Journal of Instrumentation

Toward sub-Kelvin resistive cooling and non destructive detection of trapped non-neutral electron plasma

S. Di Domizio, D. Krasnický, Vittorio Lagomarsino, G. Testera, Renzo Vaccarone, Sandra Zavatarelli

Published Jan 1, 2015• 4 citations
Publisher

Abstract

A resonant circuit tuned to a particular frequency of the motion of charged particles stored in a Penning trap and connected to a low noise amplifier allows, at the same time, cooling and non destructive detection of the particles. Its use is widely diffused when single or few particles are stored near the centre of a hyperbolic Penning trap. We present a consistent model that predicts the shape of the induced signal when the tuned circuit is used to detect and cool the axial motion of a cold non neutral plasma stored in an open-ended cylindrical Penning trap. The model correctly accounts for the not negligible axial plasma size. We show that the power spectrum of the signal measured across the tuned circuit provides information about the particle number and insights about the plasma temperature. We report on the design of a HEMT-based cryogenic amplifier working at 14.4 MHz and 4.2 K and the results of the noise measurements. We have measured a drain current noise in the range from 6 to 17 pA/√Hz, which corresponds to an increase of the tuned circuit equivalent temperature of at maximum 0.35 K. The cryogenic amplifier has a very low power consumption from few tens to few hundreds of μW corresponding to a drain current in the range 100–800 μ A . An additional contribution due to the gate noise has been identified when the drain current is below 300 μA; above that value an upper limit of the increase of the equivalent tuned circuit temperature due to this contribution of 0.02 K has been obtained. These features make the tuned circuit connected to this amplifier a promising device for detecting and cooling the axial motion of an electron plasma when the Penning trap is mounted inside a dilution refrigerator.

DOI
10.1088/1748-0221/10/01/p01009

Related papers

Cooling and self-excitation of a one-electron oscillator

Brian D’Urso

With a superposition of static electric and magnetic fields, a Penning trap can confine a single electron indefinitely. A weak coupling of the electron to its environment makes the system ideal for precision measurements but renders detection of the electron’s motion difficult. In order to measure fundamental properties of an electron, we employ a Penning trap cooled by a dilution refrigerator. This experimental setup requires extremely low-noise and low-power cryogenic amplifiers for detection of the electron’s axial motion. We design and demonstrate new amplifiers, resulting in a significant decrease of the axial temperature and in an improved signal to noise ratio. We use feedback to cool the axial motion and to self-excite the axial motion for continuous monitoring of the axial frequency. These techniques for cooling and measuring the axial frequency may contribute to an improved measurement of the electron g factor, a more stringent CPT test in leptons, and a more precise measurement of the proton-electron mass ratio.

•Jan 1, 2003•11 citations

Highly sensitive superconducting circuits at ∼700 kHz with tunable quality factors for image-current detection of single trapped antiprotons

Hiroki Nagahama, Schneider Georg, G., Andreas H. Mooser +14 more

We developed highly sensitive image-current detection systems based on superconducting toroidal coils and ultra-low noise amplifiers for non-destructive measurements of the axial frequencies (550–800 kHz) of single antiprotons stored in a cryogenic multi-Penning-trap system. The unloaded superconducting tuned circuits show quality factors of up to 500 000, which corresponds to a factor of 10 improvement compared to our previously used solenoidal designs. Connected to ultra-low noise amplifiers and the trap system, signal-to-noise-ratios of 30 dB at quality factors of >20 000 are achieved. In addition, we have developed a superconducting switch which allows continuous tuning of the detector’s quality factor and to sensitively tune the particle-detector interaction. This allowed us to improve frequency resolution at constant averaging time, which is crucial for single antiproton spin-transition spectroscopy experiments, as well as improved measurements of the proton-to-antiproton charge-to-mass ratio.

•Nov 15, 2016•24 citations

Cryogenic Resonant Amplifier for Electron-on-Helium Image Charge Readout

M. A. Belyanchikov, Jakob A. Kraus, Denis Konstantinov

Abstract An electron-on-helium qubit is a promising physical platform for quantum information technologies. Among all the “blueprints” for the qubit realization, a hybrid Rydberg-spin qubit seems to be a promising one toward quantum computing using electron spins. The main technological challenge on the way to such qubits is a detection of fA range image current induced by a Rydberg transition of a single electron. To address this problem, we aim to use a tank LC-circuit in conjunction with a high-impedance and low power dissipation cryogenic amplifier. Here, we report our progress toward realization of a resonant image current detector with a homemade cryogenic amplifier based on FHX13LG HEMT. We present a detailed characterization of the transistor at room and cryogenic temperatures, as well as details of the amplifier design and performance. At the power dissipation level of amplifier well below 100 μW, the measured voltage and current noise level are $$0.6~{\hbox {nV}}/\sqrt{\textrm{Hz}}$$ 0.6 nV / Hz and below 1.5 fA/ $$\sqrt{\textrm{Hz}}$$ Hz , respectively. Based on the actual image current measurements of the Rydberg transition in a many-electron system on liquid helium, we estimate an SNR = 8 with a measurement bandwidth 1 Hz for the detection of a single-electron transition, providing the noise level at the output is solely determined by the noise of the amplifier.

