Recent Developments and Proposed Schemes for Trapped Ion Frequency Standards

Lute Maleki

Published Dec 1, 1981• 1 citation
OpenAlex

Abstract

Ion traps are exciting candidates as future precision frequency sources. Recent developments demonstrate that mercury ion frequency standards are capable of a stability performance comparable to commercial cesium standards. There is, however, considerable room for improvement with regard to the signal to noise problem. The 40 GHz microwave frequency implies that a careful design should be implemented to ensure the elimination of the unwanted side bands in the microwave pump signal. A long life, high performance light source to be used in a trapped mercury ion microwave standard must be developed and the long term performance of a trapped mercury ion microwave standard must be investigated. While newly proposed two photon pumping schemes in conjuction with mercury ions promise exciting developments for both microwave and optical frequency standards, other ions that may be potential candidates should be evaluated for their usefulness.

Related papers

Performance of Micro Mercury Trapped Ion Clock

Thai M. Hoang, Sang Kun Chung, Thanh Le +4 more

Trapped mercury ion is a promising candidate for portable and operational atomic clocks. It operates at the 40.5GHz ground state hyperfine transition of 199Hg ions. Mercury ions are confined in a linear RF trap, buffer-gas cooled, with a long ion storage lifetime in an ultra-high vacuum enclosure without an active pump. The clock transition is optically-pumped and detected with a Hg discharge lamp without need of a laser light source. This approach has proven to provide simplicity in design and implementation and robustness in operation. In this paper, we report our recently effort on developing a micro and low-power mercury trapped ion clock in the DARPA ACES program. In a 5 mm x 10 mm linear trap (0.25 cc), we have achieved the clock fractional frequency stability of .7 × 10-12τ-1/2and reached down to 7 × 10-14after 1000 seconds of averaging. We will discuss the clock performance under different conditions and implications toward a micro mercury trapped ion clock with low power and high stability.

Joint Conference of the European Frequency and Time Forum and IEEE International Frequency Control Symposium (EFTF/IFCS)•Apr 1, 2019•10 citations

Mercury Trapped-Ion Frequency Standard for the Global Positioning System

Robert L. Tjoelker, Eric A. Burt, Sang‒Ho Chung +11 more

We report on progress towards the development of a small, low-mass and -power, high-stability mercury trapped-ion frequency standard for the Global Positioning System. The design performance goal is a frequency stability reaching into the 10(-16) range using technologies that allow for more than 10 years of continuous operational life. Key features include using a multi-pole ion trap to minimize sensitivity to ion-number-dependent effects and a nitrogen buffer gas for long vacuum pump life. The development program is structured in three phases with the goal of gaining early flight experience and keeping development costs in check.

•Nov 27, 2001•6 citations

High-resolution, high-accuracy spectroscopy of trapped ions

Dana J. Berkeland, J.D. Miller, Flávio C. Cruz +6 more

Microwave spectroscopy using trapped and cooled ions can achieve precision and accuracy comparable to the best cesium frequency standards. We discuss standards based on 199Hg+ ions trapped in linear Paul traps: the Jet Propulsion Laboratory (JPL) standard, which uses up to 107 atoms confined near the trap axis, and the recently evaluated NIST standard, which uses approximately ten ions laser cooled and crystalized on the trap axis. We consider future directions in trapped ion frequency standard work, including the use of entangled states for achieving higher precision, and progress on trapped ion optical frequency standards. Finally, we discuss scientific and technical applications of extremely stable frequency standards.

