Conference on Lasers and Electro-Optics Pacific Rim (CLEO-PR)

Bidirectional coincidence counting of correlated photons to determine the detection efficiency of a single-photon detector without a reference standard

Akio Yoshizawa, Satoru Odate, Daiji Fukuda, Hidemi Tsuchida

Published Aug 1, 2007• 0 citations
Publisher

Abstract

Bidirectional coincidence counting of correlated photons was adopted to determine the detection efficiency of single-photon detectors operating at 1550 nm without comparing to a reference standard. A self-calibration method best suitable for single-photon detectors was established.
DOI
10.1109/cleopr.2007.4391563

Related papers

Quantum efficiency measurements by bidirectional coincidence counting of correlated photon pairs

Satoru Odate, Akio Yoshizawa, Daiji Fukuda +1 more

We propose and demonstrate a procedure for characterizing the quantum efficiency of a single-photon detector in the telecommunication wavelength band. Our procedure employs a bidirectional coincidence counting technique to distinguish optical component losses from the detection efficiency. The standard deviations of the measured quantum efficiencies were nearly identical to the standard deviations derived from a detection probability having a Poisson distribution.

Optics Letters•Oct 25, 2007•8 citations

Absolute self-calibration of the quantum efficiency of single-photon detectors

Xihao Chen, Yanhua Zhai, Da Zhang +1 more

Single-photon detectors are finding increasingly wide application in all branches of science and technology thus a precise knowledge of their quantum efficiency is an essential prerequisite. We have performed the first proof-of-principle demonstration of a means to determine the quantum efficiency by which no second detector or reference standard is required. This absolute self-calibration method is based on the time correlation of entangled photons pairs emitted in spontaneous parametric down conversion. The measured quantum efficiency of a Si avalanche photodiode single-photon detector agrees well with the product specifications.

Optics Letters•Jul 25, 2006•21 citations

Absolute calibration of single-photon and multiplexed photon-number-resolving detectors

Lior Cohen, Y. Pilnyak, D. Istrati +3 more

Single-photon detectors are widely used in modern quantum optics experiments and applications. Like all detectors, it is important for these devices to be accurately calibrated. A single-photon detector is calibrated by determining its detection efficiency; the common method to measure this quantity requires comparison to another detector. Here, we suggest a method to measure the detection efficiency of a single-photon detector without requiring an external reference detector. Our method is valid for individual single-photon detectors as well as multiplexed detectors, which are known to be photon number resolving. The method exploits the even-number photon-statistics of a nonlinear source, as well as the nonlinear loss of a single-photon detector that occurs when multiple photons are incident simultaneously. We have analytically modeled multiplexed detectors and used the results to experimentally demonstrate the calibration of a single-photon detector without the need for an external reference detector.

arXiv:1711.03594•Physical Review A•Jul 5, 2018•27 citations

Calibration of a single-photon counting detectors without the need of input photon flux calibration (Conference Presentation)

Thomas Gerrits

Calibration of fiber-coupled single-photon detectors usually requires knowledge of the input photon flux inside the fiber and/or knowledge of the linearity of a reference power meter. Many approaches have been presented in the past to accurately measure the photon detection probability of a single photon detector [1-6]. Under certain assumptions, one can utilize waveguide-coupled single photon detectors and a series of photon-counting measurements and a single-photon source to calibrate the detection efficiency of a single photon detector without the need of a reference power meter and the knowledge of the incoming photon flux. Here, this method is presented. Furthermore, if a reference detector is used, the detection efficiency of all evanescently coupled waveguide detectors can be measured, and the measurement outcome does not depend on splicing or fiber connection losses within in the setup, i.e., the measurement is setup-independent. In addition, the method, when using a reference detector, can be utilized to measure and distinguish between the absorption of a waveguide-coupled single photon detector and its internal detection efficiency. [1] A. J. Miller et al, Opt. Express 19, 9102–9110 (2011) [2] I. Muller et al., Metrologia 51, S329 (2014). [3] A. L. Migdall, Instrumentation and Measurement, IEEE Transactions on 50, 478–481 (2001). [4] S. V. Polyakov, A. L. Migdall, Optics Express 15, 1390–1407 (2007). [5] A. Avella et al., Optics Express 19, 23249–23257 (2011). [6] T. Lunghi et al., Opt. Express 22, 18078-18092 (2014)

•Jun 9, 2017•0 citations

Measurement of InGaAs single photon detector quantum efficiency at 1550nm

Kun Zhao, Changming Liu, Haidong Chen +1 more

Single photon detectors nowadays are widely used in numerous applications such as quantum cryptography, laser ranging, single molecule spectroscopy and so on. And the calibration of the detection efficiency is quite important for some of the applications. In recent years, researchers find that correlated photon pairs can be employed to calibrate the detection efficiency of single-photon detectors with quite high precision. Firstly, we calibrated the InGaAs single photon detector quantum efficiency at 1550nm by correlated photon pairs. Consistency between the measurement result using correlated photon pairs and the reference value is better than 0.8%. Secondly, the detectors are calibrated with an attenuated laser source. Thirdly, the two methods for measuring the quantum efficiency of single photon detectors, correlated photon pairs and laser attenuation, respectively, are discussed and analyzed in detail. Finally, results derived from the two methods have been compared with each other.

Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE•Feb 20, 2015•1 citation

Absolute self-calibration of the quantum efficiency of single-photon detectors

Chang Jun-Tao, Ling-An Wu, 中国科学院物理研究所光物理开放实验室,北京 100080

An absolute self-calibration method is proposed for determining the quantum efficiency of a single photon detector which does not require any standard referenc e or a second detector. The method is based on the time correlation of photon pa irs generated in parametric downconversion. A beam of frequency downconverted ph otons is split at a polarizing beamsplitter then recombined at a second beamspli tter, but with a time delay introduced between them so that the two photons of a n original pair enter the detector one by one. The photoelectric pulses from the detector are split into three paths; a compensating delay is introduced into tw o of the paths so that the two signals arrive simultaneously at a coincidence co unter; the third path goes directly to a photon counter. From the ratio of the r eadings of the two counters we can derive the quantum efficiency of the detector . Two methods are given, one for continuous wave lasers and another for high rep etition rate pulsed lasers.

Acta Physica Sinica•Jan 1, 2003•5 citations
Single-photon detectors are finding increasingly wide application in all branches of science and technology thus a precise knowledge of their quantum efficiency is an essential prerequisite. We have performed the first proof-of-principle demonstration of a means to determine the quantum efficiency by which no second detector or reference standard is required. This absolute self-calibration method is based on the time correlation of entangled photons pairs emitted in spontaneous parametric down conversion. The measured quantum efficiency of a Si avalanche photodiode single-photon detector agrees well with the product specifications.
Single-photon detectors are widely used in modern quantum optics experiments and applications. Like all detectors, it is important for these devices to be accurately calibrated. A single-photon detector is calibrated by determining its detection efficiency; the common method to measure this quantity requires comparison to another detector. Here, we suggest a method to measure the detection efficiency of a single-photon detector without requiring an external reference detector. Our method is valid for individual single-photon detectors as well as multiplexed detectors, which are known to be photon number resolving. The method exploits the even-number photon-statistics of a nonlinear source, as well as the nonlinear loss of a single-photon detector that occurs when multiple photons are incident simultaneously. We have analytically modeled multiplexed detectors and used the results to experimentally demonstrate the calibration of a single-photon detector without the need for an external reference detector.
Calibration of fiber-coupled single-photon detectors usually requires knowledge of the input photon flux inside the fiber and/or knowledge of the linearity of a reference power meter. Many approaches have been presented in the past to accurately measure the photon detection probability of a single photon detector [1-6]. Under certain assumptions, one can utilize waveguide-coupled single photon detectors and a series of photon-counting measurements and a single-photon source to calibrate the detection efficiency of a single photon detector without the need of a reference power meter and the knowledge of the incoming photon flux. Here, this method is presented. Furthermore, if a reference detector is used, the detection efficiency of all evanescently coupled waveguide detectors can be measured, and the measurement outcome does not depend on splicing or fiber connection losses within in the setup, i.e., the measurement is setup-independent. In addition, the method, when using a reference detector, can be utilized to measure and distinguish between the absorption of a waveguide-coupled single photon detector and its internal detection efficiency. [1] A. J. Miller et al, Opt. Express 19, 9102–9110 (2011) [2] I. Muller et al., Metrologia 51, S329 (2014). [3] A. L. Migdall, Instrumentation and Measurement, IEEE Transactions on 50, 478–481 (2001). [4] S. V. Polyakov, A. L. Migdall, Optics Express 15, 1390–1407 (2007). [5] A. Avella et al., Optics Express 19, 23249–23257 (2011). [6] T. Lunghi et al., Opt. Express 22, 18078-18092 (2014)
Single photon detectors nowadays are widely used in numerous applications such as quantum cryptography, laser ranging, single molecule spectroscopy and so on. And the calibration of the detection efficiency is quite important for some of the applications. In recent years, researchers find that correlated photon pairs can be employed to calibrate the detection efficiency of single-photon detectors with quite high precision. Firstly, we calibrated the InGaAs single photon detector quantum efficiency at 1550nm by correlated photon pairs. Consistency between the measurement result using correlated photon pairs and the reference value is better than 0.8%. Secondly, the detectors are calibrated with an attenuated laser source. Thirdly, the two methods for measuring the quantum efficiency of single photon detectors, correlated photon pairs and laser attenuation, respectively, are discussed and analyzed in detail. Finally, results derived from the two methods have been compared with each other.
An absolute self-calibration method is proposed for determining the quantum efficiency of a single photon detector which does not require any standard referenc e or a second detector. The method is based on the time correlation of photon pa irs generated in parametric downconversion. A beam of frequency downconverted ph otons is split at a polarizing beamsplitter then recombined at a second beamspli tter, but with a time delay introduced between them so that the two photons of a n original pair enter the detector one by one. The photoelectric pulses from the detector are split into three paths; a compensating delay is introduced into tw o of the paths so that the two signals arrive simultaneously at a coincidence co unter; the third path goes directly to a photon counter. From the ratio of the r eadings of the two counters we can derive the quantum efficiency of the detector . Two methods are given, one for continuous wave lasers and another for high rep etition rate pulsed lasers.