arXiv:2311.04437•Journal of Low Temperature Physics•Jan 19, 2024•6 citations

Highly sensitive superconducting circuits at ∼700 kHz with tunable quality factors for image-current detection of single trapped antiprotons

H. Nagahama, G. Schneider, A. Mooser +14 more

We developed highly sensitive image-current detection systems based on superconducting toroidal coils and ultra-low noise amplifiers for non-destructive measurements of the axial frequencies (550-800 kHz) of single antiprotons stored in a cryogenic multi-Penning-trap system. The unloaded superconducting tuned circuits show quality factors of up to 500 000, which corresponds to a factor of 10 improvement compared to our previously used solenoidal designs. Connected to ultra-low noise amplifiers and the trap system, signal-to-noise-ratios of 30 dB at quality factors of >20 000 are achieved. In addition, we have developed a superconducting switch which allows continuous tuning of the detector's quality factor and to sensitively tune the particle-detector interaction. This allowed us to improve frequency resolution at constant averaging time, which is crucial for single antiproton spin-transition spectroscopy experiments, as well as improved measurements of the proton-to-antiproton charge-to-mass ratio.

arXiv:1612.08481•Review of Scientific Instruments•Nov 1, 2016•50 citations

Electrostatic modes as a diagnostic in Penning trap experiments

Carl S. Weimer, J. J. Bollinger, David J. Wineland +1 more

A subset of electrostatic modes of a cold cloud of electrons trapped in a Penning trap has been observed and identified using a recent theoretical model. The detection of these modes is accomplished using electronic techniques which could apply to any ion species. The modes are observed in the low density, low rotation limit of the cloud where the electrons approach a two-dimensional charged disk. We observe both axially symmetric and asymmetric drumhead modes. By measuring the frequency of these modes the shape, rotation frequency, and density of the cloud is found in a real-time non-destructive manner. In addition, it is found that radio-frequency sideband cooling compresses the cloud increasing its density. The ability to measure and control the density of a trapped ion cloud might be useful for experiments on low-temperature ion/neutral collisions or recombination rates.

Bulletin of the American Physical Society•May 1, 1993•0 citations

Experimental observations of resistive cooling of He{sup +} ions in a Penning trap

Stephen Cornford, R.A. Keneficks

A Penning trap which is optimized for transverse resistive cooling yields a cooling time constant for trapped He{sup +} ions of as low as 3 seconds, depending on cooling duty cycle. Opposite quadrants of the ring electrode resonate (Q=1800 at 7.4 MHz) with Nb/Ti inductors at 4.2 K. The f{prime} motion is resonantly cooled by operation at 2 Tesla axial field and trap voltages from 12 to 38 Volts. Ion signals are magnetically coupled to a 4 K GaAs fet preamplifier, wide-band 300 K amplifier and linear diode detection system. The rate of cooling is set by fractional time on resonance during sweep of the trap voltage. Shifting and broadening due to space charge are demonstrated. Other interesting features of the ion spectra, such as peak structure, response to heating by f{sup +} - f{sup {minus}}, and growth/decay of the peak components, will be presented.