International Conference on Atomic Physics (ICAP)•Jan 1, 1999•1 citation

High-Performance Microwave Frequency Standard Based on Sympathetically Cooled Ions

H. R. Qin, S. N. Miao, J. Z. Han +4 more

The ion microwave frequency standard is a candidate for the next generation of microwave frequency standard with the potential for very wide applications. The Dick effect and second-order Doppler frequency shift (SODFS) limit the performance of ion microwave frequency standards. The introduction of sympathetic cooling technology can suppress the Dick effect and SODFS and improve the stability and accuracy of the frequency standard. However, the sympathetically cooled ion microwave frequency standard has seldom been studied before. This paper reports a high-performance sympathetically cooled ion microwave frequency standard in a Paul trap. Using laser-cooled ${}^{40}{\mathrm{Ca}}^{+}$ as coolant ions, ${}^{113}{\mathrm{Cd}}^{+}$ ion crystal is cooled to below 100 mK and has a coherence lifetime of over 40 s. The short-term frequency stability reached $3.48\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}13}/{\ensuremath{\tau}}^{1/2}$, which is comparable to that of the mercury ion frequency standard. Its uncertainty is $1.5\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}14}$, which is better than that of directly laser-cooled cadmium ion frequency standard.

arXiv:2204.05040•Physical Review Applied•Aug 8, 2022•15 citations

High Resolution, High-Accuracy Spectroscopy of Trapped Ions

Dana J. Berkeland, J.D. Miller, Flávio C. Cruz +6 more

Microwave spectroscopy using trapped and cooled ions can achieve precision and accuracy comparable to the best cesium frequency standards. We discuss standards based on 199Hg+ ions trapped in linear Paul traps: the Jet Propulsion Laboratory (JPL) standard, which uses up to 107 atoms confined near the trap axis, and the recently evaluated NIST standard, which uses approximately ten ions laser cooled and crystalized on the trap axis. We consider future directions in trapped ion frequency standard work, including the use of entangled states for achieving higher precision, and progress on trapped ion optical frequency standards. Finally, we discuss scientific and technical applications of extremely stable frequency standards.

•Jan 1, 1998•0 citations

Trapped-ion and trapped-atom microwave frequency standards

Peter Fisk

The stability and accuracy of atomic microwave frequency standards (atomic clocks) have been improving at a rate exceeding one order of magnitude per decade for the past 30 years. This sustained improvement has been driven mainly by the many and diverse technological applications requiring highly stable and accurate time and frequency standards. There is no reason to suspect that this rate of progress is slowing. Over the last decade, research and development in atomic frequency standards has tended increasingly towards the use of atom and ion trapping technologies to approach the supposedly ideal unperturbed atomic frequency reference consisting of a single atom or ion either motionless or in free-fall in a perfect, field-free vacuum. This work has been facilitated by the relatively recent development of techniques whereby laser light is used to cool, confine and manipulate atoms or ions. This paper reviews the current status of atomic microwave frequency standards based on trapped atoms and trapped ions, and attempts to explain the motivation for their development and the principles of their operation, with particular emphasis on the physics of factors which limit their performance.

