Acta Physica Sinica

Absolute self-calibration of the quantum efficiency of single-photon detectors

Chang Jun-Tao, Ling-An Wu, 中国科学院物理研究所光物理开放实验室,北京 100080

Published Jan 1, 2003• 5 citations
Publisher

Abstract

An absolute self-calibration method is proposed for determining the quantum efficiency of a single photon detector which does not require any standard referenc e or a second detector. The method is based on the time correlation of photon pa irs generated in parametric downconversion. A beam of frequency downconverted ph otons is split at a polarizing beamsplitter then recombined at a second beamspli tter, but with a time delay introduced between them so that the two photons of a n original pair enter the detector one by one. The photoelectric pulses from the detector are split into three paths; a compensating delay is introduced into tw o of the paths so that the two signals arrive simultaneously at a coincidence co unter; the third path goes directly to a photon counter. From the ratio of the r eadings of the two counters we can derive the quantum efficiency of the detector . Two methods are given, one for continuous wave lasers and another for high rep etition rate pulsed lasers.

DOI
10.7498/aps.52.1132

Related papers

Absolute self-calibration of single-photon detectors with cw and pulsed spontaneous parametric down-conversion

Ling-An Wu, Xi-Hao Chen, Yanhua Zhai +2 more

Single-photon detectors are of great importance in spectroscopy, laser ranging, astronomy and wherever weak light detection is required, so a precise knowledge of their quantum efficiency is essential, but their calibration is by no means trivial. Traditional methods need a black body radiation source or a highly attenuated laser beam as a reference standard, or an already calibrated reference detector. It is well known that absolute measurement of the quantum efficiency, absolute in the sense that the calibration is independent of any radiometric standard or the properties of another detector, can be performed using entangled twin photons generated in optical spontaneous parametric down-conversion together with a reference detector. We have performed the first proof-of-principle experiment to determine the quantum efficiency by which no second detector or reference standard is required. This absolute self-calibration method is also based on the time correlation of photons emitted in parametric down-conversion, but with the introduction of a suitable time delay between the arrival times of the twin photons. Furthermore, we describe a scheme for absolute self-calibration of the quantum efficiency at different wavelengths by means of a pulsed source of nondegenerate down-converted photons. This method should have even greater potential for applications.

Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE•Aug 31, 2006•0 citations

Absolute self-calibration of the quantum efficiency of single-photon detectors

Xihao Chen, Yanhua Zhai, Da Zhang +1 more

Single-photon detectors are finding increasingly wide application in all branches of science and technology thus a precise knowledge of their quantum efficiency is an essential prerequisite. We have performed the first proof-of-principle demonstration of a means to determine the quantum efficiency by which no second detector or reference standard is required. This absolute self-calibration method is based on the time correlation of entangled photons pairs emitted in spontaneous parametric down conversion. The measured quantum efficiency of a Si avalanche photodiode single-photon detector agrees well with the product specifications.

Optics Letters•Jul 25, 2006•21 citations

Absolute self-calib ration of single-photon detectors with cw and pulsed spontaneous parametric down-conversion - art. no. 630505

La Wu, Xiangdong Chen, YH Zhai +2 more

Single-photon detectors are of great importance in spectroscopy, laser ranging, astronomy and wherever weak light detection is required, so a precise knowledge of their quantum efficiency is essential, but their calibration is by no means trivial. Traditional methods need a black body radiation source or a highly attenuated laser beam as a reference standard, or an already calibrated reference detector. It is well known that absolute measurement of the quantum efficiency, absolute in the sense that the calibration is independent of any radiometric standard or the properties of another detector, can be performed using entangled twin photons generated in optical spontaneous parametric down-conversion together with a reference detector. We have performed the first proof-of-principle experiment to determine the quantum efficiency by which no second detector or reference standard is required. This absolute self-calibration method is also based on the time correlation of photons emitted in parametric down-conversion, but with the introduction of a suitable time delay between the arrival times of the twin photons. Furthermore, we describe a scheme for absolute self-calibration of the quantum efficiency at different wavelengths by means of a pulsed source of nondegenerate down-converted photons. This method should have even greater potential for applications.

•Jan 1, 2006•0 citations

Simplified Method of Absolute Self-calibration of theQuantum Efficiency of Single-photon Detectors

Feng Shaohui

Based on former protocols, we propose a simpler way to realize the absolute self-calibration of quantum efficiency of a single photon detector. We replace coincidence counter system with coaxial transmission line and DAQ (data acquisition) device to reduce the complexity. As we know, a pulse will reverse at the short circuit termination. Using this character and setting the coaxial transmission line in precise length, we can distinguish between single pulse and two continuous pulses in one sampling period of DAQ. After recording and analyzing the data, we will get the quantum efficiency of the detector. This protocol need neither any other reference standard nor another detector.

Acta Sinica Quantum Optica•Jan 1, 2004•0 citations

Investigation of absolute measurement in quantum efficiency of photoelectric detectors by means of correlated photons

Yu Feng, Xiaobing Zheng, Yanli Qiao +1 more

There are two general methods for radiometric calibration of detectors. One is based on radiometric sources and the other is based on detectors. Because two methods need to establish primary standard of high precision and transfer chain of standard, the precision of standard will be reduced by extension of the chain. An absolute calibration of quantum efficiency of detectors can be realized by using correlated photons generated in spontaneous parametric down-conversion (SPDC) effect of nonlinear crystal. Generation mechanism of SPDC is interpreted with theories of quantum mechanics and classical theories. The theory of calibration based correlated photons are demonstrated. The project for proof of principle of absolute calibration of quantum efficiency of radiometric detectors is given in this paper. Comparing with traditional methods, the advantages of it are pointed out. A scheme using to calibrate quantum efficiency of infrared detectors is described here. The measurement of spatial and spectral response of detectors is mentioned too.

Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE•Feb 23, 2006•0 citations

Measuring the absolute photodetection efficiency using photon number correlations

Michael Lindenthal, Johannes Kofler

We present two methods for determining the absolute detection efficiency of photon-counting detectors directly from their singles rates under illumination from a nonclassical light source. One method is based on a continuous variable analog to coincidence counting in discrete photon experiments, but it does not actually rely on high detector time resolutions. The second method is based on difference detection, which is a typical detection scheme in continuous variable quantum optics experiments. Since no coincidence detection is required with either method, they are useful for detection efficiency measurements of photodetectors with detector time resolutions far too low to resolve coincidence events.

arXiv:quant-ph/0606041•Applied Optics•Aug 20, 2006•10 citations
An absolute self-calibration method is proposed for determining the quantum efficiency of a single photon detector which does not require any standard referenc e or a second detector. The method is based on the time correlation of photon pa irs generated in parametric downconversion. A beam of frequency downconverted ph otons is split at a polarizing beamsplitter then recombined at a second beamspli tter, but with a time delay introduced between them so that the two photons of a n original pair enter the detector one by one. The photoelectric pulses from the detector are split into three paths; a compensating delay is introduced into tw o of the paths so that the two signals arrive simultaneously at a coincidence co unter; the third path goes directly to a photon counter. From the ratio of the r eadings of the two counters we can derive the quantum efficiency of the detector . Two methods are given, one for continuous wave lasers and another for high rep etition rate pulsed lasers.
Single-photon detectors are of great importance in spectroscopy, laser ranging, astronomy and wherever weak light detection is required, so a precise knowledge of their quantum efficiency is essential, but their calibration is by no means trivial. Traditional methods need a black body radiation source or a highly attenuated laser beam as a reference standard, or an already calibrated reference detector. It is well known that absolute measurement of the quantum efficiency, absolute in the sense that the calibration is independent of any radiometric standard or the properties of another detector, can be performed using entangled twin photons generated in optical spontaneous parametric down-conversion together with a reference detector. We have performed the first proof-of-principle experiment to determine the quantum efficiency by which no second detector or reference standard is required. This absolute self-calibration method is also based on the time correlation of photons emitted in parametric down-conversion, but with the introduction of a suitable time delay between the arrival times of the twin photons. Furthermore, we describe a scheme for absolute self-calibration of the quantum efficiency at different wavelengths by means of a pulsed source of nondegenerate down-converted photons. This method should have even greater potential for applications.
Single-photon detectors are finding increasingly wide application in all branches of science and technology thus a precise knowledge of their quantum efficiency is an essential prerequisite. We have performed the first proof-of-principle demonstration of a means to determine the quantum efficiency by which no second detector or reference standard is required. This absolute self-calibration method is based on the time correlation of entangled photons pairs emitted in spontaneous parametric down conversion. The measured quantum efficiency of a Si avalanche photodiode single-photon detector agrees well with the product specifications.
Single-photon detectors are of great importance in spectroscopy, laser ranging, astronomy and wherever weak light detection is required, so a precise knowledge of their quantum efficiency is essential, but their calibration is by no means trivial. Traditional methods need a black body radiation source or a highly attenuated laser beam as a reference standard, or an already calibrated reference detector. It is well known that absolute measurement of the quantum efficiency, absolute in the sense that the calibration is independent of any radiometric standard or the properties of another detector, can be performed using entangled twin photons generated in optical spontaneous parametric down-conversion together with a reference detector. We have performed the first proof-of-principle experiment to determine the quantum efficiency by which no second detector or reference standard is required. This absolute self-calibration method is also based on the time correlation of photons emitted in parametric down-conversion, but with the introduction of a suitable time delay between the arrival times of the twin photons. Furthermore, we describe a scheme for absolute self-calibration of the quantum efficiency at different wavelengths by means of a pulsed source of nondegenerate down-converted photons. This method should have even greater potential for applications.
Based on former protocols, we propose a simpler way to realize the absolute self-calibration of quantum efficiency of a single photon detector. We replace coincidence counter system with coaxial transmission line and DAQ (data acquisition) device to reduce the complexity. As we know, a pulse will reverse at the short circuit termination. Using this character and setting the coaxial transmission line in precise length, we can distinguish between single pulse and two continuous pulses in one sampling period of DAQ. After recording and analyzing the data, we will get the quantum efficiency of the detector. This protocol need neither any other reference standard nor another detector.
There are two general methods for radiometric calibration of detectors. One is based on radiometric sources and the other is based on detectors. Because two methods need to establish primary standard of high precision and transfer chain of standard, the precision of standard will be reduced by extension of the chain. An absolute calibration of quantum efficiency of detectors can be realized by using correlated photons generated in spontaneous parametric down-conversion (SPDC) effect of nonlinear crystal. Generation mechanism of SPDC is interpreted with theories of quantum mechanics and classical theories. The theory of calibration based correlated photons are demonstrated. The project for proof of principle of absolute calibration of quantum efficiency of radiometric detectors is given in this paper. Comparing with traditional methods, the advantages of it are pointed out. A scheme using to calibrate quantum efficiency of infrared detectors is described here. The measurement of spatial and spectral response of detectors is mentioned too.
We present two methods for determining the absolute detection efficiency of photon-counting detectors directly from their singles rates under illumination from a nonclassical light source. One method is based on a continuous variable analog to coincidence counting in discrete photon experiments, but it does not actually rely on high detector time resolutions. The second method is based on difference detection, which is a typical detection scheme in continuous variable quantum optics experiments. Since no coincidence detection is required with either method, they are useful for detection efficiency measurements of photodetectors with detector time resolutions far too low to resolve coincidence events.