Large-scale multipass two-chamber rf atomic magnetometer

D.J. Heilman, Karen L. Sauer, David W. Prescott, C.Z. Motamedi, Nezih Dural, M. V. Romalis, Thomas W. Kornack

Published Nov 8, 2024• 11 citations
arXivHTMLPDF

Abstract

Optically pumped atomic magnetometers are capable of measuring radio-frequency magnetic fields with high sensitivity, but their narrow bandwidth, of the order of 1 kHz, restricts their applications. Further, their full potential is often not realized, with noisy environments or quantum noise curtailing their sensitivity. This study presents a two-chamber atomic magnetometer, with independent control over the tuning field and pumping helicity in each chamber, which addresses both bandwidth and sensitivity limitations. It demonstrates a practical instance of bandwidth expansion, by simultaneous measurement of signals spaced more than 100 kHz apart, at frequencies corresponding to nuclear quadrupole resonance of ammonium nitrate and potassium chlorate, two materials commonly found in improvised explosive devices. It further demonstrates: the intrinsic reduction of common-mode noise, including environmental and probe light-shift noise, down to 2.4 $\mathrm{fT}/\sqrt{\mathrm{Hz}}$, with the use of resonant, but oppositely pumped, atomic chambers; the reduction of photon shot noise down to 28 $\mathrm{aT}/\sqrt{\mathrm{Hz}}$ with the use of 44 probe passes through the two chambers; and the reduction of atom noise down to 34 $\mathrm{aT}/\sqrt{\mathrm{Hz}}$, in agreement with prediction, using a large combined volume, 128 ${\mathrm{cm}}^{3}$, of ${}^{87}\mathrm{Rb}$ atoms. This work underlines the potential sensitivity of these quantum sensors, while exploring the technical limitation, namely the high pump power required, that stands as a barrier to that sensitivity. Further, it marks a significant step towards utilizing atomic magnetometers in applications demanding broader bandwidth or multiplexing, notably simultaneous measurements at far-spaced radio frequencies.

DOI
10.1103/physrevapplied.22.054024

Related papers

Spin-Damping in an Ultra-Sensitive Tunable RF Atomic Magnetometer

Orang Alem

Optically pumped radio frequency (RF) atomic magnetometers have been shown to have an improved sensitivity over standard tuned coils for frequencies less than 50 MHz, making these RF magnetometers attractive for the detection of nuclear quadrupole resonance (NQR) and low-field nuclear magnetic resonance (NMR) signals. In an atomic magnetometer a linearly polarized probe beam measures, through Faraday rotation, the transverse atomic magnetization induced by the resonant RF signal. The resonance, or Larmor, frequency of the magnetometer is easily tuned with a small magnetic field. We construct an atomic magnetometer based on a potassium vapor cell with a measured sensitivity of 0.22 ± 0.02 fT/ √ Hz and a detection bandwidth of 334 ± 11 Hz when tuned to 423 kHz, giving a Q of over 1200. While high Q magnetometers are sensitive to weak magnetic fields, they are also sensitive to magnetic transients, such as those associated with the excitation pulses needed for magnetic resonance. The ringing created by such transients can obscure the signal of interest. This is particularly detrimental for magnetic resonance signals that decay faster than the transverse relaxation rate. We demonstrate that by feeding back part of the optical signal to orthogonal electromagnetic coils, this unwanted ringing can be quickly damped out leaving the magnetometer ready for the detection of the signal. This negative feedback as applied to the K spins is called spin-damping and can be used to relax the K atoms faster by more than two orders of magnitude. Using spin-damping we reduce the dead-time before data acquisition from 0.8 to 0.2 ms in the detection of NQR signals from ammonium nitrate, one of the so-called fertilizer bomb explosives with an NQR frequency of 423 kHz. Furthermore, we discover that spin-damping can not only be used to suppress signals, it can also be used to suppress noise. We show that it can be used to suppress both environmental noise and more fundamental quantum noise, such as photon shot and spin projection noise. The latter suppression opens the door for its use in quantum control to initialize spin systems before measurements. For phase-sensitive detection, as is often used in magnetic resonance, spin-damping suppresses signal and noise in such a way as to increase the sensitivity bandwidth of the magnetometer. We prove, both theoretically and experimentally, that it is possible to increase the magnetometer detection bandwidth by a factor of 3 or more without significant loss in sensitivity. Such gains in bandwidth for a high Q magnetometer translate into savings in detection time, particularly for substance detection where the frequency of the signal of interest may not be completely known.

•Aug 18, 2011•1 citation

Continuous High-Sensitivity and High-Bandwidth Atomic Magnetometer

Rujie Li, Fred N. Baynes, Andre Nicholas Luiten +1 more

Here we demonstrate an atomic magnetometer that has a bandwidth of more than 100 kHz and a sensitivity of $180\phantom{\rule{0.2em}{0ex}}\mathrm{fT/}\ensuremath{\surd}\mathrm{Hz}$ at a Fourier frequency of 8 Hz, and $0.7\phantom{\rule{0.2em}{0ex}}\mathrm{nT/}\ensuremath{\surd}\mathrm{Hz}$ at 100 kHz. These sensitivity measurements are achieved at geophysically useful magnetic field magnitudes (approximately $50\phantom{\rule{0.2em}{0ex}}\ensuremath{\mu}\mathrm{T}$ inside a three-layer $\ensuremath{\mu}$-metal shield) and are limited by the photon shot noise for frequencies above 8 Hz. This device is based on a nonlinear magneto-optical rotation sensor that is operated with an active feedback to the pump modulation frequency. We present a theoretical description of the response functions of the components in the magnetometer. We show that the range of operation of atomic magnetometers can thus be expanded beyond the conventional high-sensitivity, low-bandwidth domain, to provide a high-linearity, high-bandwidth, and high-sensitivity sensor.

Physical Review Applied•Dec 23, 2020•75 citations

High-sensitivity operation of an unshielded single cell radio-frequency atomic magnetometer

Han Yao, Benjamin Maddox, Ferruccio Renzoni

Real-world applications of atomic magnetometers require the ability to operate them with high-sensitivity in the presence of magnetic noise. In the present work, high-sensitivity operation of unshielded atomic magnetometers in a magnetically noisy environment is demonstrated. The distinguishing feature of the demonstrated approach is the implementation of active in-situ bias field stabilization using multiple fluxgate magnetometers. This is combined with the use of a counter-propagating pump and triple-pass probe configuration, to maximize the atomic polarization and the probe rotation respectively, so to reach high-sensitivity. The improvement in sensitivity of the unshielded system with respect to previous realizations is fully characterized, with the contributions of the different modifications of the apparatus individually quantified. The presented set-up is suitable for the detection of long-range magnetic fields, where shielding or differential measurements using multi-sensor set-ups do not constitute viable options.

Optics Express•Oct 21, 2022•29 citations

All-Optical Broadband Atomic Magnetometer Utilizing Raman-Driven Spin Noise Spectroscopy

Sayari Majumder, Saptarishi Chaudhuri

We report the development and performance of an atomic magnetometer (AM) utilizing the Raman-driven spin noise spectroscopy (RDSNS) technique, which measures the intrinsic spin fluctuations of an atomic medium subjected to a coherent two-photon optical drive. The RDSNS signal is noninvasively detected using a far-off resonant weak probe light utilizing a polarimetric detection scheme. This overcomes the limitations of the small dynamic range and low bandwidth of traditional optically pumped magnetometers (OPMs) or AMs operating in the spin exchange relaxation-free (SERF) regime. We experimentally demonstrate, using both paraffin-coated and buffer-gas-filled vapor cell containing neutral rubidium atoms, that the dynamic range of such magnetometers can reach beyond 150 G while maintaining a dc magnetometer resolution of around 30 pT and a magnetic noise sensitivity of around$30~\text {pT}/\sqrt {\text {Hz}}$at a bandwidth of 100 Hz and a magnetic field of about 0.6 mT. Being an all-optical setup, this instrument is largely immune to interference with stray radio frequency fields and has the potential to be miniaturized and field deployable. One of the practical applications would be to measure variations in geomagnetic fields precisely without having to shield the magnetometer.

IEEE Transactions on Instrumentation and Measurement•Jan 1, 2025•6 citations

Ultra-sensitive high-density Rb-87 radio-frequency magnetometer

Igor Savukov, Todor S. Karaulanov, M. G. Boshier

Radio-frequency (RF) atomic magnetometers (AMs) can be used in many applications, such as magnetic resonance imaging and nuclear quadrupole resonance. High-density AMs provide both superior sensitivity and large bandwidth. Previously, high-density potassium AMs were demonstrated, but these magnetometers have various disadvantages, such as high-temperature of operation and bulky design. We demonstrate a rubidium-87 RF AM with 5 fT/Hz1/2 sensitivity (3 fT Hz1/2 probe noise), which is comparable to that of the best potassium magnetometers. Our magnetometer also features a simple fiber-optic design, providing maximum flexibility for magnetic-field measurements.

Applied Physics Letters•Jan 13, 2014•34 citations

Spin damping in an rf atomic magnetometer

Orang Alem, Karen L. Sauer, M. V. Romalis

Under negative feedback, the quality factor $Q$ of a radio-frequency atomic magnetometer can be decreased by more than two orders of magnitude, so that any initial perturbation of the polarized spin system can be rapidly damped, preparing the magnetometer for detection of the desired signal. We find that noise is also suppressed under such spin damping, with a characteristic spectral response corresponding to the type of noise; therefore magnetic, photon shot, and spin-projection noise can be measured distinctly. While the suppression of resonant photon shot noise implies the closed-loop production of polarization-squeezed light, the suppression of resonant spin-projection noise does not imply spin squeezing, rather simply the broadening of the noise spectrum with $Q$. Furthermore, the application of spin damping during phase-sensitive detection suppresses both signal and noise in such a way as to increase the sensitivity bandwidth. We demonstrate a threefold increase in the magnetometer's bandwidth while maintaining 0.3 fT/$\sqrt{\mathrm{Hz}}$ sensitivity.

arXiv:1207.2842•Physical Review A•Jan 16, 2013•19 citations
Optically pumped atomic magnetometers are capable of measuring radio-frequency magnetic fields with high sensitivity, but their narrow bandwidth, of the order of 1 kHz, restricts their applications. Further, their full potential is often not realized, with noisy environments or quantum noise curtailing their sensitivity. This study presents a two-chamber atomic magnetometer, with independent control over the tuning field and pumping helicity in each chamber, which addresses both bandwidth and sensitivity limitations. It demonstrates a practical instance of bandwidth expansion, by simultaneous measurement of signals spaced more than 100 kHz apart, at frequencies corresponding to nuclear quadrupole resonance of ammonium nitrate and potassium chlorate, two materials commonly found in improvised explosive devices. It further demonstrates: the intrinsic reduction of common-mode noise, including environmental and probe light-shift noise, down to 2.4 fT/Hz\mathrm{fT}/\sqrt{\mathrm{Hz}}fT/Hz​, with the use of resonant, but oppositely pumped, atomic chambers; the reduction of photon shot noise down to 28 aT/Hz\mathrm{aT}/\sqrt{\mathrm{Hz}}aT/Hz​ with the use of 44 probe passes through the two chambers; and the reduction of atom noise down to 34 aT/Hz\mathrm{aT}/\sqrt{\mathrm{Hz}}aT/Hz​, in agreement with prediction, using a large combined volume, 128 cm3{\mathrm{cm}}^{3}cm3, of 87Rb{}^{87}\mathrm{Rb}87Rb atoms. This work underlines the potential sensitivity of these quantum sensors, while exploring the technical limitation, namely the high pump power required, that stands as a barrier to that sensitivity. Further, it marks a significant step towards utilizing atomic magnetometers in applications demanding broader bandwidth or multiplexing, notably simultaneous measurements at far-spaced radio frequencies.
Optically pumped radio frequency (RF) atomic magnetometers have been shown to have an improved sensitivity over standard tuned coils for frequencies less than 50 MHz, making these RF magnetometers attractive for the detection of nuclear quadrupole resonance (NQR) and low-field nuclear magnetic resonance (NMR) signals. In an atomic magnetometer a linearly polarized probe beam measures, through Faraday rotation, the transverse atomic magnetization induced by the resonant RF signal. The resonance, or Larmor, frequency of the magnetometer is easily tuned with a small magnetic field. We construct an atomic magnetometer based on a potassium vapor cell with a measured sensitivity of 0.22 ± 0.02 fT/ √ Hz and a detection bandwidth of 334 ± 11 Hz when tuned to 423 kHz, giving a Q of over 1200. While high Q magnetometers are sensitive to weak magnetic fields, they are also sensitive to magnetic transients, such as those associated with the excitation pulses needed for magnetic resonance. The ringing created by such transients can obscure the signal of interest. This is particularly detrimental for magnetic resonance signals that decay faster than the transverse relaxation rate. We demonstrate that by feeding back part of the optical signal to orthogonal electromagnetic coils, this unwanted ringing can be quickly damped out leaving the magnetometer ready for the detection of the signal. This negative feedback as applied to the K spins is called spin-damping and can be used to relax the K atoms faster by more than two orders of magnitude. Using spin-damping we reduce the dead-time before data acquisition from 0.8 to 0.2 ms in the detection of NQR signals from ammonium nitrate, one of the so-called fertilizer bomb explosives with an NQR frequency of 423 kHz. Furthermore, we discover that spin-damping can not only be used to suppress signals, it can also be used to suppress noise. We show that it can be used to suppress both environmental noise and more fundamental quantum noise, such as photon shot and spin projection noise. The latter suppression opens the door for its use in quantum control to initialize spin systems before measurements. For phase-sensitive detection, as is often used in magnetic resonance, spin-damping suppresses signal and noise in such a way as to increase the sensitivity bandwidth of the magnetometer. We prove, both theoretically and experimentally, that it is possible to increase the magnetometer detection bandwidth by a factor of 3 or more without significant loss in sensitivity. Such gains in bandwidth for a high Q magnetometer translate into savings in detection time, particularly for substance detection where the frequency of the signal of interest may not be completely known.
Here we demonstrate an atomic magnetometer that has a bandwidth of more than 100 kHz and a sensitivity of $180\phantom{\rule{0.2em}{0ex}}\mathrm{fT/}\ensuremath{\surd}\mathrm{Hz}$ at a Fourier frequency of 8 Hz, and $0.7\phantom{\rule{0.2em}{0ex}}\mathrm{nT/}\ensuremath{\surd}\mathrm{Hz}$ at 100 kHz. These sensitivity measurements are achieved at geophysically useful magnetic field magnitudes (approximately $50\phantom{\rule{0.2em}{0ex}}\ensuremath{\mu}\mathrm{T}$ inside a three-layer $\ensuremath{\mu}$-metal shield) and are limited by the photon shot noise for frequencies above 8 Hz. This device is based on a nonlinear magneto-optical rotation sensor that is operated with an active feedback to the pump modulation frequency. We present a theoretical description of the response functions of the components in the magnetometer. We show that the range of operation of atomic magnetometers can thus be expanded beyond the conventional high-sensitivity, low-bandwidth domain, to provide a high-linearity, high-bandwidth, and high-sensitivity sensor.
Real-world applications of atomic magnetometers require the ability to operate them with high-sensitivity in the presence of magnetic noise. In the present work, high-sensitivity operation of unshielded atomic magnetometers in a magnetically noisy environment is demonstrated. The distinguishing feature of the demonstrated approach is the implementation of active in-situ bias field stabilization using multiple fluxgate magnetometers. This is combined with the use of a counter-propagating pump and triple-pass probe configuration, to maximize the atomic polarization and the probe rotation respectively, so to reach high-sensitivity. The improvement in sensitivity of the unshielded system with respect to previous realizations is fully characterized, with the contributions of the different modifications of the apparatus individually quantified. The presented set-up is suitable for the detection of long-range magnetic fields, where shielding or differential measurements using multi-sensor set-ups do not constitute viable options.
We report the development and performance of an atomic magnetometer (AM) utilizing the Raman-driven spin noise spectroscopy (RDSNS) technique, which measures the intrinsic spin fluctuations of an atomic medium subjected to a coherent two-photon optical drive. The RDSNS signal is noninvasively detected using a far-off resonant weak probe light utilizing a polarimetric detection scheme. This overcomes the limitations of the small dynamic range and low bandwidth of traditional optically pumped magnetometers (OPMs) or AMs operating in the spin exchange relaxation-free (SERF) regime. We experimentally demonstrate, using both paraffin-coated and buffer-gas-filled vapor cell containing neutral rubidium atoms, that the dynamic range of such magnetometers can reach beyond 150 G while maintaining a dc magnetometer resolution of around 30 pT and a magnetic noise sensitivity of around30 pT/Hz30~\text {pT}/\sqrt {\text {Hz}}30 pT/Hz​at a bandwidth of 100 Hz and a magnetic field of about 0.6 mT. Being an all-optical setup, this instrument is largely immune to interference with stray radio frequency fields and has the potential to be miniaturized and field deployable. One of the practical applications would be to measure variations in geomagnetic fields precisely without having to shield the magnetometer.
Radio-frequency (RF) atomic magnetometers (AMs) can be used in many applications, such as magnetic resonance imaging and nuclear quadrupole resonance. High-density AMs provide both superior sensitivity and large bandwidth. Previously, high-density potassium AMs were demonstrated, but these magnetometers have various disadvantages, such as high-temperature of operation and bulky design. We demonstrate a rubidium-87 RF AM with 5 fT/Hz1/2 sensitivity (3 fT Hz1/2 probe noise), which is comparable to that of the best potassium magnetometers. Our magnetometer also features a simple fiber-optic design, providing maximum flexibility for magnetic-field measurements.
Under negative feedback, the quality factor QQQ of a radio-frequency atomic magnetometer can be decreased by more than two orders of magnitude, so that any initial perturbation of the polarized spin system can be rapidly damped, preparing the magnetometer for detection of the desired signal. We find that noise is also suppressed under such spin damping, with a characteristic spectral response corresponding to the type of noise; therefore magnetic, photon shot, and spin-projection noise can be measured distinctly. While the suppression of resonant photon shot noise implies the closed-loop production of polarization-squeezed light, the suppression of resonant spin-projection noise does not imply spin squeezing, rather simply the broadening of the noise spectrum with QQQ. Furthermore, the application of spin damping during phase-sensitive detection suppresses both signal and noise in such a way as to increase the sensitivity bandwidth. We demonstrate a threefold increase in the magnetometer's bandwidth while maintaining 0.3 fT/Hz\sqrt{\mathrm{Hz}}Hz​ sensitivity.