Reports on Progress in Physics•Aug 1, 1997•79 citations
Ion traps are exciting candidates as future precision frequency sources. Recent developments demonstrate that mercury ion frequency standards are capable of a stability performance comparable to commercial cesium standards. There is, however, considerable room for improvement with regard to the signal to noise problem. The 40 GHz microwave frequency implies that a careful design should be implemented to ensure the elimination of the unwanted side bands in the microwave pump signal. A long life, high performance light source to be used in a trapped mercury ion microwave standard must be developed and the long term performance of a trapped mercury ion microwave standard must be investigated. While newly proposed two photon pumping schemes in conjuction with mercury ions promise exciting developments for both microwave and optical frequency standards, other ions that may be potential candidates should be evaluated for their usefulness.
Trapped mercury ion is a promising candidate for portable and operational atomic clocks. It operates at the 40.5GHz ground state hyperfine transition of 199Hg ions. Mercury ions are confined in a linear RF trap, buffer-gas cooled, with a long ion storage lifetime in an ultra-high vacuum enclosure without an active pump. The clock transition is optically-pumped and detected with a Hg discharge lamp without need of a laser light source. This approach has proven to provide simplicity in design and implementation and robustness in operation. In this paper, we report our recently effort on developing a micro and low-power mercury trapped ion clock in the DARPA ACES program. In a 5 mm x 10 mm linear trap (0.25 cc), we have achieved the clock fractional frequency stability of .7 × 10-12τ-1/2and reached down to 7 × 10-14after 1000 seconds of averaging. We will discuss the clock performance under different conditions and implications toward a micro mercury trapped ion clock with low power and high stability.
We report on progress towards the development of a small, low-mass and -power, high-stability mercury trapped-ion frequency standard for the Global Positioning System. The design performance goal is a frequency stability reaching into the 10(-16) range using technologies that allow for more than 10 years of continuous operational life. Key features include using a multi-pole ion trap to minimize sensitivity to ion-number-dependent effects and a nitrogen buffer gas for long vacuum pump life. The development program is structured in three phases with the goal of gaining early flight experience and keeping development costs in check.
Microwave spectroscopy using trapped and cooled ions can achieve precision and accuracy comparable to the best cesium frequency standards. We discuss standards based on 199Hg+ ions trapped in linear Paul traps: the Jet Propulsion Laboratory (JPL) standard, which uses up to 107 atoms confined near the trap axis, and the recently evaluated NIST standard, which uses approximately ten ions laser cooled and crystalized on the trap axis. We consider future directions in trapped ion frequency standard work, including the use of entangled states for achieving higher precision, and progress on trapped ion optical frequency standards. Finally, we discuss scientific and technical applications of extremely stable frequency standards.
The ion microwave frequency standard is a candidate for the next generation of microwave frequency standard with the potential for very wide applications. The Dick effect and second-order Doppler frequency shift (SODFS) limit the performance of ion microwave frequency standards. The introduction of sympathetic cooling technology can suppress the Dick effect and SODFS and improve the stability and accuracy of the frequency standard. However, the sympathetically cooled ion microwave frequency standard has seldom been studied before. This paper reports a high-performance sympathetically cooled ion microwave frequency standard in a Paul trap. Using laser-cooled 40Ca+{}^{40}{\mathrm{Ca}}^{+}40Ca+ as coolant ions, 113Cd+{}^{113}{\mathrm{Cd}}^{+}113Cd+ ion crystal is cooled to below 100 mK and has a coherence lifetime of over 40 s. The short-term frequency stability reached $3.48\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}13}/{\ensuremath{\tau}}^{1/2}$, which is comparable to that of the mercury ion frequency standard. Its uncertainty is $1.5\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}14}$, which is better than that of directly laser-cooled cadmium ion frequency standard.
Microwave spectroscopy using trapped and cooled ions can achieve precision and accuracy comparable to the best cesium frequency standards. We discuss standards based on 199Hg+ ions trapped in linear Paul traps: the Jet Propulsion Laboratory (JPL) standard, which uses up to 107 atoms confined near the trap axis, and the recently evaluated NIST standard, which uses approximately ten ions laser cooled and crystalized on the trap axis. We consider future directions in trapped ion frequency standard work, including the use of entangled states for achieving higher precision, and progress on trapped ion optical frequency standards. Finally, we discuss scientific and technical applications of extremely stable frequency standards.
The stability and accuracy of atomic microwave frequency standards (atomic clocks) have been improving at a rate exceeding one order of magnitude per decade for the past 30 years. This sustained improvement has been driven mainly by the many and diverse technological applications requiring highly stable and accurate time and frequency standards. There is no reason to suspect that this rate of progress is slowing. Over the last decade, research and development in atomic frequency standards has tended increasingly towards the use of atom and ion trapping technologies to approach the supposedly ideal unperturbed atomic frequency reference consisting of a single atom or ion either motionless or in free-fall in a perfect, field-free vacuum. This work has been facilitated by the relatively recent development of techniques whereby laser light is used to cool, confine and manipulate atoms or ions. This paper reviews the current status of atomic microwave frequency standards based on trapped atoms and trapped ions, and attempts to explain the motivation for their development and the principles of their operation, with particular emphasis on the physics of factors which limit their performance.