arXiv:2303.07566

Towards a transportable Ca$^+$ optical clock with a systematic uncertainty of $4.8\times 10^{-18}$

Mengyan Zeng, Yao Tien Huang, Baolin Zhang, Yanmei Hao, Zixiao 子晓 Ma 马, Ruming Hu, Huaqing Zhang, Zheng Chen, Miao Wang, Hua Guan, Kelin Gao

Published Mar 14, 2023• 0 citations
arXivHTMLPDF

Abstract

We present a compact, long-term nearly continuous operation of a room-temperature Ca$^+$ optical clock setup towards a transportable clock, achieving an overall systematic uncertainty of $4.8\times 10^{-18}$ and an uptime rate of 97.8% over an 8-day period. The active liquid-cooling scheme is adopted, combined with the precise temperature measurement with 13 temperature sensors both inside and outside the vacuum chamber to ensure the accurate evaluation of the thermal environment for the optical clock. The environmental temperature uncertainty is evaluated as 293.31(0.4) K, corresponding to a blackbody radiation (BBR) frequency shift uncertainty of $4.6\times 10^{-18}$, which is reduced more than two times compared to our previous work. Through the frequency comparison between the room temperature Ca$^+$ optical clock and a cryogenic Ca$^+$ optical clock, the overall uncertainty of the clock comparison is $7.5\times 10^{-18}$, including a statistic uncertainty of $4.9\times 10^{-18}$ and a systematic uncertainty of $5.7\times 10^{-18}$. This work provides a set of feasible implementations for high-precision transportable ion optical clocks.

DOI
10.48550/arxiv.2303.07566

Related papers

A liquid nitrogen-cooled Ca+ optical clock with systematic uncertainty of 3×10-18

Yao Huang, Baolin Zhang, Mengyan Zeng +6 more

Abstract Here we present a liquid nitrogen-cooled Ca+ optical clock with an overall systematic uncertainty of 3×10-18. In contrast with the room-temperature Ca+ optical clock that we have reported previously, the temperature of the blackbody radiation (BBR) shield in vacuum has been reduced to 82(5) K using liquid nitrogen. An ion trap with a lower heating rate and improved cooling lasers were also introduced. This allows cooling the ion temperature to the Doppler cooling limit during the clock operation, and the systematic uncertainty due to the ion’s secular (thermal) motion is reduced to < 1×10-18. The uncertainty due to the probe laser light shift and the servo error are also reduced to < 1×10-19 and 4×10-19 with the hyper-Ramsey method and the higher-order servo algorithm, respectively. By comparing the output frequency of the cryogenic clock to that of a room-temperature clock, the differential BBR shift between the two was measured with a fractional statistical uncertainty of 7×10-18. The differential BBR shift was used to calculate the static differential polarizability, and it was found in excellent agreement with our previous measurement with a different method. This work suggests that the BBR shift of optical clocks can be well suppressed in a liquid nitrogen environment. This is advantageous because conventional liquid-helium cryogenic systems for optical clocks are more expensive and complicated. Moreover, the proposed system can be used to suppress the BBR shift significantly in other types of optical clocks such as Yb+, Sr+, Yb, Sr, etc.

arXiv:2103.08913•Mar 16, 2021•10 citations

Liquid-Nitrogen-Cooled Ca + Optical Clock with Systematic Uncertainty of 3×10−18

Yao Huang, Baolin Zhang, Mengyan Zeng +7 more

We present a liquid-nitrogen-cooled $\mathrm{Ca}$${}^{+}$ optical clock with an overall systematic uncertainty of $3.0\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}18}$. In contrast to the room-temperature $\mathrm{Ca}$${}^{+}$ optical clock that we have reported previously, the cryogenic black-body radiation (BBR) shield in vacuum is cooled to $82\ifmmode\pm\else\textpm\fi{}5$ K using liquid nitrogen. We also implement an ion trap with a reduced heating rate and improved laser cooling. This allows the ion temperature to fall to the Doppler-cooling limit during the clock operation and the systematic uncertainty associated with the secular (thermal) motion of the ion is reduced to $<1\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}18}$. The uncertainty arising from the probe laser light shift and the servo error is also reduced to $<1\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}19}$ and $4\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}19}$ with the hyper-Ramsey method and the higher-order servo algorithm, respectively. By comparing the output frequency of the cryogenic clock to that of a room-temperature clock, the differential BBR shift between the two is determined with a fractional statistical uncertainty of $7\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}18}$. The differential BBR shift is used to calculate the static differential polarizability and the result is found to be in excellent agreement with our previous measurement using a different method. This work suggests that the BBR shift of optical clocks can be suppressed well in a liquid-nitrogen environment. Systems similar to what is presented here can also be used to suppress the BBR shift significantly in other types of optical clocks, such as $\mathrm{Yb}$${}^{+}$, $\mathrm{Sr}$${}^{+}$, $\mathrm{Yb}$, Sr, etc.

