance signals. Based on this structure, a liquid crystal is added in our magnetometer system and constitutes a polarization rotator. By controlling the voltage applied on the liquid crystal, the light linear polarization vector can be kept perpendicular with the ambient magnetic field direction, which in turn provides the maximum resonance signal amplitude. Moreover, the system exhibits a magnetic-field noise floor of about 2pT/√Hz, which is not degraded due to the presence of the liquid crystal and varying magnetic field direction. The experiment results prove that our method can eliminate the dead-zone effect, improve the system spatial isotropy, and thus be suitable in mobile applications.
Free-induction-decay (FID) magnetometers have evolved as simple magnetic sensors for sensitive detection of unknown magnetic fields. However, these magnetometers suffer from a fundamental problem known as a “dead zone,” making them insensitive to certain magnetic field directions. Here, we demonstrate a simple experimental scheme for the dead-zone-free operation of a FID atomic magnetometer. Using a single laser beam containing equal strength of linear- and circular-polarization components and amplitude modulation at a low-duty cycle, we have synchronously pumped the rubidium (87Rb) atoms with both first- and second-order frequency harmonics. Such a pumping scheme has enabled us to observe the free Larmor precession of atomic spins at a frequency of ΩL (orientation) and/or 2ΩL (alignment) in a single FID signal, depending on the direction of the external magnetic field. We observed that the amplitude of the FID signal does not go to zero for any magnetic field direction, proving the absence of dead zones in the magnetometer. The magnetometer has a sensitivity in the range of 3.2–8.4 pT/Hz in all directions. Our experimental scheme can be crucial in developing miniaturized atomic magnetometers for various practical applications, including geomagnetic applications.
The dead zones of atomic magnetometers bring great challenges for the mobile applications in high-sensitivity magnetic field measurement. To solve this fundamental problem, we present a dead-zone-free orientation-based atomic magnetometer employing free-induction-decay (FID) scheme. A modulation scheme combining synchronous optical pumping and radio frequency pulse manipulation is utilized to achieve the free precession of spin polarization in all magnetic field orientations, and the FID signals are detected in two directions with an orthogonal pump-probe light configuration. We perform a comprehensive theoretical analysis of spin polarization evolution for arbitrary magnetic field orientations, and both simulation and experimental results demonstrate that the fusion of information from the pump and probe beams enables the dead-zone-free measurement of the magnetometer, with the sensor exhibiting a sensitivity in the range of 0.6 – 1.4 pT / Hz 1 / 2 across all directions. The proposed scheme is of great significance for the advancement of pulsed atomic magnetometers and their various applications in mobile platforms.
We investigate an all-optical scheme for high-sensitivity measurements of the Earth's scale magnetic field (BBB). The scheme is based on the Bell-Bloom technique with frequency-modulated light. A single elliptically polarized light beam is used both for pumping the alkali-metal atoms and probing Larmor precession of their spins under the external magnetic field. In contrast to many other high-sensitivity magnetometry schemes based on nonlinear polarization rotation of the light, the proposed approach enables the observation of magnetic resonance via changes in the light-beam ellipticity parameter. The approach has been validated in experiments with a vertical-cavity surface-emitting laser, irradiating a 0.1250.1250.125 cm3^33 cesium vapor cell filled with a buffer gas. In the current experimental conditions, the achieved sensitivity is estimated at 220220220 fT/√\surd√Hz under BBB ≈ \,\approx\,≈505050 μμμT, while the shot-noise floor corresponds to ≈ \approx\,≈202020 fT/√\surd√Hz. The bandwidth is estimated at ≈ \approx\,≈111 kHz. The proposed simple and robust single-beam scheme is well suited for miniaturization and is therefore particularly promising for the development of compact, highly sensitive, low-power-consumption atomic magnetometers for a wide range of applications in the geomagnetic field.
In this paper, we present the experiment and the theory scheme of light-atom interaction in atomic magnetometers by using a hybrid Poincaré beam (HPB) to solve an annoying problem, named “dead zone.” This kind of magnetometer can be sensitive to arbitrary directions of external magnetic fields. The HPB has a complex polarization distribution, consisting of a vector radially polarized beam and a scalar circularly polarized beam in our experiment. These two kinds of beams have different directions of dead zones of external magnetic fields; thereby, the atomic magnetometer with an HPB can avoid the non-signal area when the direction of the external magnetic field is in the plane perpendicular to the light polarization plane. Furthermore, the optical magnetic resonance (OMR) signal using an HPB still has no dead zones even when the direction of the external magnetic field is in the plane parallel to the polarization plane in our scheme. Our work has the potential to simplify and optimize dead-zone-free atomic magnetometers.
We present a single-beam Bell-Bloom magnetometer using a compact, square optical path in a vapor cell with intensity modulation of the light beam. This design achieves dead-zone-free three-axis magnetic field measurements and, additionally, can determine the magnitude and direction of the magnetic field gradient. Experimental results demonstrate the dead-zone-free operation, with sensitivity in each axis scaling with the effective optical path length along that direction. Strong magnetic field gradients induce distinct resonance peak splitting, whereby the splitting width is proportional to the gradient magnitude. By applying a reference gradient, the system can infer the direction of an external gradient, effectively functioning as a dual-axis gradiometer. Rotating the vapor cell allows optimization of the optical path to minimize gradient-induced measurement distortions by aligning it perpendicular to the gradient direction.
We demonstrate an all-optical magnetometer capable of measuring the magnitude and direction of a magnetic field using nonlinear magneto-optical rotation in a cesium vapor. Vector capability is added by effective modulation of the field along orthogonal axes and subsequent demodulation of the magnetic-resonance frequency. This modulation is provided by the AC Stark shift induced by circularly polarized laser beams. The sensor exhibits a demonstrated rms noise floor of 50 fT/ √ Hz in measurement of the field magnitude and 0.5 mrad/ √ Hz in the field direction; elimination of technical noise would improve these sensitivities to 12 fT/ √ Hz and 5 μrad/ √ Hz, respectively. Applications for a precise all-optical vector magnetometer would include magnetically sensitive fundamental physics experiments, such as the search for a permanent electric dipole moment of the neutron.