FLaser frequency stabilization and shifting by using modulation transfer spectroscopy

Bing Cheng, 王兆英, 吴彬, 许翱鹏, 王启宇, 徐云飞, 林强

Published Jan 1, 2014• 0 citations
OpenAlex

Abstract

The stabilizing and shifting of laser frequency are very important for the interaction between the laser and atoms. The modulation transfer spectroscopy for the87 Rb atom with D2 line transition F = 2 → F = 3 is used for stabilizing and shifting the frequency of the external cavity grating feedback diode laser. The resonant phase modulator with electro–optical effect is used to generate frequency sideband to lock the laser frequency. In the locking scheme, circularly polarized pump- and probe-beams are used. By optimizing the temperature of the vapor, the pump- and probe-beam intensity, the laser linewidth of 280 kHz is obtained. Furthermore, the magnetic field generated by a solenoid is added into the system. Therefore the system can achieve the frequency locking at any point in a range of hundreds of megahertz frequency shifting with very low power loss.

Related papers

Laser frequency stabilization and shifting by using modulation transfer spectroscopy

Bing Cheng, Zhaoying Wang, Bin Wu +4 more

The stabilizing and shifting of laser frequency are very important for the interaction between the laser and atoms. The modulation transfer spectroscopy for the 87 Rb atom with D2 line transition F = 2 → F′ = 3 is used for stabilizing and shifting the frequency of the external cavity grating feedback diode laser. The resonant phase modulator with electro—optical effect is used to generate frequency sideband to lock the laser frequency. In the locking scheme, circularly polarized pump- and probe-beams are used. By optimizing the temperature of the vapor, the pump- and probe-beam intensity, the laser linewidth of 280 kHz is obtained. Furthermore, the magnetic field generated by a solenoid is added into the system. Therefore the system can achieve the frequency locking at any point in a range of hundreds of megahertz frequency shifting with very low power loss.

Chinese Physics B•Oct 1, 2014•25 citations

Frequency Modulation Spectroscopy and Modulation Transfer Spectroscopy for Laser Stabilization to Rubidium Atomic Resonances

Jinni Wu, Ningxuan Zheng, Wenhui Li

We make a detailed comparison of frequency modulation spectroscopy (FMS) and modulation transfer spectroscopy (MTS) and of their applications in stabilizing laser frequency. In both FMS and MTS, intensity modulations of probe beam are observed in saturated absorption spectroscopy of 87Rb atoms. However, the two modulation profiles appear different due to their distinct origins. Demodulation gives rise to frequency discrimination spectra, which can serve as error signals for frequency locking. The optimal error signal of FMS has a larger linear slope but also a slanted baseline, in contrast to that of MTS. We apply FMS and MTS to stabilize two 780 nm external cavity diode lasers to D2 lines of 87Rb atoms. We measure laser linewidth and stability using beat notes of two independently locked lasers. A laser linewidth as narrow as 12.9 kHz is achieved in FMS locking, compared to 70.8 kHz in MTS locking. A laser stability at the level of 10−12 is reached in both schemes. Nevertheless, FMS locking offers a better short-term stability, whereas MTS locking yields a smaller long-term drift. Our investigation of FMS and MTS would be valuable for making a choice of these spectroscopy techniques for laser stabilization as well as other relevant applications.

Photonics•Oct 21, 2026•0 citations

Laser frequency stabilization in the 10 −14 range via optimized modulation transfer spectroscopy on the 87 Rb D 2 line

Sanglok Lee, Geol Moon, Sang Eon Park +5 more

We present a high-performance laser frequency stabilization method using modulation transfer spectroscopy (MTS) on the rubidium 87 D 2 transition line. A substantial improvement of the laser frequency stability was achieved by searching for the optimal diameter and intensity settings of the probe and pump beam. The frequency instability measured from the beat frequency of two locked external cavity diode lasers (ECDLs) reached a short-term stability of 4.5 × 10 − 14 / τ and did not exceed 2 × 10 −12 until 10 5 s, which is the best performance reported thus far with a D 2 transition. The long-term stability is limited by the offset fluctuations of the baseline induced by the residual amplitude modulation (RAM), which can be further improved by reducing the current temperature variation of about 0.2 K by means of temperature stabilization or through a further reduction of the RAM.