Physical Review Applied•Mar 15, 2022•82 citations

Liquid-Nitrogen-Cooled 40 Ca + Ion Optical Clock with a Systematic Uncertainty of 4.4 × 10 − 19

Bao-Lin 宝林 Zhang 张, Zixiao Ma, Yao Huang +8 more

We report a single-ion optical clock based on the 4S_{1/2}→3D_{5/2} transition of the ^{40}Ca^{+} ion, operated in a liquid nitrogen cryogenic environment, achieving a total systematic uncertainty of 4.4×10^{-19}. We employ a refined temperature evaluation scheme to reduce the frequency uncertainty due to blackbody radiation, and 3D sideband cooling to minimize the second-order Doppler shift. We have precisely determined the average Zeeman coefficient of the ^{40}Ca^{+} clock transition to be 14.345(15) Hz/mT^{2}, thereby significantly reducing the quadratic Zeeman shift uncertainty. Moreover, the cryogenic environment enables the lowest reported heating rate due to ambient electric field noise in trapped-ion optical clocks.

Physical Review Letters•Dec 22, 2025•13 citations

A compact, transportable optical clock with 1×10−17 uncertainty and its absolute frequency measurement

Jian Zhong Cao, Jinbo Yuan, Shaomao Wang +13 more

We report a robust, compact, and transportable optical clock (TOC-729-2) based on a trapped single 40Ca+ ion with a systematic uncertainty of 1.1×10−17, which is limited by the black-body radiation shift uncertainty at room temperature. By comparing it with the previous transportable optical clock (TOC-729-1) similar but completely independent, the instability was measured to be better than 1.2×10−14/τ. Benefiting from the modular and integrated design, this TOC was constructed in a volume of ∼0.33 m3 excluding the controlling electronics in 19-in. racks. After being moved ∼1200 km away by express delivery, the single-ion signal was restored within 24 h. With the TOC uptime of 92% in 35-day period, the absolute frequency of the 729 nm transition of 40Ca+ was measured using a satellite link to International Atomic Time (TAI) to provide traceability to the SI second, and the result is 411 042 129 776 400.15(22) Hz, corresponding to a relative uncertainty of 5.3×10−16.

Applied Physics Letters•Jan 31, 2022•22 citations

Nearly continuous Ca^+ optical clocks with stability at the 10^-18 level

Yao Song Huang, Baolin Zhang, Mengyan Zeng +6 more

Abstract Optical clocks are important for precise measurements in the field of physics. As reported, both the instability and uncertainty of optical lattice clocks are more than two orders of magnitude smaller than those of the best microwave clocks. Therefore, in the near future, optical clocks could be used to redefine the second. Nevertheless, an optical clock with reliability comparable to microwave clocks has not been achieved thus far. In this paper, we compared the frequencies of two Ca^+ optical clocks that were nearly continuously operated for 31 days. Through the comparison experiment, the frequency stability of a single clocks was found to be 6.3×10^-18 at an averaging time of 520 000 s and 7.9×10^-18 at averaging time of 262000 s, while the operating uptime rate reached more than 90% in the period of around 5 days. Thus, our experiment demonstrated that it is possible to increase the stability of single-ion optical clocks to the 10^-18 level, while still maintaining quasi-continuous operation with a high operating rate. This result further confirms that optical clocks can potentially be used to redefine the second.

