arXiv:2610.40952
Characteristics of VHF and UHF band detection using Stark shifting in Rydberg-atom sensors
Rajavardhan Talashila, Noah Schlossberger, David S. La Mantia, Abby Halasi-Kun, Baran Kayim, Michael A. Viray, Robert Wyllie, Christopher L. Holloway, Nikunjkumar Prajapati
Abstract
We present a characterization of the sensitivity of broadband electric field measurements in the Very High Frequency (VHF) and Ultra High Frequency (UHF) bands using AC stark shifting of Rydberg states. The dependence of the sensitivity and dynamic range of the field measurements using nD5/2 states of cesium for principal quantum numbers n= 60, 70, 80, 90 and 100 is presented, detailing the intricacies involved. The frequencies considered for the measurement range from 50 MHz to 900 MHz for which a transverse electromagnetic (TEM) cell is used to excite the Rydberg atoms with the RF fields. The corresponding sensitivities achieved are in the $(μV/m)/\sqrt{Hz}$ range. A general trend of better sensitivity for higher principal quantum numbers is observed. This work paves the way for a better understanding of the AC Stark dependent field measurements using Rydberg sensors.
AI Summary
The paper characterizes how principal quantum number affects sensitivity, calibration, and dynamic range in broadband Rydberg-atom RF sensing based on AC Stark shifts rather than resonant Autler–Townes splitting. It studies cesium $nD_{5/2}$ states with $n=60,70,80,90,100$ over 50–900 MHz using two-photon EIT in a vapor cell inside a transverse-electromagnetic cell. Applied fields are calibrated by fitting state-dependent Stark shifts using calculated polarizabilities. Sensitivity measurements use an RF local oscillator and a signal offset by 200 kHz, with local-oscillator power and coupling-laser detuning optimized for maximum optical beat-note amplitude. Minimum detectable fields are inferred from the intersection of the signal response with the measured noise floor and normalized by the square root of the 1 kHz resolution bandwidth; measurements are repeated five times. ARC polarizability and Floquet calculations support interpretation of resonances and spectral shifts. Reported sensitivities are in the $\mu\mathrm{V}\,\mathrm{m}^{-1}/\sqrt{\mathrm{Hz}}$ range, generally improving with increasing $n$, with $n=100$ giving the best sensitivity. At 200 MHz, dynamic range decreases from approximately 60 dB at lower $n$ to 40 dB at higher $n$. Calibration factors generally agree across states, with deviations below 5% near resonant features. The reported sensitivities are local-field sensitivities. Nearby transitions and avoided crossings complicate high-$n$ operation and require appropriate laser lockpoints. Field-induced broadening limits useful local-oscillator strength; available coupling-laser power and TEM-cell frequency coverage also constrain the measurements. The quantum-number dependence of the noise floor remains unresolved.
Error found: The displayed calibration-factor equation incorrectly places the polarizability outside the square root: the preceding Stark relation instead requires E/√P_in = √[−2Δf/(αP_in)], although whether this error affected the reported measurements is not established.
Written by gpt-6.1-sol from the paper's full text. Automated: it can be wrong.