Optics Letters•Jan 17, 2023•47 citations

Frequency Stabilization Technology of 1560 nm Fiber Laser Based on Rubidium Modulation Transfer Spectroscopy

Yang Bai, Jun Ruan, Dandan Liu +3 more

The modulation transfer spectroscopy (MTS) technique used to stabilize two different transition lines of rubidium, required for laser cooling of atoms, is investigated. The laser source is based on a 1560 nm semiconductor laser diode that is amplified by an erbium doped fiber amplifier (EDFA). The laser output is frequency doubled with a fiber coupled periodically poled Lithium Niobate (PPLN) crystal to 780 nm. To address the weak spectral features of repumping laser, a theoretical study is conducted followed by experimental verification. The 87Rb D2 line F = 1 → F' = 0 transition temperature-enhanced MTS signal is observed, which provides a choice to stabilize the repumping laser frquency on the F = 1 → F ' = 2 transition using frequency shifting with acousto-optic modulators (AOMs). Compared with the orignal MTS signal, this method is shown to be more stable and accurate for laser frequency locking. Laer frequency locking is realized by controlling the driving current of the seed laser using digital PID feedback. After locking, the laser linewidth is 35.36 kHz.

IEEE photonics journal•May 20, 2024•5 citations

Frequency stabilization and linewidth narrowing with modulation transfer spectroscopy

Xi Zhang, Hui Liu, Kunliang Jiang +4 more

We report laser frequency stabilization with modulation transfer spectroscopy (MTS) on 85 Rb atoms. With both PZT (piezo-electric transducer) slow-loop feedback and current fastloop feedback to the laser head, we get a linewidth narrowing less than 5 kHz simultaneously. Laser injection to a laser diode and frequency beating with another polarization spectroscopy based stabilization setup are also employed to check the narrow linewidth property. With the help of the technique, a linewidth around kHz-level laser is obtained and pave the way for the locking of the lattice laser of ytterbium clock with transfer cavity technique. The setup can be used as a frequency reference for precise frequency control of atomic clock system.

Wuhan University Journal of Natural Sciences•Jan 12, 2018•0 citations

Study on modulation transfer spectroscopy of Rb atoms

Shengzhan Chen, Qu Liu, Yong Shen +2 more

Closed-loop frequency stabilization-'laser locking'-is required for laser cooling, precision spectroscopy and atomic clocks. We compare the modulation transfer with the direct modulation in signal intensity and noise. By calculating, we get the conclusion that the ratio of signal intensity between modulation transfer scheme and the direct modulation scheme is 1.13:1 whereas the SNR of the modulation transfer scheme and the direct modulation scheme is almost 1.17:1, when the power of the laser is equal and without considering the details of interaction between lights and atoms. And we experimentally realized a stable external-cavity diode laser via MTS referencing on the Rb D2 lines at 780nm.