•Mar 11, 2021•0 citations

Progress towards a 40Ca+ optical clock with a fractional uncertainty at the 10−18 level

Yao Huang, Hua Guan, Kelin Gao

We recently reported a 5×10-17level frequency comparison of two40Ca+optical clocks, one of which has fractional frequency uncertainty of 3.4×10-17, limited by the BBR effect. Here we present progress towards a40Ca+optical clock with uncertainty at the 10-18level. The differential static scalar polarizability Δα0 of the40Ca+ion clock transition is obtained by precisely measuring the “magic” rf drive frequency. Δα0 is measured as -7.2653(44)x10-40J m2/V2, the blackbody radiation shift is then calculated to be 0.3790(2) Hz at 300 K. The contribution of the blackbody shift coefficient to the uncertainty of the optical clock at room temperature is reduced to 5.6×10-19, the excess micromotion induced clock uncertainty is also reduced to the 10-19level by choosing the appropriate trap drive frequency. With a suppression of the BBR field temperature uncertainty to the sub-Kelvin level and improved clock stability, a robust, relatively low-cost40Ca+ion clock with uncertainty at 10-18level can be made.

Joint Conference of the European Frequency and Time Forum and IEEE International Frequency Control Symposium (EFTF/IFCS)•Jul 1, 2017•0 citations
Towards a transportable Ca+^++ optical clock with a systematic uncertainty of 4.8×10−184.8\times 10^{-18}4.8×10−18
We present a compact, long-term nearly continuous operation of a room-temperature Ca+^++ optical clock setup towards a transportable clock, achieving an overall systematic uncertainty of 4.8×10−184.8\times 10^{-18}4.8×10−18 and an uptime rate of 97.8% over an 8-day period. The active liquid-cooling scheme is adopted, combined with the precise temperature measurement with 13 temperature sensors both inside and outside the vacuum chamber to ensure the accurate evaluation of the thermal environment for the optical clock. The environmental temperature uncertainty is evaluated as 293.31(0.4) K, corresponding to a blackbody radiation (BBR) frequency shift uncertainty of 4.6×10−184.6\times 10^{-18}4.6×10−18, which is reduced more than two times compared to our previous work. Through the frequency comparison between the room temperature Ca+^++ optical clock and a cryogenic Ca+^++ optical clock, the overall uncertainty of the clock comparison is 7.5×10−187.5\times 10^{-18}7.5×10−18, including a statistic uncertainty of 4.9×10−184.9\times 10^{-18}4.9×10−18 and a systematic uncertainty of 5.7×10−185.7\times 10^{-18}5.7×10−18. This work provides a set of feasible implementations for high-precision transportable ion optical clocks.
Abstract Here we present a liquid nitrogen-cooled Ca+ optical clock with an overall systematic uncertainty of 3×10-18. In contrast with the room-temperature Ca+ optical clock that we have reported previously, the temperature of the blackbody radiation (BBR) shield in vacuum has been reduced to 82(5) K using liquid nitrogen. An ion trap with a lower heating rate and improved cooling lasers were also introduced. This allows cooling the ion temperature to the Doppler cooling limit during the clock operation, and the systematic uncertainty due to the ion’s secular (thermal) motion is reduced to < 1×10-18. The uncertainty due to the probe laser light shift and the servo error are also reduced to < 1×10-19 and 4×10-19 with the hyper-Ramsey method and the higher-order servo algorithm, respectively. By comparing the output frequency of the cryogenic clock to that of a room-temperature clock, the differential BBR shift between the two was measured with a fractional statistical uncertainty of 7×10-18. The differential BBR shift was used to calculate the static differential polarizability, and it was found in excellent agreement with our previous measurement with a different method. This work suggests that the BBR shift of optical clocks can be well suppressed in a liquid nitrogen environment. This is advantageous because conventional liquid-helium cryogenic systems for optical clocks are more expensive and complicated. Moreover, the proposed system can be used to suppress the BBR shift significantly in other types of optical clocks such as Yb+, Sr+, Yb, Sr, etc.