Bulletin of the American Physical Society•Dec 1, 1992•0 citations
A resonant circuit tuned to a particular frequency of the motion of charged particles stored in a Penning trap and connected to a low noise amplifier allows, at the same time, cooling and non destructive detection of the particles. Its use is widely diffused when single or few particles are stored near the centre of a hyperbolic Penning trap. We present a consistent model that predicts the shape of the induced signal when the tuned circuit is used to detect and cool the axial motion of a cold non neutral plasma stored in an open-ended cylindrical Penning trap. The model correctly accounts for the not negligible axial plasma size. We show that the power spectrum of the signal measured across the tuned circuit provides information about the particle number and insights about the plasma temperature. We report on the design of a HEMT-based cryogenic amplifier working at 14.4 MHz and 4.2 K and the results of the noise measurements. We have measured a drain current noise in the range from 6 to 17 pA/√Hz, which corresponds to an increase of the tuned circuit equivalent temperature of at maximum 0.35 K. The cryogenic amplifier has a very low power consumption from few tens to few hundreds of μW corresponding to a drain current in the range 100–800 μ A . An additional contribution due to the gate noise has been identified when the drain current is below 300 μA; above that value an upper limit of the increase of the equivalent tuned circuit temperature due to this contribution of 0.02 K has been obtained. These features make the tuned circuit connected to this amplifier a promising device for detecting and cooling the axial motion of an electron plasma when the Penning trap is mounted inside a dilution refrigerator.
With a superposition of static electric and magnetic fields, a Penning trap can confine a single electron indefinitely. A weak coupling of the electron to its environment makes the system ideal for precision measurements but renders detection of the electron’s motion difficult. In order to measure fundamental properties of an electron, we employ a Penning trap cooled by a dilution refrigerator. This experimental setup requires extremely low-noise and low-power cryogenic amplifiers for detection of the electron’s axial motion. We design and demonstrate new amplifiers, resulting in a significant decrease of the axial temperature and in an improved signal to noise ratio. We use feedback to cool the axial motion and to self-excite the axial motion for continuous monitoring of the axial frequency. These techniques for cooling and measuring the axial frequency may contribute to an improved measurement of the electron g factor, a more stringent CPT test in leptons, and a more precise measurement of the proton-electron mass ratio.
We developed highly sensitive image-current detection systems based on superconducting toroidal coils and ultra-low noise amplifiers for non-destructive measurements of the axial frequencies (550–800 kHz) of single antiprotons stored in a cryogenic multi-Penning-trap system. The unloaded superconducting tuned circuits show quality factors of up to 500 000, which corresponds to a factor of 10 improvement compared to our previously used solenoidal designs. Connected to ultra-low noise amplifiers and the trap system, signal-to-noise-ratios of 30 dB at quality factors of >20 000 are achieved. In addition, we have developed a superconducting switch which allows continuous tuning of the detector’s quality factor and to sensitively tune the particle-detector interaction. This allowed us to improve frequency resolution at constant averaging time, which is crucial for single antiproton spin-transition spectroscopy experiments, as well as improved measurements of the proton-to-antiproton charge-to-mass ratio.
Abstract An electron-on-helium qubit is a promising physical platform for quantum information technologies. Among all the “blueprints” for the qubit realization, a hybrid Rydberg-spin qubit seems to be a promising one toward quantum computing using electron spins. The main technological challenge on the way to such qubits is a detection of fA range image current induced by a Rydberg transition of a single electron. To address this problem, we aim to use a tank LC-circuit in conjunction with a high-impedance and low power dissipation cryogenic amplifier. Here, we report our progress toward realization of a resonant image current detector with a homemade cryogenic amplifier based on FHX13LG HEMT. We present a detailed characterization of the transistor at room and cryogenic temperatures, as well as details of the amplifier design and performance. At the power dissipation level of amplifier well below 100 μW, the measured voltage and current noise level are 0.6 nV/Hz0.6~{\hbox {nV}}/\sqrt{\textrm{Hz}}0.6 nV/Hz​ 0.6 nV / Hz and below 1.5 fA/ Hz\sqrt{\textrm{Hz}}Hz​ Hz , respectively. Based on the actual image current measurements of the Rydberg transition in a many-electron system on liquid helium, we estimate an SNR = 8 with a measurement bandwidth 1 Hz for the detection of a single-electron transition, providing the noise level at the output is solely determined by the noise of the amplifier.
We developed highly sensitive image-current detection systems based on superconducting toroidal coils and ultra-low noise amplifiers for non-destructive measurements of the axial frequencies (550-800 kHz) of single antiprotons stored in a cryogenic multi-Penning-trap system. The unloaded superconducting tuned circuits show quality factors of up to 500 000, which corresponds to a factor of 10 improvement compared to our previously used solenoidal designs. Connected to ultra-low noise amplifiers and the trap system, signal-to-noise-ratios of 30 dB at quality factors of >20 000 are achieved. In addition, we have developed a superconducting switch which allows continuous tuning of the detector's quality factor and to sensitively tune the particle-detector interaction. This allowed us to improve frequency resolution at constant averaging time, which is crucial for single antiproton spin-transition spectroscopy experiments, as well as improved measurements of the proton-to-antiproton charge-to-mass ratio.
A subset of electrostatic modes of a cold cloud of electrons trapped in a Penning trap has been observed and identified using a recent theoretical model. The detection of these modes is accomplished using electronic techniques which could apply to any ion species. The modes are observed in the low density, low rotation limit of the cloud where the electrons approach a two-dimensional charged disk. We observe both axially symmetric and asymmetric drumhead modes. By measuring the frequency of these modes the shape, rotation frequency, and density of the cloud is found in a real-time non-destructive manner. In addition, it is found that radio-frequency sideband cooling compresses the cloud increasing its density. The ability to measure and control the density of a trapped ion cloud might be useful for experiments on low-temperature ion/neutral collisions or recombination rates.
A Penning trap which is optimized for transverse resistive cooling yields a cooling time constant for trapped He{sup +} ions of as low as 3 seconds, depending on cooling duty cycle. Opposite quadrants of the ring electrode resonate (Q=1800 at 7.4 MHz) with Nb/Ti inductors at 4.2 K. The f{prime} motion is resonantly cooled by operation at 2 Tesla axial field and trap voltages from 12 to 38 Volts. Ion signals are magnetically coupled to a 4 K GaAs fet preamplifier, wide-band 300 K amplifier and linear diode detection system. The rate of cooling is set by fractional time on resonance during sweep of the trap voltage. Shifting and broadening due to space charge are demonstrated. Other interesting features of the ion spectra, such as peak structure, response to heating by f{sup +} - f{sup {minus}}, and growth/decay of the peak components, will be presented.