Chinese Control and Decision Conference (CCDC)•May 1, 2017•1 citation
The stabilizing and shifting of laser frequency are very important for the interaction between the laser and atoms. The modulation transfer spectroscopy for the87 Rb atom with D2 line transition F = 2 → F = 3 is used for stabilizing and shifting the frequency of the external cavity grating feedback diode laser. The resonant phase modulator with electro–optical effect is used to generate frequency sideband to lock the laser frequency. In the locking scheme, circularly polarized pump- and probe-beams are used. By optimizing the temperature of the vapor, the pump- and probe-beam intensity, the laser linewidth of 280 kHz is obtained. Furthermore, the magnetic field generated by a solenoid is added into the system. Therefore the system can achieve the frequency locking at any point in a range of hundreds of megahertz frequency shifting with very low power loss.
The stabilizing and shifting of laser frequency are very important for the interaction between the laser and atoms. The modulation transfer spectroscopy for the 87 Rb atom with D2 line transition F = 2 → F′ = 3 is used for stabilizing and shifting the frequency of the external cavity grating feedback diode laser. The resonant phase modulator with electro—optical effect is used to generate frequency sideband to lock the laser frequency. In the locking scheme, circularly polarized pump- and probe-beams are used. By optimizing the temperature of the vapor, the pump- and probe-beam intensity, the laser linewidth of 280 kHz is obtained. Furthermore, the magnetic field generated by a solenoid is added into the system. Therefore the system can achieve the frequency locking at any point in a range of hundreds of megahertz frequency shifting with very low power loss.
We make a detailed comparison of frequency modulation spectroscopy (FMS) and modulation transfer spectroscopy (MTS) and of their applications in stabilizing laser frequency. In both FMS and MTS, intensity modulations of probe beam are observed in saturated absorption spectroscopy of 87Rb atoms. However, the two modulation profiles appear different due to their distinct origins. Demodulation gives rise to frequency discrimination spectra, which can serve as error signals for frequency locking. The optimal error signal of FMS has a larger linear slope but also a slanted baseline, in contrast to that of MTS. We apply FMS and MTS to stabilize two 780 nm external cavity diode lasers to D2 lines of 87Rb atoms. We measure laser linewidth and stability using beat notes of two independently locked lasers. A laser linewidth as narrow as 12.9 kHz is achieved in FMS locking, compared to 70.8 kHz in MTS locking. A laser stability at the level of 10−12 is reached in both schemes. Nevertheless, FMS locking offers a better short-term stability, whereas MTS locking yields a smaller long-term drift. Our investigation of FMS and MTS would be valuable for making a choice of these spectroscopy techniques for laser stabilization as well as other relevant applications.
We present a high-performance laser frequency stabilization method using modulation transfer spectroscopy (MTS) on the rubidium 87 D 2 transition line. A substantial improvement of the laser frequency stability was achieved by searching for the optimal diameter and intensity settings of the probe and pump beam. The frequency instability measured from the beat frequency of two locked external cavity diode lasers (ECDLs) reached a short-term stability of 4.5 × 10 − 14 / τ and did not exceed 2 × 10 −12 until 10 5 s, which is the best performance reported thus far with a D 2 transition. The long-term stability is limited by the offset fluctuations of the baseline induced by the residual amplitude modulation (RAM), which can be further improved by reducing the current temperature variation of about 0.2 K by means of temperature stabilization or through a further reduction of the RAM.
The modulation transfer spectroscopy (MTS) technique used to stabilize two different transition lines of rubidium, required for laser cooling of atoms, is investigated. The laser source is based on a 1560 nm semiconductor laser diode that is amplified by an erbium doped fiber amplifier (EDFA). The laser output is frequency doubled with a fiber coupled periodically poled Lithium Niobate (PPLN) crystal to 780 nm. To address the weak spectral features of repumping laser, a theoretical study is conducted followed by experimental verification. The 87Rb D2 line F = 1 → F' = 0 transition temperature-enhanced MTS signal is observed, which provides a choice to stabilize the repumping laser frquency on the F = 1 → F ' = 2 transition using frequency shifting with acousto-optic modulators (AOMs). Compared with the orignal MTS signal, this method is shown to be more stable and accurate for laser frequency locking. Laer frequency locking is realized by controlling the driving current of the seed laser using digital PID feedback. After locking, the laser linewidth is 35.36 kHz.
We report laser frequency stabilization with modulation transfer spectroscopy (MTS) on 85 Rb atoms. With both PZT (piezo-electric transducer) slow-loop feedback and current fastloop feedback to the laser head, we get a linewidth narrowing less than 5 kHz simultaneously. Laser injection to a laser diode and frequency beating with another polarization spectroscopy based stabilization setup are also employed to check the narrow linewidth property. With the help of the technique, a linewidth around kHz-level laser is obtained and pave the way for the locking of the lattice laser of ytterbium clock with transfer cavity technique. The setup can be used as a frequency reference for precise frequency control of atomic clock system.
Closed-loop frequency stabilization-'laser locking'-is required for laser cooling, precision spectroscopy and atomic clocks. We compare the modulation transfer with the direct modulation in signal intensity and noise. By calculating, we get the conclusion that the ratio of signal intensity between modulation transfer scheme and the direct modulation scheme is 1.13:1 whereas the SNR of the modulation transfer scheme and the direct modulation scheme is almost 1.17:1, when the power of the laser is equal and without considering the details of interaction between lights and atoms. And we experimentally realized a stable external-cavity diode laser via MTS referencing on the Rb D2 lines at 780nm.