We present a liquid-nitrogen-cooled Ca\mathrm{Ca}Ca+{}^{+}+ optical clock with an overall systematic uncertainty of $3.0\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}18}$. In contrast to the room-temperature Ca\mathrm{Ca}Ca+{}^{+}+ optical clock that we have reported previously, the cryogenic black-body radiation (BBR) shield in vacuum is cooled to $82\ifmmode\pm\else\textpm\fi{}5$ K using liquid nitrogen. We also implement an ion trap with a reduced heating rate and improved laser cooling. This allows the ion temperature to fall to the Doppler-cooling limit during the clock operation and the systematic uncertainty associated with the secular (thermal) motion of the ion is reduced to $<1\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}18}$. The uncertainty arising from the probe laser light shift and the servo error is also reduced to $<1\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}19}$ and $4\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}19}$ with the hyper-Ramsey method and the higher-order servo algorithm, respectively. By comparing the output frequency of the cryogenic clock to that of a room-temperature clock, the differential BBR shift between the two is determined with a fractional statistical uncertainty of $7\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}18}$. The differential BBR shift is used to calculate the static differential polarizability and the result is found to be in excellent agreement with our previous measurement using a different method. This work suggests that the BBR shift of optical clocks can be suppressed well in a liquid-nitrogen environment. Systems similar to what is presented here can also be used to suppress the BBR shift significantly in other types of optical clocks, such as Yb\mathrm{Yb}Yb+{}^{+}+, Sr\mathrm{Sr}Sr+{}^{+}+, Yb\mathrm{Yb}Yb, Sr, etc.
We report a single-ion optical clock based on the 4S_{1/2}→3D_{5/2} transition of the ^{40}Ca^{+} ion, operated in a liquid nitrogen cryogenic environment, achieving a total systematic uncertainty of 4.4×10^{-19}. We employ a refined temperature evaluation scheme to reduce the frequency uncertainty due to blackbody radiation, and 3D sideband cooling to minimize the second-order Doppler shift. We have precisely determined the average Zeeman coefficient of the ^{40}Ca^{+} clock transition to be 14.345(15) Hz/mT^{2}, thereby significantly reducing the quadratic Zeeman shift uncertainty. Moreover, the cryogenic environment enables the lowest reported heating rate due to ambient electric field noise in trapped-ion optical clocks.
We report a robust, compact, and transportable optical clock (TOC-729-2) based on a trapped single 40Ca+ ion with a systematic uncertainty of 1.1×10−17, which is limited by the black-body radiation shift uncertainty at room temperature. By comparing it with the previous transportable optical clock (TOC-729-1) similar but completely independent, the instability was measured to be better than 1.2×10−14/τ. Benefiting from the modular and integrated design, this TOC was constructed in a volume of ∼0.33 m3 excluding the controlling electronics in 19-in. racks. After being moved ∼1200 km away by express delivery, the single-ion signal was restored within 24 h. With the TOC uptime of 92% in 35-day period, the absolute frequency of the 729 nm transition of 40Ca+ was measured using a satellite link to International Atomic Time (TAI) to provide traceability to the SI second, and the result is 411 042 129 776 400.15(22) Hz, corresponding to a relative uncertainty of 5.3×10−16.
Abstract Optical clocks are important for precise measurements in the field of physics. As reported, both the instability and uncertainty of optical lattice clocks are more than two orders of magnitude smaller than those of the best microwave clocks. Therefore, in the near future, optical clocks could be used to redefine the second. Nevertheless, an optical clock with reliability comparable to microwave clocks has not been achieved thus far. In this paper, we compared the frequencies of two Ca^+ optical clocks that were nearly continuously operated for 31 days. Through the comparison experiment, the frequency stability of a single clocks was found to be 6.3×10^-18 at an averaging time of 520 000 s and 7.9×10^-18 at averaging time of 262000 s, while the operating uptime rate reached more than 90% in the period of around 5 days. Thus, our experiment demonstrated that it is possible to increase the stability of single-ion optical clocks to the 10^-18 level, while still maintaining quasi-continuous operation with a high operating rate. This result further confirms that optical clocks can potentially be used to redefine the second.
We recently reported a 5×10-17level frequency comparison of two40Ca+optical clocks, one of which has fractional frequency uncertainty of 3.4×10-17, limited by the BBR effect. Here we present progress towards a40Ca+optical clock with uncertainty at the 10-18level. The differential static scalar polarizability Δα0 of the40Ca+ion clock transition is obtained by precisely measuring the “magic” rf drive frequency. Δα0 is measured as -7.2653(44)x10-40J m2/V2, the blackbody radiation shift is then calculated to be 0.3790(2) Hz at 300 K. The contribution of the blackbody shift coefficient to the uncertainty of the optical clock at room temperature is reduced to 5.6×10-19, the excess micromotion induced clock uncertainty is also reduced to the 10-19level by choosing the appropriate trap drive frequency. With a suppression of the BBR field temperature uncertainty to the sub-Kelvin level and improved clock stability, a robust, relatively low-cost40Ca+ion clock with uncertainty at 10-18